Annex 2 — 동물 진단용 방사선발생장치의 안전관리에 관한 규칙

288075 · 2026-07-09

■ Rules on Safety Management of Radiation Generating Devices for Animal Diagnostics [Appendix 2] <Amended 2012.12.11> Inspection Standards for Radiation Protection Facilities (Related to Article 4, Paragraph 4) 1. Radiation Protection Facilities Radiation protection facilities in accordance with Item (a) or (b) must be provided. However, if the radiation control device is animal-generated In cases where it is separated from a single radiation generating device, radiation shielding facilities in accordance with subparagraph (c) or (d) Must additionally equipped. a. When installing radiation shielding facilities in the examination room 1) A place isolated from the animal treatment area where an animal diagnostic X-ray generator is installed and imaging is performed (this The ceiling, floor, and walls of the "examination room" (hereinafter referred to as "examination room" in this table) shall have radiation shielding facilities (hereinafter referred to as "examination room" in this table) must be installed, and the radiation leakage dose and acid The total dose shall be 2.58 × C/kg (100 mR per week) or less. However, where persons pass or does not require the installation of barriers in non-residential directions, and The sum of leakage and scattered doses measured outside the protective wall installed in the direction of human habitation must be 2.58×C/kg (10mR per week) or less. 2) If the shielding material of the barrier is made of lead plate, it must meet the following requirements. a) The lead sheet shall not wrinkle or bend due to its own weight, and physical impact must not be directly subjected to the attack. b) The parts where lead plates connect shall be overlapped with lead plates having a width of at least 1.5 cm, and the lead plates shall be The portion connecting to shielding materials other than the plate must be overlapped with a lead plate at least 2.5 cm wide. b. In the case of installing radiation shielding facilities in the imaging area Install an animal diagnostic X-ray generator by partially opening and partitioning the animal treatment space, and perform imaging In the vicinity of the place where animal treatment is performed (hereinafter referred to as the "shooting area" in this table), there are animal treatment facilities with a height of 1.8m or more A protective partition must be installed, and radiation leaks measured outside the protective partition for animal treatment The sum of snow dose and scattered dose shall be 2.58 × C/kg (100 mR per week) or less. However, Protective partitions do not need to be installed in directions where traffic does not pass or reside, and radiation tubes Leakage line measured from the outside of a protective partition for veterinary treatment installed in the direction where persons other than staff reside The sum of the dose and scattered dose shall be 2.58 × C/kg (10 mR per week) or less. . Radiation shielding facilities of the control room The control room is isolated from the imaging room or imaging area and is equipped with a device to control animal diagnostic radiation. It refers to a place, and if there is a control room, install patient viewing windows, etc., to observe the animal's movements. It must be done, and a protective wall must be installed in the part adjacent to the examination room or examination area, and a viewing window for the patient must be installed. . d. Radiation protection facilities of the control center The control station is a device for controlling animal diagnostic radiation by partially opening and partitioning it from the imaging room or imaging area. is installed, and if a control station is present, protective partitions and circular view A window must be installed, and radiation shielding material must be used for the patient viewing window. 2. Radiation shielding facilities for mobile animal diagnostic radiation generators a. In cases where a mobile animal diagnostic X-ray generator is used for on-site medical treatment at farms, etc., subparagraph 4 Must be equipped with an X-ray protective apron for veterinary treatment in accordance with subparagraph (b). b. In cases where a mobile animal diagnostic X-ray generator is used within an animal hospital, Radiation protection facilities must be provided. 3. Number of installed X-ray generators for animal diagnosis Install one unit per shooting room or shooting area; however, if it is unavoidable to install two or more units, Two units must not be operated simultaneously by utilizing conversion switches, etc. However, animal diagnosis Among the radiation generating devices for , a dedicated scanning room must be provided for the computed tomography scanner. 4. Device for Radiation Damage Protection When using a radiation generator for animal diagnosis, the animal's owner or manager and radiation Radiation protection devices intended to protect line-related workers, etc., are [subject to] the notification by the Commissioner of the Food and Drug Safety Ministry It must comply with Korean Industrial Standards (KS) pursuant to the Industrial Standardization Act, and its types are as follows: . In this case, each animal hospital must be equipped with one radiation protective apron for animal treatment in accordance with subparagraph (b), but In combination, at least two radiation protection devices must be provided. a. Radiation shielding partition for veterinary medical use or radiation shielding partition for medical use b. Radiation protection apron for veterinary medical use or radiation protection apron for medical use . Radiation protection gloves for animal medical care or radiation protection gloves for medical care d. Other devices for protection against radiation damage