Annex 5 — 특수의료장비의 설치 및 운영에 관한 규칙
287043 · 2026-06-17
■ Rules on the Installation and Operation of Specialized Medical Equipment [Appendix 5] <Amended 2011.6.27>
Quality Control Items (Related to Article 8, Paragraph 2)
1. Magnetic Resonance Imaging Device
┏━━━┯━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━┓
┃Cycle │Film │PACS ┃
┣━━━┿━━━━━━━━━━━━━━━━━━━┿━━━━━━━━━━━━━━━━━━━┫
┃Daily │Table movement check │Table movement check ┃
┃ │ │ ┃
┃ │Hard copy device operation check │ ┃
┃ │ │ ┃
┃ │Constant Temperature and Humidity Unit Operation Check │Constant Temperature and Humidity Unit Operation Check ┃
┃ │ │ ┃
┃ │Data Storage Device Operation Check │Data Storage Device Operation Check ┃
┃ │ │ ┃
┃ │Developer Performance Management │ ┃
┃ │ │ ┃
┃ │Patient Monitor System Management │Patient Monitor System Management ┃
┠───┼───────────────────┼───────────────────┨
┃Weekly │Inspection of lighting, ventilation, temperature, and noise in the reading room │Inspection of lighting, ventilation, temperature, and noise in the reading room ┃
┃ │ │ ┃
┃ │Emergency Patient Rescue System Inspection │Emergency Patient Rescue System Inspection ┃
┠───┼───────────────────┼───────────────────┨
┃1Month │Center Frequency Check │Center Frequency Check ┃
┃ │ │ ┃
┃ │Transmitter acquisition and attenuation │Transmitter acquisition and attenuation ┃
┃ │ │ ┃
┃ │Location confirmation video (scout localization view) │Location confirmation video (scout localization view) ┃
┃ │MTQ391b40e1dfdb3a5101162Accuracy check of START │accuracy check ┃
┃ │ │ ┃
┃ │Ratio of measured values in the image to actual measured values │Comparison of measured values in video and actual measured values ┃
┃ │Teacher Evaluation │Evaluation ┃
┃ │ │ ┃
┃ │Artifact Inspection Items │ArtifactsInspectionItems ┃
┠───┼───────────────────┼───────────────────┨
┃3月 │Clinical Imaging Evaluation │Clinical Imaging Evaluation ┃
┃ │ │ ┃
┃ │ │Read Monitor Management (CRT) ┃
┠───┼───────────────────┼───────────────────┨
┃6 MONTH │ │Reader Monitor Management (LCD) ┃
┠───┼───────────────────┼───────────────────┨
┃1Year │Illumination measurement of reading desks and reading rooms │Illumination measurement of the reading room ┃
┗━━━┷━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━┛
For items with a cycle of less than 1 month, a copy of the quality control record log is not required during document inspection.
, and during precision inspection, verify whether the inspection was performed using the original quality control record ledger and the test specimen.
┏━━━┯━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━┓2. Computed Tomography System
┃Cycle │Film │PACS ┃
┣━━━┿━━━━━━━━━━━━━━━━━━━┿━━━━━━━━━━━━━━━━━━━┫
┃Daily │Table movement check │Table movement check ┃
┃ │ │ ┃
┃ │Hard copy device operation check │ ┃
┃ │ │ ┃
┃ │Data Storage Device Operation Check │Data Storage Device Operation Check ┃
┃ │ │ ┃
┃ │Developer Performance Management │ ┃
┃ │ │ ┃
┃ │Contrast agent injector operation check │Contrast agent injector operation check ┃
┠───┼───────────────────┼───────────────────┨
┃Every week │Inspection of lighting, ventilation, temperature, and noise in the reading room │Inspection of lighting, ventilation, temperature, and noise in the reading room ┃
┃ │ │ ┃
┃ │Emergency Patient Rescue System Inspection │Emergency Patient Rescue System Inspection ┃
┃ │ │ ┃
┃ │Emergency Shut-off Switch Operation Check │Emergency Shut-off Switch Operation Check ┃
┠───┼───────────────────┼───────────────────┨
┃3月 │Clinical Imaging Evaluation │Clinical Imaging Evaluation ┃
┃ │ │ ┃
┃ │ │Read Monitor Management (CRT) ┃
┠───┼───────────────────┼───────────────────┨
┃6months │Test of patient table movement interval accuracy │Test of patient table movement interval accuracy ┃
┃ │ │ ┃
┃ │Tube Voltage Test │Tube Voltage Test ┃
┃ │ │ ┃
┃ │Tube Current Test │Tube Current Test ┃
┃ │ │ ┃
┃ │Test using standard phantom │Test using standard phantom ┃
┃ │ │ ┃
┃ │ │Reader Monitor Management (LCD) ┃
┠───┼───────────────────┼───────────────────┨
┃1Year │CT Number Linearity │CT Number Linearity ┃
┃ │ │ ┃
┃ │Location confirmation video (scout localization view) │Location confirmation video (scout localization view) ┃
┃ │MTQ391b40e1dfdb3a5101162Accuracy check of START │accuracy check ┃
┃ │ │ ┃
┃ │Ratio of measured values in the image to actual measured values │Ratio of measured values in the image to actual measured values ┃
┃ │Teacher Evaluation │Teacher Evaluation ┃
┃ │ │ ┃
┃ │Illumination measurement of reading desks and reading rooms │Illumination measurement of a reading room ┃
┃ │ │ ┃
┃ │Patient Exposure Dose Test │Patient Exposure Dose Exposure Test ┃
┗━━━┷━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━┛
For items with a cycle of less than 1 month, a copy of the quality control record log is not required during document inspection.
