■ Rules on Safety Management of Radiation Generating Devices for Animal Diagnostics [Appendix 1] <Amended Aug. 26, 2019>
Inspection Standards for Radiation Generating Devices for Animal Diagnostics (Related to Article 4, Paragraph 1)
┏━━━━━━━━━┯━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━┓
┃Type of device │Exam Items │Based ┃
┠─────────┼────────────┼───────────────────────────┨
┃1. Animal Diagnostic Exfoliant │a. When verifying grounding facilities │The grounding terminal of the grounding facility is separately determined by the Minister of Trade, Industry and Energy ┃
┃Line device │Hump │In accordance with the technical standards for electrical installations, Class 3 grounding work ┃
┃ │ │You must check if it is grounded on the grounding wire. ┃
┃ ├────────────┼───────────────────────────┨
┃ │or. In case of external leakage current │Animal diagnostic radiation generating device or animal diagnostic radiation ┃
┃ │Hump │generator, a person ┃
┃ │ │human body or ┃
┃ │ │conductive parts that can be contacted by animals to the ground or other ┃
┃ │ │securely grounded, ┃
┃ │ │It must be 0.1mA or less. However, for devices other than stationary devices ┃
┃ │ │has an external leakage current of 0.5mA with the ground wire disconnected ┃
┃ │ │must be. ┃
┃ ├────────────┼───────────────────────────┨
┃ │. Reproduction of exposure dose │Evaluating the performance and reliability of a radiographic generator for animal diagnosis ┃
┃ │test (automatic exposure agent │As such, the coefficient of variation for the radiation dose is 0.05 ┃
┃ │(including fish) │must be. The coefficient of variation refers to the average value of the measured radiation dose ┃
┃ │ │It refers to the standard deviation for , and is calculated using the following formula. ┃
┃ │ │ ┃
┃ │ │ Remarks: CV=Coefficient of Variation ┃
┃ │ │ S=Standard deviation of radiation dose measurements for the population ┃
┃ │ │ =Average of measurements ┃
┃ │ │ =i-th radiation dose measurement ┃
┃ │ │ n=Number of measurements(The number of measurements must be 5 or more) ┃
┃ │ │ ┃
┃ │ │ ┃
┃ ├────────────┼───────────────────────────┨
┃ │ │ ┃
┃ │ │ ┃
┃ │. Tube voltage test │The Percentage Mean Error (PAE) of the tube voltage relative to the set value (indicated value) is ┃
┃ │ │. This is expressed by the following formula: ┃
┃ │ │calculates. ┃
┃ │ │ ┃
┃ │ │ Remarks: PAE=Percentage Mean Error ┃
┃ │ │ Xp=Setting Value(Indicator Value) ┃
┃ │ │ =Average of measured radiation dose ┃
┃ │ │ ┃
┃ │ │ ┃
┃ ├────────────┼───────────────────────────┨
┃ │E. Tube Current Test │For short-time ratings, the percentage mean error with respect to the indicated value is ┃
┃ │ │Must be within ±15%. ┃
┃ ├────────────┼───────────────────────────┨
┃ │B. X-ray irradiation field(照射 │The illuminance of the light irradiation field projected from the irradiation field control mechanism is EXELTQ1b41df6008b3ead501922END ┃
┃ │野)Test │The distance from the focal point of the SID to the surface of the water (hereinafter referred to as "SID") ┃
┃ │ 1) Roughness Test │) The average illuminance at 100 cm must be 100 Lux or higher ┃
┃ │ │. ┃
┃ │ │ ┃
┃ │ 2) X-ray irradiation field and │Difference between the center of the X-ray irradiation field and the center of the light irradiation field and each ┃
