Annex 4 — 진단용 방사선 발생장치의 안전관리에 관한 규칙
288071 · 2026-07-09
■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Appendix 4] <Amended Sep. 11, 2020>
Registration Standards for Inspection and Measurement Agencies (Related to Article 6, Paragraph 1)
1. Inspection agencies (agencies inspecting diagnostic radiation generating devices and radiation shielding facilities) for diagnostic
Inspection of corporations, institutions or organizations, or internal bodies related to the quality control of radiation generating devices
(hereinafter referred to as the “Self-Inspection Institution”) as
It must meet the criteria of each item. However, in the case of in-house testing institutions, the Commissioner of the Korea Disease Control and Prevention Agency [determined]
Some of the criteria in each of the following items may be exempted as per the .
a. An inspection agency has a conflict of interest with a manufacturer, importer, or repairer of diagnostic radiation generating devices
It will be an independent organization without START
B. Establish and maintain a quality assurance system in accordance with international standards.
. It shall be equipped with facilities necessary for inspection operations, inspection equipment storage facilities, and comparative measurement facilities.
d. Test items of the inspection standards for diagnostic radiation generating devices in Attached Table 1 and radiation in Attached Table 2
To measure and inspect radiation shielding facilities and radiation protection equipment according to the inspection standards for protection facilities
Equip with equipment capable of START
E. Inspection equipment shall, through standard materials or calibration, ensure that its accuracy and precision meet national measurement standards or national
The correlation shall be maintained to be consistent with the measurement standard
B. Appoint a Technical Officer with any of the following qualifications
1) Radiology specialist or a bachelor's degree in Physics, Electrical, Electronic, Mechanical Engineering, or Radiology from a four-year university
Persons who have graduated from the Department of Biomedical Engineering, the Department of Nuclear Engineering, or related science and engineering departments (in accordance with laws and regulations and
(including persons recognized as having the same level of educational qualifications) as a diagnostic person for 4 years or more
Person engaged in technical work related to the inspection of radiation generating devices
2) A person who has graduated from a department specified in 1) of a junior college (an academic degree of the same level in accordance with laws and regulations
(including persons recognized as having experience) as a diagnostic radiation generating facility for 6 years or more
Person engaged in technical work related to the inspection of dental values
3) At least 4 years of technical experience as a radiographer in technical duties related to the inspection of diagnostic radiation generating devices
Person who died
Company. An inspection team consisting of two or more inspectors with at least one year of practical experience in the field
Will have a team
A. Chief Technology Officers and team leader-level inspectors shall possess technical competence in the following fields and knowledge of relevant laws and regulations.
be equipped with the formula
1) Radiation Physics
2) Radiation Safety
3) Inspection of diagnostic radiation generating devices
In addition, it shall comply with the standards prescribed and notified by the Commissioner of the Korea Disease Control and Prevention Agency.
2. A measuring institution (an institution that measures the exposure dose of radiation-related workers) is [subject to] the “National Standards Framework Act”
An institution designated as a national calibration institution (radiation field) pursuant to Article 14 of the Atomic Energy Safety Act or
An institution that has registered reading services pursuant to Article 78, for the measurement of its own institution, certain measurements
An organization with the capability (hereinafter referred to as the “Self-Measurement Organization”) that meets the criteria of each of the following items
must be done. However, in the case of self-measurement institutions, as determined by the Commissioner of the Korea Disease Control and Prevention Agency, the following
Some of the criteria in each item may be exempted.
a. Establish and maintain a quality assurance system in accordance with international standards.
B. Be equipped with facilities necessary for measurement operations
. Equipped with individual exposure dose measuring devices and reading devices to synthesize, analyze, and process measurement results
Equip yourself with a device capable of replacement, but ensure that measurement operations are not interrupted due to breakdown, etc.
Possess the device
. Inspection equipment shall, through standard materials or calibration, ensure accuracy and precision according to national measurement standards or national
The correlation shall be maintained to be consistent with the measurement standard
E. Appoint a Chief Technology Officer with any of the following qualifications:
1) Radiology specialist or a bachelor's degree in Physics, Electrical, Electronic, Mechanical Engineering, or Radiology from a four-year university
Persons who have graduated from the Department of Biomedical Engineering, the Department of Nuclear Engineering, or related science and engineering departments (in accordance with laws and regulations and
(including persons recognized as having the same level of educational qualifications) as a diagnostic person for 4 years or more
Person engaged in technical work related to the inspection of radiation generating devices
2) A person who has graduated from a department specified in 1) of a junior college (an academic degree of the same level in accordance with laws and regulations
(including persons recognized as having experience) as a diagnostic radiation generating facility for 6 years or more
Person engaged in technical work related to the inspection of dental values
3) At least 4 years of technical experience as a radiographer in technical duties related to the inspection of diagnostic radiation generating devices
Person who died
B. Secure two or more measurement personnel as full-time employees
d. In addition, it shall comply with the standards prescribed and notified by the Commissioner of the Korea Disease Control and Prevention Agency