■ Rules on Safety Management of Radiation Generating Devices for Animal Diagnostics [Appendix 3] <Amended Oct. 17, 2019>
Dose Limits for Radiation Workers (Related to Article 4, Paragraph 7)
┌──────┬────────────┬─────────────────────┐
│Exposure Classification │Division │Dose Limit │
├──────┼────────────┼─────────────────────┤
│1. Effective Dose │a. If pregnant │Must be 1 mSv (0.1 rem) or less per 3 months. │
│ ├────────────┼─────────────────────┤
│ │b. Other cases of workers │must be 50 mSv (5 rem) or less per year, and 5 years │
│ │U │must be 100 mSv (10 rem) or less │
│ │ │. │
├──────┼────────────┼─────────────────────┤
│2. Equivalent Dose │a. Lens │Must be 150 mSv (15 rem) or less per year. │
│ ├────────────┼─────────────────────┤
│ │or. Skin (in case of pregnancy abdominal │Must be 500 mSv (50 rem) or less per year. │
│ │(Excluding the surface of the part)ㆍ │ │
│ │Hands and Feet │ │
│ ├────────────┼─────────────────────┤
│ │. Abdominal surface (in the case of pregnancy │Annual must be 2 mSv (0.2 rem) or less. │
│ │Right) │ │
└──────┴────────────┴─────────────────────┘
Remarks
1. "Effective dose" refers to the degree of risk resulting from the dose distribution among tissues within the human body as a single quantity
To represent the equivalent line of each tissue for all human tissues exposed to radiation
amount by the organizational weight of the corresponding organization. In this case, the entire
The sum of the weights of exposed tissues shall be 1.
2. "Equivalent dose" refers to the absorption line used to indicate the dose received by a specific organ of the human body
dose by the radiation weighting factor of the corresponding radiation. In this case, for diagnostic X-rays
The radiation weighting value of is set to 1.
3. Details necessary for measurement, such as methods for measuring dose limits for radiation workers, are...
The head of the agency shall determine this upon approval by the Director of the Animal and Plant Quarantine Agency.
4. The dose limit per 3 months is for radiation workers using T-Al badges
It shall be the amount measured at the end of each quarter, and in the case of using a film badge, during the quarter
The measured amounts shall be considered as the sum.
5. The annual dose limit is for each year, regardless of when the first exposure dose was measured during the year
Based on the last day of December, the dose for radiation workers using T-Al badges
The limit shall be the sum of the quantities measured at the last day of each quarter.
6. The 5-year cumulative dose limit is January 2008, regardless of when the initial exposure dose was measured
It shall be the cumulative dose over every 5 years based on 1 day.