Rules on Safety Management of Diagnostic Radiation Generating Devices
288071 · 2026-07-09
Article 1 (Purpose) The purpose of this Rule is to prescribe matters necessary to prevent harm to patients and radiation workers caused by radiation and to promote the appropriateness of medical treatment by safely managing diagnostic radiation generating devices installed and operated by medical institutions pursuant to Article 37 of the Medical Service Act.
Article 2 (Definitions) The definitions of terms used in this Rule are as follows. <Amended June 27, 2011>
1. The term "diagnostic radiation generating device" refers to an instrument used to diagnose diseases using radiation, and which falls under any of the following subparagraphs.
A. Diagnostic X-ray device
B. Diagnostic X-ray Generator
C. Dental diagnostic X-ray generator
D. Computed tomography (including dental computed tomography devices, otolaryngological computed tomography devices, and positron emission tomography computed tomography devices)
E. Devices that generate radiation for the diagnosis of diseases, such as mammography machines
2. "Radiation protection facilities" refers to radiation shielding facilities and radiation damage protection devices located at a place where a diagnostic radiation generating device is installed to prevent radiation exposure (exposure of the human body to radiation).
3. The term "radiation-related worker" refers to a person whose primary workplace is a place where a diagnostic radiation generating device is installed, and who is engaged in radiation-related duties such as the management, operation, and manipulation of the diagnostic radiation generating device.
4. "Safety management" refers to the management of facilities related to diagnostic imaging information, such as diagnostic radiation generators, radiation shielding facilities and darkrooms, developers, radiation film cassettes, grids for removing scattered X-rays, and X-ray observation tables, as well as the management of radiation exposure for radiation-related personnel.
5. "Radiation Zone" refers to a place where a diagnostic radiation generating device is installed and the external radiation dose is 0.3 mSv (30 mrem) or more per week, and which is partitioned by compartments such as walls or protective partitions.
Article 3 (Reporting)
① Pursuant to Article 37, Paragraph 1 of the Medical Service Act (hereinafter referred to as the "Act"), the founder or manager of a medical institution (including public health centers, health medical centers, and branch health centers pursuant to Articles 10, 12, and 13 of the Regional Health Act; regional military manpower offices conducting conscription examinations pursuant to Article 11 of the Military Service Act; military hospitals pursuant to Article 6 of the Order of the Armed Forces Medical Command and all medical facilities of each military branch and directly affiliated organizations; health rooms pursuant to Article 3 of the School Health Act; and correctional facilities pursuant to Article 2, Paragraph 4 of the Act on the Execution of Sentences and Treatment of Prisoners; hereinafter the same applies) shall, in the case of installing a diagnostic radiation generating device, notify the institution by 3 days prior to the date of use; in the case of suspending use, notify the institution by 3 days from the date of suspension; in the case of resuming use after suspension, notify the institution by 3 days prior to the date of use; and regarding transfer, disposal, or relocation [the city, county, or district where the medical institution is located (autonomous districts [Excluding cases of relocation resulting from the relocation of a medical institution within the jurisdiction of (hereinafter referred to as "competent jurisdiction")] In such cases, a report in accordance with Form No. 1 or Form No. 2 attached hereto, respectively, with the documents specified in each of the following subparagraphs attached, must be submitted to the Mayor, County Governor, or District Head having jurisdiction over the location of the relevant medical institution (referring to the head of an autonomous district; hereinafter the same applies) having jurisdiction over the location of the medical institution, within 45 days from the date the cause occurred. <Amended 2010.1.22, 2011.6.27, 2012.11.15, 2015.7.24, 2015.11.18, 2025.7.18>
1. Documents to be attached to Attached Form No. 1
A. One copy of the inspection report for the diagnostic radiation generator
B. One copy of the Radiation Protection Facility Inspection Report
C. One copy of the Report on Radiation-Related Workers in accordance with Form No. 6
D. One original copy of the Certificate of Transfer or Certificate of Relocation Report in accordance with Form No. 3 of the Attached Form for the person who transferred or relocated (to be attached only in cases of receiving a transfer or relocating and installing)
E. One copy of the Certificate of Registration for Specialized Medical Equipment (Attach only if specialized medical equipment is being installed)
F. Original Certificate of Reporting Suspension of Use of Diagnostic Radiation Generating Device in accordance with Form No. 3 (to be attached only if use resumes after suspension)
d. One copy of the Medical Device Manufacturing Permit or Import Permit (excluding cases where specialized medical equipment has been registered in accordance with Article 2 of the "Rules on the Installation and Operation of Specialized Medical Equipment")
H. One copy of proof of purchase or lease, such as a tax invoice or contract (excluding cases where specialized medical equipment has been registered in accordance with Article 2 of the "Rules on the Installation and Operation of Specialized Medical Equipment")
2. Documents to be attached to Form No. 2
A. Original Certificate of Reporting in accordance with Form No. 3
B. One copy of documents verifying the transfer or assignment (to be attached only when filing a transfer report)
C. One copy of documents confirming disposal (attach only when filing a disposal report)
D. One copy of documents verifying the transfer (attach only when filing a transfer report)
② The Mayor, County Governor, or District Head who receives a report pursuant to Paragraph 1 shall issue a certificate of report in accordance with Form No. 3.
③ The Mayor, County Governor, or District Head shall reissue a report certificate if a person who has received such certificate pursuant to Paragraph 2 applies for reissuance in accordance with Form No. 4 due to reasons such as damage or loss of the certificate.
④ The founder or manager of a medical institution that installs and operates a diagnostic radiation generating device shall submit a change notification form in accordance with Annex 5, attaching the documents specified in each of the following subparagraphs, to the Mayor, County Governor, or District Head in advance if they intend to change the founder or the name of the medical institution, or at least three days prior to the date of use of the diagnostic radiation generating device if they have relocated the medical institution within the jurisdiction. <Amended 2011.6.27, 2015.7.24>
1. Original Certificate of Reporting in accordance with Form No. 3
2. Delete <2015.1.8>
3. One copy of the inspection report for diagnostic radiation generators (applicable only if the medical institution has relocated within the jurisdiction)
4. One copy of the Radiation Protection Facility Inspection Report (Applicable only if the medical institution has been relocated within the jurisdiction)
5. One copy of documents verifying the relocation (applicable only if the medical institution was relocated within the same jurisdiction)
⑤ The Mayor, County Governor, or District Head who has received a report of change pursuant to Paragraph 4 shall amend the relevant matters of the report certificate or reissue the certificate.
⑥ If there is a change in affiliated radiation-related workers, the founder or manager of a medical institution shall, within three months from the date the cause occurred, submit a report in accordance with Form No. 6 attached hereto to the Mayor, County Governor, or District Head having jurisdiction over the location of the medical institution, attaching a copy of the health examination results report pursuant to Article 13 (in cases where a radiation worker under Article 2, subparagraph 21 of the Atomic Energy Safety Act reports engaging in diagnostic radiation-related work due to a change in affiliation, this includes a copy of the radiation exposure record confirmation issued by the Central Registration Center for Radiation Worker Information). <Amended 2010.1.22, 2012.11.15>
⑦ When a Mayor, County Governor, or District Head receives a confirmation of radiation exposure record pursuant to Paragraph 6, they shall notify the Commissioner of the Korea Disease Control and Prevention Agency of such fact within 15 days. <Newly added 2010.1.22, 2013.3.23, 2020.9.11>
⑧ The Mayor, County Governor, or District Head shall prepare and keep the Diagnostic Radiation Management Register in the form of Annex 7, and record the matters to be reported under Paragraphs 1, 4, and 6, and Article 10, Paragraph 2. <Amended 2010.1.22>
Article 3-2 (Reporting, etc. through the Integrated Healthcare Resources Reporting Portal)
① The Mayor, County Governor, or District Head may process matters concerning reports under Article 3, Paragraphs 1, 4, and 6 through the electronic civil complaint window pursuant to Article 12-2, Paragraph 1 of the Enforcement Rules of the National Health Insurance Act (hereinafter referred to as the "Integrated Health and Medical Resources Reporting Portal").
② The Mayor, County Governor, or District Head shall notify the Health Insurance Review and Assessment Service (hereinafter referred to as the "HIRA") pursuant to Article 62 of the National Health Insurance Act of matters processed under Paragraph 1 (including matters processed upon receiving a report in writing) in the manner prescribed in Article 12-2, Paragraph 3 of the Enforcement Rules of the National Health Insurance Act.
③ Notwithstanding Article 3, Paragraphs 1 and 4, if the founder or manager of a medical institution makes a report pursuant to Paragraph 1 through the Integrated Reporting Portal for Health and Medical Resources, they may submit copies of documents or omit the submission of documents according to the classifications in each of the following subparagraphs.
1. Documents that may be submitted as copies: Report of Radiation-Related Workers pursuant to Article 3, Paragraph 1, Subparagraph 1, Item (c)
2. Documents pursuant to each of the following items for which submission may be omitted
a. Original Certificate of Transfer Report or Certificate of Relocation Report pursuant to Article 3, Paragraph 1, Subparagraph 1, Item (d)
B. Copy of the Certificate of Registration for Specialized Medical Equipment pursuant to Article 3, Paragraph 1, Subparagraph 1, Item E
C. Original Certificate of Reporting Suspension of Use of Diagnostic Radiation Generating Devices Pursuant to Article 3, Paragraph 1, Subparagraph 1, Item F.
D. Copy of medical device manufacturing license or import license pursuant to Article 3, Paragraph 1, Subparagraph 1, Item (g)
E. Original of the Certificate of Reporting pursuant to Article 3, Paragraph 1, Subparagraph 2, Item A
F. Original of the Certificate of Reporting pursuant to Article 3, Paragraph 4, Subparagraph 1
④ The Mayor, County Governor, District Head, and the Review and Evaluation Institute may process data containing Resident Registration Numbers or Alien Registration Numbers pursuant to Article 19, Subparagraph 1 or 4 of the Enforcement Decree of the Personal Information Protection Act if it is unavoidable for the purposes of the duties under Paragraphs 1 through 3.