┃ │Difference in light irradiation field │The difference in the surroundings is when the center of the line of use was incident perpendicularly ┃
┃ │Test │It must be within ±2% of SID during X-ray irradiation. However, ┃
┃ │ │When the direction is constant, it must be within ±1% of SID. ┃
┃ │ │ ┃
┃ │ 3) When measuring the probe │equipped with an irradiation tube for the purpose of local area imaging 1) and ┃
┃ │experience │2) is impossible, at the irradiation field ┃
┃ │ │must be 7cm or less. ┃
┃ ├────────────┼───────────────────────────┨
┃ │Inc. Other necessary tests │ ┃
┃ │ 1) When indicating fault depth │The tolerance for the indicated scale shall be within ±1 mm. ┃
┃ │Test(For Tomography Only │ ┃
┃ │corresponding) │ ┃
┃ │ 2) mAs Test │a) In the case of a transformer type: when the energization time is 0.1 seconds or more ┃
┃ │ │has a percentage mean error of ±20% for the indicated value, 0.1 second ┃
┃ │ │, the percentage mean error for the indicated value is ± ┃
┃ │ │Make it within 20%. ┃
┃ │ │b) For condenser X-ray devices: For cases less than 10 mAs ┃
┃ │ │has an error of ±2mAs for the indicated value, and ┃
┃ │ │, the percentage mean error for the indicated value is ±20% ┃
┃ │ │must be within. ┃
┃ │ 3) Field Agreement Test │a) When SID is maximized, the minimum irradiation field is 5 cm × 5 cm. ┃
┃ │(Viewing device only │must be adjustable to lower. ┃
┃ │corresponding) │b) The X-ray irradiation field exceeds the effective area of the image plane during fluoroscopy ┃
┃ │ │must not. ┃
┃ │ │) The error between the center of the receiving plane and the center of the X-ray irradiation field is ┃
┃ │ │SID. ┃
┃ │ │ ┃
┃ │ 4) Incident radiation dose rate │Incident radiation dose at the center point of the radiation line incident on the subject ┃
┃ │test(vision device only │The START rate is 10 R/min (2.58 × 10 ) when an automatic exposure controller is present. ┃
┃ │corresponding) │-3C/kg·min), 5R/min if there is no automatic exposure controller ┃
┃ │ │Tube capable of bringing (1.29×10⁻³ C/kg·min) or more ┃
┃ │ │Does not operate under any combination of voltage and tube current ┃
┃ │ │must. ┃
┃ │ │ ┃
┃ │ 5) Cancer X-ray Test │5cm from surfaces where humans or animals may come into contact ┃
┃ │(Condenser X-ray │and 2 mR/hr (5.16 × 10 ┃
┃ │(Applies only to devices) │C/kg·hr). ┃
┃ │ │Allowable value for the indicated value of the irradiation time by timer adjustment ┃
┃ │ 6) Timer Test (Shooting │must be less than or equal to the value shown in the following table. ┃
┃ │applicable only to devices for START │ ┃
┃ │) │ ┃
┃ │ ├─────┬────────┬────────────┨
┃ │ │RectificationMethod │Display value │Tolerance ┃
┃ │ ├─────┼────────┼────────────┨
┃ │ │Single-phase device │T<10pulses │±0 pulse a) ┃
┃ │ │ │10pulse≤T │±10% ┃
┃ │ ├─────┼────────┼────────────┨
┃ │ │Dasang Stop│T<0.01 seconds │-1.5msec~+6msec ┃
┃ │ │value │0.01 sec≤T<0.04 sec│±20% ┃
┃ │ │b), c) │0.04 seconds≤T │±10% ┃
┃ │ ├─────┼────────┼────────────┨
┃ │ │Inverter type cabinet│T<0.01 seconds │±1msec ┃
┃ │ │value │0.01 seconds≤T │±10% ┃
┃ │ ├─────┴────────┴────────────┨
┃ │ │ ┃
┃ │ │ ┃
┃ │ │a) 1 pulse corresponds to 1/2 of 1 cycle of the power supply frequency ┃
┃ │ │refers to the time. In the case of a single-phase device, the effective pulse electric ┃
┃ │ │angle. The first 1 pearl ┃
