의료기기법 시행규칙

287877 · 2026-07-01

Ministry of Food and Drug Safety ( Medical Device Policy Division ), 043-719-3758, 3783 Article 1 (Purpose) The purpose of this ruleis to prescribe matters delegated by the Medical Devices Act and the Enforcement Decree of the same Act , and matters necessary for the implementation thereof. Article 2 (Standards and procedures for classification and designation of medical devices pursuant to Article 3, Paragraph 2 of the Medical Devices Act (hereinafter referred to as the “Act”) are as set forth in Table 1 . Article 3 (Procedure for Application for Manufacturing Permit) ①A person who intends to obtain a medical device manufacturing permit pursuant to the main clause of Article 6, Paragraph 1 of the Act shall submit an application in the form of Annex No. 1 (including an application in the form of an electronic document) to the head of the regional Food and Drug Administration having jurisdiction over the location of the manufacturing facility, attaching the documents specified in each of the following subparagraphs (including electronic documents). In this case, if there are two or more manufacturing facilities located at different places, the application may be submitted to the head of the regional Food and Drug Administration having jurisdiction over any one of the manufacturing facilities. <Amended Jan. 16, 2024; Apr. 1, 2025> 1. A medical certificate from a physician proving that it does not fall under the main clause of Article 6 , Paragraph 1, Subparagraph 1 of the Act , or a medical certificate from a specialist capable of proving it in cases falling under the proviso of the same subparagraph, provided that six months have not elapsed from the date of issuance (corporations shall not submit this). 2. A medical certificate from a physician proving that it does not fall under Article 6, Paragraph 1, Subparagraph 3 of the Act , provided that six months have not elapsed since the date of issuance (corporations shall not submit this). 3. The following documents regarding the Quality Manager pursuant to Article 6, Paragraph 7 of the Act (hereinafter referred to as the “Quality Manager”) a. Documents that can verify the qualification pursuant to Article 11, Paragraph 2. However, if the qualification of the quality manager has been previously verified through the Medical Device Integrated Information System pursuant to Article 31-3, Paragraph 1 of the Act (hereinafter referred to as the “Medical Device Integrated Information System”), the submission thereof shall be waived. B. Documents verifying employment status, such as a certificate of employment ② The head of the regional Food and Drug Administration who receives the application pursuant to Paragraph 1 shall verify the certificate of corporate registration (only applicable to corporations) through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act . ③ If the application for permission under Paragraph 1 meets the standards, the Regional Director of the Food and Drug Safety Administration shall issue a permit in the form of Annex 2 to the applicant and record the matters of each of the following subparagraphs in the manufacturing permit register. <Amended Oct. 22, 2019> 1. Permit Number and Date of Permit 2. Name and Resident Registration Number of the manufacturer (in the case of a corporation, the name and Resident Registration Number of its representative) 3. Name and location of the manufacturing facility 4. Name, Resident Registration Number, and Classification of Qualifications of the Quality Manager Article 4 (Subjects of Manufacturing Permit, Manufacturing Certification, and Manufacturing Notification) ①Medical devices requiring manufacturing certification or manufacturing notification by product category pursuant to Article 6, Paragraph 2, Subparagraph 1 of the Act shall be classified according to the following subparagraphs. 1. Manufacturing Certification by Product Class: Medical devices of Class 2 designated pursuant to Article 2 (hereinafter referred to as “Class”), among which medical devices prescribed and notified by the Minister of Food and Drug Safety 2. Manufacturing Notification by Product Class: Medical devices among Class 1 medical devices designated and notified by the Minister of Food and Drug Safety ② Medical devices requiring a product-specific manufacturing license or certification, or a manufacturing notification, pursuant to Article 6, Paragraph 2, Subparagraph 2 of the Act, shall be classified according to the following subparagraphs. <Amended Mar. 8, 2024; Aug. 7, 2024> 1. Manufacturing license by item: Medical devices in each of the following subparagraphs a. Class 3 and Class 4 medical devices B. Among Class 2 medical devices, medical devices excluding those under Paragraph 1, Subparagraph 1, Item C of this Subparagraph, and Subparagraph 2 of this Paragraph C. Among Class 2 medical devices, medical devices that are not substantially equivalent in structure, principle, performance, purpose of use, or method of use to medical devices that have already been licensed, certified, or reported (excluding medical devices for export) 2. Manufacturing Certification by Item: Among Class 2 medical devices, medical devices excluding those under Paragraph 1, Subparagraph 1, as designated and notified by the Minister of Food and Drug Safety 3. Manufacturing notification by item: Class 1 medical devices excluding medical devices under Paragraph 1, Subparagraph 2 Article 5 (Procedure for Manufacturing Permit) ①Any person who intends to obtain a manufacturing permit for a medical device pursuant to Article 6, Paragraph 2 of the Act shall submit to the Minister of Food and Drug Safety an application form in Annex 3 (including an application form in electronic form) with the following documents (including electronic documents) attached. 1. Documents proving that facilities and manufacturing and quality control systems have been established pursuant to Article 6, Paragraph 4 of the Act 2. Technical documents and clinical trial data, etc. (hereinafter referred to as “technical documents, etc.”) or a notice of the results of the examination of technical documents, etc. pursuant to Article 9, Paragraph 4, for which two years have not elapsed since the date of issuance. However, in cases falling under the proviso to Article 9, Paragraph 1, a notice of the results of the examination of technical documents, etc. must be submitted. 3. A copy of the consignment contract in cases where testing for manufacturing processes or quality control has been outsourced ② If the Minister of Food and Drug Safety recognizes that a product for which a manufacturing permit has been applied pursuant to Paragraph 1 is compliant with the standards, he/she shall issue a manufacturing permit in the form of Annex 4 to the applicant and record the following matters in the manufacturing permit register. 1. Permit Number and Date of Permit 2. Name (Product name, Item name, and Model name) ③ Detailed matters concerning the guidelines for preparing attached materials required for an application for a license to manufacture medical devices pursuant to Article 3 (including conditional manufacturing licenses and licenses for changes thereto) and attached materials required for an application for a manufacturing license pursuant to Paragraph 1 (including conditional manufacturing licenses and licenses for changes thereto), the requirements for each material, and the scope of exemption, etc., shall be determined and announced by the Minister of Food and Drug Safety . ④ When the Minister of Food and Drug Safety receives an application for evaluation of a new medical technology and an application for confirmation of eligibility for medical benefits or non-reimbursement pursuant to Article 3-2, Paragraph 2 of the Rules on Evaluation of New Medical Technologies and Article 9-2, Paragraph 2 of the Rules on Standards for Medical Benefits under the National Health Insurance, along with an application for manufacturing license pursuant to Paragraph 1, or receives an objection pursuant to Article 9-2, Paragraph 4 of the Rules on Standards for Medical Benefits under the National Health Insurance, the Minister shall send the relevant applications, etc. to the Minister of Health and Welfare and the President of the Health Insurance Review and Assessment Service without delay. In this case, the Minister of Food and Drug Safety may mutually share data related to the relevant application for manufacturing license with the Minister of Health and Welfare or the President of the Health Insurance Review and Assessment Service and may provide opinions regarding the evaluation of the new medical technology and eligibility for medical benefits or non-reimbursement, and may refer to the data received pursuant to Paragraph 5 when granting the manufacturing license. <Newly added Jul. 29, 2016; Apr. 13, 2020> ⑤ In accordance with Article 4 , Paragraph 2 of the Rules on the Evaluation of New Medical Technologies and Article 9-2, Paragraph 3 ( main clause) and Paragraph 4 of the Rules on Standards for Medical Benefits under the National Health Insurance, if the Minister of Food and Drug Safety receives the results of the evaluation of new medical technologies and the results of the confirmation regarding eligibility for medical benefits or non-reimbursement, etc., from the Minister of Health and Welfare and the President of the Health Insurance Review and Assessment Service, the Minister shall immediately notify the applicant thereof. <Newly added July 29, 2016> ⑥ The matters prescribed in Paragraphs 4 and 5 shall also apply in cases where a conditional manufacturing permit is obtained pursuant to Article 7, Paragraph 1 of the Act or a permit for change is obtained pursuant to Article 12, Paragraph 1 of the Act . <Newly added July 29, 2016> ⑦ In addition to the matters prescribed in Paragraphs 4 through 6, necessary detailed matters concerning the provision of data and opinions, etc., between agencies shall be determined and announced by the Minister of Food and Drug Safety . <Newly added July 29, 2016> Article 6 (Procedure for Manufacturing Certification) ①Any person who intends to obtain manufacturing certification for a medical device pursuant to Article 6, Paragraph 2 of the Act shall submit an application in the form of Annex No. 5 (including an application in the form of an electronic document)attached with the documents specified in each subparagraph of Article 5, Paragraph 1 (including electronic documents) to the head of the Korea Medical Device Safety Information Institute (hereinafter referred to as the “Information Institute”) pursuant to Article 42 of the Act , who has been entrusted with certification duties pursuant to Article 44, Paragraph 2 of the Act . < Amended June 14, 2018> ② If the head of the Information Center recognizes that a product for which manufacturing certification has been applied pursuant to Paragraph 1 conforms to the standards, he/she shall issue a manufacturing certification in the form of Annex 6 to the applicant and record the matters of each of the following subparagraphs in the manufacturing certification register. <Amended June 14, 2018> 1. Verification Number and Verification Date 2. Name (Product name, Item name, and Model name) ③ Detailed matters concerning the guidelines for preparing attached materials required for an application for manufacturing certification under Paragraph 1 (including conditional manufacturing certification and certification for changes thereto), the requirements for each material, and the scope of exemption, etc. shall be determined and announced by the Minister of Food and Drug Safety . Article 7 (Procedure for Notification of Manufacturing) ①Any person who intends to file a notification of manufacturing of a medical device pursuant to Article 6, Paragraph 2 of the Act shall submit a notification form in Annex No. 7 (including a notification form in the form of an electronic document)attached with the documents specified in Article 5, Paragraph 1, Subparagraph 3, to the head of the information center entrusted with the duties concerning notification pursuant to Article 44, Paragraph 2 of the Act . <Amended June 14, 2018> ② When the head of the Information Center accepts a manufacturing report pursuant to Paragraph 1, he/she shall issue a manufacturing report certificate in accordance with Form No. 7-2 of the Attached Form to the report and record the matters of each of the following subparagraphs in the manufacturing report ledger. <Amended June 14, 2018; August 7, 2024> 1. Report Number and Date of Report 2. Name (Product name, Item name, and Model name) ③ Detailed matters concerning the guidelines for preparing attached materials required for a manufacturing notification (including a conditional manufacturing notification and a notification of change thereto) under Paragraph 1, the requirements for each material, and the scope of exemption, etc. shall be determined and announced by the Minister of Food and Drug Safety . Article 8 (Standards for Facilities and Manufacturing and Quality Control Systems) ① The standards for facilities and manufacturing and quality control systems that a person who intends to obtain a manufacturing license pursuant to the main text of Article 6, Paragraph 4 of the Act , and a person who intends to obtain a manufacturing license or manufacturing certification or file a manufacturing report, must meet are as set forth in Attached Table 2 . ② In the proviso to Article 6, Paragraph 4 of the Act , “cases prescribed by the Decree of the Prime Minister, such as entrusting testing for quality control or manufacturing processes” refers to cases where a person seeking a manufacturing license, or a person seeking a manufacturing license or manufacturing certification, or a person filing a manufacturing report, entrusts a manufacturing process or entrusts testing for quality control to a medical device testing and inspection institution pursuant to Article 6, Paragraph 2, Subparagraph 4 of the Act on Testing and Inspection, etc. in the Food and Drug Sectors (hereinafter referred to as “medical device testing and inspection institution”), a non-clinical testing institution pursuant to Article 10-2, Paragraph 1 of the Act (hereinafter referred to as “non-clinical testing institution”), or a manufacturer recognized as compliant pursuant to Subparagraph 2, Item (f) of Annex 2. In such cases, a person seeking a manufacturing license, or a person seeking a manufacturing license or manufacturing certification, or a person filing a manufacturing report may not be required to equip the facilities of a manufacturing plant pursuant to Subparagraph 1, Item (a) of Annex 2 , according to the classifications in each of the following subparagraphs. <Amended May 1, 2017> 1. In the case of outsourcing the manufacturing process: Manufacturing facilities and equipment related to the outsourced process 2. In cases where quality control testing is entrusted to a medical device testing and inspection institution, a non-clinical testing institution, or a manufacturer recognized as compliant pursuant to Subparagraph 2, Item (f) of Annex 2 : Testing laboratories for quality control, facilities and equipment related to testing Article 9 (Examination of Technical Documents, etc.) ① A person who intends to obtain a manufacturing license or manufacturing certification may undergo an examination by the Minister of Food and Drug Safety in advance regarding the conformity of the technical documents, etc. required to be submitted pursuant to Article 6, Paragraph 5 of the Act. However, in the case of items designated by the Minister of Food and Drug Safety as medical devices that are essentially equivalent in structure, principle, performance, purpose of use, and method of use, etc. , to medical devices that have already been licensed, certified, or reported, the examination must be conducted by a technical document examination agency pursuant to Article 6-4, Paragraph 1 of the Act (hereinafter referred to as the “technical document examination agency”). <Amended May 1, 2017> ② Any person who intends to undergo an examination of technical documents, etc. pursuant to Paragraph 1 for a medical device (excluding in vitro diagnostic medical devices) shall submit a request for examination in the form of Annex No. 8 (including a request for examination in electronic form) to the Minister of Food and Drug Safety or the head of the technical document examination agency, attaching the materials corresponding to each of the following subparagraphs (including materials in electronic form). However, in the case of medical devices that are essentially equivalent in structure, principle, performance, purpose of use, and method of use to medical devices that have already received approval or certification, excluding those medical devices for which the Minister of Food and Drug Safety has determined and notified that the submission of clinical trial data is necessary , the materials under subparagraphs 5 through 7 may not be submitted. <Amended June 12, 2019; December 31, 2020; June 24, 2021; January 16, 2024> 1. Data compared with products that have already been approved or certified 2. Data regarding the purpose of use 3. Data on the operating principle 4. Data comprising the following items for verifying the performance and safety of the product, including test standards, the basis for their establishment, and actual measured values. However, if there are no test standards domestically or abroad, the test standards, the basis for their establishment, and actual measured values ​​established by the person seeking examination of technical documents, etc., to verify the performance and safety of the product. a. Data on electrical and mechanical safety B. Data on biological safety C. Safety data regarding radiation D. Data on electromagnetic safety E. Data regarding performance B. Data on physical and chemical properties D. Data regarding stability 5. Data regarding origin or discovery and development history 6. Data regarding clinical trials 7. Data on usage status abroad, etc. ③ Any person who intends to undergo an examination of technical documents, etc. pursuant to Paragraph 1 for in vitro diagnostic medical devices shall submit a request for examination in the form of Annex 8 (including a request for examination in electronic form) to the Minister of Food and Drug Safety or the head of the technical document examination agency, attaching the materials corresponding to each of the following subparagraphs (including materials in electronic form). <Amended Jan. 4, 2017; Jun. 24, 2021; Jan. 16, 2024> 1. Data on development history, measurement principles and methods, and current usage status at home and abroad 2. Data on raw materials and manufacturing methods 3. Data regarding the purpose of use 4. Data regarding storage method and usage period or validity period 5. The following data to verify product performance A. Data on Analytical Performance Testing B. Data regarding clinical performance testing C. Data regarding quality control testing D. Data regarding standard materials and specimen storage, etc. 6. Data on Handler Safety for In Vitro Diagnostic Medical Devices 7. Data comparing with products that have already been approved or certified ④ Notwithstanding Paragraphs 2 and 3, if a person who has received a prior review of a manufacturing permit, notification, etc., pursuant to Article 11 of the Act submits a notice of prior review results pursuant to Article 25 , Paragraph 3, they may not submit data regarding matters for which a judgment of conformity was received as a result of the prior review among the data specified in each subparagraph of Paragraph 2 or Paragraph 3. <Newly added Jan. 16, 2024> ⑤ The Minister of Food and Drug Safety or the head of the technical document review agency who has received a request for review pursuant to Paragraphs 2 and 3 shall review the request and notify the applicant of the results in accordance with Form No. 9. <Amended Jan. 16, 2024> ⑥ Detailed matters concerning the guidelines for preparing attachments required for the examination of technical documents, etc., the requirements, exemption procedures, and scope of each attachment, the scope, criteria, and procedures of the examination, and the management standards for clinical performance tests, etc., shall be determined and announced by the Minister of Food and Drug Safety . <Amended Jan. 4, 2017; Jan. 16, 2024> Article 10 (Exemption from Medical Device Manufacturing Permit, etc.) Medical devices that may be manufactured without obtaining a manufacturing permit pursuant to Article 6, Paragraph 1 of the Act , or without obtaining or filing a manufacturing permit, manufacturing certification, or manufacturing notification pursuant to Paragraph 2 of the same Article , are as follows: <Amended June 15, 2016; June 14, 2018; December 31, 2018> 1. Medical devices subject to clinical trials for which a clinical trial plan has been approved pursuant to the main text of Article 10, Paragraph 1 of the Act 2. Separate medical devices necessary to conduct a clinical trial that does not require approval of the clinical trial plan pursuant to the proviso of Article 10, Paragraph 1 of the Act 3. Separate medical devices necessary to conduct clinical trials concerning pharmaceuticals, etc., for which approval of the clinical trial plan is required pursuant to Article 34 of the Pharmaceutical Affairs Act 4. Medical devices manufactured for the purpose of obtaining a medical device manufacturing license or certification, or filing a manufacturing report. 5. Medical devices used for research purposes, such as product development 6. Medical devices for personal use or relief purposes, etc., as designated and notified by the Minister of Food and Drug Safety 7. Deleted <December 31, 2018> 8. Medical devices used in clinical laboratories (referring to places where disease diagnosis, testing, etc. is performed; hereinafter the same) that the Minister of Food and Drug Safety recognizes as suitable after evaluating quality control systems, testing performance, proficiency, etc., and which are designated and notified by the Minister of Food and Drug Safety Article 11 (Qualifications of Quality Manager, etc.) ① Any person who intends to obtain a manufacturing license pursuant to Article 6, Paragraph 1 of the Act shall have at least one quality manager for each manufacturing facility. ② The duties of the Quality Manager may be performed by a person holding any of the qualifications set forth in the following subparagraphs. <Amended June 15, 2016; April 13, 2020; December 19, 2022> 1. Persons holding a license classified under each of the following items pursuant to the Act on Medical Technicians, etc. a. In the case of manufacturing or importing eyeglass lenses or contact lenses: Optician B. In the case of manufacturing or importing dental materials: Dental technicians and dental hygienists C. In the case of manufacturing or importing radiation generating devices: Radiologist D. Deleted <May 1, 2020> E. In the case of manufacturing or importing medical devices necessary for physical therapy or rehabilitation training: Physical therapist 2. Persons holding the Biomedical Engineer or Quality Management Engineer qualification under the National Technical Qualifications Act 2-2. A person holding the qualification of a medical device RA (Regulatory Affairs) expert recognized by the Minister of Food and Drug Safety pursuant to Article 19, Paragraph 1 of the Framework Act on Qualifications 3. A person who has obtained a bachelor's degree from a school pursuant to each subparagraph of Article 2 of the Higher Education Act (excluding junior colleges pursuant to subparagraph 4 of the same article ; hereinafter referred to as “universities, etc.” in this article) (including a person recognized by laws and regulations as having an academic qualification of an equivalent or higher level; hereinafter the same applies in this article) and who has majored in the fields of natural science, engineering, or medicine pursuant to Article 2, Paragraph 9 of the Regulations on the Establishment and Operation of Universities (hereinafter referred to as “medical device-related fields” in this article). 4. A person who has obtained a bachelor's degree from a university or similar institution in a field other than medical devices, and who has obtained a master's degree or higher in a medical device-related field from a graduate school pursuant to Article 29 of the Higher Education Act. 5. A person who has obtained a bachelor's degree from a university or similar institution in a field unrelated to medical devices and has at least one year of experience in quality control at a medical device manufacturing or importing company (including experience prior to obtaining the degree or academic recognition). 6. A person who has graduated from a junior college pursuant to Article 2, Subparagraph 4 of the Higher Education Act (including a person recognized as having an equivalent or higher level of education in accordance with relevant laws and regulations; hereinafter the same applies in this Article) who has majored in a medical device-related field and has at least one year of experience in quality control at a medical device manufacturing or importing company (including experience prior to graduation or recognition of educational qualifications). 7. A person who has graduated from a junior college pursuant to Article 2, Subparagraph 4 of the Higher Education Act, majored in a field other than medical device-related fields, and has at least 3 years of experience (including experience prior to graduation or recognition of academic qualifications) in quality control work at a medical device manufacturing or importing company ( 2 years in the case of a graduate of a junior college with a 3-year period of study pursuant to Article 48 , Paragraph 1 of the Higher Education Act). 8. A graduate of a high school or vocational high school pursuant to Article 2 , Subparagraph 3 of the Elementary and Secondary Education Act (including persons recognized as having an equivalent level of education or higher in accordance with relevant laws and regulations, but excluding cases falling under Subparagraph 9) who has at least 5 years of experience in quality control work at a medical device manufacturing or importing company. 9. A graduate of an industry-demand-tailored high school in a medical device-related field pursuant to Article 90, Paragraph 1, Subparagraph 10 of the Enforcement Decree of the Elementary and Secondary Education Act, who has at least 3 years of experience in quality control at a medical device manufacturing or importing company. 10. Persons with at least 6 years of experience in quality control at a medical device manufacturing or importing company. ③ In cases where a manufacturer appoints two or more quality managers pursuant to Paragraph 1, the duties of the quality managers shall be divided to clearly define the limits of responsibility for each quality manager. Article 12 (Scope of Duties of Quality Manager, etc.) ① The scope of duties to be performed by a quality manager pursuant to Article 6-2, Paragraph 1 of the Act shall be as follows: 1. The task of thoroughly inspecting the hygiene status of employees and providing them with the education and training necessary for the production and import of high-quality medical devices. 2. Duties of supervising whether employees receive education and training pursuant to Subparagraph 1 3. Tasks to prepare Standard Operating Procedures for manufacturing medical devices in accordance with the standards for the manufacturing and quality control system specified in Subparagraph 2 of Annex 2 , and to manufacture medical devices in accordance with the prepared Standard Operating Procedures. 4. The task of thoroughly conducting necessary tests, inspections, or verifications from the receipt of raw materials to the shipment of finished products, preparing and maintaining manufacturing control records and quality control records for each manufacturing unit, and preserving them for 5 years from the date of manufacture (or a period corresponding to the product lifespan if the product lifespan exceeds 5 years). 5. Task of evaluating the quality control results of the manufacturing plant and deciding whether to ship products. 6. Tasks related to establishing, implementing, and maintaining a quality management system in accordance with the standards for the manufacturing and quality control system set forth in Item 2 of Annex 2 7. Hygienically managing the facilities of the manufacturing plant to prevent health and hygiene hazards, and preventing cross-contamination or contamination from the outside. 8. Managing and supervising to prevent the placement of objects that may cause harm in the workplace, and preventing the generation of substances harmful to public health in the workplace. 9. Other duties related to manufacturing management, quality management, and safety management among the compliance requirements of manufacturers under Article 27, Paragraph 1, as determined and notified by the Minister of Food and Drug Safety ② If a quality manager ceases to perform the duties of a quality manager at the establishment concerned, they may submit a report of non-work of quality manager in the form of Annex 10 (including a report in the form of an electronic document) to the competent regional Food and Drug Administration. ③ The Quality Manager shall not hold any other duties in addition to the duties under Paragraph 1. However, this shall not apply in any of the following cases: 1. Cases where tasks are performed that do not affect the duties of the Quality Manager 2. Cases where a manufacturer also engages in importing, and the manufacturer's quality manager performs the duties of the importer's quality manager. Article 13 (Contents and Hours of Training for Quality Managers, etc.) ① The contents of the training pursuant to Article 6-2, Paragraph 2 of the Act shall be as follows: 1. Latest standard specifications for medical devices 2. Laws, systems, and technologies necessary for safety management, such as ensuring the safety and efficacy of medical devices 3. Laws, regulations, and technologies required for the manufacturing and quality control of medical devices 4. Other matters determined by the Minister of Food and Drug Safety for the manufacturing, quality, and safety management of medical devices ② The training time shall be at least 8 hours per year. ③ A Quality Manager shall receive training conducted by a Quality Manager training institution pursuant to Article 14, Paragraph 1, within six months from the date of commencement of employment . However, this shall not apply if the training corresponding to the relevant year has been completed prior to commencing employment as a Quality Manager. <Amended June 24, 2021> Article 14 (Designation of Quality Manager Training Institutions, etc.) ① Any person who intends to be designated as a training institution pursuant to Article 6-2, Paragraph 5 of the Act (hereinafter referred to as the “Training Institution”) shall meet the requirements of each of the following subparagraphs. <Amended Jan. 21, 2022> 1. It must be an information source or a medical device-related institution or organization announced by the Minister of Food and Drug Safety. 