Annex 1 — 진단용 방사선 발생장치의 안전관리에 관한 규칙

288071 · 2026-07-09

■ Rules on Safety Management of Diagnostic Radiation Generating Devices [Appendix 1] <Amended 2012.11.15> Inspection Standards for Diagnostic Radiation Generating Devices (Related to Article 4, Paragraph 1) ┏━━━━━━━━┯━━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━━┓ ┃MTQ97051fb4bce345a000140Type of START device │TestItem │Based ┃ ┠────────┼─────────┼─────────────────────┨ ┃1. Diagnostic Radiation │a. Check Grounding Equipment │The grounding terminal of the grounding facility shall be as per the Minister of Knowledge Economy ┃ ┃Line generator │Test │, the 3rd ┃ ┃(Diagnostic X-ray │ │grounded to the grounding wire by the grounding work ┃ ┃line device, true │ │. ┃ ┃Single-use X-ray ├─────────┼─────────────────────┨ ┃generator, breast │B. External leakage current │Diagnostic radiation generating device or diagnostic radiation ┃ ┃Imaging device) │Test │Each unit device and component constituting the START generator ┃ ┃ │ │Conductive parts that may be in contact with humans during START ┃ ┃ │ │From the ground or other contacts through the human body ┃ ┃ │ │External leakage current flowing in contactable conductive parts ┃ ┃ │ │is 0.1mA or less with secure grounding ┃ ┃ │ │must be connected. However, devices other than mounted devices┃ ┃ │ │With the ground wire removed, the external leakage current is 0.5mA ┃ ┃ │ │must be less than or equal to. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │. The re- │Performance and reliability of diagnostic radiation generators ┃ ┃ │formation test(automatic │As for evaluating the variation meter for the radiation dose ┃ ┃ │Exposure Controller Form │number must be 0.05 or less. The coefficient of variation refers to the ┃ ┃ │) │The standard deviation for the average value of the determined radiation dose ┃ ┃ │ │stating, it is calculated using the following formula. ┃ ┃ │ │ ┃ ┃ │ │ Remarks: CV=Coefficient of Variation ┃ ┃ │ │ S=For the population of radiation dose measurements ┃ ┃ │ │StandardDeviation ┃ ┃ │ │ =Average of measurements ┃ ┃ │ │ =i-th radiation dose measurement ┃ ┃ │ │ n=Number of measurements(Number of measurements must be 5 or more times ┃ ┃ │ │must) ┃ ┃ │ │ ┃ ┃ │ │ ┃ ┃ ├─────────┼─────────────────────┨ ┃ │ │ ┃ ┃ │ │ ┃ ┃ │. Tube voltage test │The Percentage Mean Error (PAE) of the tube voltage is the set value ( ┃ ┃ │ │It must be within ±10% for the value). ┃ ┃ │ │is is calculated using the following formula. ┃ ┃ │ │ ┃ ┃ │ │ ┃ ┃ │ │ Remarks: PAE=Percentage Mean Error ┃ ┃ │ │ Xp=Setting Value(Indicator Value) ┃ ┃ │ │ =Average of measured radiation dose ┃ ┃ ├─────────┼─────────────────────┨ ┃ │E. Tube Current Test │For short-time ratings, the percentage mean error is the indicated value ┃ ┃ │ │. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │Bar. X-ray irradiation │ ┃ ┃ │Test │ ┃ ┃ │ 1) Roughness Test │of the light irradiation field projected from the irradiation field control mechanism ┃ ┃ │ │Illuminance from the focal point of the X-ray tube to the image plane ┃ ┃ │ │average distance (hereinafter referred to as "SID") of 100 cm ┃ ┃ │ │The temperature must be 100 Lux or higher. ┃ ┃ │ 2) X-ray irradiation │Difference between the center of the X-ray irradiation field and the center of the optical irradiation field ┃ ┃ │and Gwangsajaya's │The difference between this and each surrounding area is that the center of the usage line is ┃ ┃ │DifferenceTest │When hired as a START employee, within ±2% of SID ┃ ┃ │ │must be constant. However, the direction of X-ray irradiation must be constant ┃ ┃ │ │, it must be within ±1% of SID. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │ │ ┃ ┃ │ │ ┃ ┃ │Inc. Other necessary │Tolerance for the indicated scale is within ±1 mm ┃ ┃ │test │should. ┃ ┃ │ 1) Fault Depth Display │ ┃ ┃ │test(Tomography │(1) In the case of a transformer type: the current duration is 0.1 seconds ┃ ┃ │MTQ97051fb4bce345a000216Applies only to START) │When it is a START, the percentage average error relative to the indicated value ┃ ┃ │ 2) mAsTest │The difference is ±20%, and when it is less than 0.1 seconds, the indication is ┃ ┃ │ │value is within ±20% ┃ ┃ │ │. ┃ ┃ │ │(2) For a condenser X-ray device: 10mAs ┃ ┃ │ │, the error regarding the indicated value is ┃ ┃ │ │±2mAs, if 10mAs or more, indicate ┃ ┃ │ │The percentage mean error for the value is within ±20% ┃ ┃ │ │must be. ┃ ┃ │ 3) Investigation_Agreement_Time │(1) When SID is maximized, the minimum irradiation field is 5 cm ┃ ┃ │Test(VisualizationDevice │It must be adjustable to 5 cm or less. ┃ ┃ │only) │(2) The X-ray irradiation field is the effective ┃ ┃ │ │It shall not exceed the area. ┃ ┃ │ │(3) center of the receiving plane and center of the X-ray irradiation field ┃ ┃ │ │The error shall be within ±2% of the SID. ┃ ┃ │ 4) On the chest wall side │The X-ray irradiation area must include the entire surface of the image ┃ ┃ │X-ray irradiation │but, exceeding 2% of SID, towards the chest wall ┃ ┃ │test(mammography │The investigation period must not be extended. ┃ ┃ │MTQ97051fb4bce345a000242 Devices for START only │ ┃ ┃ │) │ ┃ ┃ │ 5) When adjusting compression bandage size │The chest wall length of the compression bandage is the breast support's ┃ ┃ │(for mammography │Length of the chest wall direction of the wound surface at a height of 4.2 cm ┃ ┃ │device only │. ┃ ┃ │) │ ┃ ┃ │ │ ┃ ┃ │ │ ┃ ┃ │ │ ┃ ┃ │ 6) Incident radiation dose │At the center point of the line of use incident on the subject ┃ ┃ │Rate Test(Transparency Market │The irradiation dose rate is when there is an automatic exposure controller ┃ ┃ │value only applies) │10R/min (2.58×10-3C/kg·min), automatic exposure ┃ ┃ │ │If there is no controller, 5R/min(1.29×10⁻³C/k ┃ ┃ │ │g·min) and ┃ ┃ │ │Does not operate in any combination of tube currents ┃ ┃ │ │must. ┃ ┃ │ 7) Dark X-ray │5cm away from surfaces where people may come into contact ┃ ┃ │Hum (Capacitor type Ex │and 2mR/hr(5.1 ┃ ┃ │Only the wire device │6×10-7C/kg·hr) shall not be exceeded ┃ ┃ │is treated) │. ┃ ┃ │ │ ┃ ┃ │ 8) Timer Test │ ┃ ┃ │(Image device only │Regarding the indicated value of the investigation time by timer adjustment ┃ ┃ │corresponding) │The allowable values ​​are less than or equal to the values ​​shown in the following table. ┃ ┃ │ │must. ┃ ┃ │ ├────┬───────┬────────┨ ┃ │ │RectificationMethod│Display value │Tolerance ┃ ┃ │ ├────┼───────┼────────┨ ┃ │ │Single-phase device│T<10pulses │±0 pulse(2) ┃ ┃ │ │ │10pulse≤T │±10% ┃ ┃ │ ├────┼───────┼────────┨ ┃ │ │Multiphase Rectification│T<0.01 seconds │-1.5msec~+6ms ┃ ┃ │ │Device │0.01 sec≤T<0.04│ec ┃ ┃ │ │(1), (3)│second │±20% ┃ ┃ │ │ │0.04 seconds≤T │±10% ┃ ┃ │ ├────┼───────┼────────┨ ┃ │ │Inverter Type│T<0.01 seconds │±1msec ┃ ┃ │ │device │0.01 sec≤T │±10% ┃ ┃ │ ├────┴───────┴────────┨ ┃ │ │(1) Depending on the input method of each device, at the time of input ┃ ┃ │ │MTQ97051fb4bce345a000296 Initial input section from START to regular operation ┃ ┃ │ │, the initial part ┃ ┃ │ │is not included in the