Annex 8 — 동물용 의약품등 취급규칙
279531 · 2025-10-31
■ Rules on the Handling of Veterinary Medicinal Products, etc. [Appendix 8] <Amended Jan. 5, 2024>
Standards for the Classification and Designation of Veterinary Medical Devices and Veterinary In Vitro Diagnostic Medical Devices and
Procedure (Related to Article 5, Paragraph 6)
1. Classification Standards for Veterinary Medical Devices
a. The Director General of the Animal and Plant Quarantine Agency [regarding] the purpose of use and the effects on animals and handlers at the time of use of veterinary medical devices
classifies veterinary medical devices into the following four classes according to the degree of potential risk
class. In this case, if it falls under two or more classes, the highest risk
Classified by grade according to the degree.
1) Class 1: Veterinary medical device with minimal potential risk
2) Class 2: Veterinary medical device with low potential risk
3) Class 3: Veterinary medical devices with potential risk of severe symptoms
4) Class 4: Veterinary medical device with high risk
B. The criteria for determining the potential risk under Item A are as follows:
1) Duration of contact with animals
2) Degree of invasion
3) Whether veterinary medicines or energy, etc. are delivered to sick livestock or handlers
Part
. The detailed criteria regarding the risks of Items A and B are as follows.
Class 1: Not in direct contact with animals or handlers or still
There is almost no risk to materials, and due to failure or abnormality, animals or handlers...
values for veterinary medical devices with minor impact
2) Class 2: Risk to animals or handlers due to malfunction or abnormality during use
exists, but the likelihood of facing life-threatening danger or significant dysfunction is low, so potential
Veterinary medical device with low risk
3) Class 3: Used by insertion into animals for a certain period or has high potential risk
Veterinary Medical Device
Class 4: Direct contact with the heart, central nervous system, central circulatory system, etc., in animals [can lead to]
Veterinary medical devices used, utilizing animal tissues or extracts, or safety, etc.
Veterinary medical devices using raw materials with insufficient information for verification, high risk
Veterinary medical device with START properties
2. Classification Standards for In Vitro Diagnostic Medical Devices for Animals
a. The Director General of the Animal and Plant Quarantine Agency [regarding] in vitro diagnostic medical devices for animals, the purpose of use and the animal and handler at the time of use
, veterinary in vitro diagnostic medical devices are classified into the following 4
It is classified into the START grade. In this case, if it falls under two or more grades, the most
Classified by grade according to high risk.
1) Class 1: In vitro diagnostic medical device for animals with minimal potential risk
2) Class 2: In vitro diagnostic medical devices for animals with low potential risk
3) Class 3: In vitro diagnostic medical devices for animals with potential risk of moderate to severe harm
4) Class 4: Veterinary in vitro diagnostic medical device with high risk
나. 가목의 잠재적 위해성에 대한 판단기준은 다음과 같다.
1) Purpose of Use and Precautions for Use
2) Influence of Diagnostic Test Results on Animals or Handlers
3) Whether diagnostic information can be used independently or in combination with other diagnostic information
The importance of diagnostic information, such as whether it can be done
4) User Clinical Experience
3. Classification Procedure
a. The Director General of the Animal and Plant Quarantine Agency classifies veterinary medical devices by instruments/machinery, devices, and materials, and
eaching each classification group considering the similarity of raw materials, manufacturing processes, and quality control systems, functions
Classified in detail by item that exhibits independent action, and the grade for each item is determined
and announce it. In this case, if necessary, determined through deliberation by the Animal Pharmacist Deliberation Committee
.
a. The Director General of the Animal and Plant Quarantine Agency shall classify in vitro diagnostic medical devices for animals by diagnostic testing field, and each
classification groups considering the similarity of raw materials, manufacturing processes, and quality control systems functionally distinct
Classify in detail by item that is independently exerted, and determine the grade for each item.
is hereby notified. In this case, if necessary, it shall be determined after deliberation by the Animal Pharmacist Deliberation Committee.
4. Application for Reclassification and Designation Procedure
a. The Director of the Quarantine and Inspection Agency, upon application by interested parties, etc., or when it is recognized that reclassification is necessary
In cases where this occurs, the grade by item may be reclassified, and if necessary, the Animal Pharmacist Review Committee
Classify after deliberation.
b. When reclassifying, the degree of potential risk and validity based on the following criteria
must be reviewed.
1) Description by Item and Use of Relevant Veterinary Medical Devices and Veterinary In Vitro Diagnostic Medical Devices
Whether they are identical items with similar purpose, use, principle, characteristics, and function
2) Safety and performance are sufficient compared to items already classified, designated, and managed
Whether it is secured
. I would like to apply for reclassification of the grade must attach the following documents and the Quarantine Headquarters
Must be submitted to the head.
1) Materials regarding technical documents, etc.
2) Other animals similar to veterinary medical devices and veterinary in vitro diagnostic medical devices subject to reclassification
Structure, principles, performance, purpose of use, and use of medical devices for START and veterinary in vitro diagnostic medical devices
Data on the comparison and analysis of technical characteristics such as methods
(d) Pursuant to the provisions of subparagraph (c), the Director of the Quarantine and Inspection Agency who receives an application for reclassification of grade shall [apply for] it within 60 days
After reviewing and deciding within the START period, the result must be notified to the applicant.