2. Secure professional personnel, including education managers to manage and supervise overall education operations and staff necessary for education operations. 3. Equip the training facilities and equipment necessary for conducting the training. ② Any person who intends to apply for designation as a training institution pursuant to Article 6-2, Paragraph 5 of the Act shall submit to the Minister of Food and Drug Safety the Application for Designation as a Medical Device Quality Manager Training Institution (Form No. 11) , attaching the materials specified in each of the following subparagraphs. <Amended Jan. 21, 2022> 1. Data on the operational organization and personnel status to conduct the training 2. Data on the current status of educational facilities and equipment 3. Supporting data for tuition fee calculation 4. Education Implementation Regulations 5. Basic Education Plan ③ Upon receiving an application under Paragraph 2, the Minister of Food and Drug Safety may conduct an on-site inspection to evaluate whether the contents of the application comply with the designation requirements under Paragraph 1. <Amended Jan. 21, 2022> ④ When the Minister of Food and Drug Safety designates an educational institution, he/she shall issue a certificate of designation in the form of Annex 12 to the applicant. <Amended June 14, 2018> ⑤ If a training institution intends to change the name, address, or representative designated pursuant to Paragraph 4, it shall submit an Application for Change of Designation as a Medical Device Quality Manager Training Institution (Form No. 11-2) to the Minister of Food and Drug Safety within 30 days from the date the cause for the change occurred, attaching the Certificate of Designation as a Medical Device Quality Manager Training Institution and documents verifying the changes. <Newly added Jun. 14, 2018; Jan. 21, 2022> ⑥ The Minister of Food and Drug Safety may change the designation of a training institution after verifying the application, etc. under Paragraph 5. In this case, the details of the change shall be noted in the designation certificate for the medical device quality manager training institution. <Newly added June 14, 2018> ⑦ In cases where the Minister of Food and Drug Safety designates an educational institution pursuant to Paragraph 4 or modifies its designation pursuant to Paragraph 6, the Minister shall publicize the fact of the designation or modification of the educational institution, as well as the name, address, and name of the representative of the educational institution, through the internet homepage of the Ministry of Food and Drug Safety, etc. <Newly added June 14, 2018> ⑧ The educational implementing institution may collect tuition fees from trainees, taking into consideration expenses necessary for the education, such as textbook fees, field practice fees, and instructor allowances. In this case, the tuition fee shall be determined by the head of the educational implementing institution at the level of actual costs. <Amended June 14, 2018> ⑨ In addition to the matters prescribed in Paragraphs 1 through 8, detailed matters regarding the requirements for designation of educational implementing institutions and matters necessary for designation management, etc., shall be determined and announced by the Minister of Food and Drug Safety . <Amended Jun. 14, 2018; Jan. 21, 2022> Article 15 (Matters to be Complied with by Educational Institutions, etc.) ① Pursuant to Article 6-2, Paragraph 6 of the Act , educational institutions shall comply with the following matters: 1. In accordance with the basic education plan under Article 14, Paragraph 2, Subparagraph 5, establish an education plan each year that includes the target audience and content of education for the following year, and submit it to the Minister of Food and Drug Safety for approval before the commencement of the following fiscal year. 2. Conduct education in accordance with the annual education plan under Item 1. 3. Produce educational materials and provide them to the trainees. 4. Issue a certificate of completion to those who have completed the training, record matters regarding the implementation of the training such as the list of completers, and retain the records for two years from the date of recording. 5. Report records regarding the implementation of training in the previous year to the Minister of Food and Drug Safety by January 31 of the following year. ② The Minister of Food and Drug Safety shall determine and announce the necessary detailed matters regarding the establishment of education plans and methods of conducting education as prescribed in each subparagraph of Paragraph 1 . [Wholly Amended Jan. 21, 2022] Article 15-2 (Designation of Technical Document Review Agency, etc.) ① Any person who intends to be designated as a technical document review agency pursuant to Article 6-4, Paragraph 1 of the Act shall meet the designation standards prescribed and notified by the Minister of Food and Drug Safety, such as the organization, personnel, and operational standards required for each field of review of technical documents, etc. ② Any person who intends to be designated as a technical document review agency shall submit to the Minister of Food and Drug Safety the Application for Designation or Renewal of Designation as a Medical Device Technical Document Review Agency (including applications submitted electronically) in the form of Annex 12-2, attaching the documents specified in each of the following subparagraphs (including electronic documents). <Amended July 1, 2026> 1. Documents regarding the organization and personnel status (in the case of personnel status, include documents proving qualifications and experience) 2. Documents regarding operating standards 3. Documents regarding the field of examination of technical documents, etc. 4. Documents proving the financial capacity to review technical documents, etc., as prescribed and notified by the Minister of Food and Drug Safety ③ Upon receiving an application pursuant to Paragraph 2 or 7, the Minister of Food and Drug Safety may conduct an on-site inspection to evaluate whether the contents of the application comply with the designation criteria. <Amended July 1, 2026> ④ When the Minister of Food and Drug Safety designates a technical document review agency, he/she shall issue a certificate of designation as a medical device technical document review agency in the form of Annex 12-3 to the applicant. ⑤ If a technical document review agency intends to change any of the matters falling under any of the following subparagraphs, it must submit an application for change of designation of a medical device technical document review agency (including an application in the form of electronic document) in Annex 12-4 to the Minister of Food and Drug Safety within 30 days from the date the cause for change occurred, attaching the designation certificate of the medical device technical document review agency and documents (including electronic documents) that can verify the changes. 1. Representative of the technical document review agency 2. Name and location of the technical document review agency 3. Fields of Examination for Technical Documents, etc. ⑥ When the Minister of Food and Drug Safety changes the designation of a technical document review agency, he/she shall issue a designation certificate for a medical device technical document review agency stating the changes. ⑦ If a technical document review agency intends to renew its designation pursuant to Article 6-4, Paragraph 4 of the Act , it shall submit to the Minister of Food and Drug Safety , at least 180 days prior to the expiration of the validity period, an application for designation or renewal of designation as a medical device technical document review agency in the form of Annex 12-2 (including an application in electronic form), attached with the documents specified in each subparagraph of Paragraph 2 (including electronic documents). <Newly added July 1, 2026> ⑧ If the Minister of Food and Drug Safety recognizes that a technical document review agency that has applied for renewal pursuant to Paragraph 7 is suitable for performing the duties of reviewing the conformity of technical documents, etc., he/she shall issue a new Certificate of Designation as a Medical Device Technical Document Review Agency in the form of Annex 12-3 no later than 30 days before the expiration of the validity period. <Newly added July 1, 2026> ⑨ When the Minister of Food and Drug Safety designates, modifies, or renews the designation of a technical document review agency, he/she shall publicize the name, address, representative, validity period, etc., of the technical document review agency on the internet homepage of the Ministry of Food and Drug Safety, etc. <Amended July 1, 2026> 10. Detailed matters concerning the scope of documents to be submitted pursuant to Paragraph 2 shall be determined and announced by the Minister of Food and Drug Safety . <Amended July 1, 2026> [This Article Newly Inserted May 1, 2017] Article 15-3 (Compliance Requirements for Technical Document Review Agencies) The term “matters prescribed by Decree of the Prime Minister” in Article 6-4, Paragraph 5 of the Act refers to the matters in each of the following subparagraphs. <Amended July 1, 2026> 1. Comply with the operating standards for technical document review institutions prescribed and notified by the Minister of Food and Drug Safety. 2. Report the results of the technical document review to the Minister of Food and Drug Safety in accordance with the regulations prescribed and notified by the Minister of Food and Drug Safety. 3. Retain a copy of the Notice of Technical Document Examination Results and the records regarding the technical document examination for 5 years from the date of notification of the technical document examination results. 4. Comply with other matters prescribed and notified by the Minister of Food and Drug Safety regarding the preparation and issuance of the Notice of Technical Document Examination Results, etc. [This Article Newly Inserted May 1, 2017] Article 16 (Application for Conditional Permit, etc.) ①Any person who intends to obtain a conditional manufacturing permit pursuant to Article 7, Paragraph 1 of the Act shall submit an application in the form of Annex No. 1, attached with the documents set forth in each of the following subparagraphs, to the head of the Regional Food and Drug Administration having jurisdiction over the location of the manufacturing facility. In this case, the head of the Regional Food and Drug Administration shall verify the Certificate of Corporate Registration (applicable only to corporations), the Certificate of Land Registration, and the Certificate of Building Registration through the joint use of administrative information pursuant to Article 36 , Paragraph 1 of the Electronic Government Act. <Amended Jan. 21, 2022> 1. In the case of installing manufacturing facilities by constructing a new building, documents confirming land ownership (if confirmed by a Certificate of Registered Land Matters, such confirmation shall substitute for the attached documents) or a copy of the lease agreement 2. In the case of installing manufacturing facilities in an existing building, documents verifying ownership of the building (if verification is possible through a Certificate of Registered Matters, such verification shall substitute for the attached documents) or a copy of the lease agreement 3. Documents under each subparagraph of Article 3, Paragraph 1 ② Any person who intends to obtain a conditional manufacturing license pursuant to Article 7, Paragraph 1 of the Act shall submit to the Minister of Food and Drug Safety an application form in accordance with Form No. 3 (including an application form in electronic form), attached with the documents referred to in Article 5 , Paragraph 1 , Subparagraph 2 (including electronic documents). <Amended Aug. 7, 2024> ③ Any person who intends to obtain conditional manufacturing certification pursuant to Article 7, Paragraph 1 of the Act shall submit an application in the form of Annex 5 (including an application in the form of an electronic document) to the head of the Information Center, attaching the documents referred to in Article 5 , Paragraph 1, Subparagraph 2 (including electronic documents). <Amended June 14, 2018; August 7, 2024> ④ Any person who intends to file a conditional manufacturing report pursuant to Article 7 , Paragraph 1 of the Act shall submit a report in the form of Annex 7 (including a report in the form of an electronic document) to the head of the Information Center. <Amended June 14, 2018; Aug. 7, 2024> ⑤ When the Minister of Food and Drug Safety, the head of a regional Food and Drug Safety Administration, or the head of an Information Center grants a conditional manufacturing license, conditional manufacturing license, or manufacturing certification, or receives a conditional manufacturing notification, they shall record the matters prescribed in Article 3 , Paragraph 3 , Article 5 , Paragraph 2 , Article 6 , Paragraph 2 , or Article 7, Paragraph 2 , and the conditions of the license, certification, or notification, in the Manufacturing License Register, Manufacturing License Register , Manufacturing Certification Register, or Manufacturing Notification Register, and issue a Conditional Manufacturing License in Form No. 13 , a Conditional Manufacturing License in Form No. 14 , a Conditional Manufacturing Certification in Form No. 15 , or a Conditional Manufacturing Notification in Form No. 15-2 . <Amended June 14, 2018; August 7, 2024> Article 17 (Notification and Verification of Fulfillment of Conditions) ① When a person who has obtained conditional approval or certification or filed a report pursuant to Article 16 has fulfilled the conditions of such approval, certification, or report, such person shall submit documents proving such fact to the Minister of Food and Drug Safety, the head of the regional Food and Drug Safety Administration having jurisdiction over the location of the manufacturing facility, or the head of the Information Center. <Amended June 14, 2018> ② Upon receiving the documents pursuant to Paragraph 1, the Minister of Food and Drug Safety, the head of a regional Food and Drug Safety Administration, or the head of the Information Center shall verify whether the conditions have been fulfilled within 20 days from the date of submission, and then replace and reissue the conditional manufacturing license, conditional manufacturing permit, conditional manufacturing certificate, and conditional manufacturing notification certificate pursuant to Article 16 , Paragraph 5 , with the manufacturing license pursuant to Article 3, Paragraph 3 ; the manufacturing permit pursuant to Article 5, Paragraph 2 ; the manufacturing certificate pursuant to Article 6, Paragraph 2; and the manufacturing notification certificate pursuant to Article 7, Paragraph 2, respectively, and record the date of fulfillment of the conditions in the notification ledger. <Amended June 14, 2018; August 7, 2024> Article 18 (Subjects of Exemption from Post-Marketing Surveillance, etc.) ① Where the Minister of Food and Drug Safety intendsto require the conduct of an investigation into safety and efficacy (hereinafter referred to as “post-marketing surveillance”) when granting a manufacturing license pursuant to Article 8, Paragraph 1 of the Act , the Minister shall indicate the period of post-marketing surveillance on the manufacturing license. <Amended Jan. 21, 2022> ② Pursuant to the proviso to the part other than the subparagraphs of Article 8, Paragraph 1 of the Act , the Minister of Food and Drug Safety may exempt any of the medical devices listed in the following subparagraphs from post-market surveillance following deliberation by the Medical Device Committee under Article 5 , Paragraph 1 of the Act (hereinafter referred to as the “Medical Device Committee”). <Amended Jan. 21, 2022; July 1, 2026> 1. A newly developed medical device under Article 8, Paragraph 1 , Subparagraph 1 of the Act corresponding to Class 1 or Class 2 (hereinafter referred to as “newly developed medical device” in this paragraph) 2. Medical devices that have received authorization, etc. in a foreign country recognized by the Minister of Food and Drug Safety as implementing medical device safety management at a level equivalent to or higher than that of Korea, and which have experience of use abroad in accordance with the standards notified by the Minister of Food and Drug Safety, as specified in each of the following subparagraphs A. Newly developed medical devices B. Rare medical devices pursuant to Article 8, Paragraph 1, Subparagraph 2 of the Act ③ Deleted <2022. 1. 21.> ④ Deleted <2022. 1. 21.> ⑤ Deleted <2022. 1. 21.> [Title amended Jan. 21, 2022] Article 18-2 (Approval of Post-Marketing Surveillance Plan, etc.) ①person who intends to obtain approval for a post-marketing surveillance plan pursuant to the first part of Article 8, Paragraph 3 of the Act shall submit to the Minister of Food and Drug Safety , at least 30 days prior to the market launch of the medical device, an application for approval of a medical device post-marketing surveillance plan in the form of Annex 16, with a post-marketing surveillance plan attached that includes information on the purpose, period, method, number of subjects, and implementing agency of the post-marketing surveillance. ② Any person seeking approval for a change to a post-marketing surveillance plan pursuant to the latter part of Article 8, Paragraph 3 of the Act shall submit an application for approval of a change to a medical device post-marketing surveillance plan (Form No. 16) to the Minister of Food and Drug Safety, attaching the documents specified in each of the following subparagraphs: a. A plan for a change to the post-marketing surveillance plan b. Data verifying the changes ③ When the Minister of Food and Drug Safety receives an application for approval or modification of a post-marketing surveillance plan pursuant to Paragraph 1 or 2, he or she shall decide whether to approve or modify the plan within 30 days and notify the applicant of the decision. ④ When the Minister of Food and Drug Safety approves a post-marketing surveillance plan pursuant to Paragraph 3, he/she shall issue a medical device post-marketing surveillance plan approval certificate in the form of Annex 16-2 to the applicant , and when approving a change to the post-marketing surveillance plan, he/she shall record the changes in the section for changes and dispositions, etc. of the previously issued approval certificate. ⑤ In addition to the matters specified in paragraphs 1 through 4, detailed matters concerning the guidelines for preparing materials to be submitted when applying for approval or modification of a post-marketing surveillance plan shall be determined and announced by the Minister of Food and Drug Safety . [This Article Newly Inserted Jan. 21, 2022] Article 18-3 (Method and Procedure, etc. for Periodic Reporting of Post-Marketing Surveillance) ①Any person who intends to report data collected and implementation status, etc., in post-marketing surveillance pursuant to Article 8, Paragraph 4 of the Act shall report to the Minister of Food and Drug Safety within the deadline set forth in Paragraph 2, attaching the documents in each of the following subparagraphs to the periodic report on post-marketing surveillance of medical devices in Form No. 16-3 of the Appendix . 1. Basic data, such as observation records of subjects in post-marketing surveillance 2. Data on the analysis and evaluation results of post-marketing surveillance 3. Data on adverse effects of medical devices in post-marketing surveillance ② Reports under Paragraph 1 shall be made every six months for the first year from the date of approval of the post-market surveillance plan for the medical device, and every year thereafter, provided that the reporting deadline is two months from the end of each reporting period. ③ Notwithstanding Paragraph 2, the last report during the post-marketing surveillance period may be substituted with the data submitted when applying for a review under Paragraph 1 of Article 8-2 of the Act . ④ Detailed matters concerning the method of preparation, etc. of materials to be submitted when making a report pursuant to Paragraphs 1 and 3 shall be determined and announced by the Minister of Food and Drug Safety . [This Article Newly Inserted Jan. 21, 2022] Article 18-4 (Submission of results from post-market surveillance) ① The term “data prescribed by the Decree of the Prime Minister” in Article 8-2, Paragraph 1 of the Act refers to data regarding the status of domestic and foreign sales and the status of approval in foreign countries. ② Any person who wishes to receive a review pursuant to Article 8-2, Paragraph 1 of the Act shall submit to the Minister of Food and Drug Safety an application for review of post-market surveillance results of medical devices in the form of Annex 16-4, attaching the following materials. 1. Data on usage performance based on post-marketing surveillance results, specifically surveillance data on safety and efficacy after domestic market launch. 2. Domestic and international data on adverse effects and safety as cases of adverse effects of medical devices subject to post-marketing surveillance 3. Data on domestic and international sales status and foreign licensing status ③ The Minister of Food and Drug Safety shall notify the applicant of the results of the review, following deliberation by the Medical Devices Committee, within six months from the date of receiving the application for review pursuant to Paragraph 2, in accordance with the notification form in Annex 17 . ④ Detailed matters concerning the guidelines for preparing materials submitted pursuant to Paragraph 2 shall be determined and announced by the Minister of Food and Drug Safety . [This Article Newly Inserted Jan. 21, 2022] Article 18-5 (Measures for recall, disposal, etc. based on results of post-market surveillance) ① When the Minister of Food and Drug Safety intends to order necessary measures such as suspension of sales, recall, or disposal pursuant to Article 8-2, Paragraph 2 of the Act , he/she shall notify the manufacturer of a medical device subject to post-market surveillance of the matters in each of the following subparagraphs. 1. Name of the medical device 2. Details and Reasons for the Measure ② A manufacturer of a medical device subject to post-market surveillance who has received a notice pursuant to Paragraph 1 shall take the appropriate measures within 30 days from the date of receiving the notice and notify the Minister of Food and Drug Safety of the results. [This Article Newly Inserted Jan. 21, 2022] Article 19 (Method and Procedure for Re-evaluation, etc.) ① When the Minister of Food and Drug Safety intends to conduct a re-evaluation of a medical device pursuant to Article 9 of the Act , he/she shall determine the items subject to re-evaluation after deliberation by the Medical Device Committee and then announce the matters in each of the following subparagraphs on the internet homepage of the Ministry of Food and Drug Safety. 1. Items Subject to Re-evaluation 2. Re-evaluation Application Period 3. Contents of the materials required for re-evaluation ② Manufacturers of medical devices subject to re-evaluation shall submit to the Minister of Food and Drug Safety, during the re-evaluation application period under Paragraph 1, Subparagraph 2, an application for medical device re-evaluation in Form No. 18 (including an application in electronic form), with the data under Paragraph 1, Subparagraph 3 (including data in electronic form) attached. ③ The Minister of Food and Drug Safety who has received an application for re-evaluation pursuant to Paragraph 2 shall re-evaluate the relevant item and, after making the draft of the re-evaluation results available for viewing for at least one month so that the manufacturer of the item subject to re-evaluation can submit an opinion on the results, shall announce the re-evaluation results on the internet homepage of the Ministry of Food and Drug Safety. ④ Detailed matters concerning the guidelines for preparing submitted materials pursuant to Paragraphs 2 and 3, the requirements and scope of exemption for each material, and the scope of re-evaluation and implementation standards, etc. shall be determined and announced by the Minister of Food and Drug Safety . Article 20 (Approval of Clinical Trial Plan, etc.) ①Any person seeking approval of a clinical trial plan pursuant to Article 10 , Paragraph 1 of the Act shall submit to the Minister of Food and Drug Safety an Application for Approval of a Clinical Trial Plan (including an application in the form of an electronic document) in Annex No. 19 , and a person seeking to amend an approved clinical trial plan shall submit to the Minister an Application for Approval of Amendment to a Clinical Trial Plan (including an application in the form of an electronic document) in Annex No. 20, attaching the Certificate of Approval of the Clinical Trial Plan and the documents (including electronic documents) and data (including data in the form of electronic documents) specified in each of the following subparagraphs. However, in the case of applying for approval of an amendment to a clinical trial plan,the data specified in subparagraphs 2 and 3 may not be submitted in accordance with the provisions prescribed and notified by the Minister of Food and Drug Safety . <Amended Oct. 22, 2019; May 1, 2020> 1. Clinical trial protocol or clinical trial amendment protocol 2. Data proving that the medical device for clinical trials is being manufactured in accordance with the standards for facilities, manufacturing, and quality control systems set forth in Annex 2 3. Data under Article 9, Paragraph 2 , Subparagraphs 2 through 5 ② The matters to be included in the clinical trial protocol pursuant to Paragraph 1, Subparagraph 1 are as follows: <Amended Jan. 16, 2024; Aug. 7, 2024> 1. Title of the clinical trial 2. Name and address of the clinical trial institution [in cases where there is an institution other than the clinical trial institution conducting the clinical trial pursuant to the proviso of Article 10, Paragraph 4, Subparagraph 1 of the Act (hereinafter referred to as the “clinical trial participating institution”), the reason thereof, the name and address of said institution, and matters concerning the management and supervision of said institution] 3. Names and titles of the Principal Investigator, Person in Charge, and Co-investigators of the Clinical Trial 4. Name and title of the manager responsible for medical devices for clinical trials 5. Name and address of the person intending to conduct the clinical trial 6. Purpose and Background of the Clinical Trial 7. Overview of the medical device for investigational study (including purpose of use, target disease, or indication) 8. Selection criteria, exclusion criteria, number of subjects, and grounds for clinical trial participants (hereinafter referred to as “subjects”) 9. Clinical trial period 10. Clinical trial methods (including dosage, method of use, duration of use, combination therapy, etc.) 11. Observation Items, Clinical Examination Items, and Observation Examination Methods 12. Predicted Side Effects and Precautions for Use 13. Criteria for Suspension and Disqualification 14. Evaluation criteria, evaluation methods, and interpretation methods for validity (based on statistical analysis methods) 15. Evaluation Criteria, Evaluation Methods, and Reporting Methods for Safety, Including Adverse Effects 16. Subject Consent Form 17. Regulations on Victim Compensation 18. Matters concerning the treatment of subjects after the clinical trial 19. Measures for the Safety Protection of Subjects 20. Other matters necessary to conduct clinical trials safely and scientifically ③ Notwithstanding Paragraph 1, if a person who has received a prior review of a manufacturing permit, notification, etc., pursuant to Article 11 of the Act submits a notice of prior review results pursuant to Article 25, Paragraph 3, they may not submit data regarding matters for which a judgment of conformity was received as a result of the prior review among the data specified in each subparagraph of Paragraph 1. <Newly added Jan. 16, 2024> ④ Cases falling under any of the following subparagraphs shall be excluded from the subjects of approval by the Minister of Food and Drug Safety pursuant to the proviso to Article 10, Paragraph 1 of the Act . <Amended Jan. 16, 2024; Aug. 7, 2024> 1. Clinical trials using a commercially available medical device falling under any of the following subparagraphs a. Clinical trials conducted to observe clinical efficacy and investigate adverse events regarding the approved specifications of medical devices currently on the market B. Clinical trials aimed at collecting safety and efficacy data regarding the approved performance and intended use of medical devices currently on the market C. Other clinical trials using commercially available medical devices that are not directly related to safety or are determined by the Minister of Food and Drug Safety to have no risk of ethical issues. 2. Clinical trials falling under any of the following items that pose a low risk of harm to clinical trial subjects a. Clinical trials conducted in accordance with the clinical trial methods prescribed in the standards and specifications under Article 19 of the Act B. Clinical trials for medical devices that measure and display biosignals, etc., outside the body or on the body surface C. Clinical trials conducted independently by a researcher without a sponsor, which the Minister of Food and Drug Safety recognizes as having a low risk of harm to subjects. ⑤ If the Minister of Food and Drug Safety reviews the application for approval of a clinical trial plan submitted pursuant to Paragraph 1 and deems it appropriate, he/she shall issue a Certificate of Approval for a Clinical Trial Plan in accordance with Form No. 21 to the applicant; and in the case of approving a change to the clinical trial plan, he/she shall record the changes in the section for changes and dispositions of the Certificate of Approval. <Amended Jan. 16, 2024> ⑥ Detailed matters concerning the guidelines for preparing materials to be submitted, the scope of materials exempted, and approval requirements, standards, and procedures, etc., when applying for approval or amendment approval of a clinical trial plan pursuant to the provisions of Paragraphs 1 through 5 shall be determined and announced by the Minister of Food and Drug Safety . <Amended Jan. 16, 2024> Article 21 (Criteria and Procedures for Designation of Clinical Trial Institutions, etc.) ① A clinical trial institution pursuant to Article 10, Paragraph 3 of the Act (hereinafter referred to as “clinical trial institution”) shall be a medical institution falling under any one of the following subparagraphs. <Amended Apr. 13, 2020> 1. A general hospital pursuant to Article 3, Paragraph 2, Subparagraph 3, Item ( e) of the Medical Service Act 2. Training hospitals, training dental hospitals, and training Korean medicine hospitals pursuant to the "Regulations on Training and Qualification Recognition, etc. of Specialists , " the "Regulations on Training and Qualification Recognition, etc. of Dental Specialists," and the "Regulations on Training and Qualification Recognition , etc. of Korean Medicine Specialists." 