■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Appendix 3] <Amended Jan. 22, 2010>
Dose Limits for Radiation Workers (Related to Article 4, Paragraph 6)
┏━━━━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━━━━━┓
┃Exposure Classification │Dose Limit ┃
┠────────────┼─────────────────────────┨
┃Effective Dose │It must be 50 mSv (5 rem) or less per year, and the cumulative amount over 5 years ┃
┃ │The dose must be 100 mSv (10 rem) or less. ┃
┠────────────┼─────────────────────────┨
┃Equivalent Dose (Lens) │Must be 150 mSv (15 rem) or less per year. ┃
┠────────────┼─────────────────────────┨
┃Equivalent Dose (Skin, Hands and Feet)│Must be 500 mSv (50 rem) or less per year. ┃
┗━━━━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━━━━━┛
1. Effective dose refers to the degree of risk based on the dose distribution among tissues within the human body as a single quantity
To indicate the equal amount of each tissue for all tissues of the human body exposed to radiation
It refers to the amount obtained by summing the results of multiplying the radiation dose by the tissue weighting factor of the relevant tissue.
The sum of the weights for whole-body exposed tissues shall be 1.
2. Equivalent dose refers to the absorbed dose of the corresponding radiation to indicate the exposure dose to the human body
It refers to the amount multiplied by the radiation weighting factor. In this case, the radiation weighting factor for diagnostic X-rays is
1.
3. Detailed information necessary for measurement, such as methods for measuring dose limits for radiation workers, is in formula
It is determined by the head of the measuring institution upon approval by the Director of the Korea Institute of Food and Drug Safety Evaluation.
4. The annual dose limit is for each year regardless of when the first exposure dose was measured during the year
Based on the last day of December, but for radiation workers using T.L. badges
The quantity limit shall be the sum of the quantities measured at the last day of each quarter.
5. The 5-year cumulative dose limit is January 2008, regardless of when the initial exposure dose was measured
It shall be the cumulative dose over every 5 years based on 1 day.