Annex 1 — 동물 진단용 방사선발생장치의 안전관리에 관한 규칙

288075 · 2026-07-09

■ Rules on Safety Management of Radiation Generating Devices for Animal Diagnostics [Appendix 1] <Amended Aug. 26, 2019> Inspection Standards for Radiation Generating Devices for Animal Diagnostics (Related to Article 4, Paragraph 1) ┏━━━━━━━━━┯━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃Type of device │Exam Items │Based ┃ ┠─────────┼────────────┼───────────────────────────┨ ┃1. Animal Diagnostic Exfoliant │a. When verifying grounding facilities │The grounding terminal of the grounding facility is separately determined by the Minister of Trade, Industry and Energy ┃ ┃Line device │Hump │In accordance with the technical standards for electrical installations, Class 3 grounding work ┃ ┃ │ │You must check if it is grounded on the grounding wire. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │or. In case of external leakage current │Animal diagnostic radiation generating device or animal diagnostic radiation ┃ ┃ │Hump │generator, a person ┃ ┃ │ │human body or ┃ ┃ │ │conductive parts that can be contacted by animals to the ground or other ┃ ┃ │ │securely grounded, ┃ ┃ │ │It must be 0.1mA or less. However, for devices other than stationary devices ┃ ┃ │ │has an external leakage current of 0.5mA with the ground wire disconnected ┃ ┃ │ │must be. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │. Reproduction of exposure dose │Evaluating the performance and reliability of a radiographic generator for animal diagnosis ┃ ┃ │test (automatic exposure agent │As such, the coefficient of variation for the radiation dose is 0.05 ┃ ┃ │(including fish) │must be. The coefficient of variation refers to the average value of the measured radiation dose ┃ ┃ │ │It refers to the standard deviation for , and is calculated using the following formula. ┃ ┃ │ │ ┃ ┃ │ │ Remarks: CV=Coefficient of Variation ┃ ┃ │ │ S=Standard deviation of radiation dose measurements for the population ┃ ┃ │ │ =Average of measurements ┃ ┃ │ │ =i-th radiation dose measurement ┃ ┃ │ │ n=Number of measurements(The number of measurements must be 5 or more) ┃ ┃ │ │ ┃ ┃ │ │ ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │ │ ┃ ┃ │ │ ┃ ┃ │. Tube voltage test │The Percentage Mean Error (PAE) of the tube voltage relative to the set value (indicated value) is ┃ ┃ │ │. This is expressed by the following formula: ┃ ┃ │ │calculates. ┃ ┃ │ │ ┃ ┃ │ │ Remarks: PAE=Percentage Mean Error ┃ ┃ │ │ Xp=Setting Value(Indicator Value) ┃ ┃ │ │ =Average of measured radiation dose ┃ ┃ │ │ ┃ ┃ │ │ ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │E. Tube Current Test │For short-time ratings, the percentage mean error with respect to the indicated value is ┃ ┃ │ │Must be within ±15%. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │B. X-ray irradiation field(照射 │The illuminance of the light irradiation field projected from the irradiation field control mechanism is EXELTQ1b41df6008b3ead501922END ┃ ┃ │野)Test │The distance from the focal point of the SID to the surface of the water (hereinafter referred to as "SID") ┃ ┃ │ 1) Roughness Test │) The average illuminance at 100 cm must be 100 Lux or higher ┃ ┃ │ │. ┃ ┃ │ │ ┃ ┃ │ 2) X-ray irradiation field and │Difference between the center of the X-ray irradiation field and the center of the light irradiation field and each ┃ ┃ │Difference in light irradiation field │The difference in the surroundings is when the center of the line of use was incident perpendicularly ┃ ┃ │Test │It must be within ±2% of SID during X-ray irradiation. However, ┃ ┃ │ │When the direction is constant, it must be within ±1% of SID. ┃ ┃ │ │ ┃ ┃ │ 3) When measuring the probe │equipped with an irradiation tube for the purpose of local area imaging 1) and ┃ ┃ │experience │2) is impossible, at the irradiation field ┃ ┃ │ │must be 7cm or less. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │Inc. Other necessary tests │ ┃ ┃ │ 1) When indicating fault depth │The tolerance for the indicated scale shall be within ±1 mm. ┃ ┃ │Test(For Tomography Only │ ┃ ┃ │corresponding) │ ┃ ┃ │ 2) mAs Test │a) In the case of a transformer type: when the energization time is 0.1 seconds or more ┃ ┃ │ │has a percentage mean error of ±20% for the indicated value, 0.1 second ┃ ┃ │ │, the percentage mean error for the indicated value is ± ┃ ┃ │ │Make it within 20%. ┃ ┃ │ │b) For condenser X-ray devices: For cases less than 10 mAs ┃ ┃ │ │has an error of ±2mAs for the indicated value, and ┃ ┃ │ │, the percentage mean error for the indicated value is ±20% ┃ ┃ │ │must be within. ┃ ┃ │ 3) Field Agreement Test │a) When SID is maximized, the minimum irradiation field is 5 cm × 5 cm. ┃ ┃ │(Viewing device only │must be adjustable to lower. ┃ ┃ │corresponding) │b) The X-ray irradiation field exceeds the effective area of ​​the image plane during fluoroscopy ┃ ┃ │ │must not. ┃ ┃ │ │) The error between the center of the receiving plane and the center of the X-ray irradiation field is ┃ ┃ │ │SID. ┃ ┃ │ │ ┃ ┃ │ 4) Incident radiation dose rate │Incident radiation dose at the center point of the radiation line incident on the subject ┃ ┃ │test(vision device only │The START rate is 10 R/min (2.58 × 10 ) when an automatic exposure controller is present. ┃ ┃ │corresponding) │-3C/kg·min), 5R/min if there is no automatic exposure controller ┃ ┃ │ │Tube capable of bringing (1.29×10⁻³ C/kg·min) or more ┃ ┃ │ │Does not operate under any combination of voltage and tube current ┃ ┃ │ │must. ┃ ┃ │ │ ┃ ┃ │ 5) Cancer X-ray Test │5cm from surfaces where humans or animals may come into contact ┃ ┃ │(Condenser X-ray │and 2 mR/hr (5.16 × 10 ┃ ┃ │(Applies only to devices) │C/kg·hr). ┃ ┃ │ │Allowable value for the indicated value of the irradiation time by timer adjustment ┃ ┃ │ 6) Timer Test (Shooting │must be less than or equal to the value shown in the following table. ┃ ┃ │applicable only to devices for START │ ┃ ┃ │) │ ┃ ┃ │ ├─────┬────────┬────────────┨ ┃ │ │RectificationMethod │Display value │Tolerance ┃ ┃ │ ├─────┼────────┼────────────┨ ┃ │ │Single-phase device │T<10pulses │±0 pulse a) ┃ ┃ │ │ │10pulse≤T │±10% ┃ ┃ │ ├─────┼────────┼────────────┨ ┃ │ │Dasang Stop│T<0.01 seconds │-1.5msec~+6msec ┃ ┃ │ │value │0.01 sec≤T<0.04 sec│±20% ┃ ┃ │ │b), c) │0.04 seconds≤T │±10% ┃ ┃ │ ├─────┼────────┼────────────┨ ┃ │ │Inverter type cabinet│T<0.01 seconds │±1msec ┃ ┃ │ │value │0.01 seconds≤T │±10% ┃ ┃ │ ├─────┴────────┴────────────┨ ┃ │ │ ┃ ┃ │ │ ┃ ┃ │ │a) 1 pulse corresponds to 1/2 of 1 cycle of the power supply frequency ┃ ┃ │ │refers to the time. In the case of a single-phase device, the effective pulse electric ┃ ┃ │ │angle. The first 1 pearl ┃ ┃ │ │A range greater than that electrical angle in the last pulse ┃ ┃ │ │is also considered a regular pulse. ┃ ┃ │ │ Regarding the timing of injection, the initial injection section within -45° to 0° ┃ ┃ │ │The subsequent portion within 0°–45° at the time of START and termination was filmed ┃ ┃ │ │is not included in the time. ┃ ┃ │ │b) Depending on the input method of each device, from the input timing, regular operation ┃ ┃ │ │MTQ8d21810701d06d5d00386 In the case of a device having an initial input section leading up to START ┃ ┃ │ │, the initial part is not included in the shooting time ┃ ┃ │ │. ┃ ┃ │ │d) In a multiphase rectifier, the rising portion of the high-voltage waveform and the predetermined pipe ┃ ┃ │ │The projection time and cutoff at each 87% time for the voltage value ┃ ┃ │ │point, with the time in between, as the shooting time ┃ ┃ │ │. ┃ ┃ │ ├──────┬──────────┬─────────┨ ┃ │7) half-value layer(half-value layer: │Classification │Measurement tube voltage(kV) │Minimum half-value layer (mm Al)┃ ┃ │The intensity of radiation half ├──────┼──────────┼─────────┨ ┃ │Water needed to reduce │For animal diagnosis │60 │1.3 ┃ ┃ │Quality Thickness) Test │X-ray device │(Maximum tube voltage is │ ┃ ┃ │ │ │70 kV or less device │ ┃ ┃ │ │ │value) │ ┃ ┃ │ │ ├──────────┼─────────┨ ┃ │ │ │80 │2.3 ┃ ┃ │ │ │(Maximum tube voltage is 70 │ ┃ ┃ │ │ │Device exceeding kV) │ ┃ ┃ │ ├──────┴──────────┴─────────┨ ┃ │ │measured tube voltage after correcting the error following the tube voltage test ┃ ┃ │ │. ┃ ┠─────────┼────────────┼───────────────────────────┨ ┃2. Computed Tomography │A. Artifacts │Artifacts images must not appear conspicuously ┃ ┃AreaDevice │Test │and visually checked whether the centerline is properly aligned ┃ ┃ │ │. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │B. Slice Thickness Test │The tolerance between the slice thickness set value and the measured value is ┃ ┃ │ │Must be within ±1mm. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │. Tube voltage test │The percentage mean error of the tube voltage is within ±7% of the indicated value. ┃ ┃ │ │must. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │. Animal exposure dose measurement │Complies with the manufacturer's self-standards. ┃ ┃ │Test │ ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │E. mAs Test │Complies with the manufacturer's self-standards. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │bar. CT Number of │Complies with the manufacturer's self-standards. ┃ ┃ │LinearityTest │ ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │Company. Noise Test │Complies with the manufacturer's self-standards. ┃ ┃ ├────────────┼───────────────────────────┨ ┃ │Ah. High illuminance (high illuminance │Complies with the manufacturer's self-standards. ┃ ┃ │Degree) Spatial resolution │ ┃ ┃ │experiment (at small size │ ┃ ┃ │High contrast difference │ ┃ ┃ │is a material adjacent │ ┃ ┃ │Formula when │ ┃ ┃ │Special Ability Test) │ ┃ ┠─────────┼────────────┼───────────────────────────┨ ┃3. Other devices │As determined by the Director of the Quarantine and Inspection Agency │Standards determined by the Head of the Quarantine and Inspection Agency, standards determined by the manufacturer, and also ┃ ┃ │Test │applies internationally recommended standards. ┃ ┗━━━━━━━━━┷━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━┛