3. Specialized hospitals designated pursuant to Article 3-5 of the Medical Service Act 4. A hospital equipped with personnel and facilities, etc., that meet the designation criteria for training hospitals, etc. and training dental hospitals for training in a single specialty pursuant to Article 4, Paragraph 2, Subparagraph 1, Item (a) of the Enforcement Decree of the Act on the Improvement of Training Environments and Status of Residents and Article 7, Paragraph 2, Subparagraph 2 of the Regulations on Training and Recognition of Qualifications , etc. of Dental Specialists. 5. Clinical trial center within an advanced medical complex designated pursuant to the "Special Act on the Designation and Support of Advanced Medical Complexes" ② A person who intends to be designated as a clinical trial institution shall submit to the Minister of Food and Drug Safety an application for designation as a medical device clinical trial institution in the form of Annex 22 (including an application in the form of an electronic document) the documents specified in each of the following subparagraphs (including electronic documents). 1. Deleted <December 31, 2018> 2. Regulations concerning administrative procedures, etc. required for clinical trials 3. Regulations on the operation of the Clinical Trial Review Committee pursuant to Item 6 of Annex 3 and documents concerning its composition (provided, however, that if the review is to be entrusted to an external designated review committee pursuant to Subparagraph (e) of the same item, a consignment contract must be attached) 4. In cases where a designated review committee is to be operated pursuant to Item 6, Subparagraph (e) of Annex 3 for the purpose of reviewing clinical trials entrusted to and conducted by other clinical trial institutions, the regulations concerning entrusted review 5. Documents proving that it falls under any of the subparagraphs of Paragraph 1 ③ The public official in charge who receives an application pursuant to Paragraph 2 shall verify the Certificate of Corporate Registration (applicable only to corporations) and the Permit for the Establishment of a Medical Institution through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act. However, if the applicant does not consent to the verification of the Permit for the Establishment of a Medical Institution, a copy thereof shall be attached. <Amended Dec. 31, 2018> ④ When the Minister of Food and Drug Safety receives an application pursuant to Paragraph 2, he/she may conduct an investigation to evaluate whether the contents of the application meet the designation criteria. ⑤ When the Minister of Food and Drug Safety designates a clinical trial institution, he/she shall issue a Medical Device Clinical Trial Institution Designation Certificate in the form of Annex 23 to the applicant. ⑥ If a clinical trial institution intends to change its designated matters, it shall submit an Application for Change of Designation as a Medical Device Clinical Trial Institution ( Form No. 24) to the Minister of Food and Drug Safety within 30 days from the date the cause for the change occurred, attaching the Certificate of Designation as a Medical Device Clinical Trial Institution and documents verifying the changes. However, if the cause of the change is a change in the head or founder of the medical institution, or a change in location due to administrative district reorganization, only documents verifying the changes may be submitted. <Amended July 8, 2024> ⑦ When the Minister of Food and Drug Safety changes the designation of a clinical trial institution, he/she shall record the details of the change in the medical device clinical trial institution designation certificate. ⑧ When the Minister of Food and Drug Safety designates or changes the designation of a clinical trial institution pursuant to Paragraph 5 or 7, he/she shall announce the name, location, representative, etc. of the clinical trial institution on the internet homepage, etc. ⑨ Detailed matters necessary for the designation, operation, and management of clinical trial institutions shall be determined and announced by the Minister of Food and Drug Safety . Article 21-2 (Clinical Trials by Institutions Other than Clinical Trial Institutions) The term "clinical trials requiring the participation of institutions other than clinical trial institutions due to the nature of the clinical trial" pursuant to the proviso of Article 10, Paragraph 4, Subparagraph 1 of the Act refers to any of the following clinical trials: 1. Clinical trials involving medical devices that are permanently installed in buildings or facilities and cannot be moved 2. Clinical trials that may recruit subjects only from institutions falling under any of the following subparagraphs a. Infectious disease management institutions designated pursuant to Article 36, Paragraphs 1 and 2 of the Act on the Prevention and Management of Infectious Diseases B. Medical institutions or isolation facilities, sanatoriums, or clinics designated as or established and operated as infectious disease management institutions pursuant to each subparagraph of Article 37, Paragraph 1 of the Act on the Prevention and Management of Infectious Diseases [This Article Newly Inserted Aug. 7, 2024] Article 22 (Group Facilities) In the main text of Article 10, Paragraph 4, Subparagraph 2 of the Act , the term “group facilities prescribed by Decree of the Prime Minister, such as social welfare facilities” refers to the facilities listed in each of the following subparagraphs. <Amended June 12, 2019> 1. Child welfare facilities pursuant to Article 52 , Paragraph 1 , Subparagraphs 1 , 2 , 3 , and 5 of the Child Welfare Act 2. Residential facilities for persons with disabilities pursuant to Article 58, Paragraph 1, Subparagraph 1 of the Welfare Act for Persons with Disabilities 3. Mental health promotion facilities pursuant to Article 3, Subparagraph 4 of the Act on Promotion of Mental Health and Support for Welfare Services for Persons with Mental Illness (in the case of psychiatric medical institutions, only those equipped with a protection room) 4. Elderly residential welfare facilities (excluding elderly welfare housing) and elderly medical welfare facilities (excluding specialized elderly hospitals) established pursuant to Article 34 of the Social Welfare Services Act 5. Elderly welfare facilities pursuant to Article 32, Paragraph 1 , Subparagraphs 1 and 2 and Article 34, Paragraph 1 , Subparagraphs 1 and 2 of the Elderly Welfare Act 6. Welfare facilities for mother-child families (basic living support, independent living support), welfare facilities for father-child families (basic living support, independent living support) , welfare facilities for unmarried mother-child families (basic living support), and temporary support welfare facilities pursuant to Article 19 of the Single-Parent Family Support Act 7. Support facilities for victims of prostitution, etc. pursuant to Article 9, Paragraph 1 , Subparagraphs 1 and 2 of the Act on the Prevention of Prostitution and Protection of Victims, etc. 8. Sexual violence victim protection facility pursuant to Article 12 of the Act on the Prevention of Sexual Violence and Protection of Victims, etc. 9. Domestic violence victim protection facility pursuant to Article 7 of the Act on Prevention of Domestic Violence and Protection of Victims, etc. 10. Facilities established by a person who has obtained permission for rehabilitation protection business pursuant to Article 67 of the Act on Probation, etc. (only those equipped with detention facilities) 11. Correctional facilities pursuant to the Act on the Execution of Sentences and Treatment of Prisoners and the Act on the Execution of Sentences and Treatment of Military Prisoners in the Military 12. Juvenile detention centers and juvenile classification and examination centers pursuant to the Act on the Treatment of Juveniles under Protection, etc. 13. Detention facilities pursuant to Article 52, Paragraph 2 of the Immigration Control Act Article 23 Deleted <June 15, 2016> Article 24 (Standards for Conducting Clinical Trials, etc.) ① Pursuant to Article 10, Paragraphs 5 and 7 of the Act , clinical trialsshall be conducted in accordance with the standards set forth in each of the following subparagraphs and the standards for the management of clinical trials for medical devices set forth in Annex 3. <Amended Jan. 16, 2024; Aug. 7, 2024> 1. Conduct in a safe and scientific manner in accordance with the clinical trial protocol. 2. It shall be conducted at a clinical trial institution designated by the Minister of Food and Drug Safety or at a clinical trial participating institution pursuant to Article 20, Paragraph 2, Subparagraph 2. If the clinical trial is conducted at a clinical trial participating institution, the clinical trial institution shall manage and supervise said institution. 3. The Principal Investigator shall be selected from among those who possess professional knowledge and ethical competence and have sufficient experience to conduct clinical trials of the relevant medical device. 4. Explain to the subject the details of the clinical trial and the details and procedures for compensation for any health damage that may occur to the subject during the clinical trial, and obtain a consent form ( including non-face-to-face consent through an electronic document containing an electronic signature in accordance with the Electronic Signature Act; hereinafter the same applies in this Article). However, if consent cannot be obtained due to reasons such as the subject's lack of understanding or ability to express their will, consent must be obtained from a parent or guardian, etc. 5. Implement safety measures for the subject. 6. Medical devices for clinical trials shall not be used for purposes other than clinical trials. However, this shall not apply in cases determined by the Minister of Food and Drug Safety, such as when used on patients with life-threatening serious diseases, such as terminal cancer . 7. The clinical trial shall commence within two years from the date of approval or approval of the amendment of the clinical trial plan. 8. Provide the clinical trial data booklet to the investigator prior to the clinical trial in accordance with the regulations set by the Minister of Food and Drug Safety. 9. Upon obtaining new data or information related to safety and efficacy, notify the investigator without delay and review whether to incorporate it. 10. Medical devices for clinical trials shall be manufactured in accordance with the standards for facilities, manufacturing, and quality control systems set forth in Annex 2. 11. A person who has received approval for a clinical trial plan shall report any adverse events occurring during the use of the medical device to the Minister of Food and Drug Safety in accordance with the Medical Device Clinical Trial Management Standards in Annex 3. 12. In the case of a clinical trial under Article 20, Paragraph 4, Subparagraph 2 that has not been approved pursuant to the proviso of Article 10, Paragraph 1 of the Act , the performance and safety of the product shall be verified through the data specified in each item of Article 9, Paragraph 2, Subparagraph 4 before conducting the clinical trial. 13. Comply with other matters determined by the Minister of Food and Drug Safety for the proper conduct of clinical trials. ② A person who has received approval for a clinical trial plan shall submit a report on the status of the clinical trial implementation in the form of Annex No. 25 (including reports in electronic form) to the Minister of Food and Drug Safety by the end of February of each year, attaching summary data related to the safety evaluation of medical devices for investigational purposes pursuant to Item 8, Subparagraph D, Item 1) of Annex 3; and upon completion of the clinical trial, shall submit a report in the form of Annex No. 26 (including reports in electronic form) to the Minister of Food and Drug Safety within 20 days from the date of completion . < Amended Oct. 22, 2019> ③ The clinical trial sponsor and the head of the clinical trial institution shall preserve the clinical trial protocol and records and data (including electronic documents) regarding the conduct of the clinical trial for the period specified in each of the following subparagraphs. 1. Data related to clinical trials for manufacturing, import, or amendment of licenses: 3 years from the date of approval 2. Other clinical trial-related data: 3 years from the date the clinical trial was completed ④ Notwithstanding the main text of Paragraph 1, Subparagraph 4, in the case of a clinical trial using data such as a patient's medical records where personal information regarding the patient has been processed as pseudonymized information pursuant to Article 28-2 of the Personal Information Protection Act , the clinical trial may be conducted without the subject's consent upon obtaining approval from the Institutional Review Board pursuant to Subparagraph 6 of Annex 3. In this case, matters concerning the processing of pseudonymized information, etc., shall be in accordance with the Personal Information Protection Act . <Amended Jan. 16, 2024> ⑤ Detailed matters concerning the designation criteria, etc. of clinical trial institutions pursuant to Paragraph 1 shall be determined and announced by the Minister of Food and Drug Safety . Article 24-2 (Designation of Non-Clinical Testing Institutions, etc.) ① Any person who intends to be designated as a non-clinical testing institution pursuant to Article 10-2, Paragraph 1 of the Act shall meet the designation standards prescribed and notified by the Minister of Food and Drug Safety, such as the specialized personnel, facilities, and equipment required for each non-clinical testing field ② Any person who intends to be designated as a non-clinical testing institution shall submit to the Minister of Food and Drug Safety an application for designation as a medical device non-clinical testing institution in the form of Annex 26-2 (including an application in electronic form) the documents listed in each of the following subparagraphs (including electronic documents). 1. Documents regarding the organization and personnel status (in the case of personnel status, include documents proving qualifications and experience) 2. Documents regarding the status of equipment, tools, and facilities 3. Documents proving that non-clinical trials have been conducted in accordance with the regulations prescribed and notified by the Minister of Food and Drug Safety to prove the capability to conduct non-clinical trials ③ Upon receiving an application pursuant to Paragraph 2, the Minister of Food and Drug Safety may conduct an on-site inspection to evaluate whether the contents of the application comply with the designation criteria. ④ When the Minister of Food and Drug Safety designates a non-clinical testing institution, he/she shall issue a certificate of designation for a medical device non-clinical testing institution in the form of Annex 26-3 to the applicant. ⑤ If a non-clinical testing institution intends to change any of the matters falling under any of the following subparagraphs, it must submit an application for change of designation details for a medical device non-clinical testing institution in the form of Annex 26-4 (including an application in the form of an electronic document) to the Minister of Food and Drug Safety within 30 days from the date the cause for change occurred, attaching the designation certificate for the medical device non-clinical testing institution and documents (including electronic documents) that can verify the changes. 1. Representative and Chief Operating Officer of the non-clinical testing institution 2. Name and location of the non-clinical testing institution 3. Non-clinical testing field ⑥ When the Minister of Food and Drug Safety changes the designation of a non-clinical testing institution, he/she shall issue a designation certificate for a medical device non-clinical testing institution stating the changes. ⑦ When the Minister of Food and Drug Safety designates or changes the designation of a non-clinical testing institution, he/she shall announce the name, location, and representative, etc. of the non-clinical testing institution on the internet homepage of the Ministry of Food and Drug Safety, etc. ⑧ Detailed matters concerning the scope of documents to be submitted pursuant to Paragraph 2 shall be determined and announced by the Minister of Food and Drug Safety . [This Article Newly Inserted May 1, 2017] Article 24-3 (Compliance Requirements for Non-Clinical Testing Institutions) The term “matters prescribed by the Decree of the Prime Minister” in Article 10-2, Paragraph 3 of the Act refers to the following matters: 1. Prepare a nonclinical trial protocol and conduct the nonclinical trial in accordance with the protocol. 2. Inspections and audits to ensure the reliability of non-clinical trials shall be performed by persons who have no conflict of interest with the relevant non-clinical trials. 3. It must be possible to prove that the non-clinical trial was conducted in accordance with the management standards for non-clinical trials prescribed and notified by the Minister of Food and Drug Safety. 4. Retain data and records related to non-clinical trials for the periods classified in each of the following subparagraphs. a. Testing for medical device approval/amendment approval or medical device certification/amendment certification: 5 years from the date of issuance of the non-clinical test report B. Tests for other purposes: 3 years from the date of issuance of the non-clinical test report [This Article Newly Inserted May 1, 2017] Article 25 (Subjects of Prior Review, etc.) ① The subjects of prior review pursuant to Article 11, Paragraph 1 of the Act are as follows: <Amended Jan. 16, 2024> 1. Newly developed medical devices or rare medical devices pursuant to each subparagraph of Article 8, Paragraph 1 of the Act 2. Medical devices or in vitro diagnostic medical devices requiring the submission of data regarding clinical trials or clinical performance tests pursuant to Article 9, Paragraph 2 , Subparagraph 6 or Article 9 , Paragraph 3, Subparagraph 5, Item B. 3. Medical devices designated as innovative medical devices pursuant to Article 21 of the Act on Promotion of the Medical Device Industry and Support for Innovative Medical Devices (excluding cases where the designation has been revoked after designation) 4. Medical devices for clinical trials or in vitro diagnostic medical devices for clinical performance testing pursuant to Article 7 of the In Vitro Diagnostic Medical Devices Act 5. Other medical devices or in vitro diagnostic medical devices that the Minister of Food and Drug Safety deems require prior review ② The scope of prior review pursuant to Article 11, Paragraph 1 of the Act shall be as follows: <Newly added Jan. 16, 2024> 1. Matters concerning technical documents, clinical trial data, and clinical performance test data corresponding to Article 9, Paragraphs 2 and 3 2. Matters concerning plans for clinical trials and clinical performance tests corresponding to Article 20, Paragraph 1 , Subparagraphs 1 and 3, Paragraph 2 of the Act on the In Vitro Diagnostic Medical Devices, and Article 13, Paragraph 1 , Subparagraphs 1 and 3, Paragraph 2 of the Enforcement Rule of the Act on In Vitro Diagnostic Medical Devices. ③ Any person who intends to request a preliminary review pursuant to Article 11 of the Act shall submit an application for preliminary review of medical devices in the form of Annex 27 (including an application in the form of an electronic document) to the Minister of Food and Drug Safety or the head of the Information Center, in accordance with the classifications in each of the following subparagraphs. <Amended June 14, 2018; Jan 16, 2024> 1. In the case of a person seeking a manufacturing license for a medical device pursuant to Article 6, Paragraph 2 of the Act and a clinical trial pursuant to Article 10 of the Act : Minister of Food and Drug Safety 2. In the case of a person intending to obtain manufacturing certification or file a manufacturing report: the head of the information center ④ The Minister of Food and Drug Safety or the head of the Information Center shall notify the applicant of the results of the preliminary review in accordance with the Notification of Results of Preliminary Medical Device Review ( Form No. 28) . <Amended June 14, 2018; Jan. 16, 2024> ⑤ In addition to the matters prescribed in Paragraphs 1 through 4, necessary detailed matters concerning the procedures, methods, etc. of prior review shall be determined and announced by the Minister of Food and Drug Safety . <Amended Jan. 16, 2024> Article 26 (Application for Permission to Change Permit Matters, etc.) ① Pursuant to Article 12, Paragraph 1 of the Act, if there is a change in the matters for which a manufacturing business has been permitted, a manufacturer shall submit an application for permission to change (Form No. 29) to the head of the regional Food and Drug Administration within 30 days from the date of the change, attaching the data classified according to each of the following subparagraphs. <Amended Jan. 21, 2022; Jan. 16, 2024; Apr. 1, 2025> 1. In the event of a change in the location of a manufacturing facility or the addition of a manufacturing facility, a copy of the consignment contract (limited to cases where the manufacturing process or testing is outsourced) and documents proving compliance with the standards for facilities, manufacturing, and quality control systems pursuant to Annex 2 [excluding manufacturing facilities for medical devices developed and manufactured solely as software that have received authorization, certification, or notification (hereinafter referred to as “software medical devices”)]. 2. In the event of a change in the Quality Manager, the following materials regarding the changed Quality Manager a. Materials capable of verifying qualifications pursuant to Article 11, Paragraph 2. However, submission thereof shall be waived if the qualification of the Quality Manager has been previously verified through the Integrated Medical Device Information System. B. Documents verifying employment status, such as a certificate of employment ② The head of the Regional Food and Drug Administration who receives an application pursuant to Paragraph 1 shall verify the Certificate of Corporate Registration (applicable only to corporations) or the Certificate of Business Registration through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act. However, if the applicant does not consent to the verification of the Certificate of Business Registration, the applicant shall be required to attach the relevant document. <Amended Apr. 1, 2025> ③ Pursuant to Article 12, Paragraph 1 of the Act, when there is a change in matters for which a manufacturing permit or certification has been obtained or a manufacturing report has been filed, a manufacturer shall submit an application for a change permit (Form No. 30), an application for a change certification (Form No. 31), or a report (Form No. 7), attached with the documents specified in each of the following subparagraphs, to the Minister of Food and Drug Safety in the case of a change permit, and to the head of the Information Center in the case of a change certification or change report, respectively. <Amended June 14, 2018; Jan. 21, 2022> 1. Documents that can verify the changes (including electronic documents) 2. When important matters affecting the safety or efficacy of a product, such as the design, materials, chemical components, energy source, or manufacturing process of a medical device, are changed, the documents pursuant to Article 5, Paragraph 1, Subparagraph 2 3. In the event of a change in the location of the manufacturing facility for the relevant item or the addition of a manufacturing facility, a copy of the consignment contract (limited to cases where the manufacturing process or testing is outsourced) and documents proving compliance with the standards for facilities and manufacturing and quality control systems pursuant to Annex 2 (excluding manufacturing facilities for software medical devices). ④ In cases where a person who has acquired the relevant manufacturing facilities, manufacturing methods, etc., pursuant to the separation or merger of enterprises, etc., obtains a modified license or modified certification of a manufactured item or files a modified notification pursuant to Paragraph 3, he/she shall submit to the Minister of Food and Drug Safety or the head of the Information Center, attaching the transfer and acquisition agreement regarding the manufacturing facilities, manufacturing methods, etc. of the item. In such cases, when granting a modified license or modified certification of the relevant item, conditions equivalent to those attached to the license or certification of the transferor may be attached. <Amended June 14, 2018> ⑤ Notwithstanding the provisions of Paragraph 3, if the matters changed pursuant to Paragraph 3 are minor changes to the appearance, packaging materials, packaging units, etc., of manufactured products as prescribed and notified by the Minister of Food and Drug Safety, a document (including electronic documents) stating the details of the change shall be submitted to the Minister of Food and Drug Safety or the head of the Information Center. In this case, it shall be deemed that a change permit or change certification has been obtained or a change notification has been filed pursuant to Article 12 of the Act. <Amended June 14, 2018> ⑥ In any of the following cases where there is a change to matters for which a permit or certification has been obtained or a report has been filed, it shall be deemed that a permit or certification for change has been obtained or a report for change has been filed pursuant to Article 12 of the Act. However, in the case of Subparagraphs 3, 4, and 6, this shall be limited to instances where the change has been made by ordering it within a specified period pursuant to Article 36, Paragraph 2 of the Act. <Amended Jan. 21, 2022> 1. Changes to the standards for classification and designation pursuant to Article 3 of the Act 2. Changes to technical documents or clinical trial data pursuant to Article 6, Paragraph 5 of the Act 3. Follow-up measures based on the review of post-market surveillance results of newly developed medical devices, etc. pursuant to Article 8-2, Paragraph 2 of the Act 4. Re-evaluation pursuant to Article 9 of the Act 5. Change of standards and specifications pursuant to Article 19 of the Act 6. Measures following the report of adverse effects pursuant to Article 31 of the Act ⑦ When the Minister of Food and Drug Safety, the head of a regional Food and Drug Safety Administration, or the head of an Information Center grants a modified license or a modified certification, or accepts a notification of modification, in accordance with the provisions of paragraphs 1 through 6, they shall record the details of the changes in the relevant register. <Amended June 14, 2018> ⑧ Detailed matters concerning the standards, procedures, data requirements, scope of exemption, and management, etc., for change permits, change certifications, and change notifications pursuant to the provisions of paragraphs 1 through 7 shall be determined and announced by the Minister of Food and Drug Safety . Article 27 (Matters to be Complied with by Manufacturers, etc.) ① The matters to be complied with by a manufacturer of a medical device pursuant to Article 13, Paragraph 1 of the Act are as follows: <Amended Dec. 4, 2020; Jan. 21, 2022> 1. Hygienically manage the facilities of the manufacturing plant to prevent health and hygiene hazards, and prevent cross-contamination or contamination from the outside. 2. No objects that pose a risk of causing harm shall be placed in the workplace, and measures shall be taken to prevent the leakage or release of substances harmful to public health from the workplace. 3. Prepare and maintain documents regarding the receipt and shipment of raw materials and finished products, manufacturing processes, and quality control (including test standards and methods, etc.), and thoroughly conduct manufacturing and quality inspections in accordance with them. 4. Prepare and maintain records for each manufacturing unit regarding manufacturing and quality inspection pursuant to Subparagraph 3, as well as records of customer complaint handling, and preserve them for 5 years from the date of manufacture (or a period corresponding to the product lifespan if the product lifespan exceeds 5 years). 5. If released medical devices compromise safety and efficacy or are of poor quality, take corrective measures, such as recall, without delay. 6. For sterile products, be sure to use new containers for packaging and release only after verifying that they have been sterilized. 7. For electrical and mechanical products, release them after ensuring electrical and mechanical safety and electromagnetic safety. 8. For products that come into direct or indirect contact with the human body, release them only after ensuring biological safety. 9. If new data or information regarding the safety and efficacy of a medical device that has been authorized, certified, or registered (including cases of adverse effects caused by the use of the medical device) is discovered, report it in accordance with the procedures prescribed by the Minister of Food and Drug Safety and implement necessary safety measures. 10. Maintain facilities and manufacturing and quality control standards in accordance with Attached Table 2. 11. Sell medical devices manufactured in accordance with the standards of Subparagraph 2 of Annex 2. In this case, a medical device manufactured for the purpose of obtaining recognition of conformity pursuant to Subparagraph (f) of Subparagraph 2 of Annex 2, and which has been recognized as conformity pursuant to the same subparagraph, shall be deemed to have been manufactured after obtaining recognition of conformity pursuant to the same subparagraph. 12. Conduct manufacturing and quality control or production control by reflecting the latest standards and specifications established by the Minister of Food and Drug Safety pursuant to Article 19 of the Act . 13. Thoroughly inspect the hygiene status of employees and provide them with the education and training necessary for the production of high-quality medical devices. 14. Thoroughly carry out post-marketing safety management duties, including post-marketing surveillance, re-evaluation, management of medical devices subject to tracking and management, and management of safety information (including management of adverse event reports). 15. When purchasing medical devices manufactured by your company from a medical institution, comply with the following: a. Inspect whether the medical device complies with the standards for the manufacturing and quality management system specified in Item 2 of Annex 2 , and attach an inspection certificate and release the device only if it is found to be compliant. B. Prepare and keep records regarding the contents and results of the inspection pursuant to Subparagraph A, the date of issuance of the inspection certificate, etc., and preserve them for two years from the date of shipment. 