Annex 3 — 특수의료장비의 설치 및 운영에 관한 규칙

287043 · 2026-06-17

■ Rules on the Installation and Operation of Specialized Medical Equipment [Appendix 3] <Amended July 7, 2021> Standards for Quality Control Inspection of Specialized Medical Equipment (Related to Article 5, Paragraph 2) 1. Manpower Inspection a. Inspection regarding compliance with the operational personnel standards in Attached Table 1 B. Submitted Documents: One copy of the Specialized Medical Equipment Personnel Status Table (Form No. 11) 2. Facility Inspection a. Inspection regarding compliance with the facility standards of Attached Table 1 B. Submitted Documents: Consent for Joint Use of Specialized Medical Equipment (Form No. 12) Medical Institution Facility Status Table 1 Copy 3. Quality Control Record Inspection ┌───────────────────────┰──────────────────────────────────────────────┬───────────────────────────┐ │InspectionItem ┃Qa jun │Submission Documents │ ┝━━━━━━━━━━━━━━━━━━━━━━━╋━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┿━━━━━━━━━━━━━━━━━━━━━━━━━━━┥ │Quality Control ┃According to the quality control items of Attached Table 5 │ 1 copy of Quality Control Log │ │ ┃Other quality control and quality control band │ │ │ ┃Quality control records by chapter │ │ ├───────────────────────╂──────────────────────────────────────────────┼───────────────────────────┤ │ChangeDevice ┃Special medical care in Attached Form No. 10 │Repair·Calibration·Modification of Specialized Medical Equipment │ │ ┃Equipment Repair·Calibration·Change History │1 copy of History Register │ │ ┃Number of specialized medical devices according to chapter │ │ │ ┃Record of correction, adjustment, and change history │ │ ├───────────────────────╂──────────────────────────────────────────────┴───────────────────────────┤ │non GO ┃ Among the quality control items, regarding the "safety management of diagnostic radiation generating devices" │ │ ┃Items inspected in accordance with the “rules” are the inspection items of the relevant inspection items │ │ ┃Quality control record inspections may be waived only within the validity period. │ │ ┃In this case, the test results are recorded in the quality control ledger, and the test inspection results │ │ ┃A copy of the document must be kept. │ └───────────────────────┸──────────────────────────────────────────────────────────────────────────┘ 4. Phantom Image Inspection a. Magnetic Resonance Imaging Device ┏━━━━━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃StandardPhantom │ACR-PH1 MRI phantom and as determined by the Minister of Health and Welfare ┃ ┃(Medical Device │Phantom ┃ ┃For Performance Evaluation │ ┃ ┃Human Tissue Model) │ ┃ ┠────────────────────────┼─────────────────────────────────────────────────────────────────────────┨ ┃Phantom Filming Jo │Standard spin echo T1 and T2 weighted images, 25 cm scan field (scan ┃ ┃gun │FOV, for contrast resolution, self-head T1-, T2-weighted images ┃ ┃ │), 11 slices, slice thickness 5mm, slice spacing 5mm, ┃ ┃ │Number of times (NEX) 1, number of grid cells (matrix number) 256 × 256 ┃ ┠────────────────────────┼─────────────────────────────────────────────────────────────────────────┨ ┃SubmissionVideo │Phantom Footage Part 1 (The video includes the equipment serial number and shooting date ┃ ┃ │mustStart) ┃ ┗━━━━━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃port Thursday │Passing Criteria ┃ ┠───────────────────────────────────────────┼──────────────────────────────────────────────────────┨ ┃Geometric Accuracy(T1) │Sagittal (frontal vertical section) positioning image(sagittal ┃ ┃ │localization view) Vertical length is 146 ~ 150 mm ┃ ┃ │must. ┃ ┃ ├──────────────────────────────────────────────────────┨ ┃ │T1 #5 Intercept lengths (vertical, horizontal, and diagonal) are 188 ∼ ┃ ┃ │Must be 192 mm. ┃ ┠───────────────────────────────────────────┼──────────────────────────────────────────────────────┨ ┃SpatialResolution(T1,T2) │Must be able to identify down to 1.0 mm or less. ┃ ┠───────────────────────────────────────────┼──────────────────────────────────────────────────────┨ ┃Section Thickness Accuracy(T1,T2) │4.3 ? 5.7㎜ must be within. ┃ ┠───────────────────────────────────────────┼──────────────────────────────────────────────────────┨ ┃Intercept position accuracy │Must be 5㎜ or less. ┃ ┃(T1,T2) │ ┃ ┠───────────────────────────────────────────┼──────────────────────────────────────────────────────┨ ┃Image Intensity Uniformity(T1,T2) │Equipment less than 3.0T: Must be 87.5% or higher. ┃ ┃ │? Equipment 3.0T or higher: Must be 82% or higher. ┃ ┠───────────────────────────────────────────┼──────────────────────────────────────────────────────┨ ┃Ghost signal percentage(T1) │Must be 2.5% or less. ┃ ┠───────────────────────────────────────────┼──────────────────────────────────────────────────────┨ ┃Contrast Resolution(T1,T2) │? Equipment less than 3.0T: Total 9 ┃ ┃ │dog Must be able to identify above. ┃ ┃ │? Equipment 3.0T or higher: Total from 4 segments (#8-11) ┃ ┃ │Must be able to identify at least 37. ┃ ┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ B. Computed Tomography Scanner ┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃StandardPhantom │AAPM CT Performance Evaluation Phantom 76-410 and ┃ ┃(Medical Device │Phantom verified as equivalent to this according to the START bar ┃ ┃For performance evaluation │ ┃ ┃Human Tissue Model) │ ┃ ┠─────────────────────────────┼────────────────────────────────────────────────────────────────────┨ ┃Phantom Shooting Conditions │120 kV p, 250 mA s, adjustable width 10 mm, Field of View (FOV) 25 cm or larger, 25 cm ┃ ┃ │Display FOV, Standard Reconstruction Algorithm(standard ┃ ┃ │reconstruction algorithm. ┃ ┠─────────────────────────────┼────────────────────────────────────────────────────────────────────┨ ┃SubmissionVideo │Phantom Footage Part 1 (Equipment serial number and shooting date are displayed in the video ┃ ┃ │must be done) ┃ ┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃port Thursday │sum Specification 기 jun ┃ ┠───────────────────────────────────────────────┼──────────────────────────────────────────────────┨ ┃CT attenuation coefficient of water (X-ray │0 must be ± 7HU. ┃ ┃Absorption Coefficient) │ ┃ ┠───────────────────────────────────────────────┼──────────────────────────────────────────────────┨ ┃Noise │5HU must be less than or equal to. ┃ ┠───────────────────────────────────────────────┼──────────────────────────────────────────────────┨ ┃Uniform Degree │The distance between the center and the periphery must be 5HU or less. ┃ ┠───────────────────────────────────────────────┼──────────────────────────────────────────────────┨ ┃Spatial resolution (recorded in mm) │Must be able to identify down to 1.0 mm or less. ┃ ┠───────────────────────────────────────────────┼──────────────────────────────────────────────────┨ ┃Contrast Resolution (Recorded in mm) │Must be able to identify down to 6.4 mm or less. ┃ ┠───────────────────────────────────────────────┼──────────────────────────────────────────────────┨ ┃Section Thickness │Must be 1 mm or less, or -1 mm or more. ┃ ┃(Error when measuring at 5mm and 10mm │ ┃ ┃Range Record) │ ┃ ┠───────────────────────────────────────────────┼──────────────────────────────────────────────────┨ ┃Presence or absence of artifacts (indicate as present or absent) │shouldnot exist. ┃ ┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ . Mammography device ┏━━━━━━━━━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃StandardPhantom │Nuclear Associates 18-220 Breast Phantom, RMI 156 Breast Phantom, CIRS ┃ ┃(Medical Device │Company model 15 breast phantom and as determined by the Minister of Health and Welfare ┃ ┃For performance evaluation │, and the Phantom ┃ ┃Human Tissue Model) │ ┃ ┠────────────────────────────┼─────────────────────────────────────────────────────────────────────┨ ┃Phantom Shooting Conditions │1. Breast composed of 50% breast parenchyma and 50% fat, compressed to 4.2 cm ┃ ┃ │. ┃ ┃ │2. The average wired dose shall be 3.0 mGy or less. ┃ ┃ │3. Film background optical density: 1.4 or higher ┃ ┠────────────────────────────┼─────────────────────────────────────────────────────────────────────┨ ┃SubmissionVideo │Phantom Footage Part 1 (Equipment serial number and shooting date are displayed in the video ┃ ┃ │must be done) ┃ ┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┏━━━━━━━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃port Thursday │sum Specification 기 jun ┃ ┃ ├────────────────────────────────────────────┬──────────────────────────┨ ┃ │Film mammography device │CR·DR Mammography Device ┃ ┠──────────────────────────┼────────────────────────────────────────────┼──────────────────────────┨ ┃Seom Yu So │At least 4 must be observable │At least 4 must be observable ┃ ┃ │. │. ┃ ┠──────────────────────────┼────────────────────────────────────────────┼──────────────────────────┨ ┃Small calcification group │Must be able to observe 3 or more │Must be able to observe 3 or more ┃ ┃ │. │. ┃ ┠──────────────────────────┼────────────────────────────────────────────┼──────────────────────────┨ ┃species Monster(Lump) │At least 3 must be observable │Must be able to observe 3 or more ┃ ┃ │. │. ┃ ┠──────────────────────────┼────────────────────────────────────────────┼──────────────────────────┨ ┃Pseudo-lesion │Must be able to observe at least 10 │Must be able to observe at least 10 ┃ ┃ │. │. ┃ ┗━━━━━━━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━┛ 5. Clinical Imaging Tests a. Magnetic Resonance Imaging Device 1) Inspection Method ┏━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. The use registered in the Certificate of Registration of Special Medical Equipment pursuant to Article 2, Paragraph 3 is for full-body use ┃ ┃ │: 1 head image, 1 spine image (select one of lumbar or cervical vertebrae ┃ ┃ │can), 1 joint image (shoulder joint, knee joint, one of the ankle joints ┃ ┃ │can select) and one torso image (heart, breast, liver, prostate, female pelvic cavity ┃ ┃ │) ┃ ┃ │ ┃ ┃ │2. The use registered in the Certificate of Registration of Special Medical Equipment pursuant to Article 2, Paragraph 3 is specific ┃ ┃ │For site-specific cases: Images according to the following classifications ┃ ┃ │ 1) For head, spine, and joint use only: 1 head image, 1 spine image (lumbar vertebrae, ┃ ┃ │You can select one of the cervical vertebrae) and one joint image (shoulder joint, knee joint ┃ ┃ │You can select either the joint or the ankle joint) ┃ ┃ │ 2) If tofu only: 1 tofu video ┃ ┃ │ 3) For spine only: 1 spine image (select one of lumbar or cervical vertebrae ┃ ┃ │can) ┃ ┃ │ 4) For joints only: 2 joint images (shoulder joint, knee joint and ankle joint ┃ ┃ │) ┃ ┃ │ 5) For spine and joint use only: 1 spine image (select one of the lumbar or cervical vertebrae ┃ ┃ │can be selected), 1 joint image (among shoulder, knee, and ankle joints ┃ ┃ │You can select one) ┃ ┃ │ 6) For head and spine only: 1 head image, 1 spine image (lumbar vertebrae, cervical vertebrae ┃ ┃ │) ┃ ┃ │ 7) For head and joint only: 1 head image, 1 joint image (shoulder joint, no ┃ ┃ │You can select either the knee joint or the ankle joint) ┃ ┠─────────────────┼────────────────────────────────────────────────────────────────────────────────┨ ┃ScoreCalculation │Total Score 100 points ┃ ┠─────────────────┼────────────────────────────────────────────────────────────────────────────────┨ ┃PassingCriteria │All submitted videos must each score 60 points or higher to be recognized as passing. ┃ ┗━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ 2) Tofu ┏━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃OfTheSubmittedVideo │1. Submit the following pulse sequences. ┃ ┃Preceding Conditions │- Sagittal (Frontal Vertical Section) T1 Weighted Image ┃ ┃ │- Axial (Frontal Horizontal Section) T1 Highlighted Image ┃ ┃ │- Axial T2 Weighted Image ┃ ┃ │- Axial T2 Fluid Attenuated Inversion Recovery (FLAIR, fluid attenuated inversion ┃ ┃ │recovery)Video ┃ ┃ │- If tested including other pulsation sequences, submit all together ┃ ┃ │. ┃ ┃ │2. Submit normal or near-normal images. ┃ ┃ │3. Submit the best video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. Submitting either diffusion-weighted imaging or cerebral angiography as the special imaging technique ┃ ┃ │. To obtain additional points for special video techniques, the video in question was executed ┃ ┃ │Images from other patients can be submitted. ┃ ┃ │5. On the general examination cover sheet, patient name, patient number, gender, age, date of imaging, and imaging institution ┃ ┃ │. ┃ ┗━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌───────────────────┬──────────────────────────────────────────────────────────┬───────────────────┐ │Item │EvaluationContent │Score  │ ├───────────────────┴──────────────────────────────────────────────────────────┴───────────────────┤ │Inspection Labels and General Information Items (10 points) │ ├───────────────────┬────────────┬─────────────────────────────────────────────┬───────────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├─────────────────────────────────────────────┼───────────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├─────────────────────────────────────────────┼───────────────────┤ │ │ │ImageThickness │1 │ │ │ ├─────────────────────────────────────────────┼───────────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├─────────────────────────────────────────────┼───────────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├─────────────────────────────────────────────┼───────────────────┤ │ │ │matrix number │1 │ │ ├────────────┼─────────────────────────────────────────────┼───────────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├─────────────────────────────────────────────┼───────────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├─────────────────────────────────────────────┼───────────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├─────────────────────────────────────────────┼───────────────────┤ │ │ │having positional identification images or being a Picture Archiving and Communication System (PACS) │1 │ │ │ │, the location can be confirmed in the reconnaissance video. │ │ ├───────────────────┴────────────┴─────────────────────────────────────────────┴───────────────────┤ │Equipment (10 points) │ ├───────────────────┬────────────┬─────────────────────────────────────────────┬───────────────────┤ │Equipment Standard │Tesla │3T or higher(1.5T or higher, 8 points if less than 3T, 5 points if less than 1.5T) │10 │ │(10) │(tesla) │ │ │ ├───────────────────┴────────────┴─────────────────────────────────────────────┴───────────────────┤ │Video Information Items(80 points) │ │Evaluating submitted images from magnetic resonance cerebral angiography or diffusion-weighted magnetic resonance imaging │ ├───────────────────┬──────────────────────────────────────────────────────────┬───────────────────┤ │Magnetic Resonance Brain │angiography Reconstructed image for START (maximal intensity projection, volume │2 │ │Angiography │rendered images) exist (if there are rotated images in two or more directions, 2 points, │ │ │(10) │1 point if only one-directional image is available). │ │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │There is a source image prior to vascular reconstruction. │2 │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │The entire V4 segment (intracranial segment) of the vertebral artery is included and clearly observed │2 │ │ │(1 point if only part of V4 is included or unclear). │ │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │The middle cerebral artery is clearly observed from its origin to its branches and cortical region (M4). │2 │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │The 3D pixel (voxel) is 1 mm³ or less (cannot be calculated using image markers │2 │ │ │If you do not receive a score). │ │ ├───────────────────┼──────────────────────────────────────────────────────────┼───────────────────┤ │DiffusionEmulator │There are images and diffusion-weighted images (b≥1000 arcseconds/mm²). │2 │ │Mechanical Resonance Imaging │(1 point if one item is not met, 0 points if b is less than 1000 sec/mm²) │ │ │(10) ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │The matrix number is 128 × 128 or greater, and the image pixel is 2 mm² or more │2 │ │ │High(1 point for one item not met). │ │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │Segment thickness 5 mm or less and segment spacing 2 mm or less (1 point for each item not satisfied) │2 │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │There is an Apparent Diffusion Coefficient (ADC) map. │2 │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │In diffusion-weighted images (b≥1000 sec/mm²), white matter and gray matter are well distinguished (in some parts │2 │ │ │1 point if white matter and gray matter are not distinguished). │ │ ├───────────────────┼──────────────────────────────────────────────────────────┼───────────────────┤ │Artifact (10) │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │ │if 2 points). │ │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │No aliasing artifacts (if artifacts are present but do not impede image interpretation │3 │ │ │2 points). │ │ │ ├──────────────────────────────────────────────────────────┼───────────────────┤ │ │There are no other artifacts originating from the device itself or caused by operator manipulation (Artifacts │3 │ │ │is present but does not impede image interpretation (2 points). │ │ ├───────────────────┼─────────────────────┬────────────────────────────────────┼───────────────────┤ │IncludedRange │Axis T2 Weighted Image │Includes from the crown to the foramen