Annex 2 — 진단용 방사선 발생장치의 안전관리에 관한 규칙

288071 · 2026-07-09

■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Appendix 2] <Amended July 18, 2025> Inspection Standards for Radiation Protection Facilities (Related to Article 4, Paragraph 3) 1. Radiation Protection Facilities A. Radiation protection facilities in the examination room 1) A place where a diagnostic radiation generating device is installed to perform imaging for diagnostic purposes (hereinafter referred to as "imaging The ceiling, floor, and walls of the room referred to as the "Yeongsil" (movement) cannot be moved without using tools. or unremovable radiation shielding facilities (hereinafter referred to as "protective walls") must be installed and the sum of the radiation leakage dose and scattered dose measured outside the barrier is It must be 2.58 × C/kg (100 mR per week) or less. However, where persons pass or reside In directions where it is not directed, protective walls do not need to be installed, and radiation personnel other than Radiation leakage dose measured outside the protective wall installed in the direction of self-occupancy The sum of the START and scattered doses shall be 2.58 × C/kg (10 mR per week) or less. 2) 1), dental diagnostic X-ray generators (excluding panoramic and cephalothrics or a radiology room equipped solely with a diagnostic X-ray generator meets the following requirements For , the installation of a protective wall can be substituted with a medical X-ray shielding partition. a) X-ray generator for dental diagnosis (excluding panoramic and cephalometric) or diagnostic The X-ray generator must have a maximum weekly operating load of 10 mA·min or less. b) The leakage dose outside the medical X-ray shielding partition is 2.58×C/kg(per week 100mR) must be less than or equal to. ) The size of medical X-ray shielding partitions must be at least 0.8m in width × 1.8m in height . 3) Secondary defensive walls shall be installed on all walls, ceilings, and floors of the shooting room, excluding the primary defensive wall must be installed. In this case, the height of the protective wall must be at least 2m. 4) If the shielding material of the barrier is made of lead plate, it must meet the following requirements. a) It shall not wrinkle or bend due to the weight of the lead sheet itself, and Must not be directly subjected to physical impact. b) The parts where lead plates connect shall be overlapped with lead plates at least 1.5 cm wide. , and the part connecting to shielding material other than lead plate must be lead plate with a width of 2.5 cm or more Must be overlapped. 5) The film door must have a yield of at least 1mm. B. Radiation protection facilities in the control room The control room is a place partitioned from the imaging room and equipped with a control device for controlling diagnostic radiation, as follows: A patient viewing window that allows observing patient movement, etc., must be installed. . Radiation protection facilities of the control center 1) The control station is separated from the imaging room by a partially open barrier and uses diagnostic radiology is installed. 2) Floor surface of a control room where the control device is not attached to the X-ray high-voltage generator The area must be at least 1.5 m², and the distance between the protective wall and the inspection table must be at least 2 m. and, of the shielding wall and patient viewing window according to the maximum tube voltage of the diagnostic radiation generator is as follows. a) If the maximum tube voltage exceeds 100KV · The barrier wall shall have a yield equivalent of at least 1.5 mm, and its arrangement structure shall be as shown in Attached Table 9. The patient viewing window must be at least 1.5 mm soft equivalent, and the patient viewing window at 100KV The lead equivalent amount in must be marked so that it cannot be erased. b) If the maximum tube voltage is 100KV or less The barrier must be at least 1 mm thick, and its dimensions must be 1 m wide × 2 m long. Must be. · The patient viewing window must be at least 1 mm soft equivalent. 