Annex 3 — 의약품 등의 안전에 관한 규칙
284019 · 2026-03-05
■ Rules on the Safety of Pharmaceuticals, etc. [Appendix 3-3] <Newly Added Aug. 21, 2014>
Standards for Manufacturing and Quality Control of Medical High-Pressure Gases
(Related to Article 4, Paragraph 1, Subparagraph 6, Item E, Article 48, Subparagraph 5, Item E and Subparagraph 9 of the same article)
1. Definition of Terms
The meanings of the terms used in this standard are as follows.
a. "Medical high-pressure gas" refers to a gas or mixture of gases corresponding to a medical product
says.
B. "Major component gas" refers to a gas that is the main component of medical high-pressure gas.
"Cryogenic gas" refers to a temperature of minus 150 degrees Celsius or lower and a pressure of 1.013 bar.
.
. "Cylinder" refers to a cylindrical vessel suitable for compressed, liquefied, or dissolved gases.
, a device for controlling the natural outflow of gas at atmospheric pressure and room temperature
refers to what is installed.
E. "Manifold" refers to a device for emptying or filling one or more gas containers simultaneously
can It refers to equipment or devices designed to be able to.
bar. "Cylinder Bundle" refers to connecting cylinders to each other via a manifold
An assembly of cylinders fixed together and transported and used as a single device END
says.
Company. "Home Cryogenic Vessel" refers to liquid oxygen for home use
It refers to a portable cryogenic container designed to vaporize and use .
A. "Mobile Cryogenic Vessel" refers to a mobile insulated vessel
It refers to something designed to maintain contents in a liquid state, and a tanker is
This does not apply.
. A "backflow prevention valve" refers to a valve designed to allow flow in only one direction.
. A "tank" is a fixed, insulated container for the storage of liquefied or cryogenic gases
It refers to one designed for this purpose, and is also called a "stationary cryogenic vessel."
K. "Tanker" refers to an insulated vessel for the transportation of liquefied or cryogenic gases
refers to a container fixedly installed on a vehicle.
2. Purpose
This standard establishes the manufacturing and quality control standards for main component gases and medical high-pressure gases
aims to do that.
2.1 Preparation of Principal Component Gases
a. The main component gas can be produced by chemical synthesis or by refining natural resources.
a. To ensure that changes in the quality of external air do not affect the quality of the main component gas
must.
. If the initial stability test is replaced with literature data, the post-marketing stability test is substituted
can do.
d. Matters concerning specimens for storage do not apply.
E. When producing major component gases via a continuous process, continuously monitor for quality
turning.
B. When transporting main component gases, the provisions regarding medical high-pressure gases shall apply.
. In the case of filling the main component gas into a cylinder or portable cryogenic vessel, medical
Complies with matters concerning high-pressure gas.
2.2 Manufacture of Medical High-Pressure Gases and Manufacture Hygiene Management
a. Medical high-pressure gases are generally manufactured in enclosed facilities.
B. Matters applicable to cylinders also apply to cylinder bundles.
Workers, workplaces, and manufacturing facilities, etc., must be managed in a clean and hygienic manner.
3. Worker
3.1 Workers must receive appropriate education and training.
3.2 Work by external contractors, such as maintenance, that may affect the quality of medical high-pressure gases
Members must receive appropriate education and training.
4. Facility and Equipment Management
4.1 Facility Management
a. Cylinders and portable cryogenic vessels shall be inspected in an area separated from non-medical high-pressure gases.
must be prepared, filled, and stored. However, non-medical high-pressure gas and medical high-pressure
If the gas specifications are complied with and manufacturing operations are carried out in accordance with this standard, then the
It may not be so.
b. Sufficient for manufacturing, testing, and storage within the facility to avoid the risk of confusion
Space must be provided and the following requirements must be met.
1) Areas separately marked by gas type
2) Clear identification and disassembly of cylinders and mobile cryogenic vessels at various process steps
ri
. Empty cylinders/household cryogenic containers and filled cylinders/household cryogenic containers are
To protect from inappropriate external conditions, store with the cover closed.
must.
4.2 Facility Management
a. The equipment shall be designed so that suitable gas can be filled into suitable containers.
Generally, there should be no cross-connections between pipes transporting different gases.
However, if cross-connection is required, it is guaranteed that there is no risk of cross-contamination.
must.
B. Tanks and tankers shall be used exclusively for a single gas of the specified quality. However,
The quality of non-medical high-pressure gas is equivalent to that of medical high-pressure gas and complies with this standard
, this may not be the case.
. It is the time for supplying gas to the manifolds of medical high-pressure gas and non-medical high-pressure gas
system, from non-medical high-pressure gas pipelines to medical high-pressure gas pipelines
It must be verified that backflow into the START pipe is prevented.
. The charging manifold is for a single medical high-pressure gas or a specific mixture of medical high-pressure gases
It must be used for medical high-pressure gas. However, the quality of non-medical high-pressure gas must not be the same as that of medical high-pressure gas.
and may not be so if subject to this standard, in which case the
The process must be carried out by continuous production (Campaign).
E. Repair and maintenance work on facilities has an adverse effect on the quality of medical high-pressure gas
must not be given.
B. After the tanker transports non-medical high-pressure gas or after maintenance work, again
When used in the operation of medical high-pressure gases, measures such as analysis must be documented.
.