magnum (two │4 │ │(18) │ │3 points deducted if government or anti-air is not included). │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │The entire anterior, posterior, left, and right sides of the skull are included (anterior, posterior, left, and right sides of the skull │4 │ │ │ │is not included). │ │ │ ├─────────────────────┼────────────────────────────────────┼───────────────────┤ │ │Awards Ceremony T1 Highlight Video │Includes from the crown to the foramen magnum. │4 │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │Includes the center image. │3 │ │ ├─────────────────────┴────────────────────────────────────┼───────────────────┤ │ │Axial T2-weighted and Fluid-attenuated Inversion Recovery (FLAIR) images of both cerebral hemispheres │3 │ │ │Left-right symmetry is achieved. │ │ ├───────────────────┼─────────────────────┬────────────────────────────────────┼───────────────────┤ │SectionThickness │SectionThickness │Axial image: 5 mm or less[Each pulsating wave sequence(Axial T1-weighted image, │3 │ │Adequacy │ │The images of the axial T2 weighted image and axial T2 FLAIR image are insufficient │ │ │(12) │ │MTQ34834741d0717ad201719Deduct 1 point for every START] │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │Award Ceremony Video: 5mm or less │1 │ │ ├─────────────────────┼────────────────────────────────────┼───────────────────┤ │ │Intercept Interval │Axial image: 2 mm or less[(Each pulsating wave sequence(Axial T1-weighted image, Axial │3 │ │ │ │T2-weighted image, axial T2 liquid attenuation inversion recovery (FLAIR) image) image │ │ │ │ │1 point deducted for each time the award is not met] │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │Award Ceremony Video: 2mm or less │1 │ │ ├─────────────────────┼────────────────────────────────────┼───────────────────┤ │ │Image Grid Count │Axis-based T1-weighted image: The image grid count is 256 × 196 or greater and the pixels are │1 │ │ │(matrix number) │1㎟이다. │ │ │ │and pixel ├────────────────────────────────────┼───────────────────┤ │ │ │Axis-based T2-weighted image: The image grid count is 256 × 196 or greater and the pixels are │1 │ │ │ │1㎟이다. │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │Axis-based T2 FLAIR image: The image grid number is 256 × 196 or greater and the pixels │1 │ │ │ │is 1 mm². │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │Sagittal image: The image grid size is 256 × 196 or greater and the pixel size is 1 mm². │1 │ ├───────────────────┼─────────────────────┼────────────────────────────────────┼───────────────────┤ │ │Axis T2 Weighted Image │Differentiation of brain gray matter is possible (in some parts, brain gray matter is differentiated │3 │ │Image Contrast │ │1 point if not done). │ │ │(30) │ ├────────────────────────────────────┼───────────────────┤ │ │ │The detailed structures of the basal ganglia are distinguished (in some parts they are not distinguished │3 │ │ │ │1 point if either or one side is not distinguished). │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │The entire proximal middle cerebral artery (M1) can be traced (partial area │3 │ │ │ │MTQ34834741d0717ad2017431 point if START cannot be traced). │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │The entire internal cerebral vein can be traced. │3 │ │ │ │(1 point if part of the area cannot be traced). │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │The cisternal portion of the fifth cranial nerve was identified │3 │ │ │ │(1 point if only one side is visible). │ │ │ ├─────────────────────┼────────────────────────────────────┼───────────────────┤ │ │Awards Ceremony T1 Highlight Video │At the central level of the eyeball, the gray matter of both intraocular regions can be distinguished (distinguishing │3 │ │ │ │works but is not clear, 1 point). │ │ │ ├─────────────────────┼────────────────────────────────────┼───────────────────┤ │ │Axis T1 Weighted Image │The white and gray matter of the cerebrum are well distinguished (in some parts, distinction is not possible │3 │ │ │ │1 point if or boundary is ambiguous). │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │The detailed structures of the basal ganglia are distinguished. │3 │ │ ├─────────────────────┼────────────────────────────────────┼───────────────────┤ │ │Axis T2 │The sulci, ventricles, and cisterns are significantly lower and uniform compared to the white matter. │3 │ │ │Liquid Degradation Reversal Recovery │shows signal strength (signal strength is not uniform in some parts │ │ │ │(FLAIR)Video │1 point if failed). │ │ │ │ ├────────────────────────────────────┼───────────────────┤ │ │ │White matter and gray matter are well distinguished (in some parts, white matter and gray matter are │3 │ │ │ │1 point if not separated). │ │ └───────────────────┴─────────────────────┴────────────────────────────────────┴───────────────────┘ 3) Lumbar vertebra ┏━━━━━━━━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃OfTheSubmittedVideo │1. Submit the following pulse sequences. ┃ ┃Preceding Conditions │- Awards Ceremony T1 Highlight Video ┃ ┃ │- Awards Ceremony T2 Highlight Video ┃ ┃ │- One of axial T1-weighted or T2-weighted images ┃ ┃ │- In cases where the inspection included other pulsating wave sequences, submit all together ┃ ┃ │. ┃ ┃ │2. Submit normal or near-normal images. ┃ ┃ │3. Submit the best video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. Submit routine lumbar spine images performed to evaluate herniated disc ┃ ┃ │. ┃ ┃ │5. Submit non-contrast enhanced images. ┃ ┃ │6. On the general examination cover sheet, patient name, patient number, gender, age, date of imaging, and imaging institution ┃ ┃ │. ┃ ┗━━━━━━━━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌───────────────────┬────────────────────────────────────────────────────────────┬─────────────────┐ │Item │EvaluationContent │Score  │ ├───────────────────┴────────────────────────────────────────────────────────────┴─────────────────┤ │Inspection Labels and General Information Items (10 points) │ ├───────────────────┬──────────┬─────────────────────────────────────────────────┬─────────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │ImageThickness │1 │ │ │ ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │matrix number │1 │ │ ├──────────┼─────────────────────────────────────────────────┼─────────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├─────────────────────────────────────────────────┼─────────────────┤ │ │ │If positional identification images are available or it is a Picture Archiving and Communication System (PACS) reconnaissance │1 │ │ │ │The location can be confirmed in the video. │ │ ├───────────────────┴──────────┴─────────────────────────────────────────────────┴─────────────────┤ │Equipment (10 points) │ ├───────────────────┬──────────┬─────────────────────────────────────────────────┬─────────────────┤ │Equipment Standard │Tesla │3T or higher(1.5T or higher, 5 points if less than 3T, 2 points if less than 1.5T) │10 │ │(10) │(tesla) │ │ │ ├───────────────────┴──────────┴─────────────────────────────────────────────────┴─────────────────┤ │Video Information Items (80 points) │ ├───────────────────┬────────────────────────────────────────────────────────────┬─────────────────┤ │Special Video Techniques │fat suppression imaging or 3D imaging or whole-satin imaging including cervical vertebrae included │10 │ │(10) │is present and the image quality is excellent (5 points if image quality is poor). │ │ ├───────────────────┼────────────────────────────────────────────────────────────┼─────────────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │(10) │if 2 points). │ │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │No aliasing artifacts (if artifacts are present but do not impede image interpretation │3 │ │ │2 points). │ │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │Otherwise, there are no artifacts originating from the device itself or caused by operator manipulation (artifacts │3 │ │ │If present but does not impede image interpretation, 2 points). │ │ ├───────────────────┼────────────────────────────────────────────────────────────┼─────────────────┤ │IncludedRange │Includes both bilateral foramina on thalamic T1 and T2-weighted images. │3 │ │(12) ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │Includes at least one thoracic vertebra and sacral vertebra in sagittal T1 and T2-weighted images. │3 │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │There are axial images of four or more intervertebral discs starting from the lowest lumbar vertebra. │3 │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │In the axial image, there are three or more intervertebral disc levels containing superior and inferior endplates. │3 │ ├───────────────────┼────────────────────────────────────────────────────────────┼─────────────────┤ │SectionThickness │The slice thickness of the sagittal image is 4 mm or less. │4 │ │Adequacy ├────────────────────────────────────────────────────────────┼─────────────────┤ │(16) │The slice thickness of the axial image is 4 mm or less. │4 │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │The slice interval of the sagittal image is 20% or less of the slice thickness. │4 │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │The slice interval of the axial image is 20% or less of the slice thickness. │4 │ ├───────────────────┼────────────────────────────────────────────────────────────┼─────────────────┤ │Video │In thalamic T2-weighted images, the annulus fibrosus and nucleus pulposus are clearly distinguishable (in some cases, they are not distinguished │7 │ │Contrast diagram (32) │if 4 points). │ │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │The neuromuscular root within the foramen is clearly visible on the thalamic T1-weighted image (within the foramen │7 │ │ │4 points if the neuromuscular is visible but not clear). │ │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │In the axial image, the entire posterior boundary of the intervertebral disc and the vertebral endplate is visible (part of which is old │6 │ │ │If not separated, 3 points). │ │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │In thalamic T2-weighted images, the cerebrospinal fluid, spinal cord, and spinal plexuses are clearly distinguished (cerebrospinal fluid │6 │ │ │If the spinal cord and spinal cord are distinguishable, but the spinal plexus is not, 3 points). │ │ │ ├────────────────────────────────────────────────────────────┼─────────────────┤ │ │In axial images, the cerebrospinal fluid and the spinal cord or spinal plexus are clearly distinguishable (cerebrospinal fluid and │6 │ │ │If the spinal cord is distinct but the spinal nerve plexus is not, 3 points). │ │ └───────────────────┴────────────────────────────────────────────────────────────┴─────────────────┘ 4) Cervical spine ┏━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃ │1. Submit the following pulse sequences. ┃ ┃SubmissionVideo │- Awards Ceremony T1 Highlight Video ┃ ┃of │- Awards Ceremony T2 Highlight Video ┃ ┃Leading Row │- Axial T1-weighted, T2-weighted, or gradient recovery echo (GRE, gradient ┃ ┃gun │recalled echo) 1 of the videos ┃ ┃ │- If the inspection includes other pulsation sequences, submit them all together. ┃ ┃ │2. Submit normal or near-normal images. ┃ ┃ │3. Submit the best video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. Submit routine cervical spine images performed to evaluate herniated disc. ┃ ┃ │5. Submit non-contrast enhanced images. ┃ ┃ │6. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌───────────────────┬───────────────────────────────────────────────────────────┬──────────────────┐ │Item │EvaluationContent │Score  │ ├───────────────────┴───────────────────────────────────────────────────────────┴──────────────────┤ │Inspection Labels and General Information Items (10 points) │ ├───────────────────┬──────────────────┬────────────────────────────────────────┬──────────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├────────────────────────────────────────┼──────────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├────────────────────────────────────────┼──────────────────┤ │ │ │ImageThickness │1 │ │ │ ├────────────────────────────────────────┼──────────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├────────────────────────────────────────┼──────────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├────────────────────────────────────────┼──────────────────┤ │ │ │matrix number │1 │ │ ├──────────────────┼────────────────────────────────────────┼──────────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├────────────────────────────────────────┼──────────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├────────────────────────────────────────┼──────────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├────────────────────────────────────────┼──────────────────┤ │ │ │If positional images exist or it is a Picture Archiving and Communication System (PACS) │1 │ │ │ │The location can be confirmed in the reconnaissance video. │ │ ├───────────────────┴──────────────────┴────────────────────────────────────────┴──────────────────┤ │Equipment (10 points) │ ├───────────────────┬──────────────────┬────────────────────────────────────────┬──────────────────┤ │Equipment Standard │Tesla │3T or higher(1.5T or higher, 5 points if less than 3T, 2 points if less than 1.5T) │10 │ │(10) │(tesla) │ │ │ ├───────────────────┴──────────────────┴────────────────────────────────────────┴──────────────────┤ │Video Information Items (80 points) │ ├───────────────────┬───────────────────────────────────────────────────────────┬──────────────────┤ │Special Video Techniques │fat suppression imaging or 3D imaging or whole-satin imaging including the lumbar spine and young │10 │ │(10) │Product quality is excellent (5 points if video quality is poor). │ │ ├───────────────────┼───────────────────────────────────────────────────────────┼──────────────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │(10) │if 2 points). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │No aliasing artifacts (if artifacts are present but do not impede image interpretation │3 │ │ │2 points). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │There are no other artifacts originating from the device itself or caused by operator manipulation (Artifacts │3 │ │ │is present but does not impede image interpretation (2 points). │ │ ├───────────────────┼───────────────────────────────────────────────────────────┼──────────────────┤ │IncludedRange │Includes bilateral foramina in thalamic images. │3 │ │(12) ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │Includes the atlanto-occipital joint and one or more thoracic vertebrae in sagittal images │3 │ │ │. │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │There are axial images of four or more cervical intervertebral discs. │3 │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │In the axial image, there are three or more intervertebral disc levels containing superior and inferior endplates. │3 │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │Includes the cerebrum or 7 or more thoracic vertebrae in thalamic images. │-3  │ ├───────────────────┼───────────────────────────────────────────────────────────┼──────────────────┤ │SectionThickness │The slice thickness of the sagittal image is 3 mm or less. │4 │ │Adequacy ├───────────────────────────────────────────────────────────┼──────────────────┤ │(16) │The slice thickness of the axial image is 3 mm or less. │4 │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │The slice interval of the sagittal image is 20% or less of the slice thickness. │4 │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │The slice interval of the axial image is 20% or less of the slice thickness. │4 │ ├───────────────────┼───────────────────────────────────────────────────────────┼──────────────────┤ │Video │The annulus fibrosus and nucleus pulposus are differentiated on thalamic T2-weighted images (4 points if not differentiated in some areas). │7 │ │Contrast ├───────────────────────────────────────────────────────────┼──────────────────┤ │(32) │The neuromuscular root within the foramen is clearly visible on the thalamic T1-weighted image (within the foramen │7 │ │ │4 points if the neuromuscular is visible but not clear). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │In the axial image, the entire posterior boundary of the intervertebral disc and the vertebral endplate is visible (part of which is old │6 │ │ │4 points if not separated). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │In thalamic T2-weighted images, cerebrospinal fluid and spinal cord are distinguished (the distinction between cerebrospinal fluid and spinal cord is │6 │ │ │(4 points if only the part appears complete). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │In axial images, cerebrospinal fluid and spinal cord are distinguished (the distinction between cerebrospinal fluid and spinal cord is only partially │6 │ │ │If seen, 3 points). │ │ └───────────────────┴───────────────────────────────────────────────────────────┴──────────────────┘ 5) Shoulder Joint ┏━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit the following pulse sequences. ┃ ┃of │ - One or more of thial image quantum density-weighted imaging, T2-weighted, and T1-weighted imaging ┃ ┃Leading Row │ - Coronal (lateral vertical section) quantum density-weighted image, T2-weighted, T1-weighted image, 2 of which ┃ ┃gun │This is the end of ┃ ┃ │ - At least one of axial quantum density-weighted imaging, T2-weighted, and T1-weighted imaging ┃ ┃ │ - If the inspection includes other pulsation sequences, submit them all together. ┃ ┃ │2. Submit normal or near-normal images. ┃ ┃ │3. Submit the best video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. Submit routine shoulder joint images performed to evaluate the shoulder joint. ┃ ┃ │5. Submit non-contrast enhanced images. ┃ ┃ │6. On the general examination cover sheet, patient name, patient number, gender, age, date of imaging, and imaging institution ┃ ┃ │. ┃ ┗━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌───────────────────┬───────────────────────────────────────────────────────────┬──────────────────┐ │Item │EvaluationContent │Score  │ ├───────────────────┴───────────────────────────────────────────────────────────┴──────────────────┤ │Inspection Labels and General Information Items (10 points) │ ├───────────────────┬─────────────┬─────────────────────────────────────────────┬──────────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├─────────────────────────────────────────────┼──────────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├─────────────────────────────────────────────┼──────────────────┤ │ │ │ImageThickness │1 │ │ │ ├─────────────────────────────────────────────┼──────────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├─────────────────────────────────────────────┼──────────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├─────────────────────────────────────────────┼──────────────────┤ │ │ │matrix number │1 │ │ ├─────────────┼─────────────────────────────────────────────┼──────────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├─────────────────────────────────────────────┼──────────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├─────────────────────────────────────────────┼──────────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├─────────────────────────────────────────────┼──────────────────┤ │ │ │If positional identification images exist or it is a Picture Archiving and Communication System (PACS), then │1 │ │ │ │The location can be confirmed in the captured video. │ │ ├───────────────────┴─────────────┴─────────────────────────────────────────────┴──────────────────┤ │Equipment (10 points) │ ├───────────────────┬─────────────┬─────────────────────────────────────────────┬──────────────────┤ │Equipment Standard │Tesla │3T or higher(1.5T or higher, 5 points if less than 3T, 2 points if less than 1.5T) │10 │ │(10) │(tesla) │ │ │ ├───────────────────┴─────────────┴─────────────────────────────────────────────┴──────────────────┤ │Video Information Items (80 points) │ ├───────────────────┬───────────────────────────────────────────────────────────┬──────────────────┤ │Special Video Techniques │Includes fat suppression images or 3D images, and the image quality is excellent (Image quality │10 │ │(10) │5 points if quality is poor). │ │ ├───────────────────┼───────────────────────────────────────────────────────────┼──────────────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │(10) │if 2 points). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │No aliasing artifacts (if artifacts are present but do not impede image interpretation │3 │ │ │2 points). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │Otherwise, there are no artifacts originating from the device itself or caused by operator manipulation (artifacts │3 │ │ │If present but does not impede image interpretation, 2 points). │ │ ├───────────────────┼───────────────────────────────────────────────────────────┼──────────────────┤ │Included Range (9) │In the axial image, the acromioclavicular joint and the axillary fossa vertically │3 │ │ │(axillary recess) includes all. │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │In the coronal image, the coracoid process and scapular spine () anteriorly │3 │ │ │spine). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │In sagittal images, the greater tuberosity of the humerus and the spinoglenoid notch medially and extralaterally │3 │ │ │also includes. │ │ ├───────────────────┼────────────────────────────────┬──────────────────────────┼──────────────────┤ │SectionThickness │Axial image slice thickness │3㎜ or less (2 points if greater than 3㎜ and less than or equal to 4㎜). │4 │ │Adequacy ├────────────────────────────────┼──────────────────────────┼──────────────────┤ │(23) │Axial image intercept interval │is 10% or less of the section thickness (exceeding 10% of the section thickness │4 │ │ │ │20% or less, 2 points). │ │ │ ├────────────────────────────────┼──────────────────────────┼──────────────────┤ │ │Self-slice thickness of coronal image │3㎜ or less (2 points if greater than 3㎜ and less than or equal to 4㎜). │4 │ │ ├────────────────────────────────┼──────────────────────────┼──────────────────┤ │ │Intercept interval of coronal image │is 10% or less of the section thickness (exceeding 10% of the section thickness │4 │ │ │ │20% or less, 2 points). │ │ │ ├────────────────────────────────┴──────────────────────────┼──────────────────┤ │ │The field of view (FOV) is 18 cm or less. │3 │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │The matrix number is 256 × 192 or greater. │4 │ ├───────────────────┼───────────────────────────────────────────────────────────┼──────────────────┤ │Video │The labrum is clearly visible (2 points if partially indistinct). │4 │ │Contrast diagram (28) ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │The rotator interval is distinguishable (2 points if partially indistinct). │4 │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │Evaluation of the bone attachment site of the supraspinatus tendon is possible (partially │4 │ │ │(2 points if unclear). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │Evaluation of the bony attachment site of the infraspinatus tendon is possible (partially │4 │ │ │2 points if unclear). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │Evaluation of the bony attachment site of the subscapularis tendon is possible (partially │4 │ │ │(2 points if unclear). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │The entire biceps tendon is clearly visible (partially indistinct) │4 │ │ │2 points if seen). │ │ │ ├───────────────────────────────────────────────────────────┼──────────────────┤ │ │Evaluation of the axillary recess is possible (partially indistinctly visible │4 │ │ │if 2 points). │ │ └───────────────────┴───────────────────────────────────────────────────────────┴──────────────────┘ 6) Knee ┏━━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃OfTheSubmittedVideo │1. Submit the following pulse sequences. ┃ ┃Preceding Conditions │ - One or more of thalamic quantum density-weighted imaging, T2-weighted, and T1-weighted imaging ┃ ┃ │ - At least two of the following: Quantum Density Weighted Imaging, T2 Weighted, and T1 Weighted Imaging ┃ ┃ │ - At least one of axial quantum density-weighted imaging, T2-weighted, and T1-weighted imaging ┃ ┃ │ - In cases where the inspection included other pulsating wave sequences, all submitted together ┃ ┃ │. ┃ ┃ │2. Submit normal or near-normal images. ┃ ┃ │3. Submit the best video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. Submit routine knee joint images performed to evaluate the knee joint. ┃ ┃ │5. Submit non-contrast enhanced images. ┃ ┃ │6. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌───────────────────┬────────────────────────────────────────────────────┬─────────────────────────┐ │Item │EvaluationContent │Score  │ ├───────────────────┴────────────────────────────────────────────────────┴─────────────────────────┤ │Inspection Labels and General Information Items (10 points) │ ├───────────────────┬─────────────┬──────────────────────────────────────┬─────────────────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├──────────────────────────────────────┼─────────────────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├──────────────────────────────────────┼─────────────────────────┤ │ │ │ImageThickness │1 │ │ │ ├──────────────────────────────────────┼─────────────────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├──────────────────────────────────────┼─────────────────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├──────────────────────────────────────┼─────────────────────────┤ │ │ │matrix number │1 │ │ ├─────────────┼──────────────────────────────────────┼─────────────────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├──────────────────────────────────────┼─────────────────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├──────────────────────────────────────┼─────────────────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├──────────────────────────────────────┼─────────────────────────┤ │ │ │If positional images exist or it is a Picture Archiving and Communication System (PACS) │1 │ │ │ │The location can be confirmed in the reconnaissance video. │ │ ├───────────────────┴─────────────┴──────────────────────────────────────┴─────────────────────────┤ │Equipment(10 points) │ ├───────────────────┬─────────────┬──────────────────────────────────────┬─────────────────────────┤ │Equipment Standard │Tesla │3T or higher(1.5T or higher, 5 points if less than 3T, 2 points if less than 1.5T) │10 │ │(10) │(tesla) │ │ │ ├───────────────────┴─────────────┴──────────────────────────────────────┴─────────────────────────┤ │Video Information Items(80 points) │ ├───────────────────┬────────────────────────────────────────────────────┬─────────────────────────┤ │Special Video Techniques │Includes fat suppression images or 3D images, and the image quality is excellent (Image product │10 │ │(10) │5 points if quality is poor). │ │ ├───────────────────┼────────────────────────────────────────────────────┼─────────────────────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │(10) │if 2 points). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │No aliasing artifacts (artifacts are present but do not impede image interpretation │3 │ │ │if 2 points). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │There are no other artifacts originating from the device itself or caused by operator manipulation (Artifacts │3 │ │ │is present but does not impede image interpretation (2 points). │ │ ├───────────────────┼────────────────────────────────────────────────────┼─────────────────────────┤ │IncludedRange │Includes the tibial head and medial collateral ligament in sagittal images. │3 │ │(9) ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │Includes popliteal artery and patella in coronal images. │3 │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │Includes the entire patella and tibial process in the axial image. │3 │ ├───────────────────┼──────────────────────────────┬─────────────────────┼─────────────────────────┤ │SectionThickness │Self-slice thickness of sagittal image │3㎜ or less (3 points if greater than 3㎜ and less than or equal to 4㎜). │4 │ │Adequacy ├──────────────────────────────┼─────────────────────┼─────────────────────────┤ │(23) │Intercept interval of sagittal image │is 10% or less of the section thickness (exceeding 10% of the section thickness │4 │ │ │ │20% or less, 2 points). │ │ │ ├──────────────────────────────┼─────────────────────┼─────────────────────────┤ │ │Self-slice thickness of coronal image │3㎜ or less (3 points if greater than 3㎜ and less than or equal to 4㎜). │4 │ │ ├──────────────────────────────┼─────────────────────┼─────────────────────────┤ │ │Intercept interval of coronal image │10% or less of slice thickness (more than 10% of slice thickness │4 │ │ │ │20% or less, 2 points) │ │ │ ├──────────────────────────────┴─────────────────────┼─────────────────────────┤ │ │The field of view (FOV) is 16 cm or less. │3 │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │The matrix number is 256 × 192 or greater. │4 │ ├───────────────────┼────────────────────────────────────────────────────┼─────────────────────────┤ │Image Contrast │The entire anterior cruciate ligament in sagittal and coronal images │4 │ │(28) │The system is clear (2 points if the boundary is partially unclear). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │The boundary of the entire meniscus is clear in sagittal and coronal images (boundary │4 │ │ │is partially unclear, 2 points). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │The entire posterior cruciate ligament in sagittal and coronal images │4 │ │ │The system is clear (2 points if the boundary is partially unclear). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │In coronal images, the entire boundary of the medial collateral ligament is divided │4 │ │ │Clear(2 points if boundary is partially unclear). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │In coronal images, the entire boundary of the lateral collateral ligament is divided │4 │ │ │Clear(2 points if boundary is partially unclear). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │In axial images, signal intensity of cartilage and synovial fluid is distinguished (signal intensity is partially │4 │ │ │). │ │ │ ├────────────────────────────────────────────────────┼─────────────────────────┤ │ │In sagittal images, signal intensity of cartilage and synovial fluid is distinguished (signal intensity is partially │4 │ │ │MTQ34834741d0717ad2019942 points if not distinguished by START). │ │ └───────────────────┴────────────────────────────────────────────────────┴─────────────────────────┘ 7) Ankle Joint ┏━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit the following pulse sequences. ┃ ┃of │ - One or more T2 highlight videos or other award videos ┃ ┃Leading Row │ - One or more T2-weighted facial images or other facial images ┃ ┃gun │ - One or more axial T2-weighted images or other axial images ┃ ┃ │ In cases where the inspection included other pulsating wave sequences, submit all together ┃ ┃ │. ┃ ┃ │2. Submit normal or near-normal images. ┃ ┃ │3. Submit the best video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. Submit routine ankle joint images performed to evaluate the ankle joint. ┃ ┃ │5. Submit non-contrast enhanced images. ┃ ┃ │6. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────────────┬──────────────────────────────────────────────────────────┬────────────────┐ │Item │EvaluationContent │Score  │ ├──────────────────────┴──────────────────────────────────────────────────────────┴────────────────┤ │Inspection Labels and General Information Items (10 points) │ ├──────────────────────┬─────────────────┬────────────────────────────────────────┬────────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├────────────────────────────────────────┼────────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├────────────────────────────────────────┼────────────────┤ │ │ │ImageThickness │1 │ │ │ ├────────────────────────────────────────┼────────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├────────────────────────────────────────┼────────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├────────────────────────────────────────┼────────────────┤ │ │ │matrix number │1 │ │ ├─────────────────┼────────────────────────────────────────┼────────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├────────────────────────────────────────┼────────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├────────────────────────────────────────┼────────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├────────────────────────────────────────┼────────────────┤ │ │ │If positional images exist or it is a Picture Archiving and Communication System (PACS) │1 │ │ │ │The location can be confirmed in the reconnaissance video. │ │ ├──────────────────────┴─────────────────┴────────────────────────────────────────┴────────────────┤ │Equipment(10 points) │ ├──────────────────────┬─────────────────┬────────────────────────────────────────┬────────────────┤ │Equipment Standard │Tesla │3T or higher(1.5T or higher, 5 points if less than 3T, 2 points if less than 1.5T) │10 │ │(10) │(tesla) │ │ │ ├──────────────────────┴─────────────────┴────────────────────────────────────────┴────────────────┤ │Video Information Items(80 points) │ ├──────────────────────┬──────────────────────────────────────────────────────────┬────────────────┤ │Special Video Techniques │Includes fat suppression images or 3D images, and the image quality is excellent (Image quality │10 │ │(10) │If this is defective, 5 points). │ │ ├──────────────────────┼──────────────────────────────────────────────────────────┼────────────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │(10) │if 2 points). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │No aliasing artifacts (2 points if artifacts are present but do not impede image interpretation). │3 │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │Otherwise, there are no artifacts originating from the device itself or caused by operator manipulation (artifacts │3 │ │ │If present but does not impede image interpretation, 2 points). │ │ ├──────────────────────┼──────────────────────────────────────────────────────────┼────────────────┤ │IncludedRange │In sagittal images, the metatarsal base and the Achilles tendon (achilles │3 │ │(9) │tendon). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │Includes the medial and lateral malleolus of the tibia in coronal and axial images. │3 │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │Includes the distal tibia and the base of the calcaneus in sagittal, coronal, and axial images. │3 │ ├──────────────────────┼──────────────────────────┬───────────────────────────────┼────────────────┤ │SectionThickness │Self-slice thickness of sagittal image │3㎜ or less (2 points if greater than 3㎜ and less than or equal to 4㎜). │4 │ │Adequacy ├──────────────────────────┼───────────────────────────────┼────────────────┤ │(23) │Intercept interval of sagittal image │is 10% or less of the section thickness (exceeding 10% of the section thickness │4 │ │ │ │20% or less, 2 points). │ │ │ ├──────────────────────────┼───────────────────────────────┼────────────────┤ │ │Self-slice thickness of coronal image │3㎜ or less (2 points if greater than 3㎜ and less than or equal to 4㎜). │4 │ │ ├──────────────────────────┼───────────────────────────────┼────────────────┤ │ │Intercept interval of coronal image │is 10% or less of the section thickness (exceeding 10% of the section thickness │4 │ │ │ │20% or less, 2 points). │ │ │ ├──────────────────────────┴───────────────────────────────┼────────────────┤ │ │The field of view (FOV) is 16 cm or less. │3 │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │The matrix number is 256 × 192 or greater. │4 │ ├──────────────────────┼──────────────────────────────────────────────────────────┼────────────────┤ │Image Contrast │The entire deltoid ligament is clearly visible (partially indistinct) │4 │ │(28) │2 points if seen). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │The entire anterior talofibular ligament is clearly visible (partial │4 │ │ │If it appears unclear, 2 points). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │The entire calcaneofibular ligament is clearly visible (partially │4 │ │ │(2 points if unclear). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │The entire anterior inferior tibiofibular ligament is clearly visible │4 │ │ │is visible (2 points if partially unclear). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │The entire talus cartilage is clearly visible (signal intensity is heterogeneous │4 │ │ │2 points if seen). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │The entire tibialis posterior tendon is clearly visible (partially obscured │4 │ │ │2 points if clearly visible). │ │ │ ├──────────────────────────────────────────────────────────┼────────────────┤ │ │The entire peroneus longus and brevis tendon are clearly visible │4 │ │ │appears (2 points for partially unclear). │ │ └──────────────────────┴──────────────────────────────────────────────────────────┴────────────────┘ 8) Heart ┏━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit the following pulse sequences. ┃ ┃of │ - Cinema imaging ┃ ┃Leading Row │ - Delayed Contrast Enhancement Image ┃ ┃gun │ - If the inspection includes other pulsation sequences, submit them all together. ┃ ┃ │2. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │3. Submit routine imaging performed to evaluate ischemic heart disease or cardiomyopathy ┃ ┃ │. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────────────┬────────────────────────────────────────────────────────┬──────────────────┐ │Item │EvaluationContent │Score  │ ├──────────────────────┴────────────────────────────────────────────────────────┴──────────────────┤ │Inspection Labels and General Information Items (10 points) │ ├──────────────────────┬────────────────┬───────────────────────────────────────┬──────────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├───────────────────────────────────────┼──────────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├───────────────────────────────────────┼──────────────────┤ │ │ │ImageThickness │1 │ │ │ ├───────────────────────────────────────┼──────────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├───────────────────────────────────────┼──────────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├───────────────────────────────────────┼──────────────────┤ │ │ │matrix number │1 │ │ ├────────────────┼───────────────────────────────────────┼──────────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├───────────────────────────────────────┼──────────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├───────────────────────────────────────┼──────────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├───────────────────────────────────────┼──────────────────┤ │ │ │If positional images exist or it is a Picture Archiving and Communication System (PACS) │1 │ │ │ │The location can be confirmed in the reconnaissance video. │ │ ├──────────────────────┴────────────────┴───────────────────────────────────────┴──────────────────┤ │Equipment (10 points) │ ├──────────────────────┬────────────────┬───────────────────────────────────────┬──────────────────┤ │Equipment Standard │Tesla │1.5T or higher │10 │ │(10) │(tesla) │ │ │ ├──────────────────────┴────────────────┴───────────────────────────────────────┴──────────────────┤ │Video Information Items (80 points) │ ├──────────────────────┬────────────────────────────────────────────────────────┬──────────────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │3 │ │(18) │If none, 2 points). │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │No respiratory artifacts (if artifacts are present but do not impede image interpretation 2 │3 │ │ │point). │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │No artifacts caused by cardiac movement (artifacts are present but do not impede image interpretation) │3 │ │ │if 2 points). │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │No Gibbs ringing or truncation artifacts (artifacts │3 │ │ │If present but does not impede image interpretation, 2 points). │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │Aliasing or wrap-around No artifacts (Artifacts exist but zero │3 │ │ │2 points if there is no impediment to upper reading). │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │There are no other artifacts originating from the device itself or caused by operator manipulation (artifacts │3 │ │ │If present but does not impede image interpretation, 2 points). │ │ ├──────────────────────┼────────────────────────────────────────────────────────┼──────────────────┤ │IncludedRangeAnd │The vertical long axis is perpendicular to the transverse scout and the tip │4 │ │video direction │It is the surface passing between the tip of the heart and the mitral valve. │ │ │Appropriateness ├────────────────────────────────────────────────────────┼──────────────────┤ │(20) │The horizontal long axis is perpendicular to the vertical long axis, and the apex and the mitral valve are │4 │ │ │It is a surface passing through the center. │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │The left outflow tract long axis is between the apex and the midline of the mitral valve, │4 │ │ │It is the surface passing through the center of the pulse valve. │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │The short axis image is perpendicular to the long axis image. │4 │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │The short-circuit images of the delayed contrast-enhanced image have an appropriate slice interval of 10 mm. │4 │ ├──────────────────────┼────────────────────────────────────────────────────────┼──────────────────┤ │Movie Video │The temporal resolution is 45 msec or less (at least 25 image acquisitions per heartbeat). │5 │ │(20) ├────────────────────────────────────────────────────────┼──────────────────┤ │ │The slice thickness is 8 mm or less (3 points if greater than 8 mm and less than or equal to 10 mm, 1 point if greater than 10 mm). │5 │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │The short-circuited image includes the entire left ventricle. │5 │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │The section spacing is set appropriately, at 10 mm intervals. │5 │ ├──────────────────────┼────────────────────────────────────────────────────────┼──────────────────┤ │Delayed Contrast Imaging │Inversion recovery gradient echo (GRE) or steady state free wash │5 │ │Augmented Video │(SSFP) and Phase-Sensitive Inversion-Recovery (PSIR) │ │ │(20) │All videos are available. │ │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │It is the same plane (direction, slice thickness, spacing) as the movie image. │5 │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │The in-plane resolution is 2 mm or less. │5 │ │ ├────────────────────────────────────────────────────────┼──────────────────┤ │ │It can be confirmed that the signal intensity of normal myocardium is suppressed (TI reconnaissance image is available). │5 │ ├──────────────────────┼────────────────────────────────────────────────────────┼──────────────────┤ │Contrast agent indication │The name of the contrast agent is described. │2 │ │(2) │ │ │ └──────────────────────┴────────────────────────────────────────────────────────┴──────────────────┘ 9) Breast ┏━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit the following pulse sequences. ┃ ┃of │ - T2 Highlight Video ┃ ┃Leading Row │ - T1 Highlight Video ┃ ┃gun │ - Multi-period (two or more periods) contrast-enhanced T1-weighted images ┃ ┃ │ - If the inspection includes other pulsation sequences, submit them all together. ┃ ┃ │2. You must use a coil specifically designed for breasts. ┃ ┃ │3. Submit the best-shot video among the videos from a specific point in time designated by the evaluation agency ┃ ┃ │. ┃ ┃ │4. Submit the video recorded in the prone position. ┃ ┃ │5. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────────────┬─────────────────────────────────────────────────────────────────────┬─────┐ │Item │EvaluationContent │Point │ │ │ │수 │ ├──────────────────────┴─────────────────────────────────────────────────────────────────────┴─────┤ │Inspection Labels and General Information Items (10 points) │ ├──────────────────────┬──────────────┬──────────────────────────────────────────────────────┬─────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├──────────────────────────────────────────────────────┼─────┤ │ │ │Magnetic field strength │1 │ │ │ ├──────────────────────────────────────────────────────┼─────┤ │ │ │ImageThickness │1 │ │ │ ├──────────────────────────────────────────────────────┼─────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├──────────────────────────────────────────────────────┼─────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├──────────────────────────────────────────────────────┼─────┤ │ │ │matrix number │1 │ │ ├──────────────┼──────────────────────────────────────────────────────┼─────┤ │ │General Information │There is a scaler. │1 │ │ │ ├──────────────────────────────────────────────────────┼─────┤ │ │ │There is a direction indicator. │1 │ │ │ ├──────────────────────────────────────────────────────┼─────┤ │ │ │There is information regarding the image acquisition time in the contrast-enhanced image. │1 │ │ │ ├──────────────────────────────────────────────────────┼─────┤ │ │ │If positional identification images are available or it is a Picture Archiving and Communication System (PACS) reconnaissance │1 │ │ │ │The location can be confirmed in the video. │ │ ├──────────────────────┴──────────────┴──────────────────────────────────────────────────────┴─────┤ │Equipment (10 points) │ ├──────────────────────┬──────────────┬──────────────────────────────────────────────────────┬─────┤ │Equipment Standard │Tesla │3T or higher (1.5T or higher, 5 points if less than 3T) │10 │ │(10) │(tesla) │ │ │ ├──────────────────────┴──────────────┴──────────────────────────────────────────────────────┴─────┤ │Video Information Items (80 points) │ ├──────────────────────┬─────────────────────────────────────────────────────────────────────┬─────┤ │Intra-breast │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │Artifact │if 2 points). │ │ │(10) ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │No aliasing artifacts (2 points if artifacts are present but do not impede image interpretation). │3 │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │Otherwise, there are no artifacts originating from the device itself or caused by operator manipulation (artifacts │3 │ │ │If present but does not impede image interpretation, 2 points). │ │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │There are artifacts that seriously affect the reading. │-5  │ ├──────────────────────┼─────────────────────────────────────────────────────────────────────┼─────┤ │Contrast Enhancement │The name of the contrast agent is described. │3 │ │Technique ├─────────────────────┬───────────────────────────────────────────────┼─────┤ │(27) │Dynamic Contrast Enhancement │Pre-contrast and 30-second post-contrast images with fat-suppression T1-weighted imaging │5 │ │ │Image Technique │? All three or more consecutive images at 2-minute intervals are provided, and the lesion │ │ │ │ │(Computer automatic detection image replacement │ │ │ │ │ability) exists (2 points when there is no motion curve). │ │ │ ├─────────────────────┴───────────────────────────────────────────────┼─────┤ │ │T1-weighted images after contrast enhancement are fat-suppressed images. │4 │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │Contrast-enhanced T1-weighted images acquired between 60 and 120 seconds during the early phase │5 │ │ │Includes award (0 points if not included). │ │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │delayed phase contrast-enhanced T1-weighted images Shoot after 4 minutes (Adjustment │5 │ │ │If the last augmented video is within 4 minutes, 0 points). │ │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │Uses the subtraction technique. │5 │ ├──────────────────────┼─────────────────────────────────────────────────────────────────────┼─────┤ │ │Includes both breasts. │3 │ │IncludedRangeAnd ├─────────────────────────────────────────────────────────────────────┼─────┤ │Positioning │Image of both breasts from the axillary tail to the inframammary │4 │ │(13) │folds) (2 points if part is included). │ │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │There is no deformation of the breast shape (if the breast shape is partially pressed or distorted, 2 points, judgment │3 │ │ │If it interferes with reading, 0 points). │ │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │Axillary lymph nodes include levels I and II (2 points if only level I is included). │3 │ │ ├─────────────────────────────────────────────────────────────────────┼─────┤ │ │Part of breast tissue is not included. │-5  │ ├──────────────────────┼─────────────────────────────────────────────────────────────────────┼─────┤ │Special Video │Includes 3D Maximum Intensity Projection (MIP) images. │4 │ │Technique ├─────────────────────────────────────────────────────────────────────┼─────┤ │(8) │High b values ​​for diffusion-weighted images range from 750 arcseconds/mm² to 1000 arcseconds/mm² │4 │ │ │obtain. │ │ ├──────────────────────┼────────────────────────────────────┬────────────────────────────────┼─────┤ │SectionThickness, │SectionThickness │3㎜ or less (3 points if greater than 3㎜ and less than or equal to 5㎜). │5 │ │T1HighlightVideo ├────────────────────────────────────┼────────────────────────────────┼─────┤ │(10) │Intercept interval in all images │None(2 points if greater than 0 mm and less than or equal to 2 mm). │5 │ ├──────────────────────┼────────────────────────────────────┼────────────────────────────────┼─────┤ │Image Resolution· │In-plane pixel │is 1 mm or less (2 points if greater than 1 mm and less than or equal to 2 mm). │4 │ │TimeResolution ├────────────────────────────────────┴────────────────────────────────┼─────┤ │(12) │The cyst appearing bright in the enhanced image is clearly distinguishable (in some parts it is not well distinguished │4 │ │ │if 2 points). │ │ │ ├────────────────────────────────────┬────────────────────────────────┼─────┤ │ │temporal resolution │120 seconds or less (2 points if exceeding 120 seconds but less than 150 seconds) │4 │ └──────────────────────┴────────────────────────────────────┴────────────────────────────────┴─────┘ 10) Liver ┏━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit the following pulse sequences. ┃ ┃of │ - Axial T1 Weighted Image ┃ ┃Leading Row │ - Axial T2 Weighted Image ┃ ┃gun │ - Axial Diffusion-Weighted Image ┃ ┃ │ - Axial dynamic contrast-enhanced imaging ┃ ┃ │ - If the inspection includes other pulsation sequences, submit them all together. ┃ ┃ │2. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │3. Submit routine liver images performed to evaluate liver disease. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌───────────────────┬────────────────────────────────────────────────────────────────┬─────────────┐ │Item │EvaluationContent │Score  │ ├───────────────────┴────────────────────────────────────────────────────────────────┴─────────────┤ │Inspection Labels and General Information Items (10 points) │ ├───────────────────┬────────────────┬───────────────────────────────────────────────┬─────────────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├───────────────────────────────────────────────┼─────────────┤ │ │ │Magnetic field strength │1 │ │ │ ├───────────────────────────────────────────────┼─────────────┤ │ │ │ImageThickness │1 │ │ │ ├───────────────────────────────────────────────┼─────────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├───────────────────────────────────────────────┼─────────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├───────────────────────────────────────────────┼─────────────┤ │ │ │matrix number │1 │ │ ├────────────────┼───────────────────────────────────────────────┼─────────────┤ │ │General Information │There is a scaler. │1 │ │ │ ├───────────────────────────────────────────────┼─────────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├───────────────────────────────────────────────┼─────────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├───────────────────────────────────────────────┼─────────────┤ │ │ │If positional identification images exist or it is a Picture Archiving and Communication System (PACS), then │1 │ │ │ │The location can be confirmed in the captured video. │ │ ├───────────────────┴────────────────┴───────────────────────────────────────────────┴─────────────┤ │Equipment (10 points) │ ├───────────────────┬────────────────┬───────────────────────────────────────────────┬─────────────┤ │Equipment Standard │Tesla │3T or higher (1.5T or higher, 7 points if less than 3T) │10 │ │(10) │(tesla) │ │ │ ├───────────────────┴────────────────┴───────────────────────────────────────────────┴─────────────┤ │Video Information Items (80 points) │ ├───────────────────┬────────────────────────────────────────────────────────────────┬─────────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │(12) │if 2 points). │ │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │No aliasing artifacts (2 points if artifacts are present but do not impede image interpretation). │4 │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │Otherwise, there are no artifacts originating from the device itself or caused by operator manipulation (artifacts │4 │ │ │If present but does not impede image interpretation, 2 points). │ │ ├───────────────────┼──────────────────────────────────────┬─────────────────────────┼─────────────┤ │IncludedRange │Axis-type T2-weighted image and │. │4 │ │(16) │Axial Diffusion-Weighted Image ├─────────────────────────┼─────────────┤ │ │ │Includes the entire left-right and anterior-posterior abdominal cavity. │2 │ │ ├──────────────────────────────────────┼─────────────────────────┼─────────────┤ │ │on axis T1 and on axis │. │4 │ │ │Dynamic contrast-enhanced imaging ├─────────────────────────┼─────────────┤ │ │ │Includes the entire left-right and anterior-posterior abdominal cavity. │2 │ │ ├──────────────────────────────────────┴─────────────────────────┼─────────────┤ │ │There is an image including the gallbladder and extrahepatic bile ducts. │2 │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │There is a video containing the spleen. │2 │ ├───────────────────┼────────────────────────────────────────────────────────────────┼─────────────┤ │ │The name of the contrast agent is described. │3 │ │Contrast Enhancement ├────────────────────────────────────────────────────────────────┼─────────────┤ │Technique │Pre-contrast imaging, hepatic arterial phase imaging, portal venous phase imaging, and delayed phase imaging are included. │5 │ │(17) ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │Dynamic contrast-enhanced images are obtained as 3D images. │4 │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │In the arterial phase, the hepatic artery is very well contrasted, and the portal vein is partially weakly contrasted. │4 │ │ │and hepatic veins are not visualized (hepatic veins are weakly visualized, but hepatic arteries are relatively │ │ │ │2 points if clearly visible). │ │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │There is at least one subtraction image, including the arterial phase. │1 │ ├───────────────────┼─────────────────────────────────────────┬──────────────────────┼─────────────┤ │SectionThickness │Intercept thickness of T2 image on axis │8㎜ or less. │2 │ │adequacy and ├─────────────────────────────────────────┼──────────────────────┼─────────────┤ │Multiplane │Section of axial dynamic contrast-enhanced imaging │5㎜ or less (1 point if greater than 5㎜ and less than or equal to 8㎜). │2 │ │video's │One Thickness │ │ │ │Equity (14) ├─────────────────────────────────────────┼──────────────────────┼─────────────┤ │ │Section of axial dynamic contrast-enhanced imaging │No. │2 │ │ │Part Interval │ │ │ │ ├─────────────────────────────────────────┴──────────────────────┼─────────────┤ │ │It includes one or more coronal images with a slice thickness of 8 mm or less. │5 │ │ ├─────────────────────────────────┬──────────────────────────────┼─────────────┤ │ │Number of reference grid cells for axial images │The weighted image is 256 × 192 or larger, and the phase encoding is │1 │ │ │(matrix number) │direction is 192 or higher. │ │ │ │ ├──────────────────────────────┼─────────────┤ │ │ │The weighted image is 256 × 160 or higher, and the phase-coded │1 │ │ │ │direction is 160 or more. │ │ │ │ ├──────────────────────────────┼─────────────┤ │ │ │The diffusion-weighted image is 128 × 96 or larger, and the phase-coded │1 │ │ │ │direction is 96 or higher. │ │ ├───────────────────┼─────────────────────────────────┴──────────────────────────────┼─────────────┤ │Image Contrast │The liver contour is distinct on the axial T2-weighted image (although the liver contour is not distinct │2 │ │(14) │1 point if there is no impediment to image interpretation). │ │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │The portal vein is distinguishable on the axial T2-weighted image (although the outline of the portal vein is blurry, image │2 │ │ │1 point if there is no impediment to reading). │ │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │Includes heavily T2-weighted images (echo time 140 msec or more) during START. │3 │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │Includes chemical potential images (same phase, out of phase). │5 │ │ ├────────────────────────────────────────────────────────────────┼─────────────┤ │ │In axial T2-weighted images, the intrahepatic and extrahepatic bile ducts are well distinguished from surrounding structures. │2 │ │ │is visible (1 point if the outline with surrounding structures is blurry but does not impede image interpretation). │ │ ├───────────────────┼───────────────────────────────────────────────┬────────────────┼─────────────┤ │Diffusion-weighted image │b-value of axial diffusion-weighted image │ 3 or more (3 points if 2 or more).│4 │ │(7) ├───────────────────────────────────────────────┴────────────────┼─────────────┤ │ │Includes apparent diffusion coefficient (ADC) images in diffusion-weighted images. │3 │ └───────────────────┴────────────────────────────────────────────────────────────────┴─────────────┘ 11) Prostate ┏━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit the following pulse sequences. ┃ ┃of │- Awards Ceremony T2 Highlight Video ┃ ┃Leading Row │- Axial T2 Weighted Image ┃ ┃gun │- Physiognomy T2 Highlight Video ┃ ┃ │- Axial Diffusion-Weighted Image ┃ ┃ │- If the inspection includes other pulsation sequences, submit them all together. ┃ ┃ │2. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │3. Submit routine prostate images performed to evaluate prostate disease. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────────────┬──────────────────────────────────────────────────────────────────────┬────┐ │Item │EvaluationContent │Point │ │ │ │수  │ ├──────────────────────┴──────────────────────────────────────────────────────────────────────┴────┤ │Inspection Labels and General Information Items (10 points) │ ├──────────────────────┬───────────────┬──────────────────────────────────────────────────────┬────┤ │Cover │MRI test │Device Name │1 │ │(10) │Cover ├──────────────────────────────────────────────────────┼────┤ │ │ │Magnetic field strength │1 │ │ │ ├──────────────────────────────────────────────────────┼────┤ │ │ │ImageThickness │1 │ │ │ ├──────────────────────────────────────────────────────┼────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├──────────────────────────────────────────────────────┼────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├──────────────────────────────────────────────────────┼────┤ │ │ │matrix number │1 │ │ ├───────────────┼──────────────────────────────────────────────────────┼────┤ │ │General Information │There is a scaler. │1 │ │ │ ├──────────────────────────────────────────────────────┼────┤ │ │ │There is a direction indicator. │1 │ │ │ ├──────────────────────────────────────────────────────┼────┤ │ │ │The image array is appropriate. │1 │ │ │ ├──────────────────────────────────────────────────────┼────┤ │ │ │If positional identification images are available or it is a Picture Archiving and Communication System (PACS) reconnaissance │1 │ │ │ │The location can be confirmed in the video. │ │ ├──────────────────────┴───────────────┴──────────────────────────────────────────────────────┴────┤ │Equipment (10 points) │ ├──────────────────────┬───────────────┬──────────────────────────────────────────────────────┬────┤ │Equipment Standard │Tesla │3T or higher (1.5T or higher, 7 points if less than 3T) │10 │ │(10) │(tesla) │ │ │ ├──────────────────────┴───────────────┴──────────────────────────────────────────────────────┴────┤ │Video Information Items (80 points) │ ├──────────────────────┬──────────────────────────────────────────────────────────────────┬────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │3 │ │(14) │if 1 point). │ │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │No aliasing artifacts (if artifacts are present but do not impede image interpretation │3 │ │ │1 point). │ │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │No artifacts due to gastrointestinal motility (due to movement artifacts around the intestines and bladder │3 │ │ │1 point if there is blurriness but no artifacts in the prostate). │ │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │No artifacts caused by rectal air in the diffusion-weighted image (artifacts are present, but the image │3 │ │ │1 point if there is no impediment to reading). │ │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │There are no other artifacts originating from the device itself or caused by operator manipulation (artifacts │2 │ │ │If present but does not impede image interpretation, 1 point). │ │ ├──────────────────────┼───────────────────────────────────────┬──────────────────────────┼────────┤ │IncludedRange │Awards Ceremony T2 Highlight Video │Includes the entire bladder, seminal vesicles, and the entire prostate. │3 │ │(17) │ ├──────────────────────────┼────────┤ │ │ │Includes the entire left-right and front-back, posterior-posterior of the pelvic cavity. │3 │ │ ├───────────────────────────────────────┼──────────────────────────┼────────┤ │ │Axial T2 and axial diffusion-weighted images │Includes part of the bladder, seminal vesicles and the entire prostate. │3 │ │ │ ├──────────────────────────┼────────┤ │ │ │Includes the entire left-right and front-back, posterior-posterior of the pelvic cavity. │3 │ │ ├───────────────────────────────────────┴──────────────────────────┼────────┤ │ │There is an image including the aortic bifurcation. │3 │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │In axial T2-weighted images, bilateral symmetry of the lobes is achieved. │2 │ ├──────────────────────┼──────────────────────────────────────────────────────────────────┼────────┤ │Contrast Enhancement │The name of the contrast agent is described. │2 │ │Technique ├───────────────────────────────────────┬──────────────────────────┼────────┤ │(10) │Dynamic Contrast-Enhanced Imaging Techniques │Submit all dynamic contrast-enhanced images ( │2 │ │ │ │and │ │ │ │ │1 point if only created maps are included). │ │ │ │ ├──────────────────────────┼────────┤ │ │ │More than 10 images are acquired for the same plane. │2 │ │ │ ├──────────────────────────┼────────┤ │ │ │Subtraction to Dynamic Contrast-Enhanced Images │2 │ │ │ │. │ │ │ │ ├──────────────────────────┼────────┤ │ │ │Axis plane of dynamic contrast-enhanced image (thickness and angle │2 │ │ │ │Orientation) coincides with the axial T2-weighted image. │ │ ├──────────────────────┼───────────────────────────────────────┼──────────────────────────┼────────┤ │SectionThickness │Axial T2-weighted image and axial diffusion steel │is 4 mm or less (1 point if greater than 4 mm and less than or equal to 5 mm). │3 │ │adequacy and │Original Image Slice Thickness │ │ │ │multi-plane ├───────────────────────────────────────┼──────────────────────────┼────────┤ │video's │Axial T2-weighted image intercept interval │1㎜ or less. │2 │ │Adequacy ├───────────────────────────────────────┼──────────────────────────┼────────┤ │(24) │Sagittal T2-weighted image intercept interval │is 4 mm or less (1 point if greater than 4 mm and less than or equal to 5 mm). │3 │ │ ├───────────────────────────────────────┼──────────────────────────┼────────┤ │ │Sagittal T2-weighted image intercept interval │1㎜ or less. │2 │ │ ├───────────────────────────────────────┼──────────────────────────┼────────┤ │ │Coronation T2-weighted image intercept interval │is 4 mm or less (1 point if greater than 4 mm and less than or equal to 5 mm). │3 │ │ ├───────────────────────────────────────┼──────────────────────────┼────────┤ │ │Coronation T2-weighted image intercept interval │1㎜ or less. │2 │ │ ├───────────────────────────────────────┼──────────────────────────┼────────┤ │ │In-plane resolution │It is 0.7 mm or less in the axial T2-weighted image. │2 │ │ │(in-plane resolution) ├──────────────────────────┼────────┤ │ │ │It is 2.5 mm or less in axial diffusion-weighted images. │2 │ │ ├───────────────────────────────────────┼──────────────────────────┼────────┤ │ │Joya Field(FOV) │It is 180 - 220 mm in the axial image. │3 │ │ ├───────────────────────────────────────┴──────────────────────────┼────────┤ │ │The plane (thickness and orientation) of the axial T2-weighted image coincides with the axial diffusion-weighted image. │2 │ ├──────────────────────┼──────────────────────────────────────────────────────────────────┼────────┤ │Video │In axial T2-weighted images, the distinction between the periphery and the transition zone of the prostate is distinct (prostate │2 │ │Contrast │). │ │ │(10) ├──────────────────────────────────────────────────────────────────┼────────┤ │ │Non-uniform signals within the transition zone are clearly observed in the axial T2-weighted image │2 │ │ │(1 point if signal strength is non-uniform but boundaries are indistinct). │ │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │The prostatic urethra is clearly distinguishable on the axial T2-weighted image (prostatic urethra and membranous │2 │ │ │1 point if the boundary of the membranous urethra is indistinct in part). │ │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │The prostatic urethra is clearly distinguishable on sagittal T2-weighted images (prostatic urethra, membranous │2 │ │ │If the portion where the bulbar urethra is not clearly distinguished is less than half, 1 point). │ │ │ ├──────────────────────────────────────────────────────────────────┼────────┤ │ │T1. │2 │ ├──────────────────────┼───────────────────────────────────────┬──────────────────────────┼────────┤ │Diffusion-weighted image │High b values ​​ │800 sec/mm² or higher(500 sec/mm² or higher 800 sec/mm² not │3 │ │(5) │ │only if 1 point). │ │ │ ├───────────────────────────────────────┴──────────────────────────┼────────┤ │ │Includes apparent diffusion coefficient (ADC) images in diffusion-weighted images. │2 │ └──────────────────────┴──────────────────────────────────────────────────────────────────┴────────┘ 12) Female Pelvic Cavity ┏━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit the following pulse sequences. ┃ ┃of │- Awards Ceremony T2 Highlight Video ┃ ┃Leading Row │- Axial T2 Weighted Image ┃ ┃gun │- Axial T1 Weighted Image ┃ ┃ │- If the inspection includes other pulsation sequences, submit them all together. ┃ ┃ │2. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │3. Submit routine pelvic cavity images performed to evaluate uterine disease. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────────────┬─────────────────────────────────────────────────────────────────┬─────────┐ │Item │EvaluationContent │Score  │ ├──────────────────────┴─────────────────────────────────────────────────────────────────┴─────────┤ │Inspection Labels and General Information Items (10 points) │ ├──────────────────────┬─────────┬───────────────────────────────────────────────────────┬─────────┤ │Cover │MRI │Device Name │1 │ │(10) │test ├───────────────────────────────────────────────────────┼─────────┤ │ │Cover │Magnetic field strength │1 │ │ │ ├───────────────────────────────────────────────────────┼─────────┤ │ │ │ImageThickness │1 │ │ │ ├───────────────────────────────────────────────────────┼─────────┤ │ │ │Repeat Time (TR) · Echo Time (TE) · Excitation Count (NEX) │1 │ │ │ ├───────────────────────────────────────────────────────┼─────────┤ │ │ │Joya Field(FOV) │1 │ │ │ ├───────────────────────────────────────────────────────┼─────────┤ │ │ │matrix number │1 │ │ ├─────────┼───────────────────────────────────────────────────────┼─────────┤ │ │General │There is a scaler. │1 │ │ │Info ├───────────────────────────────────────────────────────┼─────────┤ │ │ │There is a direction indicator. │1 │ │ │ ├───────────────────────────────────────────────────────┼─────────┤ │ │ │The image array is appropriate. │1 │ │ │ ├───────────────────────────────────────────────────────┼─────────┤ │ │ │If positional identification images are available or it is a Picture Archiving and Communication System (PACS) reconnaissance │1 │ │ │ │The location can be confirmed in the video. │ │ ├──────────────────────┴─────────┴───────────────────────────────────────────────────────┴─────────┤ │Equipment (10 points) │ ├──────────────────────┬─────────┬───────────────────────────────────────────────────────┬─────────┤ │Equipment Standard │Tesla │3T or higher (1.5T or higher, 7 points if less than 3T) │10 │ │(10) │(tesla) │ │ │ ├──────────────────────┴─────────┴───────────────────────────────────────────────────────┴─────────┤ │Video Information Items (80 points) │ ├──────────────────────┬─────────────────────────────────────────────────────────────────┬─────────┤ │Artifact │No artifacts caused by patient movement (artifacts are present but do not impede image interpretation │4 │ │(16) │if 2 points). │ │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │No aliasing artifacts (if artifacts are present but do not impede image interpretation 2 │4 │ │ │point). │ │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │No artifacts due to gastrointestinal motility (artifacts are present but do not impede image interpretation │4 │ │ │if 2 points). │ │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │There are no other artifacts originating from the device itself or caused by operator manipulation (in │4 │ │ │2 points if there is an object but it does not impede image interpretation). │ │ ├──────────────────────┼───────────────────────┬─────────────────────────────────────────┼─────────┤ │IncludedRange │Awards Ceremony T2 Highlight Video │Includes the entire uterus and vagina. │3 │ │(15) │ ├─────────────────────────────────────────┼─────────┤ │ │ │Includes the entire left-right and front-back, posterior-posterior of the pelvic cavity. │2 │ │ ├───────────────────────┼─────────────────────────────────────────┼─────────┤ │ │Axis T2 Weighted Image │Includes the entire uterus and vagina. │3 │ │ │ ├─────────────────────────────────────────┼─────────┤ │ │ │Includes the entire left-right and front-back, posterior-posterior of the pelvic cavity. │2 │ │ ├───────────────────────┼─────────────────────────────────────────┼─────────┤ │ │Axis T1 Weighted Image │Includes the entire left-right and front-back, posterior-posterior of the pelvic cavity. │3 │ │ ├───────────────────────┴─────────────────────────────────────────┼─────────┤ │ │There is an image including the kidney and retroperitoneal space. │2 │ ├──────────────────────┼─────────────────────────────────────────────────────────────────┼─────────┤ │ │Perform contrast enhancement. │2 │ │Contrast Enhancement ├─────────────────────────────────────────────────────────────────┼─────────┤ │Technique(6) │Dynamic contrast-enhanced images are obtained in 3D. │2 │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │The name of the contrast agent is described. │2 │ ├──────────────────────┼─────────────────────────────────┬───────────────────────────────┼─────────┤ │SectionThickness │Axial T2-weighted image slice thickness │5㎜ or less (1 point if greater than 5㎜ and less than or equal to 8㎜). │3 │ │adequacy and ├─────────────────────────────────┼───────────────────────────────┼─────────┤ │Multi-plane │Axial T2-weighted image intercept interval │is 1 mm or less. │3 │ │video's ├─────────────────────────────────┼───────────────────────────────┼─────────┤ │Adequacy(23) │Number of reference grid cells for axial images │T2In the weighted image, at least one axis is 250 or greater. │1 │ │ │(matrix number) ├───────────────────────────────┼─────────┤ │ │ │T1In the weighted image, at least one axis is 250 or greater. │1 │ │ ├─────────────────────────────────┼───────────────────────────────┼─────────┤ │ │Self-slice thickness of sagittal T2 image │5㎜ or less (1 point if greater than 5㎜ and less than or equal to 8㎜). │3 │ │ ├─────────────────────────────────┼───────────────────────────────┼─────────┤ │ │Intercept interval of sagittal T2 image │is 1 mm or less. │3 │ │ ├─────────────────────────────────┼───────────────────────────────┼─────────┤ │ │Slice thickness of contrast-enhanced T1 image │5㎜ or less. │2 │ │ ├─────────────────────────────────┼───────────────────────────────┼─────────┤ │ │Slice interval of contrast-enhanced T1 image │No. │1 │ │ ├─────────────────────────────────┼───────────────────────────────┼─────────┤ │ │Joya Field(FOV) │It is 180 - 250 mm in the axial image. │2 │ │ │ ├───────────────────────────────┼─────────┤ │ │ │In the awards ceremony video, it is 180 - 250 mm. │2 │ │ ├─────────────────────────────────┴───────────────────────────────┼─────────┤ │ │Includes physiognomy video │2 │ ├──────────────────────┼─────────────────────────────────────────────────────────────────┼─────────┤ │Image Contrast │Layers of the uterus [endometrial layer, junctional zone, myometrium] are distinguished on axial T2-weighted images │3 │ │(12) │(1 point if the undifferentiated area is less than half of the uterine body). │ │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │Layers of the uterus [endometrial layer, junctional zone, myometrium] are distinguished on thalamic T2-weighted images │3 │ │ │(1 point if the undifferentiated area is less than half of the uterine body). │ │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │The cervix and periumbilical tissues are distinct (1 point if part is not distinct). │3 │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │In sagittal T2-weighted images, the vagina is distinguishable from surrounding tissues (the undistinguished area is the vagina │3 │ │ │1 point if less than half of the total length). │ │ ├──────────────────────┼─────────────────────────────────────────────────────────────────┼─────────┤ │Image Technique │Oblique coronal image performed aligned with the axis of the uterine body or cervix ( │2 │ │(4) │There is an oblique axial image. │ │ │ ├─────────────────────────────────────────────────────────────────┼─────────┤ │ │There is a T1-weighted image of fat inhibition. │2 │ ├──────────────────────┼─────────────────────────────────┬───────────────────────────────┼─────────┤ │Diffusion-weighted image │High b values ​​ │800 sec/mm² or higher (500 sec/mm² or higher 800 sec/mm² not │2 │ │(4) │ │only if 1 point). │ │ │ ├─────────────────────────────────┴───────────────────────────────┼─────────┤ │ │Includes apparent diffusion coefficient (ADC) images in diffusion-weighted images. │2 │ └──────────────────────┴─────────────────────────────────────────────────────────────────┴─────────┘ B. Computed Tomography (CT) 1) Test Method ┏━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃Submission │ 1. The purpose registered in the Certificate of Registration of Special Medical Equipment pursuant to Article 2, Paragraph 3 is for telecommunications use ┃ ┃on │: Head, standard chest, low-dose chest, dynamic contrast-enhanced liver imaging ┃ ┃ │ 2. The use registered in the Certificate of Registration of Special Medical Equipment pursuant to Article 2, Paragraph 3 is non-contrast treatment ┃ ┃ │For whole-body use: Low-dose chest, non-contrast enhanced abdominal imaging ┃ ┃ │ 3. The purpose registered in the Certificate of Registration of Special Medical Equipment pursuant to Article 2, Paragraph 3 is exclusive ┃ ┃ │: Video of the relevant area ┃ ┠──────┼───────────────────────────────────────────────────────────────────────────────────────────┨ ┃Score │Total Score 100 points ┃ ┃san │ ┃ ┠──────┼───────────────────────────────────────────────────────────────────────────────────────────┨ ┃pass │All submitted videos must each score 60 points or higher to be recognized as passing. ┃ ┃jun │ ┃ ┗━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ 2) Tofu ┏━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit based on pre-contrast CT ┃ ┃of │2. Submit CT cerebral angiography using a special imaging technique. Regarding the special imaging technique ┃ ┃Leading Row │To obtain additional points, submit images of other patients who underwent the relevant imaging procedure ┃ ┃gun │can. ┃ ┃ │3. Submit the best-recorded video among the videos from a specific point in time designated by the evaluation agency. ┃ ┃ │4. Submit routine head images performed to evaluate brain disease. ┃ ┃ │5. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌───────────────┬─────────────────────────────────────────────────────────────┬────────────────────┐ │Item │EvaluationContent │Score  │ ├───────────────┴─────────────────────────────────────────────────────────────┴────────────────────┤ │Inspection Labels and General Information Items (4 points) │ ├───────────────┬──────────────────┬──────────────────────────────────────────┬────────────────────┤ │Cover │General Information │There is a scan image (scanogram). │1 │ │(4) │ ├──────────────────────────────────────────┼────────────────────┤ │ │ │The image sequence (image number, scan time, etc.) is confirmed. │1 │ │ │ ├──────────────────────────────────────────┼────────────────────┤ │ │ │A scaler is indicated. │1 │ │ │ ├──────────────────────────────────────────┼────────────────────┤ │ │ │Left and right directions are indicated. │1 │ ├───────────────┴──────────────────┴──────────────────────────────────────────┴────────────────────┤ │Equipment and Radiation Exposure Management Standards (20 points) │ ├───────────────┬───────────────────────────────────────┬─────────────────────┬────────────────────┤ │Equipment Standard │MultipleDetectionCT │64 channels or more (16 channels or more but less than 64 channels 6 points, │10 │ │(10) │(multidetector CT) │2 channels or more, less than 16 channels (3 points) │ │ │ │Detector Channel Reference │ │ │ ├───────────────┼───────────────────────────────────────┴─────────────────────┼────────────────────┤ │Radiation │There is a dose report. │5 │ │Radiation Exposure Management ├─────────────────────────────────────────────────────────────┼────────────────────┤ │(10) │is tested at a tube voltage of 120kVp or less. │5 │ ├───────────────┴─────────────────────────────────────────────────────────────┴────────────────────┤ │Video Information Items (76 points) │ ├───────────────┬─────────────────────────────────────────────────────────────┬────────────────────┤ │Artifact │There are no respiratory artifacts or artifacts caused by patient movement. │1 │ │(10) ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │No beam-hardening artifacts on the top of the tent. │1 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The patient is located in the center of the scan. │1 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │Examine with both arms lowered. │1 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │No body ornaments, unnecessary equipment, or devices affecting the scan. │3 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │There are no ring artifacts. │3 │ ├───────────────┼─────────────────────────────────────────────────────────────┼────────────────────┤ │VideoScan │Includes from the parietal region to the base of the skull (foramen magnum). │3 │ │range ├─────────────────────────────────────────────────────────────┼────────────────────┤ │(12) │Includes the total width of the skull (up/down, front/back, left/right). │3 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The image shows the tuberculum sellar-occipital protuberance (tuberculum sellar-occipital protuberance │3 │ │ │line). │ │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │It is symmetrical left and right. │3 │ ├───────────────┼─────────────────────────────────────────────────────────────┼────────────────────┤ │VideoReconstruction │Brain and skeleton settings exist separately, or appropriate video window adjustment is possible. │2 │ │Adequacy(4) ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The brain and skeletal configurations are reconstructed using appropriate algorithms, respectively. │2 │ ├───────────────┼─────────────────────────────────────────────────────────────┼────────────────────┤ │Section thickness and │The section thickness is 5 mm or less. │4 │ │ofMultiplanarImages ├─────────────────────────────────────────────────────────────┼────────────────────┤ │Adequacy (8) │No intercept interval. │4 │ ├───────────────┼─────────────────────────────────────────────────────────────┼────────────────────┤ │ │Can distinguish between gray matter and white matter (2 points if partially unclear). │4 │ │Resolution and ├─────────────────────────────────────────────────────────────┼────────────────────┤ │Contrast │The basal ganglia are differentiated (2 points if only one side is differentiated or only a part is differentiated). │4 │ │(32) ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The Sylvian fissure is distinguished (only one side is distinguished or only a part of it │4 │ │ │2 points if distinguished). │ │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │Brain sulci are identified (2 points if the unidentified portion is less than half of the total brain). │4 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The proximal part of the middle cerebral artery is identified (2 points if only one side or only a part is identified). │4 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The cerebellar vermis is identified (2 points if indistinctly identified). │4 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The cerebellopontine cistern is identified (only one side is identified or │4 │ │ │2 points if only part is identified). │ │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │Can distinguish between the cerebellar cortex and deep white matter (distinguishes only one side or only a part │4 │ │ │If divided, 2 points). │ │ ├───────────────┼─────────────────────────────────────────────────────────────┼────────────────────┤ │Cerebrovascular │There are four or more types of reconstructed images for angiography. In this case, the reconstructed images' │4 │ │angiography │The type is maximal intensity projection, when acquired in multiple planes, each │ │ │(10) │angle recognized), bone subtraction images, volumetric visualization images (volume │ │ │ │rendered images) etc. (3 points if there are 3 types of images, 2 points if there are 2, and │ │ │ │0 points). │ │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │Contrast enhancement of the internal carotid artery at the base of the skull is adjacent to the internal │2 │ │ │Significantly higher than the internal jugular vein. │ │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │In the reconstructed image, the M4 branch of the middle cerebral artery can be clearly traced. │2 │ │ ├─────────────────────────────────────────────────────────────┼────────────────────┤ │ │The internal carotid artery is clearly observed within the carotid canal in the original image. │2 │ └───────────────┴─────────────────────────────────────────────────────────────┴────────────────────┘ 3) General Chest ┏━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit standard chest CT images. High-resolution CT images are excluded from submission ┃ ┃of │. In principle, images are submitted after contrast enhancement; however, in unavoidable cases, contrast enhancement ┃ ┃Leading Row │You can submit the previous video, but you cannot receive points for that item. ┃ ┃gun │2. For Picture Archiving and Communication System (PACS) images, changing lung and mediastinum settings ┃ ┃ │It must be possible. For film images, lung and mediastinal settings must be present respectively. ┃ ┃ │. ┃ ┃ │3. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────────────┬────────────────────────────────────────────────────────────────────┬──────┐ │Item │EvaluationContent │Score │ ├──────────────────────┴────────────────────────────────────────────────────────────────────┴──────┤ │Inspection Labels and General Information Items (4 points) │ ├──────────────────────┬────────┬───────────────────────────────────────────────────────────┬──────┤ │Cover │General │. │1 │ │(4) │Info ├───────────────────────────────────────────────────────────┼──────┤ │ │ │The video sequence (video number, scan time, etc.) is confirmed. │1 │ │ │ ├───────────────────────────────────────────────────────────┼──────┤ │ │ │A scaler is indicated. │1 │ │ │ ├───────────────────────────────────────────────────────────┼──────┤ │ │ │Left and right directions are indicated. │1 │ ├──────────────────────┴────────┴───────────────────────────────────────────────────────────┴──────┤ │Equipment and Radiation Exposure Management Standards (20 points) │ ├──────────────────────┬─────────────────────────┬──────────────────────────────────────────┬──────┤ │Equipment Standard │MultipleDetectionCT │64 channels or more (16 channels or more but less than 64 channels 6 points, 2 channels │10 │ │(10) │(multidetector CT) │or more, less than 16 channels, 3 points) │ │ │ │Detector Channel Reference │ │ │ ├──────────────────────┼─────────────────────────┴──────────────────────────────────────────┼──────┤ │Radiation │There is a dose report. │4 │ │Radiation Exposure Management ├────────────────────────────────────────────────────────────────────┼──────┤ │(10) │Uses automatic exposure control (AEC). │4 │ │ ├────────────────────────────────────────────────────────────────────┼──────┤ │ │is tested at a tube voltage of 120 kVp or less. │2 │ ├──────────────────────┴────────────────────────────────────────────────────────────────────┴──────┤ │Video Information Items (76 points) │ ├──────────────────────┬─────────────────────────┬──────────────────────────────────────────┬──────┤ │Artifact(10) │Artifacts caused by movement │There is no movement caused by breathing or disruption of the video sequence. │2 │ │ │is missing. ├──────────────────────────────────────────┼──────┤ │ │ │The boundary between the ascending aorta and the pulmonary artery due to the heartbeat │2 │ │ │ │does not blur. │ │ │ │ ├──────────────────────────────────────────┼──────┤ │ │ │The boundary between the descending aorta and the pulmonary veins due to the heartbeat │2 │ │ │ │does not blur. │ │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Line velocity interfering with reading │Linear beam hardening in the entire thoracic region excluding the shoulder joint │2 │ │ │beam hardening │(beam hardening) There are no artifacts. │ │ │ │No artifacts │ │ │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Generated by the device itself │There are no equipment artifacts. │2 │ │ │There are no artifacts. │ │ │ ├──────────────────────┼─────────────────────────┼──────────────────────────────────────────┼──────┤ │Image Scan Range │Transverse at the apex of the lung │Includes at least one image of the lung apex. │2 │ │(12) │Includes diaphragmatic-costal angle ├──────────────────────────────────────────┼──────┤ │ │. │At least one image showing the lung below the diaphragmatic-costal angle │2 │ │ │ │. │ │ │ │ ├──────────────────────────────────────────┼──────┤ │ │ │Completely includes the adrenal gland. │2 │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Includes the entire chest wall. │The chest wall completely covers the skin from the upper part of the diaphragm. │2 │ │ │ ├──────────────────────────────────────────┼──────┤ │ │ │Includes more than 50% of the diameter of both humeral heads. │2 │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │image │The left-right diameter of the torso at the upper part of the diaphragm is directly on the image screen │2 │ │ │The width occupied by │is more than 80% of the. │ │ │ │The ratio is 80% or higher. │ │ │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Excessive examination area │Includes the part above the upper part of the vocal cords. │-3 │ │ │ ├──────────────────────────────────────────┼──────┤ │ │ │Includes the part below the lower end of the kidney. │-3 │ ├──────────────────────┼─────────────────────────┴──────────────────────────────────────────┼──────┤ │Contrast agent indication │The name of the contrast agent is described. │3 │ │(5) ├────────────────────────────────────────────────────────────────────┼──────┤ │ │The total amount of contrast agent is described. │2 │ ├──────────────────────┼────────────────────────────────────────────────────────────────────┼──────┤ │Contrast-enhanced │The aorta is adequately enhanced. │2 │ │Adequacy ├────────────────────────────────────────────────────────────────────┼──────┤ │(8) │The pulmonary artery is adequately enhanced down to below the hilar region. │2 │ │ ├────────────────────────────────────────────────────────────────────┼──────┤ │ │The contrast-enhanced internal mammary artery is differentiated. │2 │ │ ├────────────────────────────────────────────────────────────────────┼──────┤ │ │The image quality remains consistent from the scan time to the end. │2 │ ├──────────────────────┼────────────────────────────────────────────────────────────────────┼──────┤ │Video Reconstruction │There is an image series of the soft tissue standard algorithm. │4 │ │Adequacy ├────────────────────────────────────────────────────────────────────┼──────┤ │(9) │There is a series of images using a closed high-resolution (sharp) algorithm. │3 │ │ ├────────────────────────────────────────────────────────────────────┼──────┤ │ │The video window can be adjusted. │2 │ ├──────────────────────┼─────────────────────────┬──────────────────────────────────────────┼──────┤ │Section thickness and │Axial image slice thickness │5㎜ or less. │4 │ │Multiplane │ ├──────────────────────────────────────────┼──────┤ │video's │ │Includes images with a thin thickness of 2 mm or less. │3 │ │Adequacy ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │(17) │Axial image intercept interval │No. │4 │ │ ├─────────────────────────┴──────────────────────────────────────────┼──────┤ │ │Includes physiognomy videos. │1 │ │ ├─────────────────────────┬──────────────────────────────────────────┼──────┤ │ │Self-slice thickness of coronal image │5㎜ or less. │1 │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Intercept interval of coronal image │No. │1 │ │ ├─────────────────────────┴──────────────────────────────────────────┼──────┤ │ │Includes the awards ceremony video. │1 │ │ ├─────────────────────────┬──────────────────────────────────────────┼──────┤ │ │Self-slice thickness of sagittal image │5㎜ or less. │1 │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Intercept interval of sagittal image │No. │1 │ │ ├─────────────────────────┴──────────────────────────────────────────┼──────┤ │ │There are artifacts (step-ladder artifact, etc.) in the coronal or sagittal images. │-3  │ ├──────────────────────┼─────────────────────────┬──────────────────────────────────────────┼──────┤ │Resolution and │Less than 1/3 of the terminal region │Pulmonary blood vessels of the entire lung excluding the region immediately below the pleura │3 │ │Contrast │Differentiating the pulmonary blood vessels of START │system. │ │ │(15) │can. │ │ │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │subsegmental │The bronchial walls below the subsegmental level are distinguished. │3 │ │ │Can distinguish bronchi. │ │ │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Mediastinum (Left and right pleural cavity │Mediastinal fat, thymic tissue, lymph nodes, pericardial fat │3 │ │ │) Old │clearly distinguishable. │ │ │ │Can distinguish the START group. │ │ │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │surrounding tissues │Can distinguish the mediastinal structures adjacent to the external wall of the trachea │2 │ │ │can be divided. │There is. │ │ │ │ ├──────────────────────────────────────────┼──────┤ │ │ │The presence of lymph nodes surrounding the trachea is confirmed. │2 │ │ ├─────────────────────────┼──────────────────────────────────────────┼──────┤ │ │Lymph nodes of the hilar region and lungs │At the hilar region, major pulmonary vessels and lymph nodes are clearly distinguished. │1 │ │ │Can distinguish blood vessels. ├──────────────────────────────────────────┼──────┤ │ │ │The course of the pulmonary artery and pulmonary vein can be separated and traced at the hilar region │1 │ │ │ │There is. │ │ └──────────────────────┴─────────────────────────┴──────────────────────────────────────────┴──────┘ 4) Low-dose chest ┏━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit non-contrast enhanced images. ┃ ┃of │2. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution ┃ ┃Leading Row │. ┃ ┃gun │3. For Picture Archiving and Communication System (PACS) images, changing lung and mediastinum settings ┃ ┃ │It must be possible. For film images, pulmonary and mediastinal settings must exist separately. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────────┬────────────────────────────────────────────────────────────────────┬──────────┐ │Item │EvaluationContent │Score  │ ├──────────────────┴────────────────────────────────────────────────────────────────────┴──────────┤ │Inspection Labels and General Information Items (4 points) │ ├──────────────────┬─────────────────────┬──────────────────────────────────────────────┬──────────┤ │Cover │General Information │There is a scan image (scanogram). │1 │ │(4) │ ├──────────────────────────────────────────────┼──────────┤ │ │ │The video sequence (video number, scan time, etc.) is confirmed. │1 │ │ │ ├──────────────────────────────────────────────┼──────────┤ │ │ │A scaler is indicated. │1 │ │ │ ├──────────────────────────────────────────────┼──────────┤ │ │ │Left and right directions are indicated. │1 │ ├──────────────────┴─────────────────────┴──────────────────────────────────────────────┴──────────┤ │Equipment and Radiation Exposure Management Standards (20 points) │ ├──────────────────┬────────────────────────────────┬───────────────────────────────────┬──────────┤ │Equipment Standard │MultipleDetectionCT │64 channels or more (16 channels or more but less than 64 channels, 6 points) │10 │ │(10) │(multidetector CT) │ │ │ │ │Detector Channel Reference │ │ │ ├──────────────────┼────────────────────────────────┴───────────────────────────────────┼──────────┤ │Radiation │There is a dose report. │4 │ │Radiation Exposure Management ├────────────────────────────────────────────────────────────────────┼──────────┤ │(10) │Imaging is performed at a tube voltage of 120 kVp or less. │2 │ │ ├────────────────────────────────────────────────────────────────────┼──────────┤ │ │Shot at 40-60 mAs or less. │2 │ │ ├────────────────────────────────────────────────────────────────────┼──────────┤ │ │The dose index (CTDIvol) is 3 mGy or less. │2 │ ├──────────────────┴────────────────────────────────────────────────────────────────────┴──────────┤ │Video Information Items (76 points) │ ├──────────────────┬─────────────────────────────────────────┬──────────────────────────┬──────────┤ │Artifact │There are no artifacts caused by movement. │In the lung settings, there is no movement due to respiration and the image │3 │ │(14) │ │There is no disorder in the order. │ │ │ │ ├──────────────────────────┼──────────┤ │ │ │ascending aorta and pulmonary artery by heartbeat │2 │ │ │ │The boundary does not blur. │ │ │ │ ├──────────────────────────┼──────────┤ │ │ │The descending aorta and pulmonary veins by heartbeats │2 │ │ │ │The boundary does not blur. │ │ │ ├─────────────────────────────────────────┼──────────────────────────┼──────────┤ │ │interfering with read │Pre-artificial structures present but not invading the lung parenchyma │2 │ │ │There are no streaking artifacts. │. │ │ │ ├─────────────────────────────────────────┼──────────────────────────┼──────────┤ │ │Image noise interfering with reading │Evaluating the presence of lung nodules despite image noise │3 │ │ │(noise) is not. │can. │ │ │ ├─────────────────────────────────────────┼──────────────────────────┼──────────┤ │ │Artifacts generated from the device itself │There are no equipment artifacts. │2 │ │ │No. │ │ │ ├──────────────────┼─────────────────────────────────────────┼──────────────────────────┼──────────┤ │VideoScan │From the apex of the lung to the diaphragm and costal angles │At least one more image is taken than the lung apex. │2 │ │range │was included. ├──────────────────────────┼──────────┤ │(12) │ │Photo showing no lungs visible below the diaphragm and costal angle │2 │ │ │ │. │ │ │ │ ├──────────────────────────┼──────────┤ │ │ │Completely includes the adrenal gland. │2 │ │ ├─────────────────────────────────────────┼──────────────────────────┼──────────┤ │ │Includes the entire chest wall. │From the upper diaphragm to the chest wall completely down to the skin │2 │ │ │ │includes. │ │ │ │ ├──────────────────────────┼──────────┤ │ │ │Includes more than 50% of the diameter of both humeral heads. │2 │ │ ├─────────────────────────────────────────┼──────────────────────────┼──────────┤ │ │Maximum left-right width of the chest wall in the START image │The left and right diameters of the torso at the upper part of the diaphragm are on the screen │2 │ │ │is 80% or more │It is more than 80% of the diameter. │ │ │ │. │ │ │ │ ├─────────────────────────────────────────┼──────────────────────────┼──────────┤ │ │Excessive examination area │Includes the part above the upper part of the vocal cords. │-3 │ │ │ ├──────────────────────────┼──────────┤ │ │ │Below the lower end of the height was included. │-3 │ ├──────────────────┼─────────────────────────────────────────┴──────────────────────────┼──────────┤ │Video │There is an image series of the soft tissue standard algorithm. │4 │ │Reconfiguration ├────────────────────────────────────────────────────────────────────┼──────────┤ │Adequacy │There is a series of images using a closed high-resolution (sharp) algorithm. │4 │ │(11) ├────────────────────────────────────────────────────────────────────┼──────────┤ │ │Apply the iterative reconstruction algorithm. │3 │ ├──────────────────┼─────────────────────────────────────────────┬──────────────────────┼──────────┤ │section thickness │Axial image slice thickness │3㎜ or less. │5 │ │and multi-plane │ ├──────────────────────┼──────────┤ │video's │ │Includes images with a thin thickness of 1.5 mm or less. │5 │ │Adequacy ├─────────────────────────────────────────────┼──────────────────────┼──────────┤ │(21) │Axial image intercept interval │No. │4 │ │ ├─────────────────────────────────────────────┴──────────────────────┼──────────┤ │ │Includes physiognomy videos. │2 │ │ ├──────────────────────────────────────────────┬─────────────────────┼──────────┤ │ │Self-slice thickness of coronal image │3㎜ or less. │1 │ │ ├──────────────────────────────────────────────┼─────────────────────┼──────────┤ │ │Intercept interval of coronal image │No. │1 │ │ ├──────────────────────────────────────────────┴─────────────────────┼──────────┤ │ │Includes the awards ceremony video. │1 │ │ ├──────────────────────────────────────────────┬─────────────────────┼──────────┤ │ │Self-slice thickness of sagittal image │3㎜ or less. │1 │ │ ├──────────────────────────────────────────────┼─────────────────────┼──────────┤ │ │Intercept interval of sagittal image │No. │1 │ │ ├──────────────────────────────────────────────┴─────────────────────┼──────────┤ │ │There are artifacts (step-ladder artifact, etc.) in the coronal or sagittal images. │-3 │ ├──────────────────┼──────────────────────────────────────────────────┬─────────────────┼──────────┤ │Resolution and │Can distinguish terminal pulmonary blood vessels. │Pneumothorax of the entire lung excluding the immediate subpleural region│3 │ │Contrast │ │The boundary of the tube is clear. │ │ │(18) ├──────────────────────────────────────────────────┼─────────────────┼──────────┤ │ │Proximal part subsegmental │Proximal subsegmental trachea│3 │ │ │Can distinguish bronchial tubes. │The branch walls are distinguished. │ │ │ ├──────────────────────────────────────────────────┼─────────────────┼──────────┤ │ │Interleaf fissures (gaps between leaves) are visible. │The interleaf fissures appear as lines. │3 │ │ ├──────────────────────────────────────────────────┼─────────────────┼──────────┤ │ │Mediastinum (the area between the left and right pleural cavities │Mediastinal fat, pericarp fat, thymus (chest│3 │ │ │분) The structure is distinguished. │Sam) Tissues are distinguished. │ │ │ ├──────────────────────────────────────────────────┼─────────────────┼──────────┤ │ │The boundaries of the major central vessels are visible. │The boundary between the main pulmonary artery and the aorta is visible. │3 │ │ ├──────────────────────────────────────────────────┼─────────────────┼──────────┤ │ │Can distinguish between organs and surrounding tissues. │The outer wall of the organ is shielded from surrounding tissues│3 │ │ │ │can be divided. │ │ └──────────────────┴──────────────────────────────────────────────────┴─────────────────┴──────────┘ 5) Dynamic contrast-enhanced liver ┏━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit contrast-enhanced dynamic liver images. Perform dynamic liver images ┃ ┃of │If not done, a single-time contrast-enhanced image may be submitted. ┃ ┃Leading Row │2. For Picture Archiving and Communication System (PACS) images, lung settings must be changeable ┃ ┃gun │Start. For film images, there must be a discard bottom closing setting. ┃ ┃ │3. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution ┃ ┃ │. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌─────────────────────┬────────────────────────────────────────────────────────────────────────┬───┐ │Item │EvaluationContent │Point │ │ │ │수  │ ├─────────────────────┴────────────────────────────────────────────────────────────────────────┴───┤ │Inspection Labels and General Information Items (4 points) │ ├─────────────────────┬─────────────┬──────────────────────────────────────────────────────────┬───┤ │Cover │General Information │There is a scan image (scanogram). │1 │ │(4) │ ├──────────────────────────────────────────────────────────┼───┤ │ │ │The video sequence (video number, scan time, etc.) is confirmed. │1 │ │ │ ├──────────────────────────────────────────────────────────┼───┤ │ │ │A scaler is indicated. │1 │ │ │ ├──────────────────────────────────────────────────────────┼───┤ │ │ │Left and right directions are indicated. │1 │ ├─────────────────────┴─────────────┴──────────────────────────────────────────────────────────┴───┤ │Equipment and Radiation Exposure Management Standards (20 points) │ ├─────────────────────┬────────────────────────────────────────────┬───────────────────────────┬───┤ │Equipment Standard │MultipleDetectionCT │64 channels or more (16 channels or more but less than 64 channels 6 │10 │ │(10) │(multidetector CT) │point, 2 channels or more but less than 16 channels 3 points) │ │ │ │Detector Channel Reference │ │ │ ├─────────────────────┼────────────────────────────────────────────┴───────────────────────────┼───┤ │Radiation │There is a dose report. │5 │ │Radiation Exposure Management ├────────────────────────────────────────────────────────────────────────┼───┤ │(10) │Uses automatic exposure control (AEC). │4 │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │Inspected at tube voltage of 100 kVp or less. │1 │ ├─────────────────────┴────────────────────────────────────────────────────────────────────────┴───┤ │Video Information Items (76 points) │ ├─────────────────────┬────────────────────────────────────────────────────────────────────────┬───┤ │Artifact │There are no ring artifacts. │3 │ │(10) ├────────────────────────────────────────────────────────────────────────┼───┤ │ │No beam-hardening artifacts. │3 │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │There are no respiratory artifacts or artifacts caused by patient movement. │1 │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │The patient is located in the center of the scan. │1 │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │Examine with both arms raised. │1 │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │No body ornaments, unnecessary equipment, or devices affecting the scan. │1 │ ├─────────────────────┼────────────────────────────────────────────────────────────────────────┼───┤ │Image Scan Range │All images include the entire liver. │5 │ │(11) ├────────────────────────────────────────────────────────────────────────┼───┤ │ │There is one or more lung field images that do not include the liver. │3 │ │ ├───────────────────────────────────┬────────────────────────────────────┼───┤ │ │Excessive examination site │Includes the lung region above the tracheobuncle. │-3 │ │ │ ├────────────────────────────────────┼───┤ │ │ │5 cm below the ischial tuberosity │-3 │ │ │ │side. │ │ │ ├───────────────────────────────────┴────────────────────────────────────┼───┤ │ │The entire abdominal wall was included, and the proportion occupied by the maximum left-right width of the abdominal wall in the image │3 │ │ │80% or more. │ │ ├─────────────────────┼────────────────────────────────────────────────────────────────────────┼───┤ │Multiple timing │It consists of pre-contrast imaging, hepatic arterial phase imaging, portal venous phase imaging, and delayed phase imaging. │5 │ │VideoComposition │ │ │ │(5) │ │ │ ├─────────────────────┼────────────────────────────────────────────────────────────────────────┼───┤ │Contrast agent indication │. │4 │ │(6) ├────────────────────────────────────────────────────────────────────────┼───┤ │ │The total amount of contrast agent is described. │2 │ ├─────────────────────┼────────────────────────────────────────────────────────────────────────┼───┤ │Video Reconstruction │Uses an algorithm (soft standard reconstruction kernel) appropriate for abdominal images. │2 │ │Adequacy ├────────────────────────────────────────────────────────────────────────┼───┤ │(4) │The video window can be adjusted. │2 │ ├─────────────────────┼────────────────────────────────────────────┬───────────────────────────┼───┤ │Section thickness and │Axial image slice thickness │1㎜ or more and 5㎜ or less (5㎜ or more and 7㎜ or less │4 │ │Multi-plane │ │if 2 points). │ │ │video's ├────────────────────────────────────────────┼───────────────────────────┼───┤ │Adequacy │Axial image intercept interval │No. │4 │ │(15) ├────────────────────────────────────────────┴───────────────────────────┼───┤ │ │Includes one or more images of contemplation or occultation. │3 │ │ ├────────────────────────────────────────────┬───────────────────────────┼───┤ │ │Self slice thickness of coronal or sagittal images │1㎜ or more and 5㎜ or less (5㎜ or more and 8㎜ or less │2 │ │ │ │if 1 point). │ │ │ ├────────────────────────────────────────────┼───────────────────────────┼───┤ │ │Intercept interval of coronal or sagittal images │No. │2 │ │ ├────────────────────────────────────────────┴───────────────────────────┼───┤ │ │There are artifacts (step-ladder artifact, etc.) in the coronal or sagittal images. │-3  │ ├─────────────────────┼───────────────────────┬────────────────────────────────────────────────┼───┤ │Video Resolution │Distinctness of right intrahepatic artery │Right intrahepatic artery greater than 50% hepatic parenchymal thickness on hepatic arterial phase imaging │4 │ │and Contrast │ │region (in the region of less than 50% of liver parenchyma thickness │ │ │(25) │ │). │ │ │ ├───────────────────────┴────────────────────────────────────────────────┼───┤ │ │In contrast-enhanced images from at least one period, the celiac artery, gastroduodenal artery, and gastrointestinal mesentery │3 │ │ │The arteries and renal arteries are clearly visible. │ │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │In at least one contrast-enhanced image, the main portal vein is clearly distinct from surrounding organs and structures │3 │ │ │is distinguished. │ │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │In hepatic portal phase imaging or single-period contrast-enhanced imaging, the hepatic portal vein in a 2 cm subcapsular region │3 │ │ │. │ │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │The entire pancreas is visible on hepatic portal phase imaging or single-phase contrast-enhanced imaging, and the... │3 │ │ │system. │ │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │In at least one time period, mesenteric (internal membrane) vessels are clearly separated from the intestinal wall │3 │ │ │can distinguish. │ │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │The first branches of the bilateral intrahepatic bile ducts can be distinguished in at least one imaging period. │3 │ │ ├────────────────────────────────────────────────────────────────────────┼───┤ │ │The common bile duct can be distinguished in at least one imaging period. │3 │ └─────────────────────┴────────────────────────────────────────────────────────────────────────┴───┘ 6) Non-contrast enhanced abdominal ┏━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Routine non-contrast enhanced images performed to evaluate urinary stones ┃ ┃of │is output. If a separate test for evaluating urinary stones is not performed, the abdomen and ┃ ┃Leading Row │Non-contrast enhanced images including the pelvis may be submitted. ┃ ┃gun │2. Submit the best-shot video among the videos from a specific point in time designated by the evaluation institution. ┃ ┃ │. ┃ ┃ │3. In the case of Picture Archiving and Communication System (PACS) images, lung settings must be changeable ┃ ┃ │. In the case of film footage, there must be a discard bottom closing setting. ┃ ┃ │4. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌─────────┬─────────────────────────────────────────────────────────────────────────────────────┬──┐ │Item │EvaluationContent │Point │ │ │ │수 │ ├─────────┴─────────────────────────────────────────────────────────────────────────────────────┴──┤ │Inspection Labels and General Information Items (4 points) │ ├─────────┬───────────────┬─────────────────────────────────────────────────────────────────────┬──┤ │Cover │General │. │1 │ │(4) │Info ├─────────────────────────────────────────────────────────────────────┼──┤ │ │ │The video sequence (video number, scan time, etc.) is confirmed. │1 │ │ │ ├─────────────────────────────────────────────────────────────────────┼──┤ │ │ │A scaler is indicated. │1 │ │ │ ├─────────────────────────────────────────────────────────────────────┼──┤ │ │ │Left and right directions are indicated. │1 │ ├─────────┴───────────────┴─────────────────────────────────────────────────────────────────────┴──┤ │Equipment and Radiation Exposure Management Standards (20 points) │ ├─────────┬────────────────────────────────────────────┬────────────────────────────────────────┬──┤ │Equipment Standard │MultipleDetectionCT │16 channels or more (2 channels or more but less than 16 channels 5 points) │10 │ │(10) │(multidetector CT) │ │ │ │ │Detector Channel Reference │ │ │ ├─────────┼────────────────────────────────────────────┴────────────────────────────────────────┼──┤ │Radiation │There is a dose report. │6 │ │Radiation Exposure Management ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │(10) │Uses automatic exposure control (AEC). │2 │ │ ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │Inspected at tube voltage of 100 kVp or less. │2 │ ├─────────┴─────────────────────────────────────────────────────────────────────────────────────┴──┤ │Video Information Items(76 points) │ ├─────────┬─────────────────────────────────────────────────────────────────────────────────────┬──┤ │Artifact │There are no ring artifacts. │5 │ │(20) ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │Beam-hardening: No artifacts (Artifacts are present but do not impede image interpretation) │5 │ │ │if 2 points). │ │ │ ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │There are no respiratory artifacts or artifacts caused by patient movement. │3 │ │ ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │The patient is located in the center of the scan. │2 │ │ ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │Examine with both arms raised. │3 │ │ ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │No body ornaments, unnecessary equipment, or devices affecting the scan. │2 │ ├─────────┼─────────────────────────────────────────────────────────────────────────────────────┼──┤ │Video Scan │Includes the entire kidney, urinary tract, and bladder. │5 │ │range ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │(10) │The entire abdominal wall is included, and the ratio of the maximum left-right width of the abdominal wall in the image is 80% │5 │ │ │This is the end. │ │ │ ├────────────────────────────────┬────────────────────────────────────────────────────┼──┤ │ │Excessive examination site │Includes the lung region above the tracheobuncle. │-3 │ │ │ ├────────────────────────────────────────────────────┼──┤ │ │ │The English area including 5 cm below the ischial tuberosity │-3 │ │ │ │award. │ │ ├─────────┼────────────────────────────────┴────────────────────────────────────────────────────┼──┤ │VideoReconstruction │Uses an algorithm (soft standard reconstruction kernel) appropriate for abdominal images. │3 │ │Adequacy ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │(6) │Video window adjustment is possible. │3 │ ├─────────┼──────────────────────────────────────────────────────────┬──────────────────────────┼──┤ │of slice thickness │Axial image slice thickness │1㎜ or more and 5㎜ or less (over 5㎜ and 7㎜ or less │4 │ │Adequacy │ │2 points). │ │ │(20) ├──────────────────────────────────────────────────────────┼──────────────────────────┼──┤ │ │Axial image intercept interval │No. │4 │ │ ├──────────────────────────────────────────────────────────┴──────────────────────────┼──┤ │ │Includes one or more physiognomy or poetry videos. │5 │ │ ├──────────────────────────────────────────────────────────┬──────────────────────────┼──┤ │ │Self slice thickness of coronal or sagittal images │1㎜ or more and 5㎜ or less (over 5㎜ and 8㎜ or less │3 │ │ │ │for 1 point) │ │ │ ├──────────────────────────────────────────────────────────┼──────────────────────────┼──┤ │ │Intercept interval of coronal or sagittal images │No. │4 │ │ ├──────────────────────────────────────────────────────────┴──────────────────────────┼──┤ │ │There are artifacts (step-ladder artifact, etc.) in the coronal or sagittal images. │-3 │ ├─────────┼─────────────────────────────────────────────────────────────────────────────────────┼──┤ │Video │Examine with a full bladder. │4 │ │Resolution ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │and │Both renal pelvises and the proximal ureter are clearly visible (only the boundary of the renal pelvis is indistinct, or the proximal │4 │ │Contrast │2 points if not clearly visible at less than half the length of the ureter). │ │ │(20) ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │In the renal sinus, the renal pelvis and blood vessels can be distinguished and traced (the boundary is │4 │ │ │2 points if undistinguished areas are less than half). │ │ │ ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │Both distal ureters and the ureterovesical junction are clearly distinguishable (of the length of the distal ureter │4 │ │ │If less than half is not distinguished, or if the ureterovesical junction cannot be distinguished, 2 points). │ │ │ ├─────────────────────────────────────────────────────────────────────────────────────┼──┤ │ │The bladder and uterus/vagina (female), and the bladder and prostate/seminal vesicles (male) are distinguished (the difference in shading over two long periods │4 │ │ │This can be distinguished, but if the boundary is unclear, 2 points). │ │ └─────────┴─────────────────────────────────────────────────────────────────────────────────────┴──┘ 7) Lumbar vertebra ┏━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┓ ┃SubmissionVideo │1. Submit images focusing on the intervertebral disc. ┃ ┃of │2. Submit non-contrast enhanced images. ┃ ┃Leading Row │3. Submit the best-recorded video among the videos from a specific point in time designated by the evaluation agency. ┃ ┃gun │4. Bone settings and soft tissue settings for Picture Archiving and Communication System (PACS) images ┃ ┃ │settings. For film images, bone and soft tissue settings must be changeable for each ┃ ┃ │Each must be present. ┃ ┃ │5. On the general examination cover sheet, the patient's name, patient number, gender, age, date of imaging, and imaging institution are ┃ ┃ │Submit the marked video. ┃ ┗━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛ ┌──────────────┬───────────────────────────────────────────────────────────────────────────┬───────┐ │Item │EvaluationContent │Point │ │ │ │수  │ ├──────────────┴───────────────────────────────────────────────────────────────────────────┴───────┤ │Inspection Labels and General Information Items (4 points) │ ├──────────────┬────────────┬──────────────────────────────────────────────────────────────┬───────┤ │Cover │General │. │1 │ │(4) │Info ├──────────────────────────────────────────────────────────────┼───────┤ │ │ │The video sequence (video number, scan time, etc.) is confirmed. │1 │ │ │ ├──────────────────────────────────────────────────────────────┼───────┤ │ │ │A scaler is indicated. │1 │ │ │ ├──────────────────────────────────────────────────────────────┼───────┤ │ │ │Left and right directions are indicated. │1 │ ├──────────────┴────────────┴──────────────────────────────────────────────────────────────┴───────┤ │Equipment and Radiation Exposure Management Standards (20 points) │ ├──────────────┬────────────────────────────────────────────────┬──────────────────────────┬───────┤ │Equipment Standard │MultipleDetectionCT │16 channels or more (2 channels or more, if less than 16 channels │10 │ │(10) │(multidetector CT) │5 points) │ │ │ │Detector Channel Reference │ │ │ ├──────────────┼────────────────────────────────────────────────┴──────────────────────────┼───────┤ │Radiation Skin │There is a dose report. │5 │ │Width Management ├───────────────────────────────────────────────────────────────────────────┼───────┤ │(10) │is tested at a tube voltage of 120 kVp or less. │5 │ ├──────────────┴───────────────────────────────────────────────────────────────────────────┴───────┤ │Video Information Items(76 points) │ ├──────────────┬───────────────────────────────────────────────────────────────────────────┬───────┤ │Artifact │There are no respiratory artifacts or artifacts caused by patient movement. │4 │ │(10) ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │There are no ring artifacts. │3 │ │ ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │There are no other artifacts. │3 │ ├──────────────┼───────────────────────────────────────────────────────────────────────────┼───────┤ │Video Scan │Includes 5 or more intervertebral discs (up to 3 points, 2 or fewer 0 points). │6 │ │range ├───────────────────────────────────────────────────────────────────────────┼───────┤ │(30) │Acquire or reconstruct images in a direction parallel to each intervertebral disc (6 points for 4 or more, │6 │ │ │Up to 3 items are 3 points). │ │ │ ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │Sufficiently includes the posterior contour of each intervertebral disc (not sufficiently includes the posterior contour │6 │ │ │3 points if there is 1 intervertebral disc, 0 points if there are 2 or more). │ │ │ ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │For three or more intervertebral discs, the superior and inferior endplates are sufficiently included (up to two: 3 points, │6 │ │ │1 or fewer is 0 points). │ │ │ ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │It is possible to identify which part of the lumbar spine each cross-sectional image corresponds to (part of the scan image │6 │ │ │If the above mark is not present or is displayed only as text in each corresponding video, 3 points). │ │ ├──────────────┼───────────────────────────────────────────────────────────────────────────┼───────┤ │Video │There is an image series of the soft tissue standard algorithm. │2 │ │Reconfiguration ├───────────────────────────────────────────────────────────────────────────┼───────┤ │Adequacy(4) │There is a series of images from the bone high-resolution (sharp) algorithm. │2 │ ├──────────────┼───────────────────────────────────────────────────────────────────────────┼───────┤ │SectionThickness │The section thickness is 3 mm or less. │4 │ │and multi-plane ├───────────────────────────────────────────────────────────────────────────┼───────┤ │video's │There is no segmental space in the intervertebral disc region. │4 │ │Adequacy ├───────────────────────────────────────────────────────────────────────────┼───────┤ │(12) │There is a video series reconstructing the awards ceremony. │4 │ ├──────────────┼───────────────────────────────────────────────────────────────────────────┼───────┤ │Resolution and │Can trace an outline along the outside of the intervertebral disc (intervertebral discs that are not outlined 1 │5 │ │Contrast │3 points if there are 3, 0 points if there are 2 or more). │ │ │(20) ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │The intervertebral disc and the dural cavity can be distinguished (there is one intervertebral disc that is not distinguished from the dural cavity │5 │ │ │3 points if it is, 0 points if 2 or more). │ │ │ ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │The thickness of the yellow ligament can be measured at the 4th and 5th lumbar vertebrae (the measurement site and measurement │5 │ │ │If parts that do not work are mixed in, 2 points). │ │ │ ├───────────────────────────────────────────────────────────────────────────┼───────┤ │ │Bilateral facet joint spaces are visible (only one facet joint space is not visible at a single intervertebral disc location │5 │ │ │2 points if visible, 0 points if not visible on both sides or not visible at two or more intervertebral disc locations). │ │ └──────────────┴───────────────────────────────────────────────────────────────────────────┴───────┘