3) In the case of a control station where the control device is attached to an X-ray high-voltage generator, 2) criteria, and the investigation switch Exterior of shooting room It must be installed in the so that it can be operated. d. Radiation when a diagnostic radiation generator is moved outside a medical institution for use Defense Facility 1) Diagnostic radiation generators must be installed in mobile medical examination vehicles, and radiation workers operates the device, provided that radiation The leakage dose shall be 2.58 × 10⁻⁵ C/kg per week (referring to 100 mR per week) or less, and amount is 5.16×10⁻⁷ C/kg per week depending on the location of use (per week 2mR) must be less than or equal to. 2) 1), the weight is 10kg or less and the maximum tube current is 20 milliampere·s Portable X-ray imaging device or portable digital diagnostic X-ray imaging device with (mAs) or less imaging device can be used without being mounted on a mobile medical examination vehicle. In this case, behind the detector must be equipped with a medical X-ray shield, and at a radius of 2m from the X-ray tube In , the radiation leakage dose must be 5.16×10⁻⁷ C/kg per week (referring to 2 mR per week) or less. . 3) In cases where used in accordance with 2), within a radius of 2m of the diagnostic radiation generating device Install access control lines to restrict access by the general public other than patients at the medical X-ray room, and By installing protective partitions or using building walls, radiation to the general public other than the patient radiation Have them wear an X-ray protective apron. E. Mobile Diagnostic Radiation Generator (Mobile Unit, C-Arm, Portable Unit, For Dental Use Mobile device and mobile X-ray bone densitometer and computed tomography device) 1) Although radiation shielding facilities are not required, the medical X-ray shielding apron under Subparagraph 3(b) (In the case of dental hospitals and dental clinics, this refers to X-ray protective aprons for dental treatment) equipped must be. 2) When taking images in a location other than the operating room, emergency room, or intensive care unit Mobile Medical Use Must be equipped with an X-ray protective partition. 3) A mobile computed tomography device manufactured for operating rooms, intensive care units, and "infectious disease Only on inpatient treatment beds pursuant to Article 41, Paragraphs 1 and 2 of the Act on the Prevention and Management of Infectious Diseases may be used. In this case, a mobile medical X-ray shield must be equipped, and The radiation leakage dose outside the relevant protective partition is 2.58 × C/kg per week (per week 10mR) must be less than or equal to. 4) 1), a mobile diagnostic radiation generator in a single examination room In the case of use, radiation protection facilities must be equipped in accordance with the standards from subparagraphs (a) to (c). must. B. Radiation shielding facilities for mammography devices: The mammography device shall have a shielding wall in accordance with subparagraph 1) , and the radiation control device installed inside the imaging room , installation of a protective wall as a medical X-ray shielding partition in accordance with subparagraph 2)b) and c) may be substituted. However, the protective partition attached to the mammography device is the floor , and the dimensions must be at least 0.6m in width and Must be 1.8m or more. 2. Installation Standards for Diagnostic Radiation Generators in Imaging Rooms a. One unit must be installed per shooting room. b. Notwithstanding subparagraph (a), in cases where two or more units are installed in a single shooting room, conversion switches, etc. A simultaneous operation prevention device must be utilized to prevent two units from operating at the same time. . Among diagnostic radiation generating devices, computed tomography (dental computed tomography device (excluding computed tomography devices for otolaryngology and otolaryngology) must be installed in a dedicated examination room . However, manufactured as a mobile type, it is used in operating rooms, intensive care units, and the "Infectious Disease Prevention and Control Computerized system used in inpatient treatment beds pursuant to Article 41, Paragraphs 1 and 2 of the Act on START Tomography devices are excluded. 3. Radiation Protection Devices: When using diagnostic radiation generators, protection from radiation Devices for protecting persons and radiation workers (hereinafter referred to as "devices for protecting against radiation damage") (hereinafter referred to as "the standard") is a standard prescribed and notified by the Minister of Food and Drug Safety or the "Industrial Standardization Method." In this case, the ex-for medical treatment per medical institution X-ray protection apron (in the case of dental hospitals and dental clinics, X-ray protection apron for dental treatment (referring to 'e') including one of the radiation protection devices according to each of the following items, 2 . a. Medical X-ray shielding partition B. Medical X-ray protective apron . X-ray protective apron for dental treatment Ra. Device for protecting patient genitals E. Other devices for protection against radiation damage