┃ │ │A range greater than that electrical angle in the last pulse ┃
┃ │ │is also considered a regular pulse. ┃
┃ │ │ Regarding the timing of injection, the initial injection section within -45° to 0° ┃
┃ │ │The subsequent portion within 0°–45° at the time of START and termination was filmed ┃
┃ │ │is not included in the time. ┃
┃ │ │b) Depending on the input method of each device, from the input timing, regular operation ┃
┃ │ │MTQ8d21810701d06d5d00386 In the case of a device having an initial input section leading up to START ┃
┃ │ │, the initial part is not included in the shooting time ┃
┃ │ │. ┃
┃ │ │d) In a multiphase rectifier, the rising portion of the high-voltage waveform and the predetermined pipe ┃
┃ │ │The projection time and cutoff at each 87% time for the voltage value ┃
┃ │ │point, with the time in between, as the shooting time ┃
┃ │ │. ┃
┃ │ ├──────┬──────────┬─────────┨
┃ │7) half-value layer(half-value layer: │Classification │Measurement tube voltage(kV) │Minimum half-value layer (mm Al)┃
┃ │The intensity of radiation half ├──────┼──────────┼─────────┨
┃ │Water needed to reduce │For animal diagnosis │60 │1.3 ┃
┃ │Quality Thickness) Test │X-ray device │(Maximum tube voltage is │ ┃
┃ │ │ │70 kV or less device │ ┃
┃ │ │ │value) │ ┃
┃ │ │ ├──────────┼─────────┨
┃ │ │ │80 │2.3 ┃
┃ │ │ │(Maximum tube voltage is 70 │ ┃
┃ │ │ │Device exceeding kV) │ ┃
┃ │ ├──────┴──────────┴─────────┨
┃ │ │measured tube voltage after correcting the error following the tube voltage test ┃
┃ │ │. ┃
┠─────────┼────────────┼───────────────────────────┨
┃2. Computed Tomography │A. Artifacts │Artifacts images must not appear conspicuously ┃
┃AreaDevice │Test │and visually checked whether the centerline is properly aligned ┃
┃ │ │. ┃
┃ ├────────────┼───────────────────────────┨
┃ │B. Slice Thickness Test │The tolerance between the slice thickness set value and the measured value is ┃
┃ │ │Must be within ±1mm. ┃
┃ ├────────────┼───────────────────────────┨
┃ │. Tube voltage test │The percentage mean error of the tube voltage is within ±7% of the indicated value. ┃
┃ │ │must. ┃
┃ ├────────────┼───────────────────────────┨
┃ │. Animal exposure dose measurement │Complies with the manufacturer's self-standards. ┃
┃ │Test │ ┃
┃ ├────────────┼───────────────────────────┨
┃ │E. mAs Test │Complies with the manufacturer's self-standards. ┃
┃ ├────────────┼───────────────────────────┨
┃ │bar. CT Number of │Complies with the manufacturer's self-standards. ┃
┃ │LinearityTest │ ┃
┃ ├────────────┼───────────────────────────┨
┃ │Company. Noise Test │Complies with the manufacturer's self-standards. ┃
┃ ├────────────┼───────────────────────────┨
┃ │Ah. High illuminance (high illuminance │Complies with the manufacturer's self-standards. ┃
┃ │Degree) Spatial resolution │ ┃
┃ │experiment (at small size │ ┃
┃ │High contrast difference │ ┃
┃ │is a material adjacent │ ┃
┃ │Formula when │ ┃
┃ │Special Ability Test) │ ┃
┠─────────┼────────────┼───────────────────────────┨
┃3. Other devices │As determined by the Director of the Quarantine and Inspection Agency │Standards determined by the Head of the Quarantine and Inspection Agency, standards determined by the manufacturer, and also ┃
┃ │Test │applies internationally recommended standards. ┃
┗━━━━━━━━━┷━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━┛