Annex 6 — 동물용 의약품등 취급규칙
279531 · 2025-10-31
■ Rules on the Handling of Veterinary Medicinal Products, etc. [Appendix 6-2] <Amended Jan. 5, 2024>
Facility and quality management system of veterinary medical devices and veterinary in vitro diagnostic medical devices
Standards (Related to Article 13-2, Paragraph 2)
Ⅰ. Facility and Quality Management System of Manufacturers of Veterinary Medical Devices and Veterinary In Vitro Diagnostic Medical Devices
1. Facilities of the manufacturing plant
a. The manufacturer (excluding software veterinary medical devices) shall have the following facilities at the manufacturing site
and equipment must be equipped and maintained, and regularly inspected veterinary medical devices
· Ensure that there is no impediment to the manufacturing and quality control of in vitro diagnostic medical devices for animals.
1) Workshop performing manufacturing work
2) A testing laboratory for quality control of raw materials, materials, and products
3) Storage facility for storing raw materials, materials, and products
4) Facilities and equipment required for manufacturing and quality control
or. The manufacturer of software animal medical devices shall, at the manufacturing facility, software animal medical
Equipped with a place, facilities, and equipment capable of manufacturing or producing devices and maintaining them
must.
2. Facility Standards for Manufacturing Plants
The manufacturing facility must have work areas and storage areas that meet the following standards.
a. Workshop
1) Install facilities to prevent rats, pests, dust, etc.
2) The ceiling of a workplace manufacturing products requiring sterilization is at risk of falling dust
It must be finished to be free of dust or dirt, and the floors and walls must be smooth.
It must be made easy to remove, and the surfaces of the ceiling, floor, and walls must be disinfected with disinfectant solution.
It shall be designed to withstand spray washing of START.
3) Workbenches shall be placed in the workplace, and in the case of manufacturing products requiring sterilization
Install sterilization facilities
a. Storage facilities shall be designed to store raw materials, materials, and products hygienically and safely
Equipment to be installed
3. Standards for Good Manufacturing Practices and Quality Management Systems
veterinary medical device manufacturers veterinary medical devices and veterinary
To meet Good Manufacturing Practice (GMP) standards for the purpose of exporting in vitro diagnostic medical devices
In order to be used, the following manufacturing and quality control systems must be established and maintained.
.
A. Quality Management System
The manufacturer, for manufacturing and quality control, has an organization, responsibilities, procedures, processes and
Establish a quality management system to efficiently manage resources, etc., and quality management
It must be written as a document such as a manual.
B. Documentation and Implementation of the Quality Management System
and
The quality management system must be implemented accordingly.
1) Quality Policy
2) 계약검토
3) 설계관리
4) 문서 및 자료관리
5) 구매관리
6) 제품의 식별 및 추적관리
7) 제조공정의 관리
8) 구매품 및 제품의 시험검사
9) 부적합품의 관리
10) 측정장비의 관리
11) 시정 및 예방조치
12) 제품의 취급, 보관, 포장, 보존 및 인도에 관한 사항
13) Education and Training
14) Other matters necessary for manufacturing and quality control
. Responsibility and Authority
The manufacturer, with respect to members who may affect quality, has responsibility and authority,
Mutual relationships among members must be defined and documented.
d. Management of Quality Records
1) The manufacturer shall identify and obtain various records necessary for implementing manufacturing and quality control
Document the procedures for collection, indexing, viewing, filing, storage, maintenance, and disposal.
Must be written and managed.
All quality records must be immediately inspected at a facility capable of preventing damage, loss, or deterioration.
Must be stored so that the color is possible.
E. Quality Inspection, etc.
1) Internal Quality Inspection
The manufacturer, in order to maintain compliance with Good Manufacturing Practice and Quality Management System standards,
Document procedures for the planning and execution of sub-quality inspections and regularly
must be performed, and the results of the performance must be utilized for manufacturing and quality control.
2) External Quality Audit
The manufacturer shall, regarding compliance with Good Manufacturing Practice (GMP) standards, for 3 years
Must undergo a regular inspection by the Director of the Quarantine and Inspection Agency.
bar. Detailed criteria
In addition to the matters specified in the regulations from Items (a) to (e), Good Manufacturing Practice and Quality Control (GMP)
Details regarding standards, application and evaluation procedures, etc., shall be determined by the Director of the Quarantine and Inspection Agency.
is determined and announced.
4. Scope of Outsourcing for Manufacturing Process and Testing
Veterinary medical device manufacturers veterinary medical devices and animal
The manufacture or testing of in vitro diagnostic medical devices for other animal medical devices · in vitro diagnostic medical devices for animals
Manufacturers of diagnostic medical devices and the Minister of Agriculture, Food and Rural Affairs pursuant to Article 51-3 of these Rules
To testing and inspection institutions for veterinary medical devices and veterinary in vitro diagnostic medical devices designated by
In the case of outsourcing, if manufacturing facilities and equipment or facilities and equipment related to testing are not equipped...
may not be required, and the manufacture and testing of veterinary medical devices and veterinary in vitro diagnostic medical devices
Detailed matters regarding the scope of the trustee and compliance requirements, etc., pursuant to the entrustment of
The provisions shall be determined and announced by the Director of the Quarantine and Inspection Agency.
Ⅱ. Quality Inspection Facilities and Quality of Importers of Veterinary Medical Devices and Veterinary In Vitro Diagnostic Medical Devices
Management System
1. Facilities of the Importer
The importer shall equip and maintain the following facilities and equipment at the import business premises:
, and by regularly inspecting this,
Ensure that there are no impediments to import and quality control.
a. Business office conducting import business
B. Warehouse for storing products
. If testing for quality control is required, the test laboratory and the test required for the relevant test
facilities. However, in the case of outsourcing testing, the test room or test site related to the outsourced test
It may not be equipped with a START.
2. Facility Standards for Import Businesses
The facilities of the import business must comply with the following standards.
a. Provide facilities capable of maintaining the condition of the product upon import.
b. Provide low-temperature storage facilities or facilities for light shielding (if low-temperature storage of the product is necessary
or
)
. When handling items for which a storage method has been specified, those conditions can be maintained
shall install the facility
Ⅲ. Facility and Quality Management System for Veterinary Medical Device Repair Business
1. Facility Management
The repair business operator shall have the facilities and equipment necessary for the repair and testing of the relevant veterinary medical device.
must.
2. Repair Management Record
The repair contractor must prepare a repair management record for each item, and the repair management record sheet shall include
The following must be included.
A. Product Name and Model Number of Veterinary Medical Device
B. Name of manufacturer (importer) and country of manufacture of veterinary medical device
. Date of manufacture and date of repair
라. 점검확인결과서
마. 주요수리내용
B. Records regarding the management of repair facilities
3. Corrective Action
If a complaint arises regarding the quality of the repaired veterinary medical device, the repairer...
and
must be implemented.
4. Record
The following records must be written and maintained.
가. 수리관리업무에 관련된 기록
나. 시험검사장비를 포함한 수리시설에 관한 기록
. Record regarding corrective action
d. Other records regarding business processing in accordance with these standards