16. In the event of receiving a request for inspection from a medical device seller or lessor pursuant to Article 39, Subparagraph 1, Item ( a) , the following matters shall be observed: a. Inspect whether the medical device complies with the standards for the manufacturing and quality management system specified in Item 2 of Annex 2 , and issue a certificate of inspection only if it is found to be compliant. B. Comply with matters prescribed and notified by the Minister of Food and Drug Safety regarding the procedures and methods for issuing inspection certificates, response periods, instructions concerning sales or leases, etc. ② Pursuant to Article 13, Paragraph 2 of the Act , a manufacturer shall report the matters set forth in each of the following subparagraphs to the Minister of Health and Welfare and the Minister of Food and Drug Safety, as prescribed and notified by the Minister of Food and Drug Safety. <Amended Apr. 13, 2020> 1. Medical device production and export performance for the relevant year 2. Reasons for discontinuation of production, quantity discontinued, and discontinuation schedule of medical devices ( applicable only to cases where production of medical devices having a significant impact on public health is discontinued, as notified by the Minister of Food and Drug Safety in consultation with the Minister of Health and Welfare ) Article 28 (Reporting of Closure, etc.) ① In accordance with Article 14 of the Act , if a manufacturer intends to report the closure or suspension of business, they shall submit a report in the form of Annex No. 32 (including a report in the form of an electronic document) attached with the documents specified in each of the following subparagraphs to the head of the regional Food and Drug Administration having jurisdiction over the location of the manufacturing facility; and if a person who has suspended business intends to report the resumption of business, they shall submit a report in the form of Annex No. 32 (including a report in the form of an electronic document) to the head of the regional Food and Drug Administration having jurisdiction over the location of the manufacturing facility. <Amended Apr. 13, 2020> 1. In case of business closure: Manufacturing license and all manufacturing licenses and manufacturing certificates. However, if the relevant license or certificate has been lost, a statement of the reason for the loss must be submitted. 2. In case of business suspension: Manufacturing license ② When a manufacturer intending to report business closure or suspension pursuant to Article 14 of the Act also intends to report business closure or suspension pursuant to Article 8, Paragraph 7 of the Value Added Tax Act, he/she shall submit a report in Form No. 32 (including a report in electronic form) to the head of the regional Food and Drug Administration having jurisdiction over the location of the manufacturing facility, attaching the documents specified in each subparagraph of Paragraph 1 and the report on suspension or closure in Form No. 9 of the Enforcement Rules of the Value Added Tax Act . In this case, the head of the regional Food and Drug Administration shall send the report on suspension or closure in Form No. 9 of the Enforcement Rules of the Value Added Tax Act to the head of the competent tax office without delay (including sending via an information and communications network; hereinafter the same applies in this Article). <Newly added Apr. 13, 2020> ③ The head of the Regional Food and Drug Administration shall record the details in the manufacturing license register upon receiving a report of business closure pursuant to Paragraphs 1 and 2, and in the manufacturing license register and the license, respectively, upon receiving a report of suspension or resumption of business. <Amended Apr. 13, 2020> ④ Where the head of the tax office having jurisdiction over the location of the manufacturing facility receives a report of suspension or closure pursuant to Article 13, Paragraph 5 of the Enforcement Decree of the Value Added Tax Act and forwards it to the head of the competent regional Food and Drug Safety Administration, the report of suspension or closure under Paragraph 1 shall be deemed to have been submitted. <Amended Apr. 13, 2020> Article 29 (Application for Import Business Permit, etc.) ①Any person who intends to obtain a permit for the import business of medical devices pursuant to Article 15, Paragraph 1 of the Act shall submit an application form in Form No. 1 (including an application form in electronic form)to the head of the regional Food and Drug Administration having jurisdiction over the location of the import business, attaching the documents specified in each subparagraph of Article 3 , Paragraph 1 (including electronic documents). ② The head of the regional Food and Drug Administration who receives the application pursuant to Paragraph 1 shall verify the certificate of corporate registration (only applicable to corporations) through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act . Article 30 (Application for Import Permit, etc.) ①Any person who intends to obtain an import permit for a medical device pursuant to Article 15, Paragraph 2 of the Act shall submit to the Minister of Food and Drug Safety an application form in Annex 3 (including an application form in electronic form) with the following documents (including electronic documents) attached. 1. The documents referred to in Article 5, Paragraph 1, Subparagraph 2. However, this may be omitted if it is proven, in accordance with the regulations set by the Minister of Food and Drug Safety, that the medical device to be imported is the same product from the same manufacturer (meaning the country of manufacture, manufacturer, and manufacturing facility are the same) as the medical device already approved. 2. Documents proving that the manufacturing facility of the imported medical device has established a manufacturing and quality control system pursuant to Article 15 , Paragraph 4 of the Act ② Any person who intends to obtain import certification for a medical device pursuant to Article 15, Paragraph 2 of the Act shall submit an application in the form of Annex 5 (including an application in the form of an electronic document) to the head of the Information Center, attaching the documents specified in each subparagraph of Paragraph 1 (including electronic documents). In this case, “authorized” in the proviso to Paragraph 1, Subparagraph 1 shall be deemed to mean “certified.” <Amended June 14, 2018> ③ Any person who intends to file an import declaration for medical devices pursuant to Article 15, Paragraph 2 of the Act shall submit a declaration form in Annex 7 (including a declaration form in electronic document format) to the head of the Information Center. <Amended June 14, 2018> Article 31 (Standards for Facilities for Quality Inspection and Manufacturing and Quality Control Systems) ①for facilities necessary for quality inspection and manufacturing and quality control systems that a person intending to obtain an import business license, or a person intending to obtain an import license or import certification or to file an import declaration, must meet pursuant to the main text of Article 15, Paragraph 4 of the Act are as set forth in Table 4 . ② In the proviso to Article 15, Paragraph 4 of the Act , the phrase “cases prescribed by the Decree of the Prime Minister, such as entrusting tests for quality control” refers to cases where a person seeking an import business license, or a person seeking an import license or certification, or a person filing an import declaration, entrusts tests for quality control to any of the persons falling under any of the following subparagraphs. In such cases, the person seeking an import business license, or a person seeking an import license or certification, or a person filing an import declaration, may not be required to possess a test laboratory or test facility pursuant to Subparagraph 1, Item (c) of Annex 4. <Amended May 1, 2017> 1. Medical Device Testing and Inspection Agency 1-2. Non-clinical testing institutions 2. Manufacturers recognized as compliant pursuant to Item 2(f) of Annex 2 Article 32 (Exemption from Medical Device Import Business Permit, etc.) ① Medical devices that may be imported without obtaining an import business permit, import permit, or import certification, or without filing an import declaration, pursuant to Article 6, Paragraph 10 of the Act which is applied mutatis mutandis pursuant to Article 15 , Paragraph 6 of the Act , are as follows: <Amended Jun. 15, 2016; Jun. 14, 2018; Dec. 31, 2018; Jan. 16, 2024> 1. Medical devices subject to clinical trials for which a clinical trial plan has been approved pursuant to the main text of Article 10, Paragraph 1 of the Act 2. Separate medical devices necessary to conduct a clinical trial that does not require approval of the clinical trial plan pursuant to the proviso of Article 10, Paragraph 1 of the Act 3. Separate medical devices necessary to conduct clinical trials concerning pharmaceuticals, etc., for which approval of the clinical trial plan is required pursuant to Article 34, Paragraph 1 of the Pharmaceutical Affairs Act 3-2. Separate medical devices necessary to conduct clinical trials concerning pharmaceuticals, etc. that do not require approval of a clinical trial plan pursuant to Article 34, Paragraph 2 of the Pharmaceutical Affairs Act 4. Medical devices imported for the purpose of obtaining import permits or certifications, or filing import declarations. 5. Medical devices used for research purposes, such as product development 5-2. Medical devices imported not for the purpose of sale, but to assist consumers in purchasing the product, such as by allowing them to inspect the actual product. 5-3. Medical devices imported for testing or inspection pursuant to Article 27 of the Act 6. Medical devices for personal use or relief purposes, etc., as designated and notified by the Minister of Food and Drug Safety 7. Deleted <December 31, 2018> 8. Medical devices used in clinical laboratories recognized as suitable by the Minister of Food and Drug Safety after evaluating their quality control systems, inspection performance, and proficiency, as specified and notified by the Minister of Food and Drug Safety ② Matters concerning the import procedures, subjects, and issuance of confirmation certificates, etc., of medical devices excluded from import business permits and import permits, import certifications, and import declarations pursuant to Paragraph 1 shall be determined and announced by the Minister of Food and Drug Safety . Article 33 (Matters to be Complied with by Importers, etc.) ① The matters to be complied with by importers of medical devices pursuant to Article 13, Paragraph 1 of the Act, which is applied mutatis mutandis pursuant to Article 15, Paragraph 6 of the Act, are as follows: <Amended Dec. 4, 2020; Jan. 21, 2022; Oct. 31, 2025> 1. Hygienically manage the facilities of the importing business to prevent health and hygiene hazards, and prevent cross-contamination or contamination from the outside. 2. Prepare and maintain documents regarding the receipt, shipment, and quality control of medical devices (including test standards and methods, labeling, packaging, etc.) and management records regarding the storage and shipment of accessories, and thoroughly conduct import and quality inspections accordingly. 3. Prepare and maintain records by import unit regarding import and quality inspection pursuant to Subparagraph 2, as well as records of customer complaint handling, and preserve them for 5 years from the date of import (or a period corresponding to the product lifespan if the product lifespan exceeds 5 years). 4. In the case of sterilized products, release after verifying that they have been sterilized. 5. For electrical and mechanical products, release them after ensuring electrical and mechanical safety and electromagnetic safety. 6. For products that come into direct or indirect contact with the human body, release them only after ensuring biological safety. 7. Prepare and keep on file a product standard document for each item that includes the following items: a. Name of the medical device (product name, item name, and model name) B. Name of manufacturer and country of manufacture of imported medical device C. Self-Quality Control Test Standards for Shape, Structure, and Finished Products D. Matters to be stated on medical device containers, etc. pursuant to Articles 20 through 23 of the Act E. Installation Method and Procedure (Applicable only to medical devices requiring installation management) F. Matters concerning sterilization methods, sterilization standards, and sterilization determination (applicable only to sterilized packaged products) d. Author and date of enactment of the product standard (in the case of revision, the reviser, date of revision, and reason for revision shall be indicated) 8. Prepare and keep on file an import management standard document containing the following items: a. Matters concerning product management and testing and inspection B. Matters concerning the determination of test and inspection results and the handling of rejected items C. Matters concerning the management of testing and inspection facilities D. Method of contacting the manufacturer of imported medical devices E. Verification items regarding the manufacturing and quality control status of the imported medical device manufacturer f. The author of the Import Management Standards and the date of enactment (in the case of amendment, the author, date of amendment, and reason for amendment shall be stated) 9. Verify the compliance of the labeling and packaging of imported medical devices and prepare a record thereof. 10. Inspect the product storage facilities and prepare a record. 11. The Quality Manager shall ensure that the following matters are implemented a. In the event that a complaint arises regarding the quality of shipped medical devices, establish and implement relevant procedures to identify the cause and take corrective measures, and record and preserve such procedures. B. To properly implement quality control of imported medical devices, keep product standards and import management standards on file and utilize them. c. Prepare work orders based on the documents in subparagraph b, and inspect and verify whether operations are being conducted in compliance with these standards. 12. If shipped medical devices compromise safety and efficacy or are of poor quality, take corrective measures, such as recall, without delay. 13. Establish an education plan for employees to ensure the quality of imported medical devices, conduct education and training regularly in accordance with the plan, and prepare and keep records thereof. 14. If new data or information regarding the safety and efficacy of a medical device that has been authorized, certified, or registered (including cases of adverse events related to the use of the medical device) is discovered, report it in accordance with the procedures prescribed by the Minister of Food and Drug Safety and implement necessary safety measures. 15. A medical device manufacturing facility shall import and sell medical devices manufactured in accordance with the standards of Subparagraph 3 of Annex 4. In this case, a medical device imported for the purpose of obtaining recognition of conformity pursuant to Subparagraph 3 of Annex 4, and which has been recognized as conformity pursuant to the same subparagraph, shall be deemed to have been imported after being recognized as conformity pursuant to the same subparagraph. 16. Comply with the import and export guidelines for medical devices announced by the Minister of Trade, Industry and Energy pursuant to Article 12 of the Foreign Trade Act, and the regulations on the management of imported medical devices determined by the Minister of Food and Drug Safety. 17. Maintain facilities, manufacturing, and quality control systems by reflecting the latest standards and specifications established by the Minister of Food and Drug Safety pursuant to Article 19 of the Act, and conduct import and quality control or import management. 18. Thoroughly carry out post-marketing safety management duties, including post-marketing surveillance, re-evaluation, management of medical devices subject to tracking and management, and management of safety information (including management of adverse event reports). 19. When importing used medical devices or purchasing medical devices imported by your company from a medical institution, you shall comply with the following: a. Inspect whether it conforms to the test standards pursuant to Subparagraph 7, Item c, and if it conforms, attach an inspection certificate and release it. B. Prepare and keep records regarding the contents and results of the inspection pursuant to Subparagraph A, the date of issuance of the inspection certificate, etc., and preserve them for two years from the date of shipment. 20. In the event of receiving a request for inspection from a medical device seller or lessor pursuant to Article 39, Subparagraph 1, Item (a), the following matters shall be observed: a. Inspect whether it complies with the standards of Item 2 of Annex 2, and issue a certificate of inspection only if it complies. B. Comply with matters prescribed and notified by the Minister of Food and Drug Safety regarding the procedures and methods for issuing inspection certificates, response periods, instructions concerning sales or leases, etc. ② Pursuant to Article 13, Paragraph 2 of the Act, which is applied mutatis mutandis under Article 15, Paragraph 6 of the Act, an importer shall report the matters set forth in each of the following subparagraphs to the Minister of Health and Welfare and the Minister of Food and Drug Safety, as prescribed and notified by the Minister of Food and Drug Safety. However, if a standard customs clearance pre-report has been submitted via an electronic trade document pursuant to the Act on the Promotion of Electronic Trade, the matters under Subparagraph 1 may not be reported. <Amended Apr. 13, 2020> 1. Medical device import performance for the relevant year 2. Reasons for suspension of medical device imports, quantity suspended, and suspension schedule ( applicable only to cases where the import of medical devices having a significant impact on public health is suspended, as notified by the Minister of Food and Drug Safety in consultation with the Minister of Health and Welfare ) Article 34 (Application Mutatis Mutandis) With respect to importers and imported items , Article 3 , Paragraph 3 ; Article 4 ; Article 5 , Paragraphs 2 through 7 ; Article 6 , Paragraphs 2 and 3; Article 7 , Paragraphs 2 and 3; Article 9 ; Articles 11 through13; Articles 16 through 19 ; Article 25 ; Article 26 (excluding Article 26 , Paragraph 1 , Item 1 ); Article 28 ; and Article 64 shall apply mutatis mutandis. In this case, “manufacturing facility” shall be deemed “import business” and “manufacturing” shall be deemed “import”; “Paragraph 1” in Article 5 , Paragraph 4 shall be deemed“ Article 30, Paragraph 1 ”; “ Article 27, Paragraph 1 ” in Article 12, Paragraph 1, Subparagraph 9 shall be deemed “ Article 33, Paragraph 1 ”; “import” in Article 12 , Paragraph 3, Subparagraph 2 shall be deemed “manufacturing”; “ Article 5 , Paragraph 1 , Subparagraph 2 ”in Article 16 , Paragraph 2 shall be deemed “ Article 30, Paragraph 1, Subparagraph 1 ”; and“ Article 5 , Paragraph 1, Subparagraph 2 ” in Article 16, Paragraph 3 shall be deemed “ Article 30 , Paragraph 1, Subparagraph 1 ”. <Amended July 29, 2016; May 1, 2017> Article 34-2 (Method of Supplying Rare and Urgently Introduced Medical Devices) ① When the Minister of Food and Drug Safety intends to supply rare and urgently introduced medical devices pursuant to the part other than each subparagraph of Article 15-2, Paragraph 1 of the Act (hereinafter referred to as “rare and urgently introduced medical devices”), he/she shall conduct a demand survey from relevant administrative agencies, medical institutions, health and medical-related institutions, organizations, experts, or patients, etc. ② When the Minister of Food and Drug Safety has conducted a demand survey pursuant to Paragraph 1, he/she shall establish a supply plan for rare and urgently needed medical devices, including the matters in each of the following subparagraphs, on a quarterly, semi-annual, and annual basis. 1. Matters concerning supply quantity 2. Matters concerning supply prices 3. Matters concerning the recipients of supply 4. Other matters deemed necessary by the Minister of Food and Drug Safety for the supply of rare or urgently needed medical devices ③ If the Minister of Food and Drug Safety intends to supply rare or urgently needed medical devices in accordance with the supply plan under Paragraph 2, he/she must enter into a supply contract with the supplier. ④ In addition to the matters prescribed in paragraphs 1 through 3, the Minister of Food and Drug Safety shall determine and announce the detailed matters necessary for the supply method of rare and urgently needed medical devices . [This Article Newly Inserted June 12, 2019] Article 34-3 (Entrustment of Supply Duties for Rare and Urgently Introduced Medical Devices) ① The Minister of Food and Drug Safety shall, pursuant to Article 15-2, Paragraph 2 of the Act , entrust the duties concerning the supply of rare and urgently introduced medical devices and the provision of related information to the Information Center. ② An information source entrusted with the duties regarding rare and urgently needed medical devices pursuant to Paragraph 1 shall submit to the Minister of Food and Drug Safety, as determined by the Minister, business operation plans, fund management plans, project implementation status, and details of fund execution, etc., regarding the entrusted duties. [This Article Newly Inserted June 12, 2019] Article 35 (Reporting of Repair Business, etc.) ①Pursuant to Article 16, Paragraph 1 of the Act , any person who intends to engage in the business of repairing medical devices shall submit a report on the repair business in the form of Annex 33 ( including a report in the form of an electronic document) to the Mayor of a Special Self-Governing City, Governor of a Special Self-Governing Province, Mayor, County Head, or District Head having jurisdiction over the location of the repair business , attaching the documents referred to in Article 3, Paragraph 1, Subparagraphs 1 and 2 (including electronic documents). In this case, the Mayor of a Special Self-Governing City, Governor of a Special Self-Governing Province, Mayor, County Head, or District Head shall verify the Certificate of Incorporated Matters (applicable only to corporations) through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act. <Amended Apr. 13, 2020; Jun. 24, 2021> ② When the Mayor of a Special Self-Governing City, the Governor of a Special Self-Governing Province, a Mayor, a County Head, or a District Head accepts a report of repair business pursuant to Paragraph 1, they shall issue a report certificate in the form of Annex 34 to the reportee and record the matters of each of the following subparagraphs in the register of reports of repair business. <Amended June 24, 2021> 1. Report Approval Number and Date of Approval 2. Types of medical devices subject to repair 3. Name and Resident Registration Number of the repairer (in the case of a corporation, the name and Resident Registration Number of its representative) 4. Name and address of the repair shop (in the case of a corporation, the address of its principal office) ③ If a repair business operator intends to change matters reported, they shall submit a change notification form ( Form No. 35, including a notification form in electronic format) to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the location of the repair business, attaching the reason for the change, supporting documents (including electronic documents), and the certificate of report. In this case, if the change of location involves a different reporting authority, the notification shall be submitted to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the new location. <Amended June 24, 2021> ④ The facilities and quality control system that a person intending to report a repair business must possess pursuant to the main text of Article 16, Paragraph 2 of the Act are as set forth in Table 5 . ⑤ In the event that a repair business operator closes or suspends business pursuant to Article 14 of the Act , which is applied mutatis mutandis pursuant to Article 16, Paragraph 4 of the Act , the operator shall submit a report in the form of Annex No. 32 (including a report in the form of an electronic document) attached with the certificate of registration (or a statement of reasons for loss if the certificate of registration was lost at the time of reporting closure) to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the location of the repair business operator; and in the event that a person who has suspended business reports the resumption of business, the operator shall submit a report in the form of Annex No. 32 (including a report in the form of an electronic document) to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the location of the repair business operator. <Amended Apr. 13, 2020; Jun. 24, 2021> ⑥ When a Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head receives a report pursuant to Paragraphs 3 and 5, they shall record the details of the report in the relevant report ledger and report certificate, respectively. <Amended June 24, 2021> ⑦ Paragraphs 2 and 4 of Article 28 shall apply mutatis mutandis to the report of suspension or closure of business pursuant to Paragraph 5. In this case, “Director of the Regional Food and Drug Administration” shall be deemed to be “Mayor of a Special Self-Governing City, Governor of a Special Self-Governing Province, Mayor, County Head, or District Head.” <Amended Apr. 13, 2020; Jun. 24, 2021> Article 36 (Compliance Requirements for Repair Operators) The matters to be complied with by a medical device repair operator pursuant to Article 13 of the Act, which is applied mutatis mutandis pursuant to Article 16, Paragraph 4 of the Act , are as follows: <Amended June 14, 2018> 1. Do not repair a medical device by altering it differently from the contents authorized, certified, or reported. However, this shall not apply in cases of minor alteration repairs pursuant to Article 46-2 . 2. If a medical device is repaired, the company name and address shall be written on the container or exterior of the medical device. 3. Notify the person who requested the repair of the medical device of the repair details in writing (including the details of minor modification repairs performed pursuant to Article 46-2). 4. Maintain the facility and quality management system pursuant to Article 35 , Paragraph 4. Article 37 (Reporting, etc. of Sales or Lease Business) ①Any person who intends to engage in the business of selling or leasing medical devices pursuant to Article 17, Paragraph 1 of the Act shall submit a report in the form of Annex No. 36 (including a report in the form of an electronic document) to the Mayor of a Special Self-Governing City, Governor of a Special Self-Governing Province, Mayor, County Head, or District Head having jurisdiction over the sales or lease business. In this case, the Mayor of a Special Self-Governing City, Governor of a Special Self-Governing Province, Mayor, County Head, or District Head shall verify the Certificate of Corporate Registration (applicable only to corporations) through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act. <Amended Apr. 13, 2020; Jun. 24, 2021> ② A person who intends to engage in the business of selling medical devices pursuant to Article 17, Paragraph 1 of the Act , and who intends to sell medical devices through e-commerce or mail-order sales pursuant to the Act on the Consumer Protection in Electronic Commerce, etc., may use a building for residential use under the Building Act or a business incubation center pursuant to Article 18-3, Paragraph 1 of the Special Measures Act on the Promotion of Venture Businesses as a business office. <Newly added July 8, 2024> ③ When the Mayor of a Special Self-Governing City, the Governor of a Special Self-Governing Province, a Mayor, a County Head, or a District Head accepts a report for the sale or rental business of medical devices pursuant to Paragraph 1, they shall issue a report certificate in the form of Annex 37 to the reporter and record the following matters in the register of sales or rental business reports. <Amended Apr. 13, 2020; Jul. 8, 2024> 1. Report Number and Date of Report 2. Name and Resident Registration Number of the medical device seller or lessor (in the case of a corporation, the name and Resident Registration Number of its representative) 3. Name of the business office and its location ④ If a seller or lessor changes the matters reported pursuant to Paragraph 1, they shall submit a report of change in the form of Annex 38 (including a report in electronic form), attached with the certificate of report and documents verifying the changes, to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the sales or rental business. In this case, if the location is changed to a different reporting authority, the report shall be submitted to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the new location. <Amended Apr. 13, 2020; Jul. 8, 2024> ⑤ If a seller or lessor intends to report the closure or suspension of business, they shall submit a report in the form of Annex No. 32 (including a report in the form of an electronic document) attached with the certificate of registration (or a statement of the reason for loss if the certificate of registration was lost at the time of reporting the closure) to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the sales or rental business; and if a person who has suspended business intends to report the resumption of business, they shall submit a report in the form of Annex No. 32 (including a report in the form of an electronic document) to the Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head having jurisdiction over the sales or rental business. <Amended Apr. 13, 2020; Jul. 8, 2024> ⑥ When a Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head receives a report pursuant to Paragraphs 3 and 4, they shall record the details of the report in the relevant report ledger and report certificate, respectively. <Amended Apr. 13, 2020; Jul. 8, 2024> ⑦ Paragraphs 2 and 4 of Article 28 shall apply mutatis mutandis to the report of suspension or closure of business pursuant to Paragraph 4. In this case, “Director of the Regional Food and Drug Administration” shall be deemed to be “Mayor of a Special Self-Governing City, Governor of a Special Self-Governing Province, Mayor, County Head, or District Head.” <Amended Apr. 13, 2020; Jul. 8, 2024> Article 38 (Exemption from Reporting of Sales Business, etc.) Medical devices that may be sold without reporting pursuant to Article 17, Paragraph 2, Subparagraph 4 of the Act are as follows: 1. Condom 2. Blood glucose meters used in which a blood glucose measurement function is included or combined with mobile phones, home appliances, etc. 3. Other medical devices notified by the Minister of Food and Drug Safety in consideration of the degree of risk and safety Article 39 (Methods for Assuring Medical Device Quality by Sellers and Lessors) The methods for assuring medical device quality pursuant to Article 18 , Paragraph 1 of the Act and Article 10-2, Subparagraph 1 of the Enforcement Decree of the Medical Devices Act (hereinafter referred to as the “Decree”)are as follows: <Amended June 15, 2016> 1. If you purchase a medical device from a medical institution, you must comply with the following: a. Request an inspection from the manufacturer or importer of the relevant medical device, or request an inspection from a medical device testing and inspection agency. B. Sell or lease only medical devices for which an inspection certificate has been obtained pursuant to Article 27, Paragraph 1, Item 15 , Subparagraph A and Article 33, Paragraph 1, Item 19, Subparagraph A, based on the inspection results under Subparagraph A. C. If instructions regarding sale or lease are received as a result of the inspection under Item A pursuant to Article 27, Paragraph 1, Item 16 , Subparagraph B and Article 33, Paragraph 1, Item 20, Subparagraph B, comply therewith. 