, and during precision inspection, verify whether the inspection was performed using the original quality control record ledger and the test specimen.
Among the quality control items, inspection in accordance with the "Rules on Safety Management of Diagnostic Radiation Generating Devices"
Items that have been performed may be exempted from quality control record inspection only during the inspection cycle of the relevant inspection item.
However, the test results shall be recorded in the quality control record ledger, and the test inspection report (or copy) shall be [received]
Must concern.
3. Mammography device
┏━━━┯━━━━━━━━━━━━┯━━━━━━━━━━━━┯━━━━━━━━━━━┓
┃Cycle │Film mammography device │CR Mammography device │DR Mammography device ┃
┠───┼────────────┼────────────┼───────────┨
┃Everyday │Darkroom Cleaning │ │ ┃
┃ │ │ │ ┃
┃ │Developer Management │ │ ┃
┃ │ │ │ ┃
┃ │ │Monitor Cleaning │ ┃
┃ │ │ │ ┃
┃ │ │Laser Printer Management │Laser Printer Management ┃
┃ │ │ │ ┃
┃ │ │Image plate, cassette cleaning│Monitor Cleaning ┃
┠───┼────────────┼────────────┼───────────┨
┃Weekly │Increase/Deep Clean │ │ ┃
┃ │ │ │ ┃
┃ │Reader cleaning │ │ ┃
┃ │ │ │ ┃
┃ │Reading room lighting, ventilation, and temperature │Reading room lighting, ventilation, and temperature │Lighting, ventilation of the reading room, ┃
┃ │also, noise check │also, noise check │Check temperature and noise ┃
┃ │ │ │ ┃
┃ │ │CR Receiver Cleaning │Flat field calibration┃
┠───┼────────────┼────────────┼───────────┨
┃1Month │ │ │Visual checklist ┃
┃ │ │ │ ┃
┃ │ │ │MTF measurement ┃
┠───┼────────────┼────────────┼───────────┨
┃3 months │Retake Analysis │Re-imagingAnalysis │Retake Analysis ┃
┃ │ │ │ ┃
┃ │Analysis of residual fixing solution in film│ │ ┃
┃ │ │ │ ┃
┃ │ │Image plate sensitivity │ ┃
┃ │ │test │ ┃
┃ │ │ │ ┃
┃ │ │Read Monitor Management (CRT) │Read Monitor Management ┃
┃ │ │ │(CRT) ┃
┃ │ │ │ ┃
┃ │Clinical Imaging Evaluation │ClinicalImagingEvaluation │ClinicalImagingEvaluation ┃
┠───┼────────────┼────────────┼───────────┨
┃6月 │DarkroomFog │ │ ┃
┃ │ │ │ ┃
┃ │Intensifying Detector - Film Adhesion Test │Breast compression device inspection │Breast Compression Device Inspection ┃
┃ │ │ │ ┃
┃ │Breast compression device inspection │CR reader check │Check Recording Monitor ┃
┃ │ │ │ ┃
┃ │ │Test using standard phantom │Test using standard phantom┃
┃ │ │ │ ┃
┃ │Test using standard phantom │Reader Monitor Management (LCD) │Read Monitor Management ┃
┃ │ │ │(LCD) ┃
┠───┼────────────┼────────────┼───────────┨
┃1Year │Investigation Check │Investigation Check │Investigation Check ┃
┃ │ │ │ ┃
┃ │Resolution Check │Resolution Check │Resolution Check ┃
┃ │ │ │ ┃
┃ │Automatic exposure device inspection and re │Automatic exposure device inspection and re │Automatic exposure device inspection and re┃
┃ │Hyeonseong │Hyeonseong │Hyeonseong ┃
┃ │ │ │ ┃
┃ │Increasing Detector Sensitivity Measurement │ │ ┃
┃ │ │ │ ┃
┃ │Artifact Inspection │Artifact Inspection │Artifact Inspection ┃
┃ │ │ │ ┃
┃ │Tube Voltage Accuracy and Reproducibility │Tube Voltage Accuracy and Reproducibility │Tube Voltage Accuracy and Reproducibility┃
┃ │ │ │ ┃
┃ │Ship Quality Check │Ship Quality Check │Ship Quality Check ┃
┃ │ │ │ ┃
┃ │Average Wired Dose Measurement │Average Wired Dose Measurement │Measurement of average wired dose ┃
┃ │ │ │ ┃
┃ │Radiation Output Rate │Radiation Output Rate │Radiation Output Rate ┃
┃ │ │ │ ┃
┃ │Illumination of the reading desk and reading room │Illumination measurement of the reading room │Illumination measurement of a reading room ┃
┃ │Measurement │ │ ┃
┗━━━┷━━━━━━━━━━━━┷━━━━━━━━━━━━┷━━━━━━━━━━━┛
For items with a cycle of less than 1 month, a copy of the quality control record log is not required during document inspection.
, and during precision inspection, verify whether the inspection was performed using the original quality control record ledger and the test specimen.
Among the quality control items, inspection in accordance with the "Rules on Safety Management of Diagnostic Radiation Generating Devices"
Items that have been performed may be exempted from quality control record inspection only during the inspection cycle of the relevant inspection item.
However, the test results shall be recorded in the quality control record ledger, and the test inspection report (or copy) shall be [received]
Must concern.