[This Article Newly Established 2015.7.24]
Article 4 (Inspection and Measurement)
① If there is any reason falling under any of the following subparagraphs, the founder or manager of a medical institution shall, in accordance with Article 37, Paragraph 2 of the Act, undergo an inspection by an inspection agency pursuant to Article 6 in accordance with the inspection standards in Annex 1 regarding the diagnostic radiation generating device before using it. However, if, when obtaining a manufacturing or import permit for a medical device pursuant to Article 5, Paragraph 1 and Article 18, Paragraph 1 of the Enforcement Rule of the Medical Devices Act, a test report including the inspection items in Annex 1 is submitted after undergoing a test inspection at a testing and inspection agency pursuant to Article 27 of the Medical Devices Act, the device may be used without undergoing the inspection under the main text. <Revised 2009.5.29, 2010.1.22, 2011.11.25, 2012.11.15>
1. When installing or relocating a diagnostic radiation generator
2. When changing the power supply facilities of a diagnostic radiation generator
3. In the case of intending to reuse a diagnostic radiation generating device for which a report of suspension of use has been filed pursuant to Article 3, Paragraph 1
4. In the case of repairing or replacing a high-voltage generator, X-ray tube, or control device that may affect the safety of a diagnostic radiation generator
② The founder or manager of a medical institution shall undergo an inspection by an inspection agency pursuant to Article 6 every three years from the date of the inspection (in the case of the proviso to Paragraph 1, this refers to an inspection by a testing and inspection agency) regarding diagnostic radiation generating devices inspected pursuant to Paragraph 1. In this case, the inspection period shall be 31 days each before and after the expiration date of the period. <Amended 2011.6.27>
③ The founder or manager of a medical institution shall, in accordance with Article 37, Paragraph 2 of the Act, undergo an inspection by an inspection agency pursuant to Article 6 regarding radiation shielding facilities before using the relevant diagnostic radiation generating device, in accordance with the inspection standards for radiation shielding facilities in Annex 2.
④ If the founder or manager of a medical institution modifies or installs radiation shielding facilities among the radiation protection facilities inspected pursuant to Paragraph 3, or if the maximum weekly operating load set at the time of designing the radiation shielding facilities is exceeded, they shall undergo an inspection of said radiation protection facilities by an inspection agency pursuant to Article 6 without delay.
⑤ In accordance with Article 37, Paragraph 2 of the Act, the founder or manager of a medical institution shall ensure that radiation-related workers undergo radiation exposure dose measurement at least once every three months when using T-L badges, and at least once every month when using film badges.
⑥ The dose limits for the measurement of radiation exposure doses of radiation-related workers pursuant to Paragraph 5 are as set forth in Attached Table 3.
7. The Commissioner of the Korea Disease Control and Prevention Agency shall establish a Radiation Exposure Dose Management Center within the Korea Disease Control and Prevention Agency to manage the personal radiation exposure doses of radiation-related workers, and shall determine and publicly announce detailed matters necessary for the operation of said Center. <Amended Jan. 22, 2010; Mar. 23, 2013; Sep. 11, 2020>
⑧ A medical institution designated as a quality certification body in accordance with the designation standards and procedures for quality certification bodies for medical radiation facilities prescribed and notified by the Commissioner of the Korea Disease Control and Prevention Agency may be exempted from inspections under the provisions of Paragraphs 1 through 4, as prescribed and notified by the Commissioner of the Korea Disease Control and Prevention Agency. <Newly added 2010.1.22, 2012.11.15, 2013.3.23, 2020.9.11>
Article 5 (Test Methods for Inspection, etc.)
① The test methods for inspections or measurements pursuant to Article 4 (hereinafter referred to as "inspections, etc.") shall be determined by the head of the inspection agency or measurement agency pursuant to Article 6 upon approval by the Commissioner of the Korea Disease Control and Prevention Agency. In this case, detailed matters regarding approval procedures, etc., shall be in accordance with the provisions separately determined and notified by the Commissioner of the Korea Disease Control and Prevention Agency. <Amended Jan. 22, 2010; Nov. 15, 2012; Mar. 23, 2013; Sep. 11, 2020>
② When conducting a re-inspection of an inspection, etc., the test may be performed using any of the following methods in addition to the test methods prescribed in Paragraph 1. <Amended 2010.1.22, 2012.11.15, 2013.3.23, 2020.9.11>
1. Test methods notified by the Commissioner of the Korea Disease Control and Prevention Agency
2. Test method determined by the manufacturer
3. Test methods recommended by the International Electrotechnical Commission (applicable only where the test methods of No. 1 or No. 2 are unavailable)
Article 6 (Inspection and Measurement Agencies)
① Any person who intends to perform duties of inspecting diagnostic radiation generating devices and radiation shielding facilities pursuant to Article 4, or duties of measuring the radiation exposure dose of radiation-related workers, shall register with the Commissioner of the Korea Disease Control and Prevention Agency in accordance with the procedures prescribed by the Commissioner. <Amended 2011.6.27, 2012.11.15, 2013.3.23, 2020.9.11>
② Any person who intends to register as an institution inspecting diagnostic radiation generating devices and radiation shielding facilities or an institution measuring the exposure dose of radiation-related workers pursuant to Paragraph 1 (hereinafter referred to as "Inspection and Measurement Institution") shall comply with the standards set forth in Annex 4. <Newly Added 2011.6.27>
③ The validity period of the registration under Paragraph 1 shall be four years from the date of registration. <Newly added Jan. 22, 2010, Jun. 27, 2011>
④ If an inspection or measurement institution whose validity period under Paragraph 3 is expiring intends to re-register, it must apply for re-registration six months prior to the date of expiration of the validity period. However, the Commissioner of the Korea Disease Control and Prevention Agency shall notify the head of the inspection or measurement institution of the re-registration procedures, etc., in advance via text message via mobile phone, e-mail, fax, telephone, document, etc., no later than two months prior to the date the re-registration application begins. <Newly added 2010.1.22, 2011.6.27, 2012.11.15, 2013.3.23, 2020.9.11>
⑤ The Commissioner of the Korea Disease Control and Prevention Agency may revoke the registration of an inspection or measurement institution, order a suspension of operations for a specified period not exceeding six months, or take necessary measures such as a corrective order if the institution falls under any of the following subparagraphs. However, if it falls under any of subparagraphs 1 through 3, the registration shall be revoked. <Newly added 2010.1.22, 2011.6.27, 2012.11.15, 2013.3.23, 2020.9.11>
1. In the case of registration obtained through false or other fraudulent means
2. In cases where a false inspection report for diagnostic radiation generating devices, etc. is issued due to intent or gross negligence
3. Cases where inspection or measurement work is performed during the period of business suspension
4. Cases where the registration standards for inspection and measurement institutions pursuant to Annex 4
5. In the event of a violation of the regulations concerning inspection and measurement duties prescribed and notified by the Commissioner of the Korea Disease Control and Prevention Agency
⑥ Matters necessary for the registration procedures and operation of the registration system for inspection and measurement institutions pursuant to Paragraph 1, and matters necessary for detailed standards, etc. for administrative dispositions pursuant to Paragraph 5, shall be separately determined and announced by the Commissioner of the Korea Disease Control and Prevention Agency. <Amended Jan. 22, 2010; Jun. 27, 2011; Nov. 15, 2012; Mar. 23, 2013; Sep. 11, 2020>
Article 7 (Application for Inspection, etc. and Prior Notification)
① Any person who intends to undergo inspection, etc. shall submit to the head of the inspection and measurement agency an application for inspection of a diagnostic radiation generating device in accordance with Form No. 8, an application for inspection of a radiation protection facility in accordance with Form No. 9, or an application for measurement of radiation exposure dose for radiation-related workers in accordance with Form No. 10. The same shall apply in the case of a re-inspection pursuant to Article 8, Paragraph 4. <Amended Jan. 22, 2010>
② The Mayor, County Governor, or District Head shall notify the founder or manager of a competent medical institution required to undergo an inspection of diagnostic radiation generating devices pursuant to Article 4, Paragraph 2, to undergo the inspection at least two months prior to the inspection date. <Newly Added 2010.1.22>
[Title revised 2010.1.22]
Article 8 (Notification of Inspection Results, etc.)
① Pursuant to Article 7, when the head of an inspection and measurement agency receives an application for inspection, etc. from the founder or manager of a medical institution, he/she shall, after conducting the inspection, etc., issue to the applicant an inspection report for diagnostic radiation generating devices in accordance with Form No. 11, an inspection report for radiation protection facilities in accordance with Form No. 12, or a radiation exposure dose measurement report for radiation-related workers in accordance with Form No. 13; and shall submit the inspection results in accordance with Form No. 14 or the measurement results in accordance with Form No. 15 to the Commissioner of the Korea Disease Control and Prevention Agency within one month from the date of inspection or measurement. <Amended 2010.1.22, 2013.3.23, 2020.9.11>
② If the results of the inspection or measurement agency are determined to be in compliance with the standards of Annex 1, Annex 2, or Annex 3, the head of the inspection or measurement agency shall notify the mayor, county governor, or district head having jurisdiction over the location of the medical institution within one month from the date of inspection or measurement; if it is determined to be in compliance or to be in compliance after re-inspection, a copy of the inspection report shall be notified without delay to the mayor, county governor, or district head having jurisdiction over the location of the medical institution, the Director of the Korea Disease Control and Prevention Agency, the President of the Health Insurance Review and Assessment Service, and the President of the National Health Insurance Service; and if it is determined to be in compliance with the dose limit, a copy of the measurement report shall be notified without delay to the Director of the Korea Disease Control and Prevention Agency and the mayor, county governor, or district head having jurisdiction over the location of the medical institution. <Revised 2010.1.22, 2013.3.23, 2020.9.11>
③ Upon receiving notification of the results of an inspection, etc., determined to be unsuitable or to have exceeded the dose limit pursuant to Paragraph 2, the Mayor, County Governor, or District Head shall, within 3 days from the date of receiving the notification, issue an order under any of the following subparagraphs to the founder or manager of a medical institution pursuant to Article 63 of the Act, and notify the Commissioner of the Korea Disease Control and Prevention Agency of such fact. <Amended Jan. 22, 2010; Mar. 23, 2013; Sep. 11, 2020>
1. Prohibition of use and re-inspection after repair and calibration of non-conforming diagnostic radiation generators or radiation shielding facilities
2. Health examinations for persons exceeding dose limits and safety measures if necessary (referring to changes in workplace or reduction of working hours, etc.; hereinafter the same applies)
④ Upon receiving an order pursuant to Paragraph 3, the founder or manager of a medical institution shall repair and calibrate diagnostic radiation generating devices or radiation shielding facilities and apply for re-inspection; for persons exceeding the dose limit, they shall immediately conduct a health examination pursuant to Article 13, and if abnormal signs are detected as a result, they shall take safety measures. In this case, if no abnormal signs appear as a result of a health examination conducted six months after the implementation of safety measures, they may request the competent Mayor, County Governor, or District Head to lift the safety measures upon obtaining the consent of the relevant person who exceeded the dose limit.