2. Prepare records regarding the contents and results of inspections pursuant to each item of Subparagraph 1, the issuer and date of issuance of the inspection certificate, instructions regarding sale or lease, etc., and preserve them for two years from the date of sale. 3. Prepare and maintain records regarding the handling of defective medical devices, and preserve them for one year from the date of the record. 4. Comply with the medical device distribution quality management standards in Annex 6. However, cases where a sales business report may be omitted pursuant to Article 17, Paragraph 2 of the Act shall be excluded. Article 40 (Matters concerning the maintenance of sales order, etc. by sellers and lessors) Matters concerning the maintenance of sales order, etc. related to medical device safety pursuant to Article 18 , Paragraph 1 of the Act and Article 10-2, Subparagraph 1 of the Enforcement Decree are as follows: 1. Do not purchase medical devices from any person other than a manufacturer, importer, or seller. However, this does not apply when purchasing from a medical institution. 2. Do not use any name or mark pursuant to any of the following subparagraphs as the name, etc. of the establishment, or use them in combination with the mark of the business establishment. a. Names or markings likely to be mistaken for the business office of a manufacturer or importer B. The name of a medical institution pursuant to Article 3 of the Medical Service Act or a name similar thereto 3. Do not sell or lease, or store or display for the purpose of sale or lease, any medical device falling under any of the following subparagraphs. a. Those without an inspection certificate attached pursuant to Article 39, Subparagraph 1, Item b . b. Items that are contaminated or damaged, or for which the Minister of Food and Drug Safety or the head of a regional Food and Drug Safety office has ordered collection or disposal. c. Items that have expired or are no longer valid 4. Do not sell, store, or display for sale any medical device that has not been authorized, certified, or reported. Article 41 (Exemption from Inspection Certificate for Used Medical Devices) Notwithstanding Article 27, Paragraph 1 , Item 15 , Article 33 , Paragraph 1 , Item 19 , Article 39 , Item 1 , Subparagraph B , and Article 40 , Item 3, Subparagraph A , medical devices with low potential risk, etc., as determined and notified by the Minister of Food and Drug Safety, may be sold without attaching an inspection certificate. Article 42 (Matters to be Labeled on Containers, etc.) Pursuant to the proviso to the part other than the subparagraphs of Article 20 of the Act , cases where matters to be labeled on the container or outer casing of a medical device are not required shall be any of the following: <Amended June 15, 2016> 1. Cases where the surface area of ​​the container or outer casing is narrow, or where it is impossible to write all the matters specified in each subparagraph of Article 20 of the Act on the container or outer casing, and the information is written on the outer container, outer packaging, or attached document. However, even in this case, the model name and the trade name of the manufacturer or importer must be written on the container or outer casing of the medical device. 2. In cases where the labeling information for medical devices intended for export is written in accordance with the standards of the importing country Article 43 (Matters to be Stated in Attached Documents) ① The term “matters prescribed by Decree of the Prime Minister” in Article 22, Paragraph 1, Subparagraph 4 of the Act refers to the matters in each of the following subparagraphs. <Amended Jan. 16, 2024> 1. Matters under Article 20 , Subparagraphs 1 through 3 and 5 through 7 of the Act 2. Purpose of use of the product 3. Storage or preservation method 4. In cases where a domestic manufacturer outsources all manufacturing processes, the trade names and addresses of the commissioner (referring to the consignor) and the manufacturer (referring to the consignee). 5. In cases where individual items are packaged for single use, the model name and manufacturer's name on the minimum unit packaging 6. In the case of medical devices that can be reused after sterilization, information on appropriate procedures for reuse, including methods for cleaning, disinfection, packaging, and re-sterilization, as well as limitations on the number of times they can be reused. 7. In the case of medical devices that emit radiation for medical treatment purposes, matters concerning the characteristics, type, intensity, and diffusion, etc. of the radiation 8. Date of creation of the attached document 9. Contact Information for Adverse Effect Reporting (Korea Medical Device Safety Information Center, Contact Info) 10. Other matters concerning technical information, such as the characteristics of the medical device ② Notwithstanding Paragraph 1, the matters to be stated in the accompanying documents for medical devices for clinical trials are as follows: 1. Marking “For clinical trials” 2. Product Name and Model Name 3. Serial number and date of manufacture (if there is an expiration date, the expiration date may be written) 4. Storage Method 5. The trade name of the manufacturer or importer (including the name of the manufacturer and country in the case of commissioned manufacturing or import) 6. Indication “Not for use other than clinical trials” ③ Where the matters specified in subparagraphs 1 through 5 and 9 of Paragraph 1 are indicated on the container, outer casing, or packaging, such indication may be omitted from the attached documents. <Amended Jan. 16, 2024> Article 44 (Method of Indicating Information) ① When indicating information on the container or outer casing, outer container or packaging, and accompanying documents of a medical device pursuant to Article 23 of the Act , the methods set forth in each of the following subparagraphs shall be followed. <Amended June 24, 2021> 1. Write in Hangul, or write Chinese characters or foreign languages ​​of the same size as the Hangul together with Hangul. However, in the following cases, Hangul may be omitted. a. Cases where medical devices for export are written in the language of the destination country B. Cases where items written in a foreign language are written in the relevant language during medical device licensing, certification, or notification. 2. In addition , comply with matters concerning font size, line spacing, and other methods of notification prescribed by the Minister of Food and Drug Safety. ② When writing the product name, the trade name of the manufacturer or importer, etc., on the container, outer casing, or packaging of a medical device, Braille marking may be used in parallel with the marking method under Paragraph 1. Article 44-2 (Provision of Information for Persons with Visual Impairments, etc.) Medical devices for which the Minister of Food and Drug Safety may recommend acts pursuant to each subparagraph of Article 23-2, Paragraph 1 of the Act shall be medical devices for self-use for the management of chronic diseases. [This Article Newly Inserted Aug. 7, 2024] Article 45 (Scope of Medical Device Advertising, etc.) The scope of advertising prohibited pursuant to Article 24 , Paragraphs 2 and 3 of the Act refers to any of the advertisements falling under any of the following subparagraphs and corresponding to Annex 7 . 1. Advertising pursuant to Article 2, Subparagraph 2 of the Act on Fair Labeling and Advertising 2. Advertising by medical device handlers (including workers, etc. employed by medical device handlers) through methods such as soliciting purchases, product descriptions, and demonstrations [Wholly Amended June 24, 2021] Article 45-2 (Objection to the result of re-examination by a self-regulatory body) Pursuant to Article 25 , Paragraph 5 of the Act and Article 10-4, Paragraph 4 of the Enforcement Decree , any person who intends to file an objection to the result of re-examination by a self-regulatory body (referring to an institution or organization reported to the Minister of Food and Drug Safety pursuant to Article 25-2, Paragraph 1 of the Act ; hereinafter the same applies) shall submit to the Minister of Food and Drug Safety an objection to the result of re-examination of medical device advertising in the form of Annex 38-2 , attaching the documents listed in each of the following subparagraphs. 1. Purpose and Grounds for Objection 2. Details of the Application for Medical Device Advertising Review 3. Copy of the Notification of Results for Medical Device Advertising Review and Re-review 4. Other reference materials that can verify the facts, etc., regarding the content of the advertisement review application [This Article Newly Inserted June 24, 2021] Article 45-3 (Reporting and Monitoring of Self-Regulatory Bodies) ①Any person who intends to report the performance of medical device advertising review duties pursuant to Article 25-2 , Paragraph 1 of the Act and Article 10-5, Paragraph 4 of the Enforcement Decree shall submit a report on a self-regulatory medical device advertising body in the form of Annex 38-3 to the Minister of Food and Drug Safety, attaching the documents specified in each of the following subparagraphs. In this case, the Minister of Food and Drug Safety shall verify the Certificate of Incorporated Matters (applicable only to corporations) through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act. <Amended Jan. 21, 2022> 1. Status of dedicated departments and full-time personnel pursuant to Article 10-5, Paragraph 3, Subparagraph 1 of the Enforcement Decree 2. Status of computer equipment and offices pursuant to Article 10-5, Paragraph 3, Subparagraph 2 of the Enforcement Decree 3. Copy of the Articles of Association of the institution or organization 4. A copy of the registration certificate of a consumer organization registered with the Fair Trade Commission pursuant to Article 29 of the Framework Act on Consumers ( applicable only in the case of Article 25-2, Paragraph 1, Subparagraph 2 of the Act ) ② The self-regulatory body shall, in accordance with Article 25-4 of the Act , submit to the Minister of Food and Drug Safety the results of monitoring whether the medical device advertisements it has reviewed comply with Article 24 , Paragraphs 2 and 3 of the Act on a quarterly basis within 30 days after the end of each quarter. [This Article Newly Inserted June 24, 2021] Article 45-4 (Medical Devices Subject to Sealing, etc.) ① In the part other than the subparagraphs of Article 25-5 of the Act , “medical devices prescribed by the Decree of the Prime Minister” means any medical device falling under any of the following subparagraphs. 1. Sterile-packaged products among medical devices subject to traceability management that are implanted in the human body for more than one year 2. The following medical devices that are at risk of contamination or deterioration if opened and distributed a. Syringes (only sterile packaged products) B. Injection needles (applicable only to sterile packaged products) C. Suture D. Wound dressings (only sterile packaged products) E. Contact lenses f. Other sterile packaged medical devices designated by the Minister of Food and Drug Safety as requiring sealing during distribution ② The seal pursuant to Article 25-5 of the Act shall be such that the container or packaging cannot be opened without breaking the seal of the container or packaging, and after opening, it shall not be easily restored to its original state. [This Article Newly Inserted Jan. 21, 2022] Article 46 (Approval for Display of Medical Devices for Exhibition Purposes, etc.) ① Any person who intends to display medical devices for exhibition purposes pursuant to the proviso to Article 26, Paragraph 1 of the Act shall obtain prior approval from the head of the regional Food and Drug Safety Administration. ② Any person who intends to display a medical device approved pursuant to Paragraph 1 shall not engage in acts other than those prescribed and notified by the Minister of Food and Drug Safety, such as simply operating the device to explain the method of use, or attaching or placing promotional materials regarding the reported contents. ③ The head of the regional Food and Drug Administration shall revoke the approval of a medical device approved pursuant to Paragraph 1 if it falls under any of the following subparagraphs, and may take measures such as sealing or suspending the display of the relevant medical device. 1. If displayed or used differently from the approved content 2. In cases where an act other than the act permitted under Paragraph 2 has been performed 3. In other cases where an act has been committed that poses a risk to public health. ④ After using a medical device approved under Paragraph 1 in accordance with the approved details, the said medical device may be used as a medical device under Article 10 , Subparagraph 4 and Article 32, Paragraph 1, Subparagraphs 5-2 and 5-3, as prescribed and notified by the Minister of Food and Drug Safety. <Newly added June 24, 2021> ⑤ Detailed matters concerning the application method and procedure for approval, approval standards, etc. pursuant to Paragraph 1 shall be determined and announced by the Minister of Food and Drug Safety . <Amended June 24, 2021> Article 46-2 (Minor Alteration Repair of Medical Devices) In accordance with the proviso to Article 26, Paragraph 3 of the Act , a medical device repair business operatormay perform repairs that alter the appearance falling under any of the following subparagraphs to the extent that the safety and efficacy of the medical device are not affected. 1. Repairs that change the color, material, etc. of the exterior that protects the medical device from external impact 2. Repairs that alter the external shape, structure, etc., such as stands, handles, and wheels, which are not directly related to the performance of the medical device. [This Article Newly Inserted June 14, 2018] Article 47 (Scope of Permissible Alteration or Modification of Medical Devices) The medical devices that manufacturers and importers may alter or modify pursuant to Article 26, Paragraph 4, Subparagraph 1 of the Act are as follows. In this case, the alteration or modification must be carried out within a scope where the performance can be improved without changing the inherent characteristics of the medical device. 1. Diagnostic radiation generating device pursuant to Article 37 of the Medical Service Act 2. Specialized medical equipment pursuant to Article 38 of the Medical Service Act 3. Other medical devices that do not compromise safety and efficacy even if altered or modified, as designated and notified by the Minister of Food and Drug Safety Article 48 (Acknowledgment of Conformity for Manufacturing and Quality Control System) ① Any person who intends to obtain an acknowledgment of conformity for a manufacturing and quality control system (hereinafter referred to as “acknowledgment of conformity”) pursuant to the first part of Article 28, Paragraph 1 of the Act shall submit to the Minister of Food and Drug Safety or the head of a quality control review agency pursuant to Article 28-3, Paragraph 1 of the Act (hereinafter referred to as “Minister of Food and Drug Safety, etc.” in this Article) an application for acknowledgment of conformity or modified acknowledgment of conformity for a medical device manufacturing and quality control system in Form No. 38-4 (including an application in electronic form), with the following documents (including electronic documents) attached. 1. Documents outlining the overview of the manufacturing facility, such as its name and location 2. Manufacturing Plant Quality Manual 3. Product Standards, Manufacturing Control Records, and Quality Control Records for the Medical Device for which Recognition of Conformity is Applied 4. Data on corrective and preventive measures necessary to prevent or improve the occurrence of non-conforming situations in the manufacturing and quality control system 5. Other matters determined and notified by the Minister of Food and Drug Safety as deemed necessary for the suitability assessment review ② In the event of an intention to change any of the matters listed in the following subparagraphs among those for which conformity has been recognized pursuant to the latter part of Article 28, Paragraph 1 of the Act, an application for recognition of conformity or recognition of modified conformity for a medical device manufacturing and quality management system in the form of Annex 38-4 (including an application in the form of an electronic document) must be submitted to the Minister of Food and Drug Safety, etc., attaching the details of the change (including materials proving the change) and the documents specified in each subparagraph of Paragraph 1 that differ as a result of the change. 1. Change of location of the manufacturing facility (limited to the place where medical device manufacturing operations are performed) 2. Other matters that affect the results of the review of the manufacturing and quality control system, as determined and notified by the Minister of Food and Drug Safety ③ In order to continue maintaining the recognition of conformity even after the expiration of the validity period pursuant to Article 28, Paragraph 4 of the Act, one must submit to the Minister of Food and Drug Safety, etc., an application for recognition of conformity or change of conformity for a medical device manufacturing and quality management system in Form No. 38-4 (including an application in electronic form), with the documents specified in each subparagraph of Paragraph 1 attached. ④ When the Minister of Food and Drug Safety, etc. receives an application pursuant to the provisions of paragraphs 1 through 3, they may conduct an on-site inspection to evaluate whether the contents of the application are in accordance with the medical device manufacturing and quality control system. ⑤ The Minister of Food and Drug Safety shall review the application pursuant to the provisions of paragraphs 1 through 3 and, if it is found to be compliant with the medical device manufacturing and quality management system, issue a certificate of conformity for the medical device manufacturing and quality management system. ⑥ Any person who intends to apply for recognition of conformity, recognition of modified conformity, or extension of the validity period thereof pursuant to the provisions of paragraphs 1 through 3 shall pay a fee determined by the head of the quality control review agency upon approval by the Minister of Food and Drug Safety (including field investigation expenses determined and notified by the Minister of Food and Drug Safety if field investigation is necessary ). ⑦ In addition to the matters prescribed in paragraphs 1 through 6, detailed matters necessary for the procedures or methods of recognition of conformity, the issuance of certificates of conformity, etc. shall be determined and announced by the Minister of Food and Drug Safety . [This Article was newly added on July 1, 2026.] [Formerly Article 48 was moved to Article 48-3 <July 1, 2026>] Article 48-2 (Inspectors of Manufacturing and Quality Management Systems, etc.) ① Persons whom the Minister of Food and Drug Safety may appoint as inspectors of medical device manufacturing and quality management systems (hereinafter referred to as “inspectors” in this Article) pursuant to Article 28-2, Paragraph 1 of the Act shall be persons who have completed specialized training regarding the inspection of medical device manufacturing and quality management systems prescribed and notified by the Minister of Food and Drug Safety . ② The content of education and training conducted periodically by the Minister of Food and Drug Safety pursuant to Article 28-2, paragraph 2 of the Act shall be matters concerning laws, systems, and technologies, etc. necessary for the examination of the medical device manufacturing and quality management system, and the detailed contents thereof shall be determined and announced by the Minister of Food and Drug Safety . ③ Pursuant to Article 28-2, Paragraph 3 of the Act, the Minister of Food and Drug Safety shall entrust the education and training of examiners to the relevant institutions or organizations listed in the following subparagraphs. In this case, the Minister of Food and Drug Safety shall publicly announce the entrusted institutions and the details of the entrusted work. 1. Informant 2. Professional personnel training institutions pursuant to Article 15 of the "Regulatory Science and Innovation Act on the Safety and Commercialization Support of Food, Drugs, etc." 3. Other institutions or organizations established to perform medical device-related duties ④ Pursuant to Article 28-2, Paragraph 4 of the Act, the Minister of Food and Drug Safety may take necessary measures, such as revoking the appointment or ordering correction, if an examiner falls under any of the following subparagraphs. However, in cases falling under subparagraphs 1 through 3, the appointment must be revoked. 1. In cases where an examiner is appointed through false or other fraudulent means 2. Cases where conformity assessment work is performed through false or other fraudulent means 3. Cases where undue money or valuables are exchanged in connection with the suitability assessment review 4. In cases where information obtained during the conformity assessment process is disclosed or used for purposes other than business purposes. 5. Failure to complete education or training pursuant to Article 28-2, Paragraph 2 of the Act without justifiable grounds ⑤ The detailed criteria for the cancellation of appointment, etc. of examiners pursuant to Paragraph 4 are as set forth in Attached Table 7-2. ⑥ In addition to the matters prescribed in paragraphs 1 through 5, matters necessary for the management and operation of examiners, etc. shall be determined and announced by the Minister of Food and Drug Safety . [This Article Newly Inserted July 1, 2026] Article 48-3 (Designation of Agency for Accreditation of Conformity, etc.) ① Any person who intends to be designated as a quality control review agency to perform the duties of accreditation of conformity pursuant to Article 28-3, Paragraph 1 of the Act shall submit to the Minister of Food and Drug Safety an application for designation or renewal of designation as a medical device quality control review agency (including an application in the form of an electronic document) in the Form of Annex 39, attaching the documents specified in each of the following subparagraphs (including electronic documents). <Amended July 1, 2026> 1. Status of Organization and Personnel of Quality Control Audit Institutions (referring to documents proving the qualifications and experience of personnel) 2. Documents regarding management and operation standards, etc. for quality audit work 3. Documents regarding the scope of quality audit work ② When the Minister of Food and Drug Safety designates a quality control review agency based on the results of the review of documents submitted pursuant to Paragraph 1 and the investigation into the actual status of quality review performance capabilities, he/she shall issue a Certificate of Designation for Medical Device Quality Control Review Agency in the form of Annex 40. ③ If a quality control review institution designated pursuant to Paragraph 2 changes any of the matters falling under any of the following subparagraphs, it shall submit to the Minister of Food and Drug Safety, within 30 days from the date the cause for change occurred, an application for change of designation matters of a medical device quality control review institution in the form of Annex 41 (including an application in the form of an electronic document), attaching the designation certificate of the review institution and materials (including electronic documents) that can verify the changes. 1. Representative of the quality control auditing agency 2. Name and location of the quality control auditing body 3. Scope of Quality Audit Services ④ If a quality control review agency intends to renew its designation pursuant to Article 28-3, Paragraph 4 of the Act, it shall submit to the Minister of Food and Drug Safety, at least 180 days prior to the expiration of the validity period, an application for designation or renewal of designation as a medical device quality control review agency in the form of Annex 39 (including an application in electronic form), attached with the documents specified in each subparagraph of Paragraph 1 (including electronic documents). <Newly added July 1, 2026> ⑤ If the Minister of Food and Drug Safety recognizes that the quality control review agency that has applied for renewal pursuant to Paragraph 4 is suitable for performing conformity recognition duties, the Minister shall issue a new Certificate of Designation as a Medical Device Quality Control Review Agency in the form of Annex 40 to the applicant no later than 30 days before the expiration of the validity period. <Newly added July 1, 2026> ⑥ The matters to be observed by a quality control auditing body pursuant to Article 28-3, Paragraph 5 of the Act are as follows: <Amended July 1, 2026> 1. Retain records regarding quality audits, including copies of the Certificate of Conformity and all audit materials related to the issuance of the Certificate of Conformity, for five years. 2. Report the results of the quality inspection in accordance with the procedures prescribed by the Minister of Food and Drug Safety. 3. Comply with the management and operation standards for medical device quality control review institutions prescribed and notified by the Minister of Food and Drug Safety. 4. Maintain the designation requirements for a quality control audit institution pursuant to Article 28-3, Paragraph 2 of the Act. 5. Comply with other matters prescribed and notified by the Minister of Food and Drug Safety, such as matters to be stated by the quality control inspection agency when issuing a certificate of conformity. ⑦ In the event that the Minister of Food and Drug Safety designates, modifies, or renews the designation of a quality control review agency, he/she shall publicize the name, address, representative, and validity period, etc., of the quality control review agency on the internet homepage of the Ministry of Food and Drug Safety, etc. <Newly added July 1, 2026> ⑧ The Minister of Food and Drug Safety may conduct regular and ad hoc inspections of quality control review agencies. <Amended July 1, 2026> ⑨ In addition to the matters prescribed in Paragraphs 1 through 5, detailed matters necessary for the designation procedures, methods, and operation of quality control review agencies shall be determined and announced by the Minister of Food and Drug Safety . <Amended July 1, 2026> [Title amended July 1, 2026] [Moved from Article 48 <July 1, 2026>] Article 48-4 (Cancellation of Recognition of Conformity, etc.) ① The phrase “failure to comply with matters prescribed by the Decree of the Prime Minister” in subparagraph 2 of Article 28-4 of the Act refers to any of the following cases: 1. Cases where medical devices are manufactured or imported due to incorrect record keeping regarding the medical device manufacturing and quality control system 2. Cases where records regarding the medical device manufacturing and quality control system are omitted 3. Cases where there is a concern that the quality of medical devices may be affected due to failure to establish detailed standards or procedures for compliance with the medical device manufacturing and quality management system, or failure to implement established detailed standards or procedures. 4. Cases where corrective and preventive measures pursuant to Article 48, Paragraph 1, Subparagraph 4 are not implemented ② The criteria for cancellation of medical device conformity recognition, etc. pursuant to Article 28-4 of the Act are as set forth in Table 7-3. [This Article Newly Inserted July 1, 2026] Article 49 (Designation and Management Standards for Medical Devices Subject to Traceability Management) ① The medical devices subject to designation under each subparagraph of Article 29 , Paragraph 1 of the Act are as follows: 1. Medical devices inserted into the human body for more than one year that fall under any of the following categories a. Implantable pacemaker B. Implantable pacemaker electrodes C. Mixed-material artificial heart valve D. Biomaterial artificial heart valve E. Non-biomaterial artificial heart valve B. Implantable defibrillator d. Electric implantable drug infusion pump Ah. Other medical devices for which the Minister of Food and Drug Safety determines and notifies the necessity of locating them. 2. Among life-sustaining medical devices, medical devices that may be used in places other than medical institutions, including the following medical devices a. Personal ventilators (limited to those worn at all times) B. Other medical devices for which the Minister of Food and Drug Safety determines and notifies the necessity of locating them. ② When the Minister of Food and Drug Safety authorizes the manufacture or import, etc. of a medical device subject to traceability management under Paragraph 1, he/she shall mark the license with “medical device subject to traceability management.” Article 49-2 (Designation Criteria, etc. for Medical Devices Subject to Long-term Follow-up Investigation) ① The Minister of Food and Drug Safety may designate medical devices subject to long-term follow-up investigation pursuant to Article 29, paragraph 2 of the Act (hereinafter referred to as “medical devices subject to long-term follow-up investigation”) by comprehensively reviewing the matters in each of the following subparagraphs. 