5. Document
5.1 In the manufacturing instructions and records for each manufacturing batch of cylinders and portable cryogenic vessels, the following
It must include the START clause MTQ5b7376ec0ca0114b00020 and be approved by the manufacturing manager.
a. Product Name
B. Manufacturing Number
. Charging work date and time
. Worker
E. Information on the manufacturing units of gases used in the filling operation
B. Equipment Used
Company. Identification Information and Manufactured Quantity
Ah. Charging pre-work
. Key Indicators for Verifying Proper Charging
Vehicle. Inspection result regarding charging completion
K. Marking and Indication Sample
T. Specifications and test results for the finished product
Part. Quantity and Reason for Non-conformity
H. Records regarding deviation management
5.2 For gas supplied to tanks in medical institutions, the manufacturing instructions and records for each manufacturing unit
In addition to the matters specified in 5.1, the following shall be included.
A. Identification Information for Tanks (Tankers)
B. Tank (Tanker) Charging Worker
. Information on supply tanks (supply tankers)
6. Manufacturing Management
6.1 Transfer and Transport of Cryogenic Gases and Liquefied Gases
a. The transfer of cryogenic gases and liquefied gases is verified to avoid all contamination
The procedure must be followed. The transfer piping must be equipped with appropriate devices, such as check valves.
.
B. Transfer hoses used for filling tanks and tankers are dedicated to the connection of specific gases
must be used. The connecting device used for connecting the tank and tanker must be the same
If not for dedicated gas use, it must be properly managed.
. When adding transported gas to a tank containing the same gas, the quality of the gas
, a sample must be collected and tested.
6.2 Filling and Marking Materials for Cylinders and Portable Cryogenic Vessels
a. Before filling, determine the manufacturing unit and in accordance with relevant standards and permit (notification) requirements
Must be managed.
a. In the case of a continuous process, process control must be performed to ensure that the gas conforms to the specified standards.
.
. Cylinders, portable cryogenic vessels, and valves are inspected before first use in production
and must be maintained and managed, and single medical high-pressure gas or specific mixed medical
Must be used exclusively for high-pressure gas.
. Inspection and maintenance work must not affect the quality and safety of the pharmaceutical product.
ni doen.
E. Install a valve to check for cylinder contamination in the following cases
An internal visual inspection must be performed before starting, and after installation, the locked state must be maintained
must be supported to prevent contamination.
1) The first cylinder used
2) If the valve is removed
3) Whenever the valve is replaced
B. Cylinders and mobile cryogenic vessels must be traceable and properly maintained.
. Before charging, the following items must be checked.
1) Positive pressure test inside the cylinder
2) Verification inspection for removal of markings related to previous manufacturing batch
3) Inspection to verify removal of damaged markings and replacement
4) Visual inspection of cylinders, portable cryogenic vessels, and valves, and discharge port extension
Connection Device Inspection
5) Confirmation of test execution in accordance with relevant laws for cylinders and portable cryogenic vessels
A. Cylinders and mobile cryogenic vessels recovered for the purpose of refilling are not contaminated.
must be managed to such an extent.
J. Must verify that the cylinder and mobile cryogenic vessel are properly filled
.
. Each filled cylinder shall be subjected to a leak test before sealing.
K. After filling, a cover must be placed on the cylinder valve to prevent contamination of the outlet
.
T. Cylinders and portable cryogenic containers shall be affixed with markings. Manufacturing Number
and expiration date may be indicated separately.
part. Medical high-pressure gas manufactured by mixing two or more different types of gases
, it must be proven that the mixture is homogeneous.
7. Quality Control
7.1 Each manufacturing unit of medical high-pressure gas shall be tested in accordance with the item approval (notification) standards
must.
7.2 Standards for Product Approval (Notification), the sample collection plan and test
The following must be followed, and unless otherwise validated, the function
A quantity test must be performed.
a. Cylinder
1) In the case of a single medical high-pressure gas charged through a manifold, each manifold
For the gas coming from at least one cylinder during every charging cycle, every
Verification and content testing must be performed whenever the cylinder of the knives is replaced.
2) When filling a single medical high-pressure gas into one cylinder at a time, respectively
, confirm
Phosphorus and content tests must be conducted.
3) Manufactured by mixing two or more gases in a cylinder with the same manifold
In the case of medical high-pressure gases, content and verification tests must be performed for each cylinder.
must.
4) For pre-mixed gases, continuous in-piping test for the mixture being filled
In cases where it is not, it follows the same standards as a single medical high-pressure gas, and otherwise
In the case of , the same standards as for mixed medical high-pressure gases must be followed.
B. Mobile cryogenic vessel
1) The final test shall include content and verification tests for each container.
2) Important characteristics of the gas remaining in each container confirmed before refilling
, testing can be performed per manufacturing batch.
. Medical institution tanks and household cryogenic containers
1) Medical institution tanks or household cryogenic containers refilled from dedicated tankers
if a test report for the tanker contents is provided at the time of shipment,
Specimen collection before and after START may be omitted.
2) In such cases, the gas specifications within the container are maintained due to continuous refilling
must prove that.
7.3 Specimens for storage shall not be stored unless otherwise specified.
7.4 Post-market safety when initial safety testing of medical high-pressure gases is replaced with literature data
Qualitative testing may be substituted.
8. Transport of packaged medical high-pressure gases
Filled cylinders and household cryogenic containers should be handled safely and kept clean.
Must be transported so that it can be used.