⑤ Upon receiving notification of a suitability judgment after a re-inspection pursuant to Paragraph 2, or upon receiving a request from the founder or manager of a medical institution to lift safety measures regarding a person who has exceeded the dose limit pursuant to Paragraph 4, the Mayor, County Governor, or District Head shall lift the order prohibiting use or the order for safety measures and notify the Commissioner of the Korea Disease Control and Prevention Agency of such fact. <Amended Jan. 22, 2010; Mar. 23, 2013; Sep. 11, 2020>
⑥ Any person who wishes to obtain a reissuance of an inspection report, etc. shall submit an application for reissuance of the inspection report, etc., in accordance with Form No. 16 attached hereto, to the head of the inspection and measurement institution.
Article 9 (Radiation Area)
① The founder or manager of a medical institution, upon installing a diagnostic radiation generating device pursuant to Article 37, Paragraph 1 of the Act, shall take measures such as establishing a radiation zone and restricting access by the general public for the safe management of diagnostic radiation.
② Radiation areas shall be marked with the radiation area designation specified in Attached Table 5.
Article 10 (Safety Manager for Diagnostic Radiation)
① The founder or manager of a medical institution shall, in accordance with Article 37, Paragraph 2 of the Act, appoint a Diagnostic Radiation Safety Manager (hereinafter referred to as the "Safety Manager") from among radiation-related personnel affiliated with the medical institution to ensure the safety management and proper use of diagnostic radiation generating devices, in accordance with the qualification standards for Diagnostic Radiation Safety Managers in Annex 6, and have the Safety Manager perform diagnostic radiation safety management duties (hereinafter referred to as "Safety Management Duties"). In this case, the Safety Manager of the medical institution may concurrently assume the duties of safety management for diagnostic radiation generating devices installed in branch hospitals or branch offices located within the same city, county, or district.
② When the founder or manager of a medical institution appoints, dismisses, or assigns a Safety Manager to concurrent duties, they shall submit a report in accordance with Form No. 6 attached hereto, attached with the documents specified in each of the following subparagraphs regarding the Safety Manager, to the Mayor, County Governor, or District Head having jurisdiction over the location of the relevant medical institution within one month. <Amended 2012.11.15>
1. One copy of final school graduation certificate (only for holders of a master's degree in science or engineering) or certificate of work experience
2. One copy of the health examination results report pursuant to Article 13
③ The Mayor, County Governor, or District Head who receives a report pursuant to Paragraph 2 shall verify the medical license of the safety manager subject to reporting through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act if the safety manager is a medical professional. However, if the employee who is a medical professional does not consent to this, a copy thereof shall be required to be attached. <Newly added 2012.11.15>
Article 11 (Duties of the Safety Manager) The duties of the Safety Manager pursuant to Article 10 are as follows.
1. Planning, Inspection, and Evaluation of Safety Management Duties
2. Implementation of in-house education and training for affiliated radiation-related personnel
3. Protective measures against radiation damage for patients and radiation workers
4. Safety Management of Facilities Related to Diagnostic Imaging Information
5. Notification to the measuring agency of changes in the affiliation of radiation workers affecting exposure dose measurement
6. Notification to the measuring agency of damage or loss of exposure dosimeters affecting the measurement of exposure doses of radiation workers
7. Matters concerning the notification pursuant to Article 3, Paragraphs 1 and 4, and the inspection or measurement pursuant to Article 4
8. Matters concerning the preparation, keeping, and preservation of documents regarding diagnostic radiation generating devices, radiation-related personnel, and radiation shielding facilities pursuant to Article 14 (hereinafter referred to as "diagnostic radiation generating devices, etc.")
Article 12 (Matters to be Complied With by Founders or Managers of Medical Institutions) Pursuant to Article 37, Paragraph 1 of the Act, the founder or manager of a medical institution shall comply with the matters set forth in each of the following subparagraphs. <Amended Jan. 22, 2010>
1. If the Safety Manager requests matters necessary for the performance of their duties, take action without delay and do not refuse without justifiable grounds.
2. If the Safety Manager fails to faithfully perform safety management duties, they shall be dismissed from their position without delay, and another employee shall be appointed as the Safety Manager.
3. For diagnostic radiation generating devices, inspections shall be completed before the expiration of the inspection validity period, and a safety manager shall participate when the inspection agency conducts the inspection.
4. When radiation workers perform duties involving a risk of radiation exposure, such as the operation, manipulation, management, inspection, and examination of diagnostic radiation generating devices, they shall wear a dosimeter, such as a film badge or a T-L badge; and when applying for the measurement of radiation exposure doses of radiation workers, those subject to measurement shall not be omitted.
5. If an order to prohibit the use of diagnostic radiation generating devices or radiation shielding facilities, and to repair, calibrate, or re-inspect, is received from the Mayor, County Governor, or District Head pursuant to Article 8, Paragraph 3, it shall be complied with without delay.
6. For those whose radiation exposure dose measurements exceed the dose limits for radiation workers specified in Attached Table 3, necessary measures, such as health examinations, shall be taken without delay.
7. Take measures to prevent radiation exposure to persons entering the radiation area other than radiation-related workers.
8. Take necessary measures for the safety management of the facility to obtain appropriate diagnostic image information.
Article 13 (Health Examinations for Radiation-Related Workers)
① Pursuant to Article 37, Paragraph 2 of the Act, the founder or manager of a medical institution shall conduct health examinations for radiation-related workers every two years in accordance with the Health Examination Checklist in Annex No. 19. However, for radiation-related workers engaging in radiation-related work for the first time, a health examination in accordance with the Health Examination Checklist in Annex No. 19 shall be conducted prior to commencing work. <Amended Jan. 22, 2010>
② In the health examination pursuant to Paragraph 1, the following matters shall be examined. <Amended 2026.7.9>
1. Medical History Questionnaire
A. Medical History and Family History
B. Occupational History and Exposure History
C. Presence or absence of symptoms of radiation exposure
D. Other Specifics
2. Clinical Examination and Blood Tests
A. Clinical examination: Symptoms of the eyes, skin, nervous system, and hematopoietic system (hemodyneogenesis system), etc.
B. Blood test: Amount of hemoglobin in peripheral blood, red blood cell count, white blood cell count, and platelet count
3. Additional examination (applicable only when it is difficult to assess the health status or a disease is suspected based on the examination results under Subparagraphs 1 and 2)
A. Peripheral Blood Smear Test
B. Slit-lamp microscopy
C. Other tests deemed necessary by a physician
③ Notwithstanding Paragraph 1, if a health examination falling under any of the following subparagraphs has been conducted for a radiation-related worker, it shall be deemed that the health examination pursuant to Paragraph 1 has been conducted. <Newly added Feb. 26, 2015; Jun. 30, 2021; Jul. 9, 2026>
1. Special health examination conducted pursuant to Article 130, Paragraph 1 of the Occupational Safety and Health Act or pre-placement health examination conducted pursuant to Paragraph 2 of the same Article
2. A health examination conducted pursuant to Article 17-3, Paragraph 3 of the Veterinarian Act, as a health examination pursuant to Article 13, Paragraph 1 of the Rules on Safety Management of Radiation Generating Devices for Animal Diagnostics
3. A health examination conducted pursuant to Article 91, Paragraph 1, Subparagraph 2 of the Atomic Energy Safety Act, as a health examination pursuant to Article 121, Paragraph 1 of the Enforcement Rule of the same Act
Article 14 (Preparation, Keeping, and Preservation of Documents) Pursuant to Article 37, Paragraph 4 of the Act, the founder or manager of a medical institution, the head of an inspection or measurement agency, and the mayor, county governor, or district head shall prepare, keep, and preserve the documents concerning diagnostic radiation generating devices, etc., specified in Annex 7. <Amended June 30, 2021>
Article 15 Deleted <2021.6.30>
Article 16 (Guidance and Supervision)
① Pursuant to Article 37, Paragraph 1 of the Act, the Mayor, County Governor, or District Head shall guide and inspect the safety management status of diagnostic radiation at medical institutions and, if necessary, order the founder or manager of the medical institution to rectify the situation.
② The Commissioner of the Korea Disease Control and Prevention Agency may provide necessary guidance and supervision to inspection and measurement agencies to ensure the fairness and efficiency of inspections, etc. <Amended Jan. 22, 2010; Nov. 15, 2012; Mar. 23, 2013; Sep. 11, 2020>
③ The Commissioner of the Korea Disease Control and Prevention Agency or the Mayor, County Governor, or District Head may, if necessary for safety management, require the founder or manager of a medical institution or the head of an inspection or measurement agency to report necessary matters or submit data. <Amended Jan. 22, 2010; Nov. 15, 2012; Mar. 23, 2013; Sep. 11, 2020>
④ The Mayor, County Governor, or District Head shall submit the status of safety management of diagnostic radiation generating devices and the status of management of radiation-related personnel as of March 31 of each year to the Commissioner of the Korea Disease Control and Prevention Agency by May 31, through the Mayor of a Special City, Metropolitan City, or Governor of a Province or Special Self-Governing Province, in accordance with Forms No. 17 and No. 18. <Amended 2010.1.22, 2013.3.23, 2020.9.11>
Article 17 (Fees)
① Any person who intends to undergo an inspection of a diagnostic radiation generating device pursuant to Article 85, Paragraph 1 of the Act shall pay the fee determined by the head of the inspection agency pursuant to Article 6 to the inspection agency. In this case, the head of the inspection agency pursuant to Article 6 shall post the amount of the fee and the method of payment, etc., on the internet homepage of the relevant agency.
② Any person who intends to receive training for the Safety Manager of a Diagnostic Radiation Generating Device pursuant to Article 85, Paragraph 1 of the Act shall pay the fee determined by the head of the Safety Manager Training Institution pursuant to Article 37, Paragraph 3 of the Act (hereinafter referred to as the "Safety Manager Training Institution" in this Paragraph) to the Safety Manager Training Institution. In this case, the head of the Safety Manager Training Institution shall post the amount of the fee and the method of payment, etc., on the institution's internet homepage, etc.