1. Whether there is a need to continuously verify the safety of the medical device due to the occurrence of adverse events resulting from the use of the medical device at least once a year. 2. Whether there is a possibility of causing serious side effects leading to death or incurable complications after implantation in the human body 3. Other matters equivalent to Subparagraphs 1 and 2 that the Minister of Food and Drug Safety deems necessary to review the designation of medical devices subject to long-term follow-up investigations. ② If the Minister of Food and Drug Safety designates a medical device subject to long-term follow-up investigation pursuant to Paragraph 1, he/she shall announce that fact . [This Article Newly Inserted Aug. 1, 2025] Article 50 (Matters Concerning Records of Medical Devices Subject to Traceability Management, etc.) ① The matters to be recorded by manufacturers, importers, sellers, lessors, and repairers (hereinafter referred to as “handlers”) handling medical devices subject to traceability management pursuant to Article 30, Paragraph 1 of the Act, and by the founders of medical institutions and physicians, Korean medicine practitioners, dentists, etc. employed at medical institutions (hereinafter referred to as “users”) handling medical devices subject to traceability management, are as follows: <Amended Aug. 1, 2025> 1. Handler's Record Items a. Manufactured/imported quantities and manufacturing/import dates by product name (or item name if no product name exists), model name, and manufacturing unit (applicable only to manufacturers and importers) B. Sales or rental quantity by product name (by item name if there is no product name), model name, and manufacturing unit, date and time of sale or rental, and the trade name and address of the seller, lessor, or medical institution founder (excluding repair businesses) C. Repair date and time by product name (by item name if no product name exists), model name, and manufacturing batch, and the business name and address of the client (applicable only to repair businesses) D. Other matters necessary to prevent the occurrence of hazards to public health and hygiene 2. User's record details a. Name and Resident Registration Number of Patient Using Medical Device Subject to Tracking Management b. The name and manufacturing number of the medical device subject to traceability management, or a substitute thereof C. Date of use of the medical device subject to tracking management D. Name and location of the medical institution using E. Other matters necessary to prevent the occurrence of hazards to public health and hygiene ② Handlers or users of medical devices subject to traceability management shall submit records and data regarding such medical devices to the Minister of Food and Drug Safety in accordance with the classifications in each of the following subparagraphs. In this case, the handler or user may submit them using a computer program prescribed by the Minister of Food and Drug Safety . <Newly added Jun. 15, 2016; Jun. 24, 2021; Aug. 1, 2025> 1. Handler: Prepare records and data regarding medical devices subject to traceability management every month in accordance with the regulations prescribed and notified by the Minister of Food and Drug Safety, and submit them by the end of the following month. 2. User a. In the event of business closure, all usage records must be submitted at the time of business closure notification. B. For medical devices subject to traceability management designated and notified by the Minister of Food and Drug Safety as potentially causing death or posing a serious threat to life , semi-annual data shall be submitted by the end of the month following the end of the semi-annual period. However, if data regarding actual use information (referring to actual use information pursuant to Article 30, Paragraph 3 of the Act; hereinafter the same) has been submitted in accordance with Article 50-2, Paragraphs 2 and 3, data related thereto may not be submitted. C. In the event of a request from the Minister of Food and Drug Safety for the submission of records and data regarding medical devices subject to traceability management, the said data shall be submitted within 10 days from the date of the request. ③ Handlers and users of medical devices subject to traceability management shall ensure that the confidentiality is protected when creating records pursuant to Paragraph 1. ④ Records regarding medical devices subject to traceability management under Paragraph 1 shall be preserved until the time specified in each of the following subparagraphs. 1. When the medical device subject to tracking management can no longer be used, such as when a patient using the device dies. 2. When the reason for preserving previous records has ceased to exist due to the new recording of Paragraph 1, Subparagraph 2, Item (a) or (c) with respect to medical devices subject to traceability management that are not disposable 3. When the need for tracking management ceases and the reason for preserving the relevant record disappears. ⑤ The Minister of Food and Drug Safety may establish and operate standard codes that can be attached to or written on the container, outer casing, or packaging of medical devices subject to traceability management for the efficient management of such devices. ⑥ Detailed matters concerning the records of handlers or users of medical devices subject to traceability management under Paragraphs 1 and 2, and their submission, shall be determined and announced by the Minister of Food and Drug Safety . <Newly added June 24, 2021> Article 50-2 (Methods and Procedures for Collection, Analysis, and Evaluation of Information on Actual Use) ① The head of an information source entrusted with the tasks of collecting, analyzing, and evaluating information on actual use pursuant to Article 44, Paragraph 2 of the Act may recruit users (hereinafter referred to as “participants”) to participate in a long-term tracking survey in order to collect, analyze, and evaluate information on actual use pursuant to Article 30, Paragraph 3 of the Act. ② The data submitted to the head of the information center pursuant to Article 30, Paragraph 3 of the Act by a participant who responded to the recruitment under Paragraph 1 shall be data regarding the actual usage information of each of the following subparagraphs generated during the process of using the medical device subject to the long-term follow-up investigation. 1. Name and Resident Registration Number of Patients Using Medical Devices Subject to Long-Term Follow-up Study 2. The name and serial number of the medical device subject to long-term follow-up investigation, or a substitute thereof 3. Date and details of use of medical devices subject to long-term follow-up study 4. Date and time of occurrence and details of the adverse event 5. Name and location of the medical institution using 6. Other matters notified by the Minister of Food and Drug Safety as deemed necessary for long-term follow-up investigations ③ When a participant submits data regarding actual usage information pursuant to Paragraph 2 to the head of the information source, they may submit it using a computer program prescribed by the Minister of Food and Drug Safety . ④ The head of the information agency shall analyze and evaluate the data regarding actual usage information submitted by participants pursuant to Paragraphs 2 and 3 on a semi-annual basis and report the results to the Minister of Food and Drug Safety by the end of the month following the end of the semi-annual period. However, in any of the following cases, the results of the analysis and evaluation must be reported without delay. 1. In cases where the frequency of adverse events differs significantly from previously reported analysis and evaluation results 2. Where the content of the adverse event falls under any of the subparagraphs of Article 51, Paragraph 1 3. Other cases equivalent to Subparagraphs 1 and 2 where the Minister of Food and Drug Safety recognizes the need to analyze and evaluate adverse events without delay and requests such analysis and evaluation. ⑤ In addition to the matters prescribed in paragraphs 1 through 4, detailed matters concerning the methods and procedures, etc. for the collection, analysis, and evaluation of information on actual use shall be determined and announced by the Minister of Food and Drug Safety . [This Article Newly Inserted Aug. 1, 2025] Article 51 (Reporting of Adverse Effects, etc.) ① Any person who intends to report matters concerning adverse effects of a medical device pursuant to Article 31, Paragraph 1 of the Act shall do so in accordance with the provisions of each of the following subparagraphs and shall preserve the relevant data for two years. 1. In the event of death or life-threatening adverse effects, within 7 days. In this case, additional details must be reported within 8 days from the date of the initial report. 2. In the event that side effects specified in each of the following subparagraphs occur or adverse events occur, within 15 days a. Cases where hospitalization or extension of the hospitalization period is necessary b. Cases where recovery is impossible or results in severe disability or functional decline c. Cases resulting in congenital malformations or abnormalities ② When the Minister of Food and Drug Safety notifies the operator of a medical institution of side effects and recall plans, etc. pursuant to Article 31, Paragraph 4 of the Act , he/she shall do so by means such as visit, mail, telephone, e-mail, or fax. ③ An operator of a medical institution who has notified a patient of side effects and recall plans, etc., pursuant to Article 31, Paragraph 5 of the Act , shall complete the Patient Notification Confirmation Form (Form No. 42 ) and submit it to the Minister of Food and Drug Safety. <Amended June 14, 2018> ④ In addition to the matters prescribed in paragraphs 1 through 3, detailed matters concerning the reporting and management of adverse effects shall be determined and announced by the Minister of Food and Drug Safety . Article 52 (Standards and Procedures for Recall of Hazardous Medical Devices, etc.) ① Pursuant to Article 31, Paragraph 2 of the Act , medical device repairers, sellers, and lessors shall, if a medical device being repaired, sold, or leased is suspected to be a medical device that causes or poses a risk of causing harm to the human body (hereinafter referred to as “medical device subject to recall”), immediately stop the repair, sale, or lease of the relevant medical device and notify the manufacturer or importer of the relevant medical device (hereinafter referred to as the “obligated party for recall”) of such fact. ② Pursuant to Article 31, Paragraph 2 of the Act , the person obligated to recall shall verify whether the medical device suspected of being subject to recall among the medical devices he or she has manufactured or imported and sold or leased, and the medical device notified pursuant to Paragraph 1, fall under any of the following subparagraphs. 1. Medical devices that cause serious side effects that cannot be cured by their use, lead to death, or pose a risk of causing such side effects or death. 2. Medical devices that cause temporary or medical side effects that can be cured by the use of the medical device, or that may cause such side effects. 3. Medical devices that rarely cause side effects from use but have issues with safety and efficacy due to non-compliance with the standards and specifications under Article 19 of the Act . ③ If, as a result of the verification under Paragraph 2, the recall obligor finds that the relevant medical device falls under any of the subparagraphs of Paragraph 2, the recall obligor shall immediately take measures such as suspending the sale of the relevant medical device, and within 5 days from the date of verification under Paragraph 2, submit a recall plan in the form of Annex 43 (including a plan in the form of an electronic document) to the head of the regional Food and Drug Administration having jurisdiction over the location of the recall obligor. In this case, the recall obligor may submit the recall plan using a computer program prescribed by the Minister of Food and Drug Safety . <Amended Dec. 31, 2018; Dec. 23, 2019> 1. Deleted <December 31, 2018> 2. Deleted <December 31, 2018> ④ Notwithstanding Paragraph 3, if the recall obligor determines that it is difficult to submit a recall plan within 5 days for a medical device under Paragraph 2, Subparagraph 3, they may request a one-time extension of the submission deadline for up to 10 days from the head of the regional Food and Drug Administration, stating the reasons therefor. <Newly added Dec. 31, 2018> ⑤ When a person obligated to recall submits a recall plan pursuant to Paragraph 3, they shall attach the documents specified in each of the following subparagraphs (including documents in the form of electronic documents). <Amended Dec. 31, 2018; Jun. 12, 2019; Dec. 23, 2019> 1. Copies of manufacturing and import records for the relevant item, and records of sales volume and date by sales outlet, rental volume and date by lessor, etc. 2. Notification of recovery plan to be notified pursuant to Article 53 , Paragraph 3 3. Deleted <December 31, 2018> 4. In cases where the medical device subject to recall falls under Paragraph 2, Item 1, information regarding the founder of the medical institution, such as the name, address, and name of the founder of the medical institution that used the said medical device ⑥ When a recall obligor prepares a recall plan pursuant to Paragraph 3, they shall determine the scheduled completion date of the recall in accordance with the classifications in each of the following subparagraphs. However, if it is determined that it is difficult to carry out the recall within that period, the date may be set beyond the recall period provided that the reasons are stated. <Amended Dec. 31, 2018> 1. Medical devices under Paragraph 2, Subparagraph 1: Within 15 days from the date the recall began. 2. Medical devices under Paragraph 2, Subparagraphs 2 and 3: Within 30 days from the date the recall began. ⑦ If the head of the Regional Food and Drug Administration determines that the recall plan submitted pursuant to paragraphs 3 through 5 is insufficient, he/she may order the relevant recall obligor to supplement the recall plan. <Amended Dec. 31, 2018> ⑧ In addition to the matters prescribed in Paragraphs 1 through 7, detailed matters concerning the assessment of the degree of risk of medical devices, the preparation and supplementation of recall plans, etc. shall be determined and announced by the Minister of Food and Drug Safety . <Newly added Dec. 31, 2018> Article 53 (Publication of Recall Plan, etc.) ① When a recall obligor receives an order to publish a recall plan from the head of a regional Food and Drug Safety Administration pursuant to Article 31, Paragraph 3 of the Act, the recall obligor shall publish the recall plan in accordance with the classifications in each of the following subparagraphs. 1. Medical devices under Article 52, Paragraph 2, Subparagraph 1 : Broadcasting pursuant to Article 2, Subparagraph 1 of the Broadcasting Act ; a general daily newspaper with a nationwide distribution area registered pursuant to Article 9, Paragraph 1 of the Act on the Promotion of Newspapers, etc. [referring to the entire edition of the relevant newspaper printed and distributed on the day]; or mass media of an equivalent or higher level (including media recognized by the Minister of Food and Drug Safety in consideration of the purpose of use, method of use, etc. of the medical device subject to recall). 2. Medical devices under Article 52, Paragraph 2, Subparagraph 2 : Announcement in a specialized journal of medicine or biomedical engineering or a medium of equivalent or higher caliber 3. Medical devices under Article 52, Paragraph 2, Subparagraph 3 : Public notice on the internet homepage of the recall obligor or on media of an equivalent or higher level. ② The head of the regional Food and Drug Safety Administration may post the name of the recall obligor, product name, manufacturing number, manufacturing date, expiration date/validity date, and reason for recall on the internet homepage. ③ The recall obligor shall notify repairers, sellers, lessors, or founders of medical institutions (hereinafter referred to as “handlers of medical devices subject to recall”) handling the medical devices subject to recall of the recall plan through visits, mail, telephone, telegram, email, fax, or public announcement through media, and shall retain materials proving such notification for two years from the end of the recall. ④ A handler of a medical device subject to recall who has been notified of the recall plan pursuant to Paragraph 3 shall take measures such as returning the medical device subject to recall, and shall complete the recall confirmation form in Annex 44 and send it to the person obligated to recall the medical device subject to recall. ⑤ Deleted <June 14, 2018> ⑥ The Minister of Food and Drug Safety may configure and operate a computer program that provides information, etc. on medical devices subject to recall, and may recommend that handlers of medical devices subject to recall install it. ⑦ In addition to the matters prescribed in Paragraphs 1 through 6, detailed matters concerning the method of public announcement, the period of public announcement, and notification, etc. of the recall plan shall be determined and announced by the Minister of Food and Drug Safety . <Newly added Dec. 31, 2018> [The amended provisions of Article 53, Paragraph 5 of Prime Minister's Decree No. 1181 (July 29, 2015) shall remain in effect until May 10, 2016, pursuant to the provisions of Article 2 of the Addenda to the same Act] Article 54 (Disposal, etc. of Medical Devices Subject to Recall) ① The person obligated to recall shall dispose of the medical devices recalled or returned, or take other measures to prevent harm, and shall prepare a Recall Evaluation Report in accordance with Form No. 45 (including a report in the form of an electronic document) regarding such disposal. <Amended Dec. 23, 2019> ② When a person obligated to recall disposes of medical devices subject to recall pursuant to Paragraph 1, they shall submit an application for disposal in the form of Annex 46 (including an application in electronic form) to the competent Special Self-Governing City Mayor, Special Self-Governing Province Governor, Mayor, County Head, or District Head, attaching the documents specified in each of the following subparagraphs (including documents in electronic form); dispose of the devices in accordance with environmental laws and regulations under the supervision of relevant public officials of the competent Special Self-Governing City, Special Self-Governing Province, City, County, or District; and prepare a confirmation of disposal in the form of Annex 47 (including a confirmation in electronic form) and retain it for two years. <Amended Dec. 23, 2019; Apr. 13, 2020; Dec. 31, 2020; Jun. 24, 2021> 1. Copy of the recovery plan in Form No. 43 2. Copy of the Retrieval Confirmation Form (Attachment No. 44) ③ When the recall is completed, the recall obligor shall submit a Recall Completion Report in accordance with Form No. 48 (including reports in electronic form) to the head of the Regional Food and Drug Administration having jurisdiction over the recall obligor's location, attaching the documents specified in each of the following subparagraphs (including documents in electronic form). In this case, the recall obligor may submit the said Recall Completion Report using a computer program prescribed by the Minister of Food and Drug Safety . <Amended Dec. 23, 2019> 1. Copy of the Retrieval Confirmation Form (Attachment No. 44) 2. Copy of the Recovery Evaluation Report in Form No. 45 3. Copy of the Confirmation of Disposal Form No. 47 (only applicable if disposed of) ④ Upon receiving the report on the completion of the recall pursuant to Paragraph 3, the head of the regional Food and Drug Safety Administration shall take measures as stipulated in each of the following subparagraphs. 1. If it is determined that the recall of the medical device subject to recall has been properly carried out in accordance with the recall plan, confirm that the recall has been completed and notify the person obligated to recall of this fact in writing. 2. If it is determined that the recall has not been carried out effectively, order the recall obligor to take additional measures necessary for the recall. ⑤ The Mayor of a Special Self-Governing City, the Governor of a Special Self-Governing Province, a Mayor, a County Governor, or a District Head shall notify the Head of the Regional Food and Drug Administration having jurisdiction over the location of the person obligated to recall of the fact that disposal pursuant to Paragraph 2 has been completed. <Amended June 24, 2021> ⑥ In addition to the matters prescribed in Paragraphs 1 through 5, detailed matters concerning measures such as disposal, preparation of recall evaluation reports, and reporting and measures regarding the completion of recall shall be determined and announced by the Minister of Food and Drug Safety . <Newly added Dec. 31, 2018> Article 54-2 (Reporting of Medical Device Supply Details) ① Medical devices for which supply details must be reported pursuant to Article 31-2, Paragraph 1 of the Act are as follows: <Newly added Jan. 16, 2024; Jul. 8, 2024> 1. Implantable medical devices pursuant to Article 12-5, Paragraph 1 of the Enforcement Decree (limited to Class 3 and Class 4 medical devices) 2. Medical devices that are medical supplies eligible for medical benefits pursuant to Article 8, Paragraph 2 of the "Rules on Standards for Medical Benefits under the National Health Insurance" (limited to cases supplied to medical institutions) 3. Other medical devices determined by the Minister of Food and Drug Safety in consultation with the Minister of Health and Welfare ② Any person who intends to report details of medical device supply pursuant to Article 31-2, Paragraph 1 of the Act shall submit a Report on Medical Device Supply Details (including electronic documents) in the form of Annex 48-2 to the Minister of Food and Drug Safety through the Integrated Medical Device Information System by the end of the month following the month in which the medical device was supplied. <Amended Jan. 16, 2024; Apr. 1, 2025> ③ Notwithstanding Paragraphs 1 and 2, the Minister of Food and Drug Safety may exclude from the reporting requirements for medical device supply details if deemed necessary for the improvement of public health or the achievement of public interest purposes. <Newly added Jan. 16, 2024> [This Article was newly added on Oct. 22, 2019.] [Formerly Article 54-2 was moved to Article 54-3 <Oct. 22, 2019>] Article 54-3 (Operation of Medical Device Integrated Information System, etc.) ① The information that medical device manufacturers and importers must register (including registration of changes; hereinafter the same applies) in the Medical Device Integrated Information System pursuant to Article 31-3, paragraph 2 of the Act (hereinafter referred to as “medical device information, etc.”) is as follows: 1. Information on Medical Device Standard Codes 2. Information regarding medical device products (including information regarding the authorization, certification, or notification of medical devices) 3. Information on medical device manufacturers and importers (including foreign manufacturers) ② The standards to be observed by medical device manufacturers and importers when registering and managing medical device information, etc. pursuant to Article 31-3, Paragraph 3 of the Act (hereinafter referred to as the “Medical Device Integrated Information Management Standards”) are as set forth in Annex 7-4. <Amended July 1, 2026> ③ The Minister of Food and Drug Safety may have the head of the Integrated Medical Device Information Center perform the following tasks: 1. Training for medical device manufacturers and importers on the registration of medical device information, etc., and methods for using the Integrated Medical Device Information System. 2. Provision of information such as medical device approval, certification, and notification details, and precautions for use ④ Detailed matters necessary for the operation and management of the Medical Device Integrated Information System, such as the subjects and scope of medical device information, etc. under Paragraph 1 and the method of providing medical device information, etc., shall be determined and announced by the Minister of Food and Drug Safety . [This Article was newly added on Dec. 31, 2018.] [Moved from Article 54-2; former Article 54-3 moved to Article 54-4 <Oct. 22, 2019>] Article 54-4 (Standards, Subjects, and Procedures for Reporting Foreign Matter) ① The scope of foreign matter (hereinafter referred to as “foreign matter”) that a medical device handler must report to the Minister of Food and Drug Safety pursuant to Article 31-5, Paragraph 1 of the Act is as follows: 1. Materials that are not normal materials of a medical device, such as metals, plastics, or fragments generated during the manufacturing process, and which may cause direct or indirect harm or injury to the human body during the use of the medical device. 2. Substances related to living organisms, such as insects and their eggs, parasites and their eggs, and animal carcasses, which may cause direct or indirect harm or injury to the human body during the use of medical devices. 3. Other substances equivalent to those in Subparagraph 1 or 2 that are designated and notified by the Minister of Food and Drug Safety to prevent harm arising from the use of medical devices. ② A medical device handler who intends to report the discovery of a foreign substance pursuant to Article 31-5, Paragraph 1 of the Act shall submit a Foreign Substance Discovery Report in the form of Annex 48-3 (including a report in the form of an electronic document) with attached evidentiary materials to the Minister of Food and Drug Safety. <Amended Oct. 22, 2019> ③ When the Minister of Food and Drug Safety receives a report on the discovery of a foreign substance pursuant to Paragraph 2, he/she shall notify the head of the regional Food and Drug Safety Administration of the details without delay. ④ When the head of the regional Food and Drug Safety Administration receives a notification pursuant to Paragraph 3, he/she shall conduct an investigation into the cause of foreign matter contamination without delay pursuant to Article 31-5 Paragraph 2 of the Act . In this case, the head of the regional Food and Drug Safety Administration may request necessary cooperation from relevant central administrative agencies, local governments, public institutions, corporations, organizations, or experts, etc., for the investigation into the cause of foreign matter contamination. ⑤ When the head of a regional Food and Drug Administration has completed an investigation into the cause of foreign matter contamination pursuant to Paragraph 4, he/she shall report the results of the investigation and the plan for action to the Minister of Food and Drug Safety, and shall take measures under each of the following subparagraphs without delay, after comprehensively considering the contents of the investigation results, the safety of the medical device, and the necessity of preventing harm. <Amended Jan. 21, 2022; Mar. 8, 2024> 1. Inspection order pursuant to Article 33 of the Act 2. An order to suspend sales, recall, dispose of, or take other measures or make public announcements pursuant to Article 34 of the Act 3. Order for necessary measures, such as suspension of use or repair, pursuant to Article 35 of the Act ⑥ In addition to the matters prescribed in paragraphs 1 through 5, the Minister of Food and Drug Safety shall determine and announce the detailed matters necessary for the procedures and methods for reporting the discovery of foreign substances, the procedures and methods for investigating foreign substance contamination, and related measures . [This Article newly inserted June 12, 2019] [Moved from Article 54-3 <October 22, 2019>] Article 54-5 (Publication of Facts of Discovery of Foreign Matter, etc.) ①If the investigation results reported pursuant to Article 54-4, Paragraph 5 fall under any of the following subparagraphs, the Minister of Food and Drug Safety may publish the fact of discovery of foreign matter, the results of the investigation, and the plan for action on the internet homepage of the Ministry of Food and Drug Safety pursuant to Article 31-5, Paragraph 3 of the Act. 1. Cases where it is recognized that harm to public health has occurred or there is a significant risk of such harm due to the use of a medical device contaminated with foreign matter 2. Cases where it is necessary to report the discovery of a foreign substance without delay to prevent harm to public health, such as when the foreign substance is found in a location where it may come into direct or indirect contact with the human body. ② In addition to the matters prescribed in Paragraph 1, detailed matters necessary for the public announcement of the discovery of foreign substances, etc. shall be determined and announced by the Minister of Food and Drug Safety . [This Article Newly Inserted Mar. 8, 2024] Article 55 (Collection, etc.) ①In cases where medical devices are collected pursuant to Article 32, Paragraph 1, Subparagraph 2 of the Act , a collection certificate in the form of Annex 49 shall be issued to the person to whom they are collected. ② If specialized inspection is required for medical devices collected pursuant to Paragraph 1, the inspection may be requested from a medical device testing and inspection agency or a non-clinical trial institution. <Amended May 1, 2017> Article 56 (Medical Device Inspector Certificate) The identification certificate proving the status of a relevant public official pursuant to Article 32 , Paragraph 2 of the Act ( including cases where it is applied mutatis mutandis pursuant to the latter part of Article 34, Paragraph 2 of the Act) is in the form of Annex 50 . Article 56-2 (On-site Inspection Procedures and Suspension of Import, etc.) ① When the Minister of Food and Drug Safety conducts an on-site inspection pursuant to the part other than each subparagraph of Article 32-2, Paragraph 1 of the Act (hereinafter referred to as “on-site inspection”), he/she shall notify the medical device manufacturer, medical device importer, manager of the overseas manufacturing facility, or the government of the exporting country of the details of the on-site inspection in writing (including electronic documents) 20 days prior to the date of the on-site inspection. ② When the Minister of Food and Drug Safety has completed an on-site inspection pursuant to Article 32-2, Paragraph 2 of the Act (including cases where the on-site inspection was refused), he/she shall notify the head of the regional Food and Drug Safety Administration of the results of the investigation and opinions on necessary measures within 30 days from the date the inspection was completed. ③ Upon receiving notification of the results of an on-site inspection pursuant to Paragraph 2, the Head of a Regional Food and Drug Administration may take any of the following measures in accordance with Article 32-2, Paragraph 2 of the Act . In such cases, when taking such measures, matters concerning the details of the measures, the reasons for the measures, and the date of the measures, etc., shall be notified in writing (including electronic documents) to the medical device manufacturer, the medical device importer, the manager of the overseas manufacturing facility, or the government of the exporting country. <Amended Jan. 21, 2022> 1. Suspension of imports 2. An