Annex 9 — 계량에 관한 법률 시행령

278291 · 2025-10-01

■ Enforcement Decree of the Act on Weights and Measures [Table 9] <Amended July 9, 2024> Standards for Testing and Inspection Facilities of Type Approval and Inspection Agencies, etc. (Related to Article 13, subparagraph 3; Article 23, subparagraph 2; and Article 24, subparagraph 2) ┏━━━━━━━┯━━━━━━━━┯━━━━━━━━━━━━━━━━━━━━┯━━┯━━┯━━┓ ┃of the meter's │EquipmentName │EquipmentSpecification │Type│Reexam│Test┃ ┃Type │ │ │Approved│jeong │Institution┃ ┃ │ │ │Institution│Execute│ㆍJa┃ ┃ │ │ │ │Fat│Check┃ ┃ │ │ │ │Autonomy│official┃ ┃ │ │ │ │Group│Producer┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃1. Non-automatic │A. Laboratory │1) Distribute the test room evenly into at least 4 locations │○ │○ │○ ┃ ┃scale │ │average of ambient temperatures measured by │ │ │ ┃ ┃ │ │Change amount is 4 oC per 12 hours, 1 hour │ │ │ ┃ ┃ │ │Not exceeding 2 oC per START │ │ │ ┃ ┃ │ │2) Relative humidity within (30 ~ 70)% │ │ │ ┃ ┃ │ │and the maximum humidity per hour │ │ │ ┃ ┃ │ │Variation within ± 5% relative humidity │ │ │ ┃ ┃ │ │'s thing │ │ │ ┃ ┃ │ │3) Regional gravitational acceleration within the test room │ │ │ ┃ ┃ │ │value │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Slope │2/1000 or more tilt possible │○ │ │ ┃ ┃ │ Tester │thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Constant temperature and humidity unit │1) Temperature: (-10 ~ 40)oC ± 2 oC │○ │ │ ┃ ┃ │ │Within │ │ │ ┃ ┃ │ │Relative Humidity: (40 ~ 85)% ± 5 │ │ │ ┃ ┃ │ │% within │ │ │ ┃ ┃ │ │3) Constant temperature and humidity area: 70 cm × 70 cm │ │ │ ┃ ┃ │ │× 50 cm or larger │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Weights │1 mg up to 20 kg, 1-2-2-5 │○ │○ │○ ┃ ┃ │ │Combined with 1-1-2-5 series │ │ │ ┃ ┃ │ │possible │ │ │ ┃ ┃ │ ├────────────────────┼──┼──┼──┨ ┃ │ │The corresponding combination weight among each of the following items │○ │○ │○ ┃ ┃ │ │Secure START │ │ │ ┃ ┃ │ │1) Visa for business scope of 50 t or less │ │ │ ┃ ┃ │ │For copper scales: 2 min of maximum capacity │ │ │ ┃ ┃ │ │and │ │ │ ┃ ┃ │ │2) Visa with a scope of work exceeding 50 t │ │ │ ┃ ┃ │ │For copper balances: for groups of 25 t or more │ │ │ ┃ ┃ │ │joint weight │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. Vehicles for Inspection │Vehicle capable of transporting inspection weights │○ │ │○ ┃ ┃ │ │Secure START │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Bar. Voltage Regulator │Minimum eye value: 5 V or less │○ │ │ ┃ ┃ │ │Adjustment range: (-15 ~ │ │ │ ┃ ┃ │ │10)% variation possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Company. Frequency │Frequency variation: ±3% of the operating frequency │○ │ │ ┃ ┃ │ TestingMachine │Changes are possible until START. │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Ah. Repetitive Loading │(0 ~ 50) kg adjustable, and │○ │ │ ┃ ┃ │ TestingMachine │Continuous repetitive loading of over 100,000 cycles │ │ │ ┃ ┃ │ │testing will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Instantaneous voltage │Descent(100% and 50%), Period(1st Company │○ │ │ ┃ ┃ │Drop tester │cycle and 2 cycles), │ │ │ ┃ ┃ │ │Measurement will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Car. Burst │The measurement range is up to (0.5 ~ 3)kV │○ │ │ ┃ ┃ │ TestingMachine │It will be possible to settle │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Car. Electrostatic │Contact