[This Article Newly Established 2021.6.30]
[Former Article 17 moved to Article 18 <2021.6.30>]
Article 18 (Exclusion of Application) Articles 4, Paragraph 5, 9, 10, and 13 shall not apply to medical institutions that use only diagnostic X-ray generators or dental diagnostic X-ray generators (excluding panoramic and cephalometric devices) among diagnostic radiation generating devices, and whose total maximum weekly operating load is 10 milliampere-minutes or less.
[Moved from Article 17 <2021.6.30>]
■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Attachment Form No. 1] <Amended July 18, 2025>
Report on Installation and Use of [ ] Diagnostic Radiation Generating Device
[ ] Reuse
(front)
───────────────┬───────────┬──────────────────┬─────────────────────
Receipt No. Receipt Date Issuance Date Processing Period 3 days
───────────────┴───────────┴──────────────────┴─────────────────────
───────┬─────────────────────────┬──────────────────────────────────
Reporter Name of Medical Institution Medical Institution Code
├─────────────────────────┴──────────────────────────────────
Location
(Phone Number/Fax Number: )
├─────────────────────────┬──────────────────────────────────
Creator's Name Creator's Date of Birth
├─────────────────────────┴──────────────────────────────────
Creator's Email
├─────────────────────────┬──────────────────────────────────
Safety Manager Name Safety Manager Date of Birth
├─────────────────────────┴──────────────────────────────────
Safety Manager Email
───────┴────────────────────────────────────────────────────────────
───────┬─────────────────────────┬──────────────────────────────────
Installer Installer's Name Company Name
├─────────────────────────┴──────────────────────────────────
Email
├─────────────────────────┬──────────────────────────────────
Installation Date Phone Number
───────┴─────────────────────────┴──────────────────────────────────
───────┬─────────────────────────┬──────────────────────────────────
Report Usage Condition [ ] New [ ] Used
Phase Device
Content Type Selling Company and Seller
[ ] Portable [ ] Mounted
(Fill out the selling company and seller only if it is a used item)
├─────────────────────────┴──────────────────────────────────
Exclusion Status [ ] Applicable [ ] Not Applicable
───────┼───────────┬──┬───┬──┬──┬──┬──┬──┬───┬──┬──┬──────────┬─────
Equipment Name Type and Model Inspection Inspection Manufacturing Manufacturing Manufacturing Manufacturing Equipment License Introduction Purchase Purchase Medical Equipment
Year/Month Company Number Year/Month Company Country Number (Report) Type Day Amount Barcode
X-ray Control High Voltage Work and Work Number
Tube device generator inspector
───────┼───┴──┴────┼──┼───┼──┼──┼──┼──┼──┼───┼──┼──┼──────────┼─────
│ │ │ │ │ │ │ │ │ │ │ │ │
├───┬──┬────┤ │ │ │ │ │ │ │ │ │ │ │
│ │ │ │ │ │ │ │ │ │ │ │ │ │ │
───────┼───┴──┴────┼──┼───┼──┼──┼──┼──┼──┼───┼──┼──┼──────────┼─────
│ │ │ │ │ │ │ │ │ │ │ │ │
├───┬──┬────┤ │ │ │ │ │ │ │ │ │ │ │
│ │ │ │ │ │ │ │ │ │ │ │ │ │ │
───────┴───┴──┴────┴──┴───┴──┴──┴──┴──┴──┴───┴──┴──┴──────────┴─────
1. Equipment Number: Enter the equipment number from Annex 1 of the "Standards on Reporting Subjects for Medical Equipment Status and Identification Coding."
2. Permit (Notification) Number: Enter the Ministry of Food and Drug Safety medical device manufacturing (import) permit number.
3. For the introduction type, select and fill out from 1. Purchase, 2. Lease, or 3. Donation.
4. The purchase amount is the amount at the time of purchase and should be written using the Korean currency unit (KRW, ₩) (if purchased abroad, convert using the exchange rate at the time).
5. Medical Equipment Barcode: A unique identification number assigned by the President of the Health Insurance Review & Assessment Service in accordance with the "Standards on Reporting Subjects for Medical Equipment Status and Identification Coding" (31 characters
Lee).
────────────────────────────────────────────────────────────────────
I hereby report as above in accordance with Article 3, Paragraph 1 of the "Rules on the Safety Management of Diagnostic Radiation Generating Devices."
Year Month Day
Declarant: (Signature or Seal)
To the Mayor, County Governor, or District Head
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
──────┬──────────────────────────────────────────────────────┬──────
Attachment 1. One copy of the inspection report for the diagnostic X-ray generator Fee
2. 1 copy of Radiation Protection Facility Inspection Report None
3. One copy of the Report on Radiation-Related Personnel in accordance with Form No. 6 (in the case of additional installation or reuse without change of personnel
(Excluding deceased cases; if reporting through the Integrated Healthcare Resources Reporting Portal, a copy may be submitted)
4. One original copy of the Certificate of Report in accordance with Form No. 3 attached by the person who transferred or assigned (to be submitted only in cases of receiving a transfer or relocating installation
...and may be omitted if reporting through the Integrated Healthcare Resources Reporting Portal)
5. One copy of the Certificate of Registration for Specialized Medical Equipment (to be submitted only when installing specialized medical equipment, and the Integrated Reporting Portal for Health and Medical Resources
)
6. Original Certificate of Reporting Suspension of Use of Diagnostic Radiation Generating Device in accordance with Form No. 3 (in case of resumption of use after suspension
; this may be omitted if reporting through the Integrated Healthcare Resources Reporting Portal)
7. One copy of the medical device manufacturing license or import license (in accordance with Article 2 of the "Rules on the Installation and Operation of Specialized Medical Equipment")
is excluded if specialized medical equipment is registered, and is omitted if reported through the Integrated Healthcare Resources Reporting Portal
can)
8. One copy of documents proving the fact of purchase or lease, such as tax invoices or contracts (「Regulations on the Installation and Operation of Specialized Medical Equipment
(Excluded in cases where specialized medical equipment has been registered in accordance with Article 2 of the [Rules])
──────┴──────────────────────────────────────────────────────┴──────
210㎜×297㎜[white paper 80g/㎡]
(Back)
━━━━━━━━━━━━━━━━━━━━━━━━
━━━━━━━━━━━━━━━━━━━━━━━━
Important Notes
────────────────────────
In cases where the installation, use, or reuse of a diagnostic radiation generating device has been reported to the Mayor, County Governor, or District Head, a report form must be submitted to the Health Insurance Review & Assessment Service.
It is deemed to have been submitted (Article 12, Paragraph 3 of the Enforcement Rule of the National Health Insurance Act).
────────────────────────
━━━━━━━━━━━━━━━━━━━━━━━━
Processing procedure
─────────────┬──────────
Reporter Processing Agency
├──────────
City County District
─────────────┼──────────
│
┌────────┐ │ ┌────┐
Fill out report ▶ Submit
│ ├──┼─ │ │
│ │ │ │ │
└────────┘ │ └──┬─┘
│ │
│ │
│
│ ▼
│ ┌────┐
Review
│ └──┬─┘
│ │
│
│ ▼
┌────────┐ │ ┌────┐
Issuance of Report Certificate ◀ Approval
│ │ ─┼──┤ │
│ │ │ │ │
└────────┘ │ └────┘
│
─────────────┴──────────
Original file: /LSW/flDownload.do?flSeq=166851937
Original file: /LSW/flDownload.do?flSeq=166851939
Original file: /LSW/flDownload.do?flSeq=166851941
Original file: /LSW/flDownload.do?flSeq=166851943
Notification Form for (Suspension of Use, Transfer, Assignment, Disposal) of Diagnostic Radiation Generating Devices
■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Attachment Form No. 2] <Amended Nov. 7, 2024>
Notification of Suspension of Use of [ ] Diagnostic Radiation Generating Device
[ ] Transfer
[ ] Previous
[ ] Discard
────────────────┬──────────────┬──────────────┬────────────────
Receipt No. Receipt Date Issue Date Processing Period 3 days
────────────────┴──────────────┴──────────────┴────────────────
────────┬───────────────────────┬──────────────────────────────
Reporter Name of Medical Institution Medical Institution Code
├───────────────────────┴──────────────────────────────
Location
(Phone Number/Fax Number: )
├───────────────────────┬──────────────────────────────
Creator's Name Creator's Date of Birth
├───────────────────────┴──────────────────────────────
Email
────────┴──────────────────────────────────────────────────────
────────┬───────────────────────┬──────────────────────────────
Name of Acquirer Organization (Company) Business Registration Number
├───────────────────────┴──────────────────────────────
Location
(Phone Number/Fax Number: )
├───────────────────────┬──────────────────────────────
Representative Name Date of Birth
├───────────────────────┴──────────────────────────────
Email
────────┴──────────────────────────────────────────────────────
(Fill out only in case of transfer or disposal.)
───────────────────────────────────────────────────────────────
Details of the reported device
───────┬─────────────┬───┬────┬───┬───┬───┬──┬───┬─────────────
Equipment Name Type and Model Inspection Serial Number Manufacturer Manufacturer Country of Manufacture Equipment Approval Medical Equipment
Date Year Month Day Number (Report) Barcode
X-ray controller high voltage number
Tube Generator
───────┼───┴────┴────┼───┼────┼───┼───┼───┼──┼───┼─────────────
│ │ │ │ │ │ │ │ │
├───┬────┬────┤ │ │ │ │ │ │ │
│ │ │ │ │ │ │ │ │ │ │
───────┼───┴────┴────┼───┼────┼───┼───┼───┼──┼───┼─────────────
│ │ │ │ │ │ │ │ │
├───┬────┬────┤ │ │ │ │ │ │ │
│ │ │ │ │ │ │ │ │ │ │
───────┴───┴────┴────┴───┴────┴───┴───┴───┴──┴───┴─────────────
1. Equipment Number: Enter the equipment number from Annex 1 of the "Standards on Reporting Subjects for Medical Equipment Status and Identification Coding."
2. Permit (Notification) Number: Enter the Ministry of Food and Drug Safety medical device manufacturing (import) permit number.
3. Medical Equipment Barcode: A unique identification number assigned by the President of the Health Insurance Review & Assessment Service in accordance with the "Standards on Reporting Subjects for Medical Equipment Status and Identification Coding" (31 characters
Lee).
───────────────────────────────────────────────────────────────
I hereby report as above in accordance with Article 3, Paragraph 1 of the "Rules on the Safety Management of Diagnostic Radiation Generating Devices."