order or public announcement regarding the suspension of sales, recall, disposal, or other treatment pursuant to Article 34 of the Act 3. Order for necessary measures, such as suspension of use or repair, pursuant to Article 35 of the Act ④ When the head of the regional Food and Drug Safety Administration lifts measures such as suspension of import pursuant to Article 32-2, Paragraph 3 of the Act , he/she shall notify the medical device manufacturer, medical device importer, manager of the overseas manufacturing facility, or the government of the exporting country of matters concerning the details of the lifting, the reason for the lifting, and the date of the lifting, etc., in writing (including electronic documents). [This Article Newly Inserted June 12, 2019] Article 57 (Order for Suspension of Sales, Recall, Disposal, and Public Announcement, etc.) ①Pursuant to Article 34, Paragraph 3 of the Act , the head of a regional Food and Drug Administration or the mayor of a special self-governing city, the governor of a special self-governing province, a mayor, a county head, or a district head may, after undergoing a verification procedure to determine whether a medical device falling under any of the subparagraphs of Article 34 , Paragraph 1 of the Act is a medical device falling under any of the subparagraphs of Article 52 , Paragraph 2, order a person obligated to recall to suspend sales, recall, dispose of, or take other measures regarding the relevant medical device and may have the fact publicly announced. <Amended Dec. 31, 2018; Apr. 13, 2020; Jan. 21, 2022> ② Paragraphs 3 through 8 of Article 52 , Article 53 , and Article 54 shall apply mutatis mutandis to the methods of suspension of sale, recall, disposal, and public announcement pursuant to Paragraph 1. <Amended Dec. 31, 2018; Jan. 21, 2022> [Title amended Jan. 21, 2022] Article 58 (Standards for Administrative Dispositions) ①The standards for administrative dispositions pursuant to Article 36 of the Act are as set forth in Attached Table 8 . ② The standards for administrative dispositions pursuant to Article 37 of the Act are as set forth in Attached Table 9 . Article 59 (Qualifications and Scope of Duties of Medical Device Inspectors) ① Medical device inspectors pursuant to Article 40 of the Act shall be appointed by the Minister of Health and Welfare, the Minister of Food and Drug Safety, the heads of regional Food and Drug Safety offices, and the mayors of special self-governing cities, governors of special self-governing provinces, mayors, county heads, and district heads from among their affiliated public officials falling under any of the following subparagraphs. <Amended June 24, 2021> 1. A person who has graduated from a department of medicine, pharmacy, dentistry, Korean medicine, veterinary medicine, biomedical engineering, or science and engineering at a junior college or university, or a person with qualifications equivalent to or higher than that. 2. Persons with at least one year of experience in health and medical administration ② The scope of duties of medical device inspectors shall be in accordance with the classifications in each of the following subparagraphs. <Amended May 1, 2017; December 31, 2018; June 24, 2021> 1. Medical device inspectors belonging to the Ministry of Health and Welfare, the Ministry of Food and Drug Safety, or regional Food and Drug Safety Administrations shall perform the duties set forth in the following subparagraphs. However, if a nationwide investigation is required or there is a need for a systematic investigation, they may perform the duties set forth in subparagraphs 2 and 3. a. Post-market surveillance of medical device manufacturers and importers, and collection and inspection of medical devices B. Post-inspection of technical document review institutions, clinical trial institutions, non-clinical trial institutions, quality control review institutions, and other places where medical devices are handled in the course of business, and collection and inspection of medical devices (limited to medical device inspectors belonging to the Ministry of Food and Drug Safety or regional Food and Drug Safety offices) 2. Deleted <June 24, 2021> 3. Medical device inspectors belonging to Special Self-Governing Cities, Special Self-Governing Provinces, Cities, Counties, and Districts conduct post-monitoring of medical device repairers, sellers, and lessors, as well as founders of medical institutions under the Medical Service Act and founders of animal hospitals under the Veterinary Medical Service Act. Article 59-2 (Qualifications, etc. of Consumer Medical Device Inspectors) ① The head of a regional Food and Drug Administration, the mayor of a special self-governing city, the governor of a special self-governing province, a mayor, a county head, or a district headmay appoint a person falling under any of the following subparagraphs as a consumer medical device inspector pursuant to Article 40-2, Paragraph 1 of the Act. < Amended Apr. 13, 2020; Jun. 24, 2021> 1. Any person falling under any of the subparagraphs of Article 59, Paragraph 1 2. Persons with at least one year of experience working at a medical device-related company or institution. 3. Persons who have completed the educational course prescribed and notified by the Minister of Food and Drug Safety 4. A member or employee of a medical device-related association or organization recommended by the head of said association or organization 5. Among the officers and employees of a consumer organization registered pursuant to Article 29 of the Framework Act on Consumers, those recommended by the head of the said organization ② The term of office for a consumer medical device inspector appointed pursuant to Paragraph 1 (hereinafter referred to as “consumer medical device inspector”) shall be two years, and they may be reappointed. ③ The “matters prescribed by the Decree of the Prime Minister” in Article 40-2, Paragraph 2, Subparagraph 3 of the Act shall be as follows: 1. Support for the disposal, sealing, or embedding of goods by relevant public officials pursuant to Article 34, Paragraph 2 of the Act 2. Reporting and Information Provision Regarding Violators of Medical Device-Related Laws 3. Other tasks such as promotion for the safe management of medical devices ④ The head of a regional Food and Drug Administration, the mayor of a special self-governing city, the governor of a special self-governing province, a mayor, a county head, or a district head who has appointed consumer medical device inspectors pursuant to Paragraph 1 shall conduct training for consumer medical device inspectors on laws and regulations related to medical devices at least once a year, in accordance with Article 40-2, Paragraph 4 of the Act . < Amended Apr. 13, 2020; Jun. 24, 2021> ⑤ The approval for exclusive entry of a consumer medical device inspector to a sales or rental business establishment pursuant to Article 40-2, Paragraph 7 of the Act and the identification card indicating the identity of the consumer medical device inspector are respectively Form No. 50-2 and Form No. 50-3 . ⑥ In addition to the matters prescribed in paragraphs 1 through 5, the Minister of Food and Drug Safety shall determine and announce the detailed matters necessary for the operation of the Consumer Medical Device Inspectorate . [This Article Newly Inserted June 15, 2016] Article 59-3 (Designation of Medical Device Safety Information Monitoring Center) ① Any person who intends to be designated as a medical device safety information (hereinafter referred to as “medical device safety information”) monitoring center (hereinafter referred to as “monitoring center”) pursuant to Article43, Paragraph 3 of the Act and Paragraph 1 , Item 7 of the same Act shall submit to the Minister of Food and Drug Safety an application for designation of a monitoring center in the form of Annex 50-4 (including an application in the form of an electronic document) with attached data (including electronic documents) regarding the matters in each of the following items. 1. Monitoring Center's Business Implementation Plan 2. Expenses required for the implementation of the Monitoring Center's projects 3. Facilities and personnel for the implementation of the Monitoring Center's projects 4. Performance records regarding the collection and analysis of medical device safety information within the last 3 years (submit only if applicable) 5. Other matters similar to the materials under Subparagraphs 1 through 4 that the Minister of Food and Drug Safety deems necessary and determines. ② When the Minister of Food and Drug Safety receives an application pursuant to Paragraph 1, he/she shall verify the applicant's medical institution establishment permit through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act . However, if the applicant does not consent to such verification, a copy of the relevant document must be attached. ③ The Minister of Food and Drug Safety may conduct an investigation into the actual conditions of facilities and personnel, etc., if deemed necessary to designate a monitoring center pursuant to Article 43, Paragraph 3 of the Act . ④ If the Minister of Food and Drug Safety deems it necessary for the efficient operation of the monitoring center, he/she may designate monitoring centers by region or area as determined by the Minister of Food and Drug Safety . ⑤ When the Minister of Food and Drug Safety designates a monitoring center pursuant to Article 43, Paragraph 3 of the Act , he/she shall issue a monitoring center designation certificate in the form of Annex 50-5 to the applicant. ⑥ In addition to the matters prescribed in paragraphs 1 through 5, the Minister of Food and Drug Safety shall determine and announce the necessary detailed matters regarding the procedures and methods for designating monitoring centers . [This Article Newly Inserted Dec. 4, 2020] Article 59-4 (Scope of Business and Operation, etc. of the Monitoring Center) ① The scope of business of the Monitoring Center shall be as follows: 1. Collection of adverse effects or safety information related to medical devices 2. Analysis and Research on Adverse Effects Related to Medical Devices 3. Exchange and cooperation with other medical institutions regarding medical device safety information 4. Education and promotion of medical device safety information 5. Businesses similar to those under Subparagraphs 1 through 4, as determined by the Minister of Food and Drug Safety as necessary ② The Monitoring Center shall, in accordance with the regulations set by the Minister of Food and Drug Safety, report annually to the Minister of Food and Drug Safety, through the information source, matters concerning business operation plans, expense management plans, and results of business execution and expense management. ③ The Monitoring Center shall notify the Information Source of matters concerning adverse effects or safety information, etc. related to medical devices every month, as determined by the Minister of Food and Drug Safety, in order to support the medical device safety information project . However, in the case of matters concerning information, etc. that have a significant impact on public health, it shall notify the Minister of Food and Drug Safety and the Information Source without delay. ④ If the Minister of Food and Drug Safety deems it necessary to ensure the appropriateness of the business operations and expense management of a monitoring center, he/she may order the relevant monitoring center to take measures such as supplementation or improvement. ⑤ If the Minister of Food and Drug Safety deems it necessary for the smooth operation of the monitoring center, he or she may support all or part of the operating expenses within the scope of the budget. ⑥ In addition to the matters prescribed in paragraphs 1 through 5, the Minister of Food and Drug Safety shall determine and announce the necessary detailed matters regarding the business, operation, and supervision of the monitoring center . [This Article Newly Inserted Dec. 4, 2020] Article 60 (Procedures and Methods for Cancellation of Medical Device Manufacturing Permit, etc.) The procedures and methods, etc. for cancellation pursuant to Article 43-2 of the Act shall be in accordance with the Administrative Procedures Act . [Title amended Jan. 21, 2022] Article 61 (Guidance and Supervision of Information Sources) Pursuant to Article 43-3, Paragraph 4 of the Act , the Minister of Food and Drug Safetyshall provide guidance and supervision to information sources regarding the matters set forth in each of the following subparagraphs at least once a year. <Amended June 14, 2018> 1. Matters concerning the appropriateness of the performance of certification or reporting tasks entrusted to the Information Center pursuant to Article 44, Paragraph 2 of the Act 2. Appropriateness of the formulation and execution of the operating budget 3. Other matters deemed necessary by the Minister of Food and Drug Safety [Title amended June 14, 2018] Article 61-2 (Qualifications, Scope of Duties, and Identification, etc. of Causation Investigators) ① Causation investigators pursuant to Article 43-5, Paragraph 1 of the Act (hereinafter referred to as “caution investigators”) shall be appointed or commissioned from among persons falling under any of the following subparagraphs. 1. An employee at the source responsible for investigating and determining the causal relationship of medical device side effects 2. Public health physician pursuant to Article 2 of the Special Measures Act for Health and Medical Care in Rural Areas, etc. 3. Physicians, dentists, doctors of Korean medicine, or nurses pursuant to Article 2, Paragraph 1 of the Medical Service Act 4. Other persons with extensive professional knowledge and experience in the field of medical devices ② The head of the information agency may pay allowances and travel expenses to the commissioned causal relationship investigators within the limits of the budget. ③ The causation investigator may investigate and determine the causal relationship between the medical device and the side effect in any of the following cases: 1. In cases where adverse effects corresponding to Article 51, Paragraph 1 , Subparagraphs 1 and 2 have occurred or are suspected to have occurred during the use of a medical device 2. In cases where multiple unintended signs, symptoms, or diseases occur in patients who used a medical device in a specific region or at a specific time 3. Other cases where the Minister of Food and Drug Safety recognizes the need to investigate and clarify the causal relationship between the medical device and the adverse effect ④ The identification card indicating the authority of the causal relationship investigator pursuant to the latter part of Article 43-5, Paragraph 3 of the Act is as shown in Form No. 50-4 of the Attached Form . ⑤ In the latter part of Article 43-5, Paragraph 3 of the Act , “related documents” refers to documents containing the matters of each of the following subparagraphs. 1. Purpose of investigation, investigation period, scope of investigation, and content of investigation 2. Name and position of the investigator 3. List of Submitted Materials 4. Legal Basis for the Investigation 5. Contents of administrative dispositions, penalties, etc., regarding refusal, obstruction, or evasion of investigation, and legal grounds 6. Matters other than those prescribed in Subparagraphs 1 through 5 that the Minister of Food and Drug Safety deems necessary for the relevant investigation [This Article Newly Inserted June 14, 2018] Article 62 (Medical Devices Excluded from Application of Validity Period) Pursuant to the proviso to Article 49, Paragraph 1 of the Act , the validity period of the manufacturing permit, etc. (hereinafter referred to as “manufacturing permit, etc.”) under the main text of Paragraph 1 of the same Article shall not apply to medical devices falling under any of the following subparagraphs. 1. Medical devices for export that are produced or imported solely for the purpose of export 2. Medical devices for which a conditional manufacturing permit or manufacturing certification has been obtained or a conditional manufacturing notification has been filed pursuant to Article 7 of the Act ( including cases applied mutatis mutandis in Article 15, Paragraph 6 of the Act ) [Wholly Amended Dec. 4, 2020] Article 62-2 (Renewal of Manufacturing Permit, etc.) ① The criteria for the renewal of a manufacturing permit, etc. pursuant to Article 49, Paragraph 3 of the Act are as follows: 1. There shall be no serious issues regarding the safety or efficacy of the medical device. 2. Faithfully submit the materials required for the renewal of manufacturing permits, etc. 3. There must be a record of production or import of the relevant medical device. 4. Must have faithfully complied with relevant medical device laws and regulations. 5. In addition , it shall comply with standards similar to those under Subparagraphs 1 through 4, as determined and notified by the Minister of Food and Drug Safety. ② Any person who intends to obtain a renewal of a manufacturing permit, etc. pursuant to Article 49, Paragraph 3 of the Act shall submit to the Minister of Food and Drug Safety [applicable only to the renewal of a manufacturing permit; hereinafter the same applies in this Article] or the Information Center [applicable only to the renewal of manufacturing (import) certification/notification ; hereinafter the same applies in this Article], attaching the following materials (including electronic documents), to the application for renewal of a manufacturing (import) permit in Form No. 51 , the application for renewal of a manufacturing (import) certification in Form No. 51-2, or the report for renewal of a manufacturing (import) notification in Form No. 51-3 (including applications and reports in electronic form) no later than 180 days before the expiration date of the validity period of the manufacturing permit, etc. 1. Original manufacturing (import) permit or certificate for the relevant medical device 2. Data proving that the safety and efficacy of the medical device were maintained during the previous validity period 3. Data regarding the production or import performance of the relevant medical device during the previous validity period 4. Other materials similar to those under Subparagraphs 1 through 3 that the Minister of Food and Drug Safety deems necessary for the renewal of manufacturing permits, etc. ③ In accordance with Article 49, Paragraph 3 of the Act , if the Minister of Food and Drug Safety or the Information Center deems it necessary for the renewal of manufacturing permits, etc., they may conduct an investigation into the actual conditions of manufacturers or importers or request relevant agencies, organizations, experts, etc., to submit data or opinions. ④ When the Minister of Food and Drug Safety or the Information Agency renews a manufacturing permit, etc. pursuant to Article 49, Paragraph 3 of the Act , they shall issue a new manufacturing (import) permit in the form of Annex No. 4 or a manufacturing (import) certificate in the form of Annex No. 6 ; in the case of a manufacturing (import) report, they shall notify the fact of such renewal in writing. In this case, the validity period of the renewed manufacturing permit, etc. shall commence on the day following the expiration of the previous validity period. ⑤ If the Minister of Food and Drug Safety or the Information Agency deems it necessary for the renewal of a manufacturing permit, etc. pursuant to Article 49, Paragraph 3 of the Act , they may notify the manufacturer or importer in advance of the fact that they cannot receive a renewal if they do not renew the manufacturing permit, etc. by the expiration of the validity period, and matters concerning the renewal procedure. ⑥ In the proviso to Article 49, Paragraph 6 of the Act , “unavoidable circumstances prescribed by Decree of the Prime Minister” refers to any of the following reasons: <Amended Jan. 21, 2022> 1. Cases where it is recognized that normal manufacturing operations could not proceed due to the failure to supply raw materials for the manufacture of medical devices 2. Cases where it is recognized that the exporting country has suspended the export of medical devices or that normal import procedures could not be carried out due to the political or economic situation of the exporting country 3. Cases where it is recognized that there was no demand for a rare medical device applicable to a small number of patients, etc. 4. Other grounds similar to those under Subparagraphs 1 through 3, as determined and notified by the Minister of Food and Drug Safety ⑦ In addition to the matters prescribed in paragraphs 1 through 6, the Minister of Food and Drug Safety shall determine and announce the necessary detailed matters regarding the renewal procedures and methods, etc. of manufacturing permits , etc. [This Article Newly Inserted Dec. 4, 2020] Article 63 (Reissuance of Licenses, etc.) ① Any person who intends to obtain a reissuance of a license, certificate, or notification certificate due to reasons recognized by the Minister of Food and Drug Safety, such as when the license, certificate, or notification certificate has been lost, rendered unusable, or has changed in its contents, shall submit an application in the form of Annex 52 (including an application in the form of an electronic document) to the person designated under each of the following subparagraphs, attaching the license, certificate, or notification certificate (excluding cases where the license, certificate, or notification certificate has been lost). <Amended June 14, 2018; June 24, 2021> 1. In the case of a medical device manufacturer or medical device importer a. Manufacturing/Import Permit: The head of the regional Food and Drug Safety Administration having jurisdiction over the location of the manufacturing or importing establishment B. Manufacturing/Import Permit: Minister of Food and Drug Safety C. Manufacturing/Import Certificate and Manufacturing/Import Declaration: Head of Information Center 2. Deleted <June 24, 2021> 3. In the case of medical device repairers, sellers, or lessors: The Mayor of a Special Self-Governing City, Governor of a Special Self-Governing Province, Mayor, County Head, or District Head having jurisdiction over the location of the repair shop, sales shop, or rental shop ② When the Minister of Food and Drug Safety, the head of a regional Food and Drug Safety Administration, the head of an information agency, or the mayor of a special self-governing city, the governor of a special self-governing province, a mayor, a county head, or a district head reissues a permit, certificate, or notification certificate pursuant to Paragraph 1, they shall record the reason for the reissuance in the relevant register. <Amended June 14, 2018; June 24, 2021> ③ Any person who intends to obtain confirmation or certification regarding matters, etc., regarding permission, certification, or notification pursuant to Articles 6 and 15 of the Act shall submit an application form in the format of Annex 53 (including applications in the form of electronic documents, and in the case of foreign languages, translations) to the Minister of Food and Drug Safety, the head of the regional Food and Drug Safety Administration having jurisdiction over the location of the manufacturing or importing establishment, or the head of the Information Center. <Amended June 14, 2018> Article 64 (Cooperation in Provision of Data) A person who has applied for a manufacturing license or manufacturing certification, or has filed a manufacturing report, pursuant to the provisions of Articles 5 through 7, may request the Minister of Food and Drug Safety or the head of the Information Center to provide documents, such as applications and reports related to the manufacturing license, etc., and data regarding the progress of the manufacturing license, etc. procedures to a specialized agency or organization entrusted with the relevant duties pursuant to Article 55 of the Medical Service Act. In this case, the Minister of Food and Drug Safety or the head of the Information Center who receives the request may provide the said documents and data to the said specialized agency or organization. <Amended June 14, 2018> Article 65 (Fees) ① The fees pursuant to Article 50 of the Act shall be as set forth in Attached Table 10. However, fees shall be waived in cases where matters of permission, certification, or notification are changed and the reason for the change is not at fault of the applicant or notifier, such as a change in location due to administrative district reorganization; and fees may be reduced or waived by up to 50 percent in any of the following cases: <Amended Sep. 20, 2024; Apr. 1, 2025> 1. Where a person subject to payment of fees under Item 1 or Item 18 of Annex 10 is a small business owner as defined in Article 2 of the Framework Act on Small Business Owners, and the Minister of Food and Drug Safety deems it necessary to reduce or exempt the fee to alleviate economic burdens or support business stability, etc. 2. In cases where a person subject to the payment of fees under Subparagraphs 3, 4, 5, 6, and 8 through 10 of Annex 10 intends to manufacture or import a medical device that includes a function for transmitting usage information pursuant to Article 23-2, Paragraph 1, Subparagraph 2 of the Act via electronic means such as voice guidance or text enlargement, or where software, devices, etc. for such purposes are provided together. 3. Where the person subject to payment of fees under Subparagraphs 3-2 and 19 of Annex 10 is a small and medium-sized enterprise as defined in Article 2 of the Framework Act on Small and Medium Enterprises, and the Minister of Food and Drug Safety deems it necessary to reduce or exempt the fees for the purpose of alleviating economic burdens or supporting business stability. ② The fees under Paragraph 1 may be paid in accordance with the classifications in each of the following subparagraphs. <Amended June 15, 2016; June 14, 2018; July 1, 2026> 1. Payment to a government agency: Methods such as electronic money or electronic payment using an information and communications network (hereinafter referred to as “electronic money, etc.” in this paragraph) or revenue stamps (cash or a certificate proving payment of cash for matters under the jurisdiction of the Minister of Food and Drug Safety) 2. Payment to local government: Electronic money, etc. or revenue stamps of the relevant local government 3. Payment to the information source: Electronic money, etc. 4. Payment to technical document review agencies and quality control review agencies: Electronic money, etc. ③ If a person who has paid the fee pursuant to Paragraphs 1 and 2 withdraws the application for permission or certification, etc. within the period prescribed and notified by the Minister of Food and Drug Safety, the Minister may refund all or part of the said fee in accordance with the standards prescribed and notified by the Minister of Food and Drug Safety . <Newly added Apr. 13, 2020> Article 66 (Review of Regulations) The Minister of Food and Drug Safety shall review the validity of the matters set forth in each of the following subparagraphs every three years based on the reference dates specified in each of the following subparagraphs (meaning the period prior to the same date as the reference date of every third year) and take measures such as improvements. <Amended Jan. 16, 2024> 1. Compliance requirements for non-clinical testing institutions pursuant to Article 24-3 : January 1, 2024 2. Manufacturer's compliance requirements pursuant to Article 27 : January 1, 2024 3. Importer's Compliance Requirements Pursuant to Article 33 : January 1, 2024 4. Notification, change notification, and notification of closure or suspension of business for medical device repair business pursuant to Article 35 ( excluding Paragraph 4 of the same article ): January 1, 2014 5. Compliance requirements for medical device repair operators pursuant to Article 35, Paragraph 4 : January 1, 2024 6. Notification, change notification, and notification of closure or suspension of business for medical device sales or rental business pursuant to Article 37 : January 1, 2014 ADDENDA <Prime Minister's Decree No. 1181, July 29, 2015> Article 1 (Enforcement Date)  This Rule shall enter into force on July 29, 2015. However, the amended provisions according to the classifications in each of the following subparagraphs shall enter into force on the date specified in each respective subparagraph. 1. Matters relating to the manufacturing and quality control system among the amended provisions of Articles 5, 6, 8, 16, 17, 26, 27, 30, 31, and 33, Annexes 2 through 4, Form No. 3, Form No. 5, and Forms No. 29 through 31: January 29, 2016 2. The part of the amended provision of the proviso to Article 9, Paragraph 2, which excludes medical devices for which the Minister of Food and Drug Safety has determined and notified that the submission of clinical trial data is necessary: ​​January 1, 2016 3. Amended provision of Subparagraph 2 of Annex 1: February 1, 2016 4. The part of the amended provisions of Article 9, Paragraph 3, excluding the subparagraphs, that requires submission to the head of the technical document review institution: January 1, 2017 Article 2 (Effective Date)  The amended provisions of Article 50, Paragraph 2 and Article 53, Paragraph 5 shall remain in effect until May 10, 2016. Article 3 (Applicability to Cooperation in Providing Data)  The amended provisions of Article 64 (including cases where they are applied mutatis mutandis pursuant to Article 34) shall also apply to persons who applied for a manufacturing or import permit or filed a manufacturing or import report prior to the enforcement of this Rule. Article 4 (Special Provisions on the Recognition of Conformity to Manufacturing and Quality Control Standards for In Vitro Diagnostic Pharmaceuticals)  Any person who, at the time of the enforcement date pursuant to Article 1, Subparagraph 1 of the Addenda to the Partial Amendment to the Enforcement Rule of the Medical Devices Act (Prime Minister's Decree No. 1081), has obtained a product license for manufacturing, sales, or importing or has filed a product notification for manufacturing, sales, or importing in accordance with Articles 4 and 5 of the “Rules on the Safety of Pharmaceuticals, etc.” shall be deemed to have been recognized as conforming to the standards of the amended provisions of Annex 2 by the date specified in each of the following subparagraphs. 