discharge 8 kV, Standby discharge 15 kV │○ │ │ ┃ ┃ │ Discharge Tester │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. High frequency │Frequency range (26 ~ 1,000)MHz, │○ │ │ ┃ ┃ │Self-testing machine │Electric field strength 3 V/m, adjustable 80 % │ │ │ ┃ ┃ │ │AM Measurement at 1 kHz will be possible │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃2. Weight │A. Laboratory │1) Distribute the test room evenly into at least 4 locations │○ │○ │○ ┃ ┃ │ │average of ambient temperatures measured by │ │ │ ┃ ┃ │ │Change amount is 4 oC per 12 hours, 1 hour │ │ │ ┃ ┃ │ │Not exceeding 2 oC per START │ │ │ ┃ ┃ │ │2) Relative humidity within (30 ~ 70)% │ │ │ ┃ ┃ │ │and maximum humidity per hour │ │ │ ┃ ┃ │ │The fluctuation range is ±5 % relative humidity this │ │ │ ┃ ┃ │ │Things inside │ │ │ ┃ ┃ │ │3) Regional gravitational acceleration within the test room │ │ │ ┃ ┃ │ │value │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Solid Density │A balance and a device for measuring solid density │○ │ │ ┃ ┃ │ Measuring Instrument │will be │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Magnetic susceptibility │Magnetization according to the type approval standards for weights │○ │ │ ┃ ┃ │ Measuring Instrument │Rate measurement will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Surface roughness │Surface according to the type approval standards for weights │○ │ │ ┃ ┃ │ Measuring Instrument │Measurement will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. Scale │1 Error measurement up to 000 kg weight │○ │ │ ┃ ┃ │ │Scales capable of START │ │ │ ┃ ┃ │ │1) Maximum capacity 1500 kg or more, Minimum │ │ │ ┃ ┃ │ │Value of 50 g or less │ │ │ ┃ ┃ │ │2) Maximum capacity 600 kg or more, minimum eye │ │ │ ┃ ┃ │ │value 10 g or less │ │ │ ┃ ┃ │ │3) Maximum capacity 150 kg or more, minimum eye │ │ │ ┃ ┃ │ │1 g or less │ │ │ ┃ ┃ │ │4) Maximum capacity 30 kg or more, minimum eye's │ │ │ ┃ ┃ │ │value 5 mg or less │ │ │ ┃ ┃ │ │5) Maximum capacity 10 kg or more, minimum eye │ │ │ ┃ ┃ │ │1 mg or less │ │ │ ┃ ┃ │ │6) Maximum capacity 3 kg or more, minimum eye's │ │ │ ┃ ┃ │ │value 0.1 mg or less │ │ │ ┃ ┃ │ │7) Maximum capacity 200 g or more, minimum eye │ │ │ ┃ ┃ │ │value 0.01 mg or less │ │ │ ┃ ┃ │ │8) Maximum dose 2 g or more, minimum eye's │ │ │ ┃ ┃ │ │value 0.000 1 mg or less │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │bar. weight │1 mg ~ up to 20 kg 1-2-2-5 Series │○ │○ │○ ┃ ┃ │ │or combination with 1-1-2-5 series │ │ │ ┃ ┃ │ │Secure available weights │ │ │ ┃ ┃ │ ├────────────────────┼──┼──┼──┨ ┃ │ │With combination weights, a total weight of 2 t or more │○ │○ │○ ┃ ┃ │ │Secure │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃3. Gas Meter │A. Laboratory │1) Distribute the test room evenly into at least 4 locations │○ │○ │○ ┃ ┃ │ │average of ambient temperatures measured by │ │ │ ┃ ┃ │ │Change amount is 4 oC per 12 hours, 1 hour │ │ │ ┃ ┃ │ │Not exceeding 2 oC per START │ │ │ ┃ ┃ │ │2) Relative humidity within (30 ~ 70)% │ │ │ ┃ ┃ │ │maintenance is possible, and the maximum │ │ │ ┃ ┃ │ │Humidity fluctuation range ±5 % Relative humidity │ │ │ ┃ ┃ │ │Within │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Error Inspection │1) Up to the maximum flow rate of the gas meter to be inspected │○ │○ │○ ┃ ┃ │device │testing is possible, and between meters │ │ │ ┃ ┃ │ │No adverse effects │ │ │ ┃ ┃ │ │ - Moving each position of the meter │ │ │ ┃ ┃ │ │When tested, the deviation rate was 0.5% │ │ │ ┃ ┃ │ │The following │ │ │ ┃ ┃ │ │2) The inspection stand measures the temperature per meter to be tested │ │ │ ┃ ┃ │ │and capable of compensating for pressure loss │ │ │ ┃ ┃ │ │3) The test air is at the ambient temperature of the test room │ │ │ ┃ ┃ │ │Inflow within 0.5 °C, and condensate │ │ │ ┃ ┃ │ │will not occur │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Leakage tester │1) Internal and external leakage of the test apparatus │○ │ │ ┃ ┃ │ │Be smaller than the larger of the following values ​​ │ │ │ ┃ ┃ │ │ a) │ │ │ ┃ ┃ │ │0.1 % of minimum flow rate │ │ │ ┃ ┃ │ │ b) 100 cm3/h │ │ │ ┃ ┃ │ │2) Twice the maximum pressure of the test gas meter │ │ │ ┃ ┃ │ │is possible, and the test meter │ │ │ ┃ ┃ │ │It will be possible to check for leakage in START. │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Pressure loss │1) Pressure at the inlet and outlet of the gas meter │○ │ │ ┃ ┃ │TestingMachine │Capable of measuring force difference │ │ │ ┃ ┃ │ │ - The pressure measuring port is at the meter inlet │ │ │ ┃ ┃ │ │a distance equal to the pipe diameter and │ │ │ ┃ ┃ │ │at the meter outlet, by the pipe diameter │ │ │ ┃ ┃ │ │will be located away │ │ │ ┃ ┃ │ │ - The pressure measuring port is in the upstream direction and down │ │ │ ┃ ┃ │ │At least pipe diameter in the direction of flow │ │ │ ┃ ┃ │ │(nominal bore) │ │ │ ┃ ┃ │ │There will be a lot of intuition │ │ │ ┃ ┃ │ │2) The pressure measuring port is attached to the pipe shaft │ │ │ ┃ ┃ │ │It must be straight, and the diameter is 3 mm │ │ │ ┃ ┃ │ │and the pressure outlet is │ │ │ ┃ ┃ │ │Must not protrude into the pipe │ │ │ ┃ ┃ │ │Things to do │ │ │ ┃ ┃ │ │3) The average pressure loss tester of the meter │ │ │ ┃ ┃ │ │The changing pressure of the meter as the average value │ │ │ ┃ ┃ │ │and 9.8 Pa │ │ │ ┃ ┃ │ │It shall be possible to measure pressures below │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. Thermometer │at the outlet of the test meter, the │○ │○ │○ ┃ ┃ │ │Temperature measurement down to 0.1 °C or less │ │ │ ┃ ┃ │ │able thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │bar. For error inspection │The minimum flow rate (Qmin) of the test meter and │○ │○ │○ ┃ ┃ │meter │Including the maximum flow rate (Qmax), respectively END │ │ │ ┃ ┃ │ │Error test at 7 different flow rates │ │ │ ┃ ┃ │ │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Company. Constant Temperature and Humidity Chamber │1) Temperature: (-25 ~ 60)oC ± 2 oC │○ │ │ ┃ ┃ │ │It must be within │ │ │ ┃ ┃ │ │2) Relative humidity: (30 ~ 96)% ± 5 │ │ │ ┃ ┃ │ │be within % │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Ah. Temperature and Humidity │Recording of changes in temperature and humidity inside the laboratory │○ │○ │○ ┃ ┃ │ Recorder │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Durability │Internal │○ │ │ ┃ ┃ │ Tester │Configuration testing shall be possible │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃4. Water Meter │a. Error Inspection │1) Volume of liquid passing through the test meter │○ │○ │○ ┃ ┃5. Hot Water Meter │device │will be able to measure │ │ │ ┃ ┃ │ │2) │ │ │ ┃ ┃ │ │An air separator must be installed. │ │ │ ┃ ┃ │ │3) Temperature of the test fluid during the test │ │ │ ┃ ┃ │ │The change will also be within 5 °C. │ │ │ ┃ ┃ │ │4) The