Year Month Day
Declarant: (Signature or Seal)
To the Mayor, County Governor, or District Head
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
──────┬───────────────────────────────────────────────┬────────
Attachment 1. One original copy of the Certificate of Reporting in accordance with Form No. 3 (in the case of reporting through the Integrated Healthcare Resources Reporting Portal, the fee
can be omitted) None
2. One copy of documents verifying the transfer/assignment (submit only when filing a transfer report)
3. One copy of documents confirming disposal (submit only when filing a disposal report)
4. One copy of documents verifying the transfer (submit only when filing a transfer report)
──────┴───────────────────────────────────────────────┴────────
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Important Notes
───────────────────────────────────────────────────────────────
If the suspension of use, transfer, relocation, or disposal of a diagnostic radiation generating device has been reported to the Mayor, County Governor, or District Head, report to the Health Insurance Review & Assessment Service
It is deemed that the document has been submitted (Article 12, Paragraph 3 of the Enforcement Rule of the National Health Insurance Act).
───────────────────────────────────────────────────────────────
210mm×297mm [white paper 80g/㎡]
Original file: /LSW/flDownload.do?flSeq=166851945
Original file: /LSW/flDownload.do?flSeq=166851947
Original file: /LSW/flDownload.do?flSeq=166851949
Certificate of Notification (Installation and Use, Transfer, Relocation, Cessation of Use) of Diagnostic Radiation Generating Devices
■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Attachment Form No. 3] <Amended Nov. 7, 2024>
Certificate of Reporting for Installation and Use of [ ] Diagnostic Radiation Generating Device
[ ] Transfer
[ ] Previous
[ ] Disable
(front)
───────┬──────────────────────┬────────────────────────
Reporter Name of Medical Institution Medical Institution Code
├──────────────────────┴────────────────────────
Location
(Phone Number/Fax Number: )
├──────────────────────┬────────────────────────
Creator's Name Creator's Date of Birth
├──────────────────────┴────────────────────────
Creator's Email
├──────────────────────┬────────────────────────
Safety Manager Name Safety Manager Date of Birth
├──────────────────────┴────────────────────────
Safety Manager Email
───────┴───────────────────────────────────────────────
───────┬──────────────────────┬────────────────────────
Name of Acquirer Organization (Company) Business Registration Number
├──────────────────────┴────────────────────────
Location
(Phone Number/Fax Number: )
├──────────────────────┬────────────────────────
Representative's Name Date of Birth
├──────────────────────┴────────────────────────
Email
───────┴───────────────────────────────────────────────
───────────────────────────────────────────────────────
Details of the reported device
──────┬───────────────┬───┬──┬───┬────┬───┬──┬───┬─────
Equipment Name Type and Model Inspection Manufacturer Manufacturer Manufacturer Country of Manufacture Equipment License Medical Equipment
Year/Month/Day Number Year/Month/Day Number (Report) Barcode
X-ray Tube Controller High Voltage Number
Generator
──────┼─────┴────┴────┼───┼──┼───┼────┼───┼──┼───┼─────
│ │ │ │ │ │ │ │ │
├─────┬────┬────┤ │ │ │ │ │ │ │
│ │ │ │ │ │ │ │ │ │ │
──────┼─────┴────┴────┼───┼──┼───┼────┼───┼──┼───┼─────
│ │ │ │ │ │ │ │ │
├─────┬────┬────┤ │ │ │ │ │ │ │
│ │ │ │ │ │ │ │ │ │ │
──────┼─────┴────┴────┼───┼──┼───┼────┼───┼──┼───┼─────
│ │ │ │ │ │ │ │ │
├─────┬────┬────┤ │ │ │ │ │ │ │
│ │ │ │ │ │ │ │ │ │ │
──────┴─────┴────┴────┴───┴──┴───┴────┴───┴──┴───┴─────
Reported as above in accordance with Article 3, Paragraph 1 of the "Rules on Safety Management of Diagnostic Radiation Generating Devices"
.
Year Month Day
┏━━┓
Official Seal of Mayor, County Governor, or District Head
┗━━┛
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
210mm×297mm [white paper 80g/㎡]
(Back)
━━━━━━━━━━━━━━━━━━━━━
───┬───┬────┬────┬───
No. Year Month Day Change Details Reason for Change Confirm
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┼───┼────┼────┼───
│ │ │ │
───┴───┴────┴────┴───
Original file: /LSW/flDownload.do?flSeq=166851951
Original file: /LSW/flDownload.do?flSeq=166851953
Original file: /LSW/flDownload.do?flSeq=166851955
Original file: /LSW/flDownload.do?flSeq=166851957
[Attachment Form No. 4] <Revised Nov. 7, 2024>
┏━━━━━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━┓
Application for Reissuance of Certificate of Registration for Diagnostic Radiation Generating Devices Processing Period
┃ ├──────────────┨
Immediately
┠─┬────────┬┬────┬───────┴──────────────┨
Phone number
Ryo
Unit Location
tube
Creator's Name Date of Birth
┃ ├────────┼┴────┴──────────────────────┨
Email
┠─┴────────┼────────────────────────────┨
Types of Medical Institutions General Hospital, Hospital, Oriental Medicine Hospital, Clinic, Dental Hospital, Dental Clinic, Other
┠──────────┼────────────────────────────┨
Report Date
┠──────────┼────────────────────────────┨
Reason for Application
(Describe in detail)
┠──────────┘ ┃
┃ ┃
┃ ┃
┃ ┃
┃ ┃
┃ ┃
┃ ┃
┠───────────────────────────────────────┨
┃ ┃
Pursuant to Article 3, Paragraph 3 of the "Rules on Safety Management of Diagnostic Radiation Generating Devices"
I hereby apply for the reissuance of the declaration certificate as above.
┃ ┃
Year Month Day
┃ ┃
Applicant: (Signature or Seal)
┃ ┃
To the Mayor, County Governor, or District Head
┃ ┃
┠───────────────────────────────────────┨
※ Required Documents
In cases where it is damaged or worn out and unusable, certificate of report stating that it became damaged or worn out and unusable
Original
┃ ┃
Remark) In the event that a lost report certificate is found, the competent Mayor, County Governor, or District Head
.
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166851959
Original file: /LSW/flDownload.do?flSeq=166851961
Original file: /LSW/flDownload.do?flSeq=166851963
■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Attachment Form No. 5] <Amended Nov. 7, 2024>
Notification of Change in Reported Details for Diagnostic Radiation Generating Devices
────────────┬───────┬──────────────────────────────────────
Receipt No. Receipt Date Processing Period Immediate
(Previous change: 3 days)
────────────┴───────┴──────────────────────────────────────
─────┬────────────┬────────────────────────────────────────
Reporter Name of Medical Institution Phone Number
├────────────┼────────────────────────────────────────
Nursing Home Code Fax Number
├────────────┴────────────────────────────────────────
Location
├────────────┬────────────────────────────────────────
Creator's Name Date of Birth
├────────────┴────────────────────────────────────────
Email
─────┴─────────────────────────────────────────────────────
─────┬────────────┬────────────────────────────────────────
Change Before Change After Change
Details
(Use separate sheet if margins are insufficient) (Use separate sheet if margins are insufficient)
─────┴────────────┴────────────────────────────────────────
I hereby report as above in accordance with Article 3, Paragraph 4 of the "Rules on the Safety Management of Diagnostic Radiation Generating Devices."
Year Month Day
Declarant (Signature or Seal)
To the Mayor, County Governor, or District Head
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
─────┬───────────────────────────────────────────────┬─────
Attached Documents 1. One original copy of the Certificate of Reporting pursuant to Form No. 3 (In the case of filing a change report through the Integrated Healthcare Resources Reporting Portal, fee
(The part above can be omitted) None
2. One copy of the inspection report for the diagnostic radiation generator (applicable only to the relocation of a medical institution within the jurisdiction)
3. One copy of the Radiation Protection Facility Inspection Report (Applicable only to the relocation of a medical institution within the jurisdiction)
4. One copy of documents verifying the relocation (applicable only to the relocation of a medical institution within the jurisdiction)
─────┴───────────────────────────────────────────────┴─────
210mm×297mm[White Paper 80g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166851965
Original file: /LSW/flDownload.do?flSeq=166851967
Original file: /LSW/flDownload.do?flSeq=166851969
Report Form for Safety Managers of Radiation-Related Workers (Change, Appointment, Dismissal, Concurrent Position)
■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Attachment Form No. 6] <Amended Nov. 7, 2024>
Radiation Worker [ ] Change Report
Safety Manager [ ] Appointment [ ] Dismissal [ ] Concurrent Position
┌──────────┬────────────┬──────────┬──────────────────────────────┐
Receipt No. Receipt Date Issue Date Processing Period 3 days
└──────────┴────────────┴──────────┴──────────────────────────────┘
────┬──────────────────────────┬───────────────────────────────────
Reporter Medical Institution Name Phone Number
├──────────────────────────┴───────────────────────────────────
Location
├──────────────────────────┬───────────────────────────────────
Creator's Name Creator's Date of Birth
├──────────────────────────┴───────────────────────────────────
Creator's Email
────┴──────────────────────────────────────────────────────────────
───────────────────────────────────────────────────────────────────
Personal Information of Radiation Workers
─────┬────┬────┬──────┬────────┬───────────────────────────────────
Name Date of Birth Start of Employment Job Responsibilities Classification of Medical Professional, etc. Reason for Reporting
(End) Date (License (Qualification) Number)
New Resignation or Suspension Existing
─────┼────┼────┼──────┼────────┼──┼────────┼───────────────────────
│ │ │ │ │ │ │
─────┼────┼────┼──────┼────────┼──┼────────┼───────────────────────
│ │ │ │ │ │ │
─────┼────┼────┼──────┼────────┼──┼────────┼───────────────────────
│ │ │ │ │ │ │
─────┼────┼────┼──────┼────────┼──┼────────┼───────────────────────
│ │ │ │ │ │ │
─────┼────┼────┼──────┼────────┼──┼────────┼───────────────────────
│ │ │ │ │ │ │
─────┼────┼────┼──────┼────────┼──┼────────┼───────────────────────
│ │ │ │ │ │ │
─────┴────┴────┴──────┴────────┴──┴────────┴───────────────────────
───────────────────────────────────────────────────────────────────
Safety Manager
─────┬────┬────┬──────┬───────────┬────────┬───────────────────────
Classification Name Date of Birth Reason for Appointment/Dismissal License (Qualification) Number Training Date
Date
─────┼────┼────┼──────┼───────────┼────────┼───────────────────────
Senior (Concurrent) Position
├────┼────┼──────┼───────────┼────────┼───────────────────────
│ │ │ │ │ │
─────┼────┼────┼──────┼───────────┼────────┼───────────────────────
Dismissal
─────┴────┴────┴──────┴───────────┴────────┴───────────────────────
Pursuant to Article 3, Paragraph 6 and Article 10, Paragraph 2 of the "Rules on Safety Management of Diagnostic Radiation Generating Devices," as above
Reporting.