shooting time. ┃ ┃ │ │(2) 1 pulse is 1/2 of 1 cycle of the power frequency ┃ ┃ │ │refers to the corresponding time. In single-phase devices, ┃ ┃ │ │, the effective pulse electrical angle is 45°–135° ┃ ┃ │ │. The first pulse or the first ┃ ┃ │ │, the range greater than that electrical angle ┃ ┃ │ │is also considered a regular pulse. ┃ ┃ │ │ Regarding the timing of injection, within -45° to 0° ┃ ┃ │ │Input section and within 0°–45° at termination ┃ ┃ │ │The subsequent part is not included in the shooting time ┃ ┃ │ │. ┃ ┃ │ │(3) In a multiphase rectifier, the rising part of the high-voltage waveform ┃ ┃ │ │and the respective 87% timing for the specified tube voltage value ┃ ┃ │ │Using the projection point and the cutoff point, the time between them ┃ ┃ │ 9) Average wired dose │Use the interval as the shooting time. ┃ ┃ │test(mammography │Compressed 4.2cm thick standard breast phantom(Standard breast ┃ ┃ │MTQ97051fb4bce345a000316 Devices for START only │Under illumination conditions when photographing a human model ┃ ┃ │) │A standard breast phantom captured using the superior-inferior imaging method ┃ ┃ │ │is 3 mGy (0.3 rad) or less ┃ ┃ │ 10) Phantom Video │should. ┃ ┃ │Evaluation Test(U │Pseudo-lesion in standard breast phantom images ┃ ┃ │For room filming │4 or more fibrous spots out of 16, 3 or more spots, ┃ ┃ │value only applies) │At least three masses must be observed. ┃ ┃ │ ├──────┬────────┬─────┨ ┃ │11) Half-value layer test │Classification │Measurement tube voltage(kV) │Minimum ┃ ┃ │ │ │ │Half-value layer(㎜ ┃ ┃ │ │ │ │Al) ┃ ┃ │ ├──────┼────────┼─────┨ ┃ │ │Diagnosis │60 │1.3 ┃ ┃ │ │X-ray device │(Maximum tube voltage is │ ┃ ┃ │ │ │70 kV or less device │ ┃ ┃ │ │Diagnosis │value) │ ┃ ┃ │ │X-ray ├────────┴┬────┨ ┃ │ │Generator │80 │2.3 ┃ ┃ │ │ │(Maximum tube voltage is 70 │ ┃ ┃ │ │ │Devices exceeding kV │ ┃ ┃ │ │ │value) │ ┃ ┃ │ ├──────┼─────────┼────┨ ┃ │ │For mammography │20 │0.20 ┃ ┃ │ │device ├─────────┼────┨ ┃ │ │ │25 │0.25 ┃ ┃ │ │ ├─────────┼────┨ ┃ │ │ │30 │0.30 ┃ ┃ │ ├──────┴─────────┴────┨ ┃ │ │The measured tube voltage is corrected for error after the tube voltage test. ┃ ┃ │ │test. ┃ ┠────────┼─────────┼─────────────────────┨ ┃2. For dental diagnosis │a. When external leakage current│Test item "B" for diagnostic radiation generators ┃ ┃X-ray generation │Hump │(External Leakage Current Test). ┃ ┃device(Panoramic ├─────────┼─────────────────────┨ ┃MawaSeparo │B. Re-evaluation of exposure dose │Test item "D" for diagnostic radiation generators ┃ ┃) │ManifestationTest │(Reproducibility test of radiation dose). ┃ ┃ ├─────────┼─────────────────────┨ ┃ │. Tube voltage test │Test Item "D" for Diagnostic Radiation Generators ┃ ┃ │ │(Tube voltage test). ┃ ┃ ├─────────┼─────────────────────┨ ┃ │. Tube current test │Test item "M" for diagnostic radiation generators ┃ ┃ │ │(Tube current test). ┃ ┃ ├─────────┼─────────────────────┨ ┃ │E. Investigation Test │1) The dental diagnostic X-ray generator is an irradiation line ┃ ┃ │ │The diameter of the irradiation field at the START stage must be 7cm or less ┃ ┃ │ │. ┃ ┃ │ │2) Dental diagnostic X-ray generator for panoramic imaging ┃ ┃ │ │The growth device is such that the center of the utilization line is in the slit on the film side. ┃ ┃ │ │It must be aligned with the core, and X-ray irradiation ┃ ┃ │ │, the difference in slit width is within ±2% ┃ ┃ │ │. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │Bar. Timer Test │The tolerance for the timer's indicated value is a percentage average ┃ ┃ │ │The standard error must be within ±10%. However, half ┃ ┃ │ │Irradiation time in a wave rectifier is less than 0.5 seconds, ┃ ┃ │ │In a single-phase full-wave rectifier, less than 0.2 seconds is ±2 pulses ┃ ┃ │ │. ┃ ┃ ├─────────┼──────────┬──────────┨ ┃ │Inc. Half-value layer test │Measurement tube voltage (kV) │Minimum half-value layer (mm Al) ┃ ┃ │ ├──────────┼──────────┨ ┃ │ │60 │1.5 ┃ ┃ │ │(Maximum tube voltage is 70 │ ┃ ┃ │ │Device with kV or less) │ ┃ ┃ │ ├──────────┼──────────┨ ┃ │ │80 │2.3 ┃ ┃ │ │(Maximum tube voltage is 70 │ ┃ ┃ │ │Device exceeding kV) │ ┃ ┃ │ ├──────────┴──────────┨ ┃ │ │The measured tube voltage is corrected for error after the tube voltage test. ┃ ┃ │ │test. ┃ ┠────────┼─────────┼─────────────────────┨ ┃3. Computed Tomography │a. Artifacts Test │ArtifactsImages should not appear noticeably ┃ ┃ImagingDevice │ │and visually inspect whether the centerline is properly aligned ┃ ┃ │ │. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │or. Slice thickness│Slice thickness error between set value and measured value ┃ ┃ │Hump │The difference must be within ±1mm. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │. Tube voltage test │The percentage mean error of the tube voltage is ±7% of the indicated value. ┃ ┃ │ │. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │d. Patient exposure dose │Complies with the manufacturer's self-standards. ┃ ┃ │Measurement Test │ ┃ ┃ ├─────────┼─────────────────────┨ ┃ │E. mAs Test │1) Transformer-type and inverter-type X-ray devices: Indicated values ​​ ┃ ┃ │ │must be within ±(10% + 0.2mAs) ┃ ┃ │ │. ┃ ┃ │ │2) Condenser X-ray device ┃ ┃ │ │ a) Less than 10mAs: The error for the indicated value is ┃ ┃ │ │Must be within ±2mAs. ┃ ┃ │ │ b) 10mAs or more: Percentage mean error relative to the indicated value┃ ┃ │ │The difference must be within ±20%. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │bar. CT Number │Complies with the manufacturer's self-standards. ┃ ┃ │Linearity test of START │ ┃ ┃ ├─────────┼─────────────────────┨ ┃ │Company. Noise Test │Complies with the manufacturer's self-standards. ┃ ┃ ├─────────┼─────────────────────┨ ┃ │Ah. Ancient Map Space │Complies with the manufacturer's self-standards. ┃ ┃ │Resolution Test │ ┃ ┃ ├─────────┼─────────────────────┨ ┃ │. Resolution Test │Must be no more than +20% of the manufacturer's self-standard. ┃ ┃ │(PET only) │ ┃ ┃ ├─────────┼─────────────────────┨ ┃ │. Sensitivity Test(PET │It must be at least 20% of the manufacturer's self-standard. ┃ ┃ │) │ ┃ ┠────────┼─────────┼─────────────────────┨ ┃4. Other │Food and Drug Safety Evaluation │Standards determined by the Director of the Korea Food and Drug Administration, No. ┃ ┃기 │determined by the Director │or ┃ ┃ │Test │. ┃ ┗━━━━━━━━┷━━━━━━━━━┷━━━━━━━━━━━━━━━━━━━━━┛ Remarks 1. Medical institutions pursuant to Article 3 of the Medical Service Act (excluding dental hospitals and dental clinics (Does), public health center pursuant to Articles 7, 8, and 10 of the "Regional Health Act" and Regarding devices for radiographic imaging, Item 1, Subparagraph (g) 10), Items 3, Subparagraphs (a), (b), and (g) Items (a) and (b) shall not apply. However, positron emission computed tomography () For imaging devices, the provisions of subparagraphs (a) through (c) of subparagraph 3 are applied. . 2. Where other laws prescribe inspection standards for Subparagraphs (j) and (j) of Item 3 If the criteria according to this table are applied to that item, may not be used.