1. Class II In Vitro Diagnostic Medical Device: November 10, 2016 2. Class 3 and Class 4 In Vitro Diagnostic Medical Devices: November 10, 2015 Article 5 (General Transitional Measures)  Any dispositions, procedures, or other acts performed prior to the enforcement of this Rule in accordance with the former provisions shall be deemed to have been performed in accordance with the corresponding provisions of this Rule. Article 6 (Transitional Measures Concerning the Application of Standards for Administrative Dispositions)  Administrative dispositions for violations committed prior to the enforcement of this Rule shall be governed by the former regulations. Article 7 (Transitional Measures concerning the Introduction of Distribution Quality Management Standards)  Any person who has reported a sales business or rental business at the time of enforcement pursuant to Article 1, subparagraph 2 of the Addendum to the Partial Amendment to the Enforcement Rule of the Medical Devices Act Decree No. 1081 shall ensure that they comply with the standards pursuant to the amended provisions of Annex 6 by December 31, 2015. Article 8 (Transitional Measures Concerning Forms)  ① A manufacturer or importer falling under Article 3 of the Addenda to Act No. 12392, the Act Partially Amending the Medical Devices Act, at the time of the enforcement of the Partial Amendment to the Enforcement Rule of the Medical Devices Act (hereinafter referred to in this Article as “manufacturer or importer under the former provisions”) may use a manufacturing or import business license issued under the former provisions until the said license is reissued pursuant to Paragraph 2. ② Manufacturers or importers under the previous regulations must obtain a reissued manufacturing or import business license in accordance with the revised regulations in Form No. 2 from the head of the regional Food and Drug Administration having jurisdiction over the location of the manufacturing or import business within three months from July 29, 2016. Article 9 (Relationship with Other Laws and Regulations)  ① Where, at the time of the enforcement of this Rule, other laws and regulations refer to the provisions of the former Enforcement Rule of the Medical Devices Act, and there are corresponding provisions in this Rule, the relevant provisions of this Rule shall be deemed to have been referred to in place of the former provisions. ② In cases where other laws and regulations refer to reagents for in vitro diagnostic analyzers or in vitro diagnostic pharmaceuticals pursuant to the "Rules on the Safety of Pharmaceuticals, etc." at the time of the enforcement of the Partial Amendment to the Enforcement Rule of the Medical Devices Act (Prime Minister's Decree No. 1081), it shall be deemed that in vitro diagnostic medical devices pursuant to these Rules have been referred to in their place. ADDENDA <Prime Minister's Decree No. 1284, June 15, 2016> This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 1307, July 29, 2016> This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 1354, Jan. 4, 2017> This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 1389, May 1, 2017> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. Article 2 (Applicability to Fee Amounts) The provisions regarding fee amounts  in the amended provisions of Annex 10, Form No. 3, Form No. 5, Form No. 7, Form No. 8, Form No. 30, and Form No. 31 shall apply from the time of application, notification, or request for examination made after the enforcement of this rule. Article 3 (Transitional Measures Concerning the Designation of Non-Clinical Testing Institutions)  Any institution designated as a medical device testing and inspection institution pursuant to Article 6, Paragraph 2, Subparagraph 4 of the Act on Testing and Inspection in Food and Drug Fields at the time of the enforcement of this Rule, and which is conducting non-clinical trials pursuant to Article 27 of the Act, shall be deemed to have been designated as a non-clinical testing institution pursuant to the amended provisions of Article 24-2. However, such institution must obtain designation as a non-clinical testing institution pursuant to the amended provisions of Article 24-2 within two years from the date of enforcement of this Rule. ADDENDA <Prime Minister's Decree No. 1469, June 14, 2018> Article 1 (Enforcement Date)  This Rule shall enter into force on June 14, 2018. However, the amended provisions of Article 10, Article 14 (paragraphs 4 through 9), Article 32, Article 51, Article 53, the part other than the subparagraphs of Article 63 (paragraph 1), Form No. 11-2 and Form No. 12 shall enter into force on the date of promulgation, and the amended provisions of Article 36 and Article 46-2 shall enter into force on September 14, 2018. Article 2 (Transitional Measures Concerning Minor Repairs of Medical Devices)  In cases where a medical device repair business operator performed minor repairs to a medical device pursuant to the amended provisions of Article 46-2 prior to the enforcement of this Rule, the former provisions shall apply notwithstanding the amended provisions of the proviso to subparagraph 1 of Article 36. ADDENDA <Prime Minister's Decree No. 1512, Dec. 31, 2018> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of Articles 52 through 54 and Article 57 shall enter into force three months after the date of promulgation, and the amended provisions of Article 54-2 and Subparagraph 29-2 of Annex 8 II shall enter into force on the date pursuant to Subparagraph 3 of Article 2, Paragraph 1 of the Addenda. Article 2 (Enforcement Date of the Medical Devices Act)  ① In the main text of Article 1 of the Addenda to Act No. 14330, the Partial Amendment to the Medical Devices Act, the term “date prescribed by Decree of the Prime Minister” refers to the enforcement date according to the classifications in each of the following subparagraphs. <Amended Oct. 22, 2019> 1. The first part of Article 16, Paragraph 4 and the amended provisions of Article 31-4 of the Partially Amended Medical Devices Act (Act No. 14330): Date of promulgation 2. Amended provisions of Article 2, Paragraph 4, Article 36, Paragraph 1, Subparagraph 9-2, and Article 53-2 of Act No. 14330, Partially Amending the Medical Devices Act: July 1, 2019 3. Amended provisions of Article 20, Article 31-3, Article 36 Paragraph 1 Subparagraph 17-3, and Article 56 Paragraph 1 Subparagraph 2-3 of Act No. 14330, Partial Amendment to the Medical Devices Act a. For Class 4 medical devices: July 1, 2019 B. For Class 3 medical devices: July 1, 2020 C. For Class 2 medical devices: July 1, 2021 D. For Class 1 medical devices: July 1, 2022 4. Amended provisions of Article 31-2, Article 36, Paragraph 1, Subparagraph 17-2, and Article 56, Paragraph 1, Subparagraph 2-2 of Act No. 14330, the Partial Amendment to the Medical Devices Act: Dates according to the classifications in each of the following subparagraphs a. Class 4 Medical Devices: July 1, 2020 B. Class 3 Medical Device: July 1, 2021 C. Class 2 Medical Device: July 1, 2022 D. Class 1 Medical Devices: July 1, 2023 ② Notwithstanding Paragraph 1, Subparagraph 3, the effective date of the amended provisions of Article 20, Subparagraph 9 and Article 22 of the Medical Devices Act, as determined pursuant to the proviso of Article 1 of the Addenda to Act No. 15279, the Partial Amendment to the Medical Devices Act, shall be July 1, 2019. Article 3 (Applicability to Recall Procedures, etc. for Hazardous Medical Devices)  The amended provisions of Articles 52 and 57 shall apply from the time a medical device is identified as a medical device subject to recall falling under any of the subparagraphs of Article 52, Paragraph 2, after the enforcement of this Rule. Article 4 (Applicability to Registration of Integrated Medical Device Information)  The amended provisions of Article 54-2 shall apply from the time a medical device manufacturer or importer supplies a medical device after the enforcement of this Rule. Article 5 (Transitional Measures Concerning Standards for Administrative Dispositions)  Notwithstanding the amended provisions of Annex 8, violations committed prior to the enforcement of this Rule shall be governed by the previous provisions. ADDENDA <Prime Minister's Decree No. 1542, June 12, 2019> This rule shall enter into force on June 12, 2019. However, the amended provisions of Article 9, Paragraph 2, Subparagraph 4 and Article 22, Subparagraph 3 shall enter into force on the date of promulgation. ADDENDA <Prime Minister's Decree No. 1567, Oct. 22, 2019> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of Article 54-2, Annex 8 II. Individual Standards No. 29-2, and Form No. 48-2 shall enter into force on the dates specified in each of the following subparagraphs, and the amended provisions of Form No. 7 shall enter into force one year after the date of promulgation. 1. Class 4 Medical Device: July 1, 2020 2. Class 3 Medical Device: July 1, 2021 3. Class II Medical Device: July 1, 2022 4. Class 1 Medical Device: July 1, 2023 Article 2 (Applicability to the recording of resident registration numbers in the manufacturing business permit register)  The amended provisions of Article 3, Paragraph 3, Subparagraph 4 (including cases where they are applied mutatis mutandis pursuant to Article 34) shall also apply to the permit register or report register of persons who obtained a manufacturing business permit or reported an import business prior to the enforcement of this rule. Article 3 (Applicability to Reporting of Medical Device Supply Details)  The amended provisions of Article 54-2 shall apply to medical devices supplied starting from the month following the month in which the enforcement date of the amended provisions pursuant to the proviso of Article 1 of the Addenda falls. Article 4 (Transitional Measures Concerning the Application of Administrative Disposition Standards)  When applying administrative disposition standards to violations committed prior to the enforcement of this Rule, the former provisions shall apply notwithstanding the amended provisions of Item 9 (g) and Item 12 (f) of Annex 8 II. Individual Standards. ADDENDA <Prime Minister's Decree No. 1580, Dec. 23, 2019> This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 1607, Apr. 13, 2020> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of Article 27, Paragraph 2, Article 33, Paragraph 2, and Subparagraph 9-2 of Annex 8 II. Individual Standards shall enter into force one year and six months after the date of promulgation, and the amended provisions of Article 65, Paragraph 3 shall enter into force six months after the date of promulgation. Article 2 (Applicability to Reporting Suspension of Production or Import)  The amended provisions of Article 27, Paragraph 2, Subparagraph 2 and Article 33, Paragraph 2, Subparagraph 2 shall apply from the date of enforcement of the said amended provisions pursuant to the proviso of Article 1 of the Addenda. Article 3 (Applicability to Refunds of Fees)  The amended provisions of Article 65, Paragraph 3 shall apply to fees paid on or after the effective date of the said amended provisions pursuant to the proviso of Article 1 of the Addenda. Article 4 (Transitional Measures Concerning the Processing Period for Notification of Change in Medical Device Repair Business)  If the procedure for notifying a change in a medical device repair business is in progress at the time of the enforcement of this Rule, the previous regulations shall apply notwithstanding the amended provisions of Form No. 35. ADDENDA <Prime Minister's Decree No. 1614, May 1, 2020> (Enforcement Rule of the In Vitro Diagnostic Medical Devices Act) Article 1 (Enforcement Date)  This rule shall enter into force on May 1, 2020. Article 2 (Amendment of Other Laws)  ① The Enforcement Rule of the Medical Devices Act is partially amended as follows: Article 11, Paragraph 2, Subparagraph 1, Item (d) is deleted. The proviso to Article 20, Paragraph 1, Subparagraph 3 and Article 20, Paragraph 3, Subparagraph 3 are each deleted. Subparagraphs c and d of Item 1 of Annex 1 are deleted, and in Subparagraph e of the same item, “in accordance with the provisions of Subparagraphs a through d” is changed to “in accordance with Subparagraphs a and b”. Delete the proviso to Item 3 in the attached documents section of Form No. 19. Delete the proviso to Item 3 in the attached documents section of Form No. 20. ② Omitted Article 3  Omitted ADDENDA <Prime Minister's Decree No. 1617, May 29, 2020> (Rules on Support and Management, etc. of Innovative Medical Devices) Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. Article 2 (Amendment of Other Laws)  Part of the Enforcement Rule of the Medical Devices Act is amended as follows. The attachment section on the front of Form No. 30 shall be as follows. The attachment section on the front of Form No. 31 shall be as follows. ADDENDA <Prime Minister's Decree No. 1658, Dec. 4, 2020> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of Article 27, Paragraph 1, Subparagraph 9, Article 33, Paragraph 1, Subparagraph 14, and Annex 8 shall enter into force one month after the date of promulgation. Article 2 (Applicability to Compliance Requirements of Manufacturers or Importers)  The amended provisions of Article 27, Paragraph 1, Subparagraph 9 and Article 33, Paragraph 1, Subparagraph 14 shall apply to data or information (including cases of adverse effects resulting from the use of medical devices) that manufacturers or importers become aware of after the effective date pursuant to the proviso of Article 1 of the Addenda. ADDENDA <Prime Minister's Decree No. 1662, Dec. 31, 2020> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. Article 2 (Transitional Measures Concerning Fees)  Notwithstanding the amended provisions of Annex 10, fees for applications for permission to manufacture or import medical devices, or requests for examination of technical documents, etc., made prior to the enforcement of this Rule shall be governed by the former provisions. Article 3 (Amendment of Other Laws)  The Enforcement Rule of the In Vitro Diagnostic Medical Devices Act is partially amended as follows: The fee section on the back of Form No. 3 shall be as follows. The fee section on the back of Form No. 5 shall be as follows. The fee section on the back of Form No. 7 shall be as follows. The fee section on the front of Form No. 18 shall be as follows. Article 4 (Transitional Measures Pursuant to Amendments to Other Laws)  Notwithstanding the amended provisions of Forms No. 3, No. 5, No. 7, and No. 18 of the Enforcement Rule of the In Vitro Diagnostic Medical Devices Act, fees for applications for permission or certification of manufacture or import of in vitro diagnostic medical devices, or for notifications, etc., made prior to the enforcement of this Rule shall be governed by the former provisions. ADDENDA <Prime Minister's Decree No. 1709, June 24, 2021> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of each of the following subparagraphs shall enter into force on the date specified in each subparagraph. 1. Amendments to Articles 45, 45-2, and 45-3, Forms 38-2 and 38-3, Annex 8 I Item 8, and Annex 10: June 24, 2021 2. Amended provision of Article 50, Paragraph 2, Subparagraph 2, Item (b): One year after the date of promulgation Article 2 (Applicability to Medical Devices for Exhibition Purposes)  The amended provisions of Article 46, Paragraph 4 shall also apply to medical devices approved as medical devices for exhibition purposes prior to the enforcement of this Rule. Article 3 (Applicability to the Submission of Records for Medical Devices Subject to Traceability Management)  The amended provisions of Article 50, Paragraph 2, Subparagraph 2, Item (b) shall apply from the time a medical institution uses a medical device subject to traceability management designated and notified by the Minister of Food and Drug Safety on or after the effective date of Article 1, Subparagraph 2 of the Addenda. ADDENDA <Prime Minister's Decree No. 1786, Jan. 21, 2022> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of each of the following subparagraphs shall enter into force on the date specified in each subparagraph. 1. Amended provisions of Article 14, Article 15, Article 45-4, Annex 8 II Subparagraphs 3-2, 3-3, 8 (d), 9 (g), 12 (f), 20-2, and 26-2, Annex 9 II Subparagraph 5, and Form No. 11: January 21, 2022 2. Amendments to Article 18, Articles 18-2 through 18-5, Article 27, Article 33, Article 54-4, Article 56-2, Article 57, Annex 8 II Subparagraphs 5 and 5-2 through 5-8, Annex 10 Subparagraph 7, Form No. 16, Forms No. 16-2 through 16-4, and Form No. 17: February 18, 2022 Article 2 (Applicability to Medical Device Sealing)  The amended provisions of Article 45-4 shall apply to medical devices manufactured or imported on or after the effective date pursuant to Article 1, subparagraph 1 of the Addenda. ADDENDA <Prime Minister's Decree No. 1819, July 20, 2022> Article 1 (Enforcement Date)  This Rule shall enter into force on July 21, 2022. However, the amended provisions of Form No. 48-2 shall enter into force on the dates specified in each of the following subparagraphs. 1. Class 2, Class 3, and Class 4 Medical Devices: July 21, 2022 2. Class 1 Medical Device: July 1, 2023 Article 2 (Applicability to Standards for Administrative Dispositions)  The amended provisions of the latter part of Item 2(a) of Subparagraph I of Annex 8 and the latter part of Item 3 of Subparagraph I of Annex 9 shall apply from the first violation committed after the enforcement of this rule. ADDENDA <Prime Minister's Decree No. 1841, Dec. 19, 2022> (Prime Minister's Decree on the Partial Amendment of Three Acts for the Expansion of the Scope of Recognition of Practical Experience Required for Qualification Acquisition, etc.) This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 1936, Jan. 16, 2024> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of Article 43, Paragraph 1 shall enter into force one year after the date of promulgation. Article 2 (Applicability to Submission of Documents Regarding Quality Managers)  The amended provisions of Article 3, Paragraph 1, Article 26, Paragraph 1, and Forms No. 1 and No. 29 shall apply from the time an application for a license or a license amendment for the manufacturing or importing of medical devices is made after the enforcement of this Rule. Article 3 (Applicability to Matters to be Stated in Attached Documents)  The amended provisions of Article 43, Paragraph 1 shall apply from the time products are shipped after the effective date pursuant to the proviso of Article 1. ADDENDA <Prime Minister's Decree No. 1946, Mar. 8, 2024> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. Article 2 (Transitional Measures Concerning Applications for Product-Specific Manufacturing Permits for Export Medical Devices)  In the case of export medical devices for which a product-specific manufacturing permit was applied for prior to the enforcement of this Rule pursuant to Article 4, Paragraph 2, Subparagraph 1, Item (c) of the former Rule and the procedure is in progress at the time of the enforcement of this Rule, the procedure for product-specific manufacturing certification or product-specific manufacturing notification pursuant to Article 4, Paragraph 2, Subparagraph 2 or 3 shall be deemed to be in progress. ADDENDA <Prime Minister's Decree No. 1971, July 8, 2024> (Prime Minister's Decree on the Partial Amendment of Five Acts to Enhance the Vitality of the People's Livelihood Economy, Including Temporary Regulatory Suspension) This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 1976, Aug. 7, 2024> Article 1 (Enforcement Date)  This Rule shall enter into force on August 7, 2024. However, the amended provisions of Article 4, Paragraph 2, Subparagraphs 1 and 2, Article 44-2, Form No. 32 and Form No. 52 shall enter into force on the date of promulgation. Article 2 (Applicability to Changes in Scope of Manufacturing Permits and Notifications by Medical Device Item, etc.)  ① The amended provisions of Article 4, Paragraph 2, Subparagraph 1, Item (c) and Subparagraph 3 of the same Paragraph shall apply from the time a manufacturing permit is obtained or a manufacturing notification is filed by a specific item after the enforcement of this Rule. ② For Class 1 medical devices for which manufacturing approval by item was obtained in accordance with the previous regulations prior to the enforcement of this rule, notwithstanding the amended provisions of Article 4, Paragraph 2, Subparagraph 1, Item (c) and Subparagraph 3 of the same paragraph, the previous regulations regarding manufacturing approval by item shall apply. ADDENDA <Prime Minister's Decree No. 1982, Sep. 20, 2024> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. Article 2 (Applicability to Fee Reductions)  The amended provisions of the proviso to Article 65, Paragraph 1 and each subparagraph of the same paragraph shall apply from the time of application for or notification of permission, certification, etc., after the enforcement of this Rule. ADDENDA <Prime Minister's Decree No. 2029, Apr. 1, 2025> Article 1 (Enforcement Date)  This Rule shall enter into force on the date of its promulgation. However, the amended provisions of Article 3, Paragraph 1, Subparagraph 3, Item (a), Article 26, Paragraph 1, Subparagraph 2, Item (a), the back of Form No. 1, and the front of Form No. 29 shall enter into force three months after the date of promulgation. Article 2 (Applicability to Exemption from Submission of Documents Verifying the Qualifications of the Quality Manager)  The amended provisions of Article 3, Paragraph 1, Subparagraph 3, Item (a), the proviso of Article 26, Paragraph 1, Subparagraph 2, Item (a), the back of Form No. 1, and the front of Form No. 29 shall apply from the date of enforcement of the proviso to Article 1 of the Addenda, when an application for a medical device manufacturing license or a license amendment is submitted. Article 3 (Applicability to Fee Reductions for Newly Developed Medical Devices, etc.)  ① The amended provisions of Article 65, Paragraph 1, Subparagraph 3 (limited to the part related to Subparagraph 19 of Annex 10) shall apply from the time an application for prior review of technical documents, etc. of newly developed medical devices is made after the enforcement of this Rule. ② Notwithstanding the amended provisions of Article 65, Paragraph 1, Subparagraph 3 (limited to the part related to Subparagraph 3-2 of Annex 10), the previous provisions shall apply to the reduction of manufacturing and import permit fees for newly developed medical devices applied for prior to the enforcement of this rule. Article 4 (Transitional Measures Concerning Fees for Advance Review of Newly Developed Medical Devices)  Notwithstanding the amended provisions of Subparagraphs 3-2 and 19 of Annex 10 and the remarks column of the same table, fees for advance review of manufacturing/import permits for newly developed medical devices or technical documents, etc. for newly developed medical devices applied for prior to the enforcement of this Rule shall be governed by the previous provisions. ADDENDA <Prime Minister's Decree No. 2039, July 24, 2025> This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 2044, Aug. 1, 2025> Article 1 (Enforcement Date)  This rule shall enter into force on August 1, 2025. Article 2 (Transitional Measures Concerning Records of Medical Devices Subject to Traceability Management) In cases where  records prepared by a user prior to the enforcement of this Rule pursuant to Article 50, Paragraph 1, Subparagraph 2 of the former Rule are submitted to the Minister of Food and Drug Safety pursuant to Paragraph 2 of the same Article, the former provisions shall apply notwithstanding the amended provisions of Article 50, Paragraph 1, Subparagraph 2. ADDENDA <Prime Minister's Decree No. 2060, Oct. 31, 2025> (Prime Minister's Decree on the Partial Amendment of Six Laws to Reflect Government Organization Reorganization) This rule shall take effect from the date of its promulgation. ADDENDA <Prime Minister's Decree No. 2127, July 1, 2026> This rule shall take effect on July 1, 2026. [Appendix 1] Standards and Procedures for Classification and Designation of Medical Devices (Related to Article 2) [Appendix 2] Standards for Facilities and Manufacturing and Quality Management Systems (Related to Article 8, Paragraph 1) [Appendix 3] Management Standards for Clinical Trials of Medical Devices (Related to Article 24, Paragraph 1) [Appendix 4] Standards for Facilities and Manufacturing and Quality Control Systems of Import Businesses (Related to Article 31, Paragraph 1) [Appendix 5] Standards for Facilities and Quality Management Systems of Repair Businesses (Related to Article 35, Paragraph 4) [Appendix 6] Standards for Quality Management of Medical Device Distribution (Related to Article 39, Subparagraph 4) [Appendix 7] Scope of Prohibited Advertisements (Related to Article 45, Paragraph 1) [Attached Table 7-2] Detailed Standards for Cancellation of Appointment, etc. of Examiners (Related to Article 48-2, Paragraph 5) [Appendix 7-3] Criteria for Cancellation, etc. of Recognition of Conformity of Medical Devices (Related to Article 48-4, Paragraph 2) [Appendix 7-4] Standards for Integrated Information Management of Medical Devices (Related to Article 54-3, Paragraph 2) [Appendix 8] Standards for Administrative Dispositions (Related to Article 58, Paragraph 1) [Appendix 9] Standards for Administrative Dispositions (Related to Article 58, Paragraph 2) [Attached Table 10] Fees (Related to Article 65, Paragraph 1) [Attachment Form No. 1] Application for Medical Device Manufacturing (Import) Business License [Attachment Form No. 2] Medical Device Manufacturing (Import) License [Attachment Form No. 3] Application for Medical Device Manufacturing (Import) Permit [Attachment Form No. 4] Medical Device Manufacturing (Import) Permit [Attachment Form No. 5] Application for Medical Device Manufacturing (Import) Certification [Attachment Form No. 6] Medical Device Manufacturing (Import) Certificate [Attachment Form No. 7] Medical Device Manufacturing (Import) Notification [Form No. 7-2] Medical Device Manufacturing (Import) Notification Certificate [Attachment Form No. 8] Request for Review of Medical Device Technical Documents, etc. [Attachment Form No. 9] Notification of Review Results for Medical Device Technical Documents, etc. [Attachment Form No. 10] Quality Manager Non-Work Report [Attachment Form No. 11] Application for Designation as a Medical Device Quality Manager Training Institution [Form No. 11-2] Application for Change of Designation of Medical Device Quality Manager Training Institution [Attachment Form No. 12] Designation Certificate for Medical Device Quality Manager Training Institution [Form No. 12-2] Application for Medical Device Technical Document Review Agency (Designation, Renewal of Designation) [Form No. 12-3] Designation Certificate for Medical Device Technical Document Review Agency [Form No. 12-4] Application for Change of Designation as Medical Device Technical Document Review Agency [Attachment Form No. 13] Conditional Medical Device Manufacturing (Import) Permit [Attachment Form No. 14] Conditional Manufacturing (Import) Permit for Medical Devices [Attachment Form No. 15] Conditional Manufacturing (Import) Certificate for Medical Devices [Form No. 15-2] Conditional Manufacturing (Import) Notification Certificate for Medical Devices [Form No. 16] Application for Approval (Amendment Approval) of Post-Market Surveillance Plan for Medical Devices [Form No. 16-2] Approval of Post-Market Surveillance Plan for Medical Devices [Form No. 16-3] Periodic Report on Post-Market Surveillance of Medical Devices [Form No. 16-4] Application for Review of Post-Market Surveillance Results of Medical Devices [Attachment No. 17] Notification of Review Results of Post-Market Surveillance for Medical Devices [Form No. 18] Application for Medical Device Re-evaluation [Attachment Form No. 19] Application for Approval of Clinical Trial Plan for Medical Devices [Attachment Form No. 20] Application for Approval of Change to Medical Device Clinical Trial Plan [Attachment Form No. 21] Medical Device Clinical Trial Plan Approval [Form No. 22] Application for Designation as a Medical Device Clinical Trial Institution [Form No. 23] Designation Certificate for Medical Device Clinical Trial Institution [Form No. 24] Application for Change of Designation of Medical Device Clinical Trial Institution [Form No. 25] Report on the Status of Medical Device Clinical Trials [Form No. 26] Medical Device Clinical Trial Completion Report [Form No. 26-2] Application for Designation of Medical Device Non-Clinical Testing Institution [Form No. 26-3] Designation Certificate for Medical Device Non-Clinical Testing Institution [Form No. 26-4] Application for Change of Designation of Medical Device Non-Clinical Testing Institution [Form No. 27] Application for Preliminary Review of Medical Devices [Form No. 28] Notification of Medical Device Preliminary Review Results [Attachment Form No. 29] Application for Permission to Change Medical Device Manufacturing (Import) Business License [Attachment Form No. 30] Application for Approval of Change in Medical Device License Details [Attachment Form No. 31] Application for Change of Certification for Medical Devices [Form No. 32] Report on Closure, Suspension, etc. of Medical Device Business [Attachment Form No. 33] Medical Device Repair Business Report [Attachment Form No. 34] Medical Device Repair Business Registration Certificate [Attachment Form No. 35] Notification of Change in Medical Device Repair Business Report Details [Attachment Form No. 36] Report on Medical Device Sales (Rental) Business [Attachment Form No. 37] Medical Device Sales (Lease) Business Registration Certificate [Attachment Form No. 38] Notification of Change in Reported Details for Medical Device Sales (Rental) Business [Form No. 38-2] Objection to the Result of Re-examination of Medical Device Advertisement [Form No. 38-3] Report to Self-Regulatory Body for Medical Device Advertising [Form No. 38-4] Application for Medical Device Manufacturing and Quality Management System (Acknowledgment of Conformity, Acknowledgment of Modified Conformity) [Attachment Form No. 39] Application for Medical Device Quality Management Inspection Agency (Designation, Renewal of Designation) [Form No. 40] Designation Certificate for Medical Device Quality Management Inspection Agency [Attachment Form No. 41] Application for Change of Designation of Medical Device Quality Management Review Agency [Attachment Form No. 42] Patient Notification Confirmation [Attachment Form No. 43] Recovery Plan [Attachment Form No. 44] Recovery Confirmation [Form No. 45] Recovery Evaluation Report [Attachment Form No. 46] Application for Disposal [Attachment Form No. 47] Disposal Confirmation [Attachment Form No. 48] Recovery Completion Report [Form No. 48-2] Medical Device Supply History Report [Form No. 48-3] Report on Discovery of Foreign Object [Attachment Form No. 49] Medical Device Collection Certificate [Attachment Form No. 50] Medical Device Inspector ID Card [Attachment No. 50-2] Approval for Exclusive Entry to Sales (Lease) Establishments by Consumer Medical Device Inspectors [Form No. 50-3] Consumer Medical Device Inspector Certificate [Form No. 50-4] Application for Designation as Medical Device Safety Information Monitoring Center [Form No. 50-5] Designation Certificate for Medical Device Safety Information Monitoring Center [Form No. 50-6] Certificate of Causation Investigator [Form No. 51] Application for Renewal of Medical Device Manufacturing (Import) Permit [Form No. 51-2] Application for Renewal of Medical Device Manufacturing (Import) Certification [Form No. 51-3] Medical Device Manufacturing (Import) Notification Renewal Report [Form No. 52] Application for Reissuance of Medical Device License, Certificate, etc. [Form No. 53] Application for Issuance of English Medical Device Certificate [Form No. 54] Application for Reclassification of Medical Devices [Form No. 55] Rapid Report on Adverse Reactions to Medical Devices [Form No. 56] Medical Device Adverse Reaction Report