Year Month Day
Declarant (Signature or Seal)
To the Mayor, County Governor, or District Head
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
──────┬────────────────────────────────────────────────────┬───────
Applicant (Reporter) 1. For the Safety Manager, one copy of the final school graduation certificate (to be submitted only by holders of a Master's degree in science or engineering) or certificate of work experience Fee
Submission Documents 2. One copy of the Health Examination Results Report in accordance with Attached Form No. 19
3. One copy of the Confirmation of Exposure Record issued by the Central Registration Center for Radiation Worker Information (Submit only if applicable) None
──────┼────────────────────────────────────────────────────┤
One copy of the license (qualification) of the responsible public official and safety manager
Checklist
──────┴────────────────────────────────────────────────────┴───────
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Consent Form for Joint Use of Administrative Information
───────────────────────────────────────────────────────────────────
I, in relation to the processing of this matter, [refer to] that the public official in charge, through the joint use of administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act, [regarding] the above
I agree to confirm the items.
※ If the applicant does not consent to the confirmation by the 담당 공무원 (responsible public official) or the joint use of computerized information processing systems and administrative information pursuant to Article 36, Paragraph 1 of the Electronic Government Act
, the applicant must submit the relevant documents in person.
Name of Safety Manager (Signature or Seal)
───────────────────────────────────────────────────────────────────
210mm×297mm [white paper 80g/㎡]
Original file: /LSW/flDownload.do?flSeq=166851971
Original file: /LSW/flDownload.do?flSeq=166851973
Original file: /LSW/flDownload.do?flSeq=166851975
[Attachment Form No. 7] <Revised 2010.1.22>
Diagnostic Radiation Management Log
┏━━━┯━━━━┯━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
1. Name of Medical Institution 2. Status of Radiation-Related Workers (Unit: persons)
Authority
Content Location Doctor Radiologist Dental Radiology Nurse Nursing Dental Industry Machinery
Department Doctor Preceptor Assistant Sanitation No Ta
Founder Specialist Assistant
Statement Jo
┃ ├────┼─┬─┬┼──┼───┼───┼──┼───┼──┼──┼─┼─┼──┨
Safety Management Surface
Manager Heo
statement and
┃ │ │ │Ja││ │ │ │ │ │ │ │ │ │ ┃
rating
┠───┴────┴─┴─┴┴──┴───┴───┴──┴───┴──┴──┴─┴─┴──┨
3. Status of Installation of Diagnostic Radiation Generators (Total number of installed units: units, Total number of shielding facilities: rooms)
┠────────────────────────────────────────────┨
Device Name: Report Number: No. Report Date:
Use: Type: □ Portable □ Stationary Condition: □ New □ Used
┠───────────────┬───┬───┬───┬────┬────┬──────┨
Type and Model Manufacturer Manufacture Number Country of Manufacture Manufacturer Installation Date
Year Month Day Issue Name
X-ray tube controller high voltage foot compatible uncompatible
Live device
┠────┴────┴─────┼───┼───┼───┼────┼────┼──┼───┨
┃ │ │ │ │ │ │ │ ┃
┠────┬────┬─────┤ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ ┃
┠────┴────┴─────┴───┴───┴───┴────┴────┴──┴───┨
Device Name: Report Number: No. Report Date:
Use: Type: □ Portable □ Stationary Condition: □ New □ Used
┠───────────────┬───┬───┬───┬────┬────┬──────┨
Type and Model Manufacturer Manufacture Number Country of Manufacture Manufacturer Installation Date
Year Month Day Issue Name
X-ray tube controller high voltage foot compatible uncompatible
Live device
┠────┴────┴─────┼───┼───┼───┼────┼────┼──┼───┨
┃ │ │ │ │ │ │ │ ┃
┠────┬────┬─────┤ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ ┃
┠────┴────┴─────┴───┴───┴───┴────┴────┴──┴───┨
Device Name: Report Number: No. Report Date:
Use: Type: □ Portable □ Stationary Condition: □ New □ Used
┠───────────────┬───┬───┬───┬────┬────┬──────┨
Type and Model Manufacturer Manufacture Number Country of Manufacture Manufacturer Installation Date
Year Month Day Issue Name
X-ray tube controller high voltage foot compatible uncompatible
Live device
┠────┴────┴─────┼───┼───┼───┼────┼────┼──┼───┨
┃ │ │ │ │ │ │ │ ┃
┠────┬────┬─────┤ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ ┃
┗━━━━┷━━━━┷━━━━━┷━━━┷━━━┷━━━┷━━━━┷━━━━┷━━┷━━━┛
210㎜×297㎜[Preservation Paper (Type 2) 70g/㎡]
Original file: /LSW/flDownload.do?flSeq=166851977
Original file: /LSW/flDownload.do?flSeq=166851979
Original file: /LSW/flDownload.do?flSeq=166851981
Application for (Initial Installation Inspection, Periodic Inspection, Re-inspection) of Diagnostic Radiation Generators
[Attachment Form No. 8] <Revised Nov. 7, 2024>
(Heads)
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
□ Application for Initial Installation Inspection
Periodic inspection of diagnostic radiation generators
□ Re-test
┠─┬────────────┬──────────┬────┬────────────┨
Name of new medical institution Phone number
Cheong
Location
┃ ├────────────┼──────────┬────┬────────────┨
Creator's Name Date of Birth
┃ ├────────────┼──────────┴────┴────────────┨
Creator's Email
┃ ├────────────┼──────────┬────┬────────────┨
Safety Manager Name Date of Birth
┃ ├────────────┼──────────┴────┴────────────┨
Safety Manager Email
┠─┴────────────┴────────────────────────────┨
Inspection Request Details
┠──────────────┬────┬────┬────┬────┬────┬───┨
Device Name Model and First Use Final Inspection Serial Number Country of Manufacture Re-inspection
Model Report Date (Year/Month/Day) and Reason for Manufacture
┃ │ │ │ │ │Sa │ ┃
┠──────────────┼────┼────┼────┼────┼────┼───┨
┃ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ ┃
┠──────────────┴────┴────┴────┴────┴────┴───┨
┃ ┃
Pursuant to Article 7 of the "Rules on Safety Management of Diagnostic Radiation Generating Devices," as above
.
┃ ┃
Year Month Day
┃ ┃
Applicant: (Signature or Seal)
┃ ┃
Your
┃ ┃
┠───────────────────────────────────────────┨
※ Required Documents
For periodic inspections and re-inspections, one copy of the inspection report for the diagnostic X-ray generator
(Applicable only to the most recent single test prior to the test application date; in the case of a re-examination, the relevant periodic test
may be substituted with a copy of the inspection report stating the failure)
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
(back)
※ This application will be processed as follows.
┏━━━━━━━━━━━━┯━━━━━━┓
Reporter Processing Agency
┃ ├──────┨
City·County·District
┠────────────┼──────┨
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃┌────────┐ │ ┌──┐┃
Fill out report ▶ Submit
┃│ │ │ │ │┃
┃│ ├──┼─ │ │┃
┃│ │ │ │ │┃
┃│ │ │ │ │┃
┃└────────┘ │ └┬─┘┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ ┃
┃ │ ▼ ┃
┃ │ ┌──┐┃
Review
┃ │ │ │┃
┃ │ │ │┃
┃ │ │ │┃
┃ │ └┬─┘┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ ┃
┃ │ ▼ ┃
┃┌────────┐ │ ┌──┐┃
Issuance of Report Certificate ◀ Approval
┃│ │ │ │ │┃
┃│ │ ─┼──┤ │┃
┃│ │ │ │ │┃
┃│ │ │ │ │┃
┃└────────┘ │ └──┘┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┗━━━━━━━━━━━━┷━━━━━━┛
Original file: /LSW/flDownload.do?flSeq=166851983
Original file: /LSW/flDownload.do?flSeq=166851985
Original file: /LSW/flDownload.do?flSeq=166851987
Original file: /LSW/flDownload.do?flSeq=166851989
[Attachment Form No. 9] <Revised Nov. 7, 2024>
(Heads)
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
Application for Inspection of Radiation Protection Facilities
┠─┬────────────┬┬────┬─────────────────┨
Name of new medical institution Phone number
Cheong
Location
┃ ├────────────┼┬────┬─────────────────┨
Creator's Name Date of Birth
┃ ├────────────┼┴────┴─────────────────┨
Creator's Email
┃ ├────────────┼┬────┬─────────────────┨
Safety Manager Name Date of Birth
┃ ├────────────┼┴────┴─────────────────┨
Safety Manager Email
┠─┴────────────┼───────────────────────┨
Contents of Radiation Protection Facilities
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┠──────────────┴───────────────────────┨
┃ ┃
Pursuant to Article 7 of the "Rules on Safety Management of Diagnostic Radiation Generating Devices," as above
I am applying for an inspection of the radiation protection facility.
┃ ┃
Year Month Day
┃ ┃
Applicant: (Signature or Seal)
┃ ┃
Your
┃ ┃
┠──────────────────────────────────────┨
※ Required Documents
A. One copy of the inspection report for radiation protection facilities (in the case of re-inspection, the rejected inspection report
Copy of the written report; in case of facility/equipment changes or increased usage, a copy of the previous inspection report.
However, for the first inspection, this is not attached)
B. Maximum rated tube voltage, tube current, and maximum energization of devices installed in radiation protection facilities
indicating the thickness and material name of the shielding wall and the location of the power supply facility of the overlooked radiation protection facility
1 copy of blueprint
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
(back)
※ This application will be processed as follows.
┏━━━━━━━━━━━━┯━━━━━━┓
Reporter Processing Agency
┃ ├──────┨
City·County·District
┠────────────┼──────┨
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃┌────────┐ │ ┌──┐┃
Fill out report ▶ Submit
┃│ │ │ │ │┃
┃│ ├──┼─ │ │┃
┃│ │ │ │ │┃
┃│ │ │ │ │┃
┃└────────┘ │ └┬─┘┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ ┃
┃ │ ▼ ┃
┃ │ ┌──┐┃
Review
┃ │ │ │┃
┃ │ │ │┃
┃ │ │ │┃
┃ │ └┬─┘┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ │ ┃
┃ │ ┃
┃ │ ▼ ┃
┃┌────────┐ │ ┌──┐┃
Issuance of Report Certificate ◀ Approval
┃│ │ │ │ │┃
┃│ │ ─┼──┤ │┃
┃│ │ │ │ │┃
┃│ │ │ │ │┃
┃└────────┘ │ └──┘┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┃ │ ┃
┗━━━━━━━━━━━━┷━━━━━━┛
Original file: /LSW/flDownload.do?flSeq=166851991
Original file: /LSW/flDownload.do?flSeq=166851993
Original file: /LSW/flDownload.do?flSeq=166851995
Original file: /LSW/flDownload.do?flSeq=166851997
[Attachment Form No. 10] <Revised Nov. 7, 2024>
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
Application for Measurement of Radiation Exposure Dose for Radiation Workers
┠─┬────────────┬─┬────┬──┬────┬─────────┨
Name of New Medical Institution Medical Insurance Phone Number
Agency Code
IN
Location
┃ ├────────────┼──┬─────┬───────────────┨
Creator's Name Date of Birth
┃ ├────────────┼──┴─────┴───────────────┨
Creator's Email
┃ ├────────────┼──┬─────┬───────────────┨
Safety Manager Name Date of Birth
┃ ├────────────┼──┴─────┴───────────────┨
Safety Manager Email
┠─┴────────────┼────────────────────────┨
Measurement Period Year Month Year Month (Monthly)
┠──────────────┴────────────────────────┨
Personal details of the measurement subject
┠──┬────────────┬──────┬────────────────┨
Name Resident Registration Number Job Responsibilities Type of Medical Professional, etc. (License Number)
┠──┼────────────┼──────┼────────────────┨
┃ │ │ │ ┃
┠──┴────────────┴──────┴────────────────┨
┃ ┃
Pursuant to Article 7 of the "Rules on Safety Management of Diagnostic Radiation Generating Devices," as above
.
┃ ┃
┃ ┃
┃ ┃
Year Month Day
┃ ┃
┃ ┃
Applicant: (Signature or Seal)
┃ ┃
Your
┃ ┃
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166851999
Source file: /LSW/flDownload.do?flSeq=166852001
Original file: /LSW/flDownload.do?flSeq=166852003
Report on (Initial Installation Inspection, Periodic Inspection, Re-inspection) of Diagnostic Radiation Generators
[Attachment Form No. 11] <Revised 2010.1.22>
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
Initial Installation Inspection Report for Diagnostic Radiation Generators
□ Periodic Inspection
□ Re-test
┃ ┃
Document Number:
Effective Date:
Receive:
Subject: Notification of Test Results
┠─────────┬──────┬────┬──────────────┨
Name of Medical Institution Location
┠─────────┼──────┼────┼──────────────┨
Item Name Usage
┠─────────┼──────┴────┴──────────────┨
Type and Model
┠─────────┼──────────────────────────┨
Country of Manufacture and Manufacturer
┠─────────┼──────────────────────────┨
Serial Number
┠─────────┼──────────────────────────┨
Date of Manufacture
┠─────────┼──────────────────────────┨
Date of initial use report
┠─────────┴──────────────────────────┨
The inspection results for the above device are hereby reported as follows.
┠─────────┬─────────────────┬──┬──┬──┨
Device Name Type and Model Inspection Inspection Judgment
Item Result
X-ray tube controller high voltage generation
device
┠─────────┼─────┴────┴──────┼──┼──┼──┨
┃ │ │ │ │ ┃
┃ ├─────┬────┬──────┤ │ │ ┃
┃ │ │ │ │ │ │ ┃
┠─────────┼─────┴────┴──────┴──┴──┴──┨
Overall Judgment
┠─────────┼──────────────────────────┨
Other
Dog
┠─────────┴──────────────────────────┨
┃┌───┬─────────┐ ┃
Description (Signature)
Manager
┃├───┼─────────┤ ┃
Examiner (Signature)
┃└───┴─────────┘ ┃
Year Month Day
┃ ┏━━┓ ┃
(Name of Inspection Agency) Chief's Seal
┃ ┗━━┛ ┃
┃ ┃
┃ ┃
Address: Phone Number:
┃ ┃
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852005
Original file: /LSW/flDownload.do?flSeq=166852007
Original file: /LSW/flDownload.do?flSeq=166852009
[Attachment Form No. 12] <Revised 2010.1.22>
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
┃ ┃
Inspection Report for Radiation Protection Facilities
┃ ┃
Document Number:
Effective Date:
Receive:
Subject: Notification of Test Results
┃ ┃
The inspection results regarding the radiation shielding facilities are hereby notified as follows.
┃ ┃
┠─────┬─────┬───────┬────────────────┨
Name of Medical Institution Location
┠─────┴─────┴───────┼────┬───────────┨
Inspection Locations and Measurement Locations of Radiation Protection Facilities Determination of Inspection Results
┠───────────────────┼────┼───────────┨
┃ │ │ ┃
┠──────────┬────────┴────┴───────────┨
Rectification method of power supply facility
and Capacity
┠──────────┼─────────────────────────┨
Maximum operating load per week mA·min/week
┠──────────┼─────────────────────────┨
Overall Judgment
┠──────────┼─────────────────────────┨
Other comments
┠──────────┴─────────────────────────┨
┃┌───┬─────────┐ ┃
Description (Signature)
Manager
┃├───┼─────────┤ ┃
Examiner (Signature)
┃└───┴─────────┘ ┃
Year Month Day
┃ ┃
┃ ┏━━┓ ┃
(Name of Inspection Agency) Chief's Seal
┃ ┗━━┛ ┃
┃ ┃
┃ ┃
Address: Phone Number:
┃ ┃
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852011
Original file: /LSW/flDownload.do?flSeq=166852013
Original file: /LSW/flDownload.do?flSeq=166852015
[Attachment Form No. 13] <Revised 2010.1.22>
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
┃ ┃
Radiation Worker Exposure Dose Measurement Report
┃ ┃
┃ ┃
Document Number:
Effective Date:
To:
Subject: Notification of Measurement Results
┃ ┃
The results of the radiation exposure dose measurement for radiation workers are hereby notified as follows.
┃ ┃
┠──────┬──────┬────┬──────────────────┨
Employee Name Resident Registration Number Exposure Period Measurement Result
(Core dose: mSv)
┠──────┼──────┼────┼──────────────────┨
┃ │ │ │ ┃
┠──────┴──────┴────┴──────────────────┨
┃┌───┬─────────┐ ┃
Description (Signature)
Manager
┃├───┼─────────┤ ┃
Measurer (Signature)
┃└───┴─────────┘ ┃
┃ ┃
Year Month Day
┃ ┃
┃ ┏━━┓ ┃
(Name of Measuring Agency) Chief's Seal
┃ ┗━━┛ ┃
┃ ┃
┃ ┃
Address: Phone Number:
┃ ┃
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852017
Original file: /LSW/flDownload.do?flSeq=166852019
Original file: /LSW/flDownload.do?flSeq=166852021
Inspection results of (diagnostic radiation generator, radiation shielding facility)
[Attachment Form No. 14] <Revised 2010.1.22>
□ Inspection results of diagnostic radiation generator
□ Radiation Protection Facility
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
Inspection Year and Inspection Date: From Year Month Day Month Day
┠───────────────────┬─────────────────────────┬───────┬─┬─┬─────┨
Reporting and Inspection of Diagnostic Radiation Generating Devices for Medical Institutions Inspector
Date Situation
┠──┬──┬──┬──┬─┬──┬──┼──┬─┬──┬─┬─┬─┬──┬──┬──┬──┼──┬──┬─┤ Ki│ ├──┬──┨
Document Enforcement Inspection Nursing Home Institution Mail Photography Equipment Use Preparation Hair Manufacturing Manufacturing Device Modification Report Recent Inspection Inspection
Number Date Classification Organization Address Number Real Name Chi Jo Jo Del Number Year Format Format Date Inspector Ja Ja
Code Name Name Country Name Day Day 1 2
┠──┼──┼──┼──┼─┼──┼──┼──┼─┼──┼─┼─┼─┼──┼──┼──┼──┼──┼──┼─┼─┼─┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┗━━┷━━┷━━┷━━┷━┷━━┷━━┷━━┷━┷━━┷━┷━┷━┷━━┷━━┷━━┷━━┷━━┷━━┷━┷━┷━┷━━┷━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852023
Original file: /LSW/flDownload.do?flSeq=166852025
Original file: /LSW/flDownload.do?flSeq=166852027
[Attachment Form No. 15] <Revised 2010.1.22>
Results of Personal Radiation Exposure Dose Measurement for Radiation Workers
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
Measurement Year and Quarter:
┠────────┬───────────┬────────┬─────────┬───┨
Measurement of radiation exposure dose for medical institutions and wearing equipment for radiation workers
Date (mSv)
Institution Name
Name Resident License Authority Agency Authority Mail Measurement Wear Wear Superficial Deep Lens
Registration Separator Name Address Number Date Start End Dose Dose Dose
No. Date Date
┠──┼──┼──┼──┼──┼──┼──┼──┼──┼──┼──┼──┼───┼───┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┗━━┷━━┷━━┷━━┷━━┷━━┷━━┷━━┷━━┷━━┷━━┷━━┷━━━┷━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852029
Original file: /LSW/flDownload.do?flSeq=166852031
Original file: /LSW/flDownload.do?flSeq=166852033
Application for Reissuance of Inspection Report (Diagnostic Radiation Generators, Radiation Protection Facilities, Measurement of Radiation Exposure Doses for Radiation Workers)
[Attachment Form No. 16] <Revised Nov. 7, 2024>
┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
□ of the inspection report, etc. of diagnostic radiation generating devices
□ Application for Reissuance of Radiation Protection Facility
□ Measurement of Radiation Exposure Doses for Radiation Workers
┃ ━─┤ ┃
┃ │ ┃
┃ ━─┘ ┃
┃ ┃
┠───┬─────┬┬────────┬───────────────┨
Applicant's Name Date of Birth
┃ ├─────┼┴────────┴───────────────┨
Address
┃ ├─────┼─────────────────────────┨
Email
┠───┴─────┼─────────────────────────┨
Name of medical institution
┠─────────┼─────────────────────────┨
Location
┠─────────┼─────────────────────────┨
Inspection (Measurement) Date
┠─────────┼─────────────────────────┨
Details of inspection (measurement)
┠─────────┼─────────────────────────┨
Application Count Tong
┠─────────┴─────────────────────────┨
┃ ┃
Pursuant to Article 8, Paragraph 6 of the "Rules on Safety Management of Diagnostic Radiation Generating Devices"
I apply as above.
┃ ┃
┃ ┃
Year Month Day
┃ ┃
Applicant: (Signature or Seal)
┃ ┃
Your
┃ ┃
┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852035
Original file: /LSW/flDownload.do?flSeq=166852037
Original file: /LSW/flDownload.do?flSeq=166852039
[Attachment Form No. 17] <Revised 2010.1.22>
Status of Safety Management for Diagnostic Radiation Generating Devices
┏━━━┯━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
Diagnostic radiation generator for regional radiation protection facilities
┠─┬─┼──┬───┬─┬──┬──┬──┼─┬─┬─┬─┬─┬──┬─┬─┬──┬──┬──┬──┨
City City Type Medical City Inspection Measures Judgment Intestinal Form Hair Preparation Manufacturing Preparation Inspection Judgment Inspection Measures
ㆍ ㆍ Organization Name Date Item Result Value Diagram Del Group Number Group Date Result Target Item
Province County Name Name Name Day National Affairs Whether
┃ │ㆍ│ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
phrase
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼───┼─┼──┼──┼──┼─┼─┼─┼─┼─┼──┼─┼─┼──┼──┼──┼──┨
Total
┗━┷━┷━━┷━━━┷━┷━━┷━━┷━━┷━┷━┷━┷━┷━┷━━┷━┷━┷━━┷━━┷━━┷━━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852041
Original file: /LSW/flDownload.do?flSeq=166852043
Original file: /LSW/flDownload.do?flSeq=166852045
[Attachment Form No. 18] <Revised 2010.1.22>
Current Status of Management of Radiation-Related Workers
┏━┯━┯━━┯━━┯━┯━━━━━━━━━━━━━━━━━━━━━━━━━━┯━┓
City City Medical Medical Radiation Workers Total
ㆍ ㆍ For institutions For institutions
Degree Group Type Name of Imaging of Dental Room Interval Nursing Dental Business Excess Group
ㆍ Trading Department Department Office Household Assistant Sanitation Non-exposure Radiation Treatment
Female Specialist Doctor Doctor Doctor Assistant
Adverb Adverb Adverb
┠─┼─┼──┼──┼─┼─┼───┼─┼─┼─┼──┼──┼─┼─┼──┼─┼─┨
┃ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ │ ┃
┠─┼─┼──┼──┼─┼─┼───┼─┼─┼─┼──┼──┼─┼─┼──┼─┼─┨
Total
┗━┷━┷━━┷━━┷━┷━┷━━━┷━┷━┷━┷━━┷━━┷━┷━┷━━┷━┷━┛
210㎜×297㎜[Standard Paper 60g/㎡ (Recycled)]
Original file: /LSW/flDownload.do?flSeq=166852047
Original file: /LSW/flDownload.do?flSeq=166852049
Original file: /LSW/flDownload.do?flSeq=166852051
(Radiation Workers, Frequent Visitors, Radiation-Related Workers) Health Examination Form
■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Attachment Form No. 19] <Amended July 9, 2026>
[ ]Radiation Workers [ ]Health Examination Form for Frequent Visitors
[ ] Radiation workers
※ Mark the appropriate box in [ ] with a √.
──────┬──┬──────────────────────┬───────────────────
Subject's Name Date of Birth Occupation (License Type)
├──┴──────────────────────┼───────────────────
Affiliation and Department Name Start Date
──────┴─────────────────────────┴───────────────────
──────┬─────────────────────────────────────────────
Inspection Date
──────┴─────────────────────────────────────────────
──────┬─────────────────────────────────────────────
Medical History
├─────────────────────────────────────────────
Family History
├─────────────────────────────────────────────
Occupational History and Exposure History (※ Please indicate workplace, content, duration, radiation dose, etc.)
├─────────────────────────────────────────────
Presence of symptoms of radiation exposure: [ ] None [ ] Present (※ If symptoms are present, please write details at the bottom)
│
├─────────────────────────────────────────────
Other Specifics
──────┴─────────────────────────────────────────────
──────┬─────────────────────────────────────────────
Clinical Examination [※ Record symptoms of the eyes, skin, nervous system, and hematopoietic system (hemodyneogenesis system), etc.]
──────┴─────────────────────────────────────────────
──────┬───────────────────────┬─────────────────────
Blood Tests Hemoglobin Level (Hb) g/dL Red Blood Cell Count (RBC) 106/µL
├───────────────────────┼─────────────────────
White blood cell count (WBC) 103/µl Platelet count (PLT) 103/µl
──────┴───────────────────────┴─────────────────────
──────┬─────────────────────────────────────────────
Additional tests: Peripheral blood smear, slit-lamp microscopy, and other tests deemed necessary by a physician.
──────┴─────────────────────────────────────────────
──────┬─────────────────────────────────────────────
Comprehensive Opinion
──────┴─────────────────────────────────────────────
Article 91, Paragraph 1, Subparagraph 2 of the Atomic Energy Safety Act and Article 121, Paragraph 1 of the Enforcement Rules of the same Act
Article 37, Paragraph 2 of the Medical Service Act and Article 13, Paragraph 1 of the Rules on Safety Management of Diagnostic Radiation Generating Devices
Article 17-3, Paragraph 3 of the Veterinary Medical Service Act and Article 13, Paragraph 1 of the Rules on Safety Management of Radiation Generating Devices for Animal Diagnostics
Health examinations were conducted for relevant employees as described above in accordance with .
Year Month Day
Attending Physician's Name:
License Number:
Contact:
┏━━┓
Medical Institution Name Official Seal
┗━━┛
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
210㎜×297㎜(Standard paper 60g/㎡(Recycled))
Original file: /LSW/flDownload.do?flSeq=166852053
Original file: /LSW/flDownload.do?flSeq=166852055
Original file: /LSW/flDownload.do?flSeq=166852057
Revision
Ministry of Health and Welfare Decree No. 1185
The Partial Amendment to the Rules on Safety Management of Diagnostic Radiation Generating Devices is hereby promulgated as follows.
July 9, 2026
Minister of Health and Welfare (Seal)
Partial Amendment to the Rules on Safety Management of Diagnostic Radiation Generating Devices
The Rules on Safety Management of Diagnostic Radiation Generating Devices are partially amended as follows.
The part other than the subparagraphs of Article 13, Paragraph 2 shall be as follows.
In the health examination pursuant to Paragraph 1, the following matters shall be examined.
Subparagraphs (a) through (c) of Article 13, Paragraph 2, Item 1 shall be amended as follows, and Item (d) shall be newly added to the same item as follows.
A. Medical History and Family History
B. Occupational History and Exposure History
C. Presence or absence of symptoms of radiation exposure
D. Other Specifics
The part other than the items of Article 13, Paragraph 2, Subparagraph 2 shall be as follows.
Clinical Examination and Blood Tests
Delete Subparagraph (b) of Article 13, Paragraph 2, Item 2; change Subparagraph (a) of the same item to Subparagraph (b); add Subparagraph (a) to the same item as follows; in Subparagraph (b) (formerly Subparagraph (a), change "peripheral blood" to "blood test: peripheral blood" and "red blood cell count and white blood cell count" to "red blood cell count, white blood cell count and platelet count"; and add Item 3 to the same paragraph as follows.
A. Clinical examination: Symptoms of the eyes, skin, nervous system, and hematopoietic system (hemodyneogenesis system), etc.
3. Additional examination (applicable only when it is difficult to assess the health status or a disease is suspected based on the examination results under Subparagraphs 1 and 2)
A. Peripheral Blood Smear Test
B. Slit-lamp microscopy
C. Other tests deemed necessary by a physician
In Article 13, Paragraph 3, replace "the main text or proviso of Paragraph 1 with respect to persons for whom a special health examination or pre-placement health examination has been conducted for radiation-related workers engaged in work involving radiation exposure pursuant to Article 130, Paragraph 1 or 2 of the Industrial Safety and Health Act" with "Paragraph 1 in cases where a health examination falling under any of the following subparagraphs has been conducted for radiation-related workers," and add the following subparagraphs to the same paragraph.
1. Special health examination conducted pursuant to Article 130, Paragraph 1 of the Occupational Safety and Health Act or pre-placement health examination conducted pursuant to Paragraph 2 of the same Article
2. A health examination conducted pursuant to Article 17-3, Paragraph 3 of the Veterinarian Act, as a health examination pursuant to Article 13, Paragraph 1 of the Rules on Safety Management of Radiation Generating Devices for Animal Diagnostics
3. A health examination conducted pursuant to Article 91, Paragraph 1, Subparagraph 2 of the Atomic Energy Safety Act, as a health examination pursuant to Article 121, Paragraph 1 of the Enforcement Rule of the same Act
Form No. 19 attached hereto is hereby attached.
Supplementary Provisions
Article 1 (Enforcement Date) This Rule shall enter into force on the date of its promulgation.
Article 2 (Applicability to Health Examinations) The amended provisions of Article 13, Paragraph 3, Subparagraphs 2 and 3 shall apply from the time the results of a health examination pursuant to the "Rules on the Safety Management of Radiation Generating Devices for Animal Diagnostics" or the "Enforcement Rules of the Atomic Energy Safety Act" are notified after the enforcement of this Rule.
Article 3 (Transitional Measures Following the Amendment of the Health Examination Form for Radiation-Related Workers) The former Health Examination Form for Radiation-Related Workers in effect at the time of the enforcement of this Rule may be used together with Form No. 19 attached pursuant to this Rule until December 31, 2026.
<img src="http://www.law.go.kr/flDownload.do?flSeq=166821959" alt="img166821959" >
</img>
Reason for Enactment/Amendment
[Partial revision]
◇ Reasons for Amendment and Key Contents
The purpose of this is to improve the system so that radiation-related workers, etc., do not have to undergo duplicate health examinations even when applicable laws change due to changes in their duties, by aligning the blood test items for health examinations that varied according to relevant laws, standardizing the format of health examination forms, and establishing mutual recognition of the conduct of health examinations.