rate of change in flow rate of the test fluid is │ │ │ ┃ ┃ │ │Requirements for Meter Type Approval Criteria │ │ │ ┃ ┃ │ │을 만족할 것 │ │ │ ┃ ┃ │ │ ) (Q1 ≤ ±5% ≤ Q2) │ │ │ ┃ ┃ │ │ b) (Q2 <±10% ≤ Q4) │ │ │ ┃ ┃ │ │5) Water Meter (Hot Water Meter) Type Approval │ │ │ ┃ ┃ │ │Error measurement based on the START standard will be possible │ │ │ ┃ ┃ │ │6) Air inside water meter (hot water meter) │ │ │ ┃ ┃ │ │(Air Hole) removal function must be present │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Pressure Loss │1) Pressure measurement points at inlet and outlet, straight pipe │○ │ │ ┃ ┃ │Tester │Part of water meter (hot water meter) type approval │ │ │ ┃ ┃ │ │shall comply with the standards │ │ │ ┃ ┃ │ │2) The average pressure loss tester of the meter │ │ │ ┃ ┃ │ │The average value of the changing pressure of the START meter │ │ │ ┃ ┃ │ │and │ │ │ ┃ ┃ │ │0.01 MPa or less │ │ │ ┃ ┃ │ │It will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Pressure tester │Maximum pressure on the water meter (hot water meter) │○ │ │ ┃ ┃ │ │Pressurization of 1.96 MPa or higher is possible, and the old │ │ │ ┃ ┃ │ │Be capable of testing up to a diameter of 350 mm │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Durability │1) Water Meter (Hot Water Meter) Type Approval │○ │ │ ┃ ┃ │ Tester │standard and │ │ │ ┃ ┃ │ │It will be possible to set the test cycle │ │ │ ┃ ┃ │ │2) Metric nominal bore 100 mm or less │ │ │ ┃ ┃ │ │Durability testing shall be possible │ │ │ ┃ ┃ │ │3) Water hammer caused by flow rate fluctuations │ │ │ ┃ ┃ │ │Will be able to digest │ │ │ ┃ ┃ │ │4) During the test, the circulating liquid temperature is kept constant │ │ │ ┃ ┃ │ │It shall be possible to maintain below the temperature │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃6. Oil Meter │a. Volume Measurement │1) Tank: 1,000 L or larger │○ │○ │○ ┃ ┃ │device │2) Scales: 1,000 kg or more │ │ │ ┃ ┃ │ (tank, scale, │ - The density of the test fluid is 0.001 │ │ │ ┃ ┃ │Flow meter │Can measure up to │ │ │ ┃ ┃ │One branch) │3) Flow meter: To be tested up to 100 m3/h │ │ │ ┃ ┃ │ │can be │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Error Inspection │1) Those with a flow rate fluctuation rate of 5% or less │○ │○ │○ ┃ ┃ │ device │2) Temperature of test fluid and tank temperature │ │ │ ┃ ┃ │ │The degree can be measured at 0.1 °C or less │ │ │ ┃ ┃ │ │able │ │ │ ┃ ┃ │ │3) According to the oil meter type approval standards │ │ │ ┃ ┃ │ │error measurement will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Durability │1) According to the oil meter type certification standards, │○ │ │ ┃ ┃ │ TestingMachine │under the condition of finding │ │ │ ┃ ┃ │ │Durability testing shall be possible │ │ │ ┃ ┃ │ │2) Water hammer caused by flow rate fluctuations │ │ │ ┃ ┃ │ │can minimize │ │ │ ┃ ┃ │ │3) During the test, the temperature of the test fluid is kept constant │ │ │ ┃ ┃ │ │It shall be possible to maintain below the temperature │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Pressure tester │Pressurization of 1.96 MPa or higher is possible 1) │○ │ │ ┃ ┃ │ │and maintaining pressurization for more than 5 minutes │ │ │ ┃ ┃ │ │able thing │ │ │ ┃ ┃ │ │2) Verification of pressurization and leakage is │ │ │ ┃ ┃ │ │It will be possible │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃7. Fuel dispenser │a. Error Inspection │1)Inspection Tank │○ │○ │○ ┃ ┃8. AdBlue Meter │device(scale │ a) Maximum capacity is 10 L or 20 L │ │ │ ┃ ┃ │or for inspection │, the minimum eye value is 10 │ │ │ ┃ ┃ │Tank) │mL or less │ │ │ ┃ ┃ │ │ b) Those with a maximum capacity of 100 L or more │ │ │ ┃ ┃ │ │, the minimum eye value is 50 mL │ │ │ ┃ ┃ │ │The following │ │ │ ┃ ┃ │ │2) Scale │ │ │ ┃ ┃ │ │ a) With a maximum capacity of 10 kg or more │ │ │ ┃ ┃ │ │Those with a minimum eye value of 1 g or less │ │ │ ┃ ┃ │ │ b) Those with a maximum capacity of 100 kg or more │ │ │ ┃ ┃ │ │, the minimum eye value is 5 g or less │ │ │ ┃ ┃ │ │'s thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Density-based liquid system │Graduation range is 0.500 g/cm3 ~ 1.000 │○ │ │ ┃ ┃ │(to the fuel dispenser │up to g/cm3 minimum eye value │ │ │ ┃ ┃ │Limited) │that is 0.001 g/cm3 or less │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Gas remover │Gas-based │○ │ │ ┃ ┃ │Tester(Fueling │Be suitable for testing of the device │ │ │ ┃ ┃ │Limited to │1) 20% or more of │ │ │ ┃ ┃ │) │air injection is possible and injection air │ │ │ ┃ ┃ │ │The amount can be adjusted │ │ │ ┃ ┃ │ │2) Can measure the amount of injected air │ │ │ ┃ ┃ │ │has a gas meter │ │ │ ┃ ┃ │ │3) Injected air test liquid storage tank │ │ │ ┃ ┃ │ │, │ │ │ ┃ ┃ │ │There shall be a stopper (valve) │ │ │ ┃ ┃ │ │4) Volumetric Tank: At maximum flow rate 1 │ │ │ ┃ ┃ │ │The measurement of the volume of lubrication supplied per minute is │ │ │ ┃ ┃ │ │able thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Hose Volume │gas products specified in the type approval standards │○ │ │ ┃ ┃ │Change Tester │Be suitable for testing of the device │ │ │ ┃ ┃ │ │ - Fuel supply system, pressurization system, pressure │ │ │ ┃ ┃ │ │System, volume change reading device, etc. │ │ │ ┃ ┃ │ │there will be │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. Thermometer │The scale range is (0 ~ 50)°C or higher and │○ │○ │○ ┃ ┃ │ │Those with a minimum eye value of 0.1 °C or less │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Test Fluid │Temperature rise effect due to durability test │○ │ │ ┃ ┃ │Storage Tank │A cooling device for minimizing START is installed │ │ │ ┃ ┃ │ │installed, or a tank of sufficient size │ │ │ ┃ ┃ │ │Hold │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃9. LPG Meter │is. That Ul │With a maximum capacity of 30 kg or more, the │○ │○ │○ ┃ ┃ │ │Those with a small scale value of 10 g or less │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Density-based liquid system │scale range is 0.500 ~ 1.000 g/cm3 │○ │○ │○ ┃ ┃ │ │and the minimum eye value is 0.001 g/cm3 │ │ │ ┃ ┃ │ │The following │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Density measurement │LPG are │○ │○ │○ ┃ ┃ │MTQ10f065e7f790ccea02360 Cylinder for START │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Thermometer │The measurement range is 0 to 30 °C or higher, and │○ │○ │○ ┃ ┃ │ │The minimum eye value is 0.1 °C or less │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃10. Nunsaegim │a. Inspection Tank │The capacity is 10 L or 20 L or less, and │○ │○ │○ ┃ ┃Tank │Q │with a minimum eye value of 10 mL or less │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Impact Test │Air with a pressure of 35 kPa or higher is put into the tank │○ │ │ ┃ ┃ │START │Press down and free from a height of 2 m or more │ │ │ ┃ ┃ │ │Will be able to drop │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Pressure test │1) Pressurize to 0.196 MPa or higher │○ │○ │○ ┃ ┃ │START │It can maintain a pressurized state for more than 5 minutes. │ │ │ ┃ ┃ │ │Will be able to maintain │ │ │ ┃ ┃ │ │2) Check for pressurization and oil leakage │ │ │ ┃ ┃ │ │Will be able to │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Constant temperature unit │Temperature (-20 to 70)°C │○ │ │ ┃ ┃ │ │Maintaining a temperature of ±2 °C within the range is possible │ │ │ ┃ ┃ │ │able │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. Vibration Test │and the amplitude is │○ │ │ ┃ ┃ │START │13.7 Hz or higher │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃11. Accumulated │a. Error Inspection │Volume of water passing through the meter (2 0001) │○ │○ │○ ┃ ┃ Calorimeter │device │L or more, or │ │ │ ┃ ┃ │ │(flow meter of 300 m3/h or more) or │ │ │ ┃ ┃ │ │Volume by mass (2 t or more) │ │ │ ┃ ┃ │ │can be measured │ │ │ ┃ ┃ │ │2) The inspection device includes an air separator, temperature │ │ │ ┃ ┃ │ │system must be attached │ │ │ ┃ ┃ │ │3) Change in water temperature during the test │ │ │ ┃ ┃ │ │within 5 oC │ │ │ ┃ ┃ │ │4) The rate of change in flow rate of the test fluid is │ │ │ ┃ ┃ │ │Requirements for Meter Type Approval Criteria │ │ │ ┃ ┃ │ │을 만족할 것 │ │ │ ┃ ┃ │ │ ga) (qi ≤ ± 5 % < qp) │ │ │ ┃ ┃ │ │ Na) (qp ≤ ± 10 % ≤ qs) │ │ │ ┃ ┃ │ │5) Required by the type approval standards │ │ │ ┃ ┃ │ │Error measurement (flow rate) shall be possible │ │ │ ┃ ┃ │ │6) Air (Air Hole) inside the flow section │ │ │ ┃ ┃ │ │There will be that device │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Flow Signal │Generates a flow rate signal identical to the flow rate section │○ │○ │○ ┃ ┃ │ Generator │, and the │ │ │ ┃ ┃ │ │Must be equipped with a counter function │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Constant temperature bath │Temperature range is (0 ~ 200)oC or higher, temperature │○ │○ │○ ┃ ┃ │ │Those with accommodative ability of 0.05 °C or less │ │ │ ┃ ┃ │ │Possess 3 or more units via START │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Pressure Tester │Pressurization of 1.96 MPa or higher is possible 1) │○ │ │ ┃ ┃ │ │and maintaining pressurization for more than 5 minutes │ │ │ ┃ ┃ │ │able thing │ │ │ ┃ ┃ │ │2) Verification of pressurization and leakage is │ │ │ ┃ ┃ │ │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. Thermometer │The measurement range is (0 ~ 200)°C or higher, the │○ │○ │○ ┃ ┃ │ │Value of cow eye 0.01 °C or less │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Bar. Durability │1) Required by the type approval standards │○ │ │ ┃ ┃ │ TestingMachine │Flow rate and test cycle can be set │ │ │ ┃ ┃ │ │Things to do │ │ │ ┃ ┃ │ │2) Temperature of the test fluid: 95 °C ± 5 │ │ │ ┃ ┃ │ │oC │ │ │ ┃ ┃ │ │3) Water hammer caused by flow rate fluctuations │ │ │ ┃ ┃ │ │can minimize │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Inc. Magnetic flux │(0 ~ 500)mT or more │○ │ │ ┃ ┃ │Density Tester │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │A. Constant Temperature and Humidity Unit │1) Temperature: (-20 ~ 60)oC ± 2 oC │○ │ │ ┃ ┃ │ │It must be within │ │ │ ┃ ┃ │ │2) Relative humidity: (30 ~ 96)% ± 5 │ │ │ ┃ ┃ │ │be within % │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Pressure loss │1) Pressure measurement points at inlet and outlet, straight pipe │○ │ │ ┃ ┃ │TestingMachine │The part shall conform to the type approval standards. │ │ │ ┃ ┃ │ │2) Average of the change in meter pressure loss │ │ │ ┃ ┃ │ │value, and │ │ │ ┃ ┃ │ │0.01 MPa or less │ │ │ ┃ ┃ │ │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Voltage Regulator │The minimum eye value is 5 V or less and use │○ │ │ ┃ ┃ │ │Voltage fluctuation of up to ±20% of the START voltage is allowed │ │ │ ┃ ┃ │ │able thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Ka. FrequencySi │Frequency variation up to ±3% of the operating frequency │○ │ │ ┃ ┃ │experience │It will be possible to move │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │T. Instantaneous voltage │Descent(100% and 50%), Period(1st Company │○ │ │ ┃ ┃ │Drop tester │cycle and 2 cycles), │ │ │ ┃ ┃ │ │Measurement will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Part. Burst │The measurement range is up to (0.5 ~ 3)kV │○ │ │ ┃ ┃ │experience │It will be possible to │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │H. Electrostatic Discharge │Contact discharge 8 kV, Standby discharge 15 kV │○ │ │ ┃ ┃ │TestingMachine │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. High frequency │Frequency range (26 ~ 1 000)MHz, │○ │ │ ┃ ┃ │Self-testing machine │Electric field strength 3 V/m, adjustable 80 % AM │ │ │ ┃ ┃ │ │Measurement of 1 kHz will be possible │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃12. Power │A. Laboratory │Temperature 23 °C ± 2 °C, Relative Humidity (40 │○ │○ │○ ┃ ┃Total │ │~ 70) % ± 10 % │ │ │ ┃ ┃ │ │thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Reference Power │1) AC Electricity Meter: 3-phase (three phase) │○ │○ │○ ┃ ┃ │meter │Accuracy of 0.05% at a power factor of 1.0 with │ │ │ ┃ ┃ │ │within, AC voltage (0 ~ 480) │ │ │ ┃ ┃ │ │V, AC current (0 ~ 150) A, Phase │ │ │ ┃ ┃ │ │(0 ~ 360)°, Frequency (45 ~ 65) │ │ │ ┃ ┃ │ │MTQ10f065e7f790ccea02495 STARTHz measurable │ │ │ ┃ ┃ │ │2) DC power meter: Accuracy is 0.05 % │ │ │ ┃ ┃ │ │within, DC voltage (0 ~ 500) V, │ │ │ ┃ ┃ │ │Measure DC current (0 ~ 160) A │ │ │ ┃ ┃ │ │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Heo Bu-ha's foot │1) AC Power Meter: AC Voltage (0 ~ │○ │○ │○ ┃ ┃ │Production( Test │480) V, AC current (0 ~ 150) A, │ │ │ ┃ ┃ │Power supply │Phase (0 ~ 360)°, Frequency (45 ~ │ │ │ ┃ ┃ │_Class_Device) │65) Hz power single phase │ │ │ ┃ ┃ │ │or three-phase generation │ │ │ ┃ ┃ │ │able to, single phase │ │ │ ┃ ┃ │ │Will include a multi-channel precision transformer │ │ │ ┃ ┃ │ │2) DC Power Meter: DC Voltage (0 ~ │ │ │ ┃ ┃ │ │500) V, direct current (0 ~ 160) A │ │ │ ┃ ┃ │ │will be able to generate power │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Harmonic generator │1) Phase 3rd and 5th harmonic AC currents │○ │ │ ┃ ┃ │animation │Rated at 0° angle (operating conditions of the device │ │ │ ┃ ┃ │ │refers to the range of performance and performance. Below │ │ │ ┃ ┃ │ │(same as), voltage │ │ │ ┃ ┃ │ │at phase angle 0° of the rated (0 ~ │ │ │ ┃ ┃ │ │10) % │ │ │ ┃ ┃ │ │will be │ │ │ ┃ ┃ │ │2) AC current phase angles 90° and 270° │ │ │ ┃ ┃ │ │에서 발생되어 홀수고조파 전류 │ │ │ ┃ ┃ │ │testing shall be possible, during every 4 cycle │ │ │ ┃ ┃ │ │At current phase angle 0° every 2 cycles │ │ │ ┃ ┃ │ │all generated harmonics [less than 1 │ │ │ ┃ ┃ │ │Non-integer multiples of harmonic orders (Non-integer multiples of harmonic orders) │ │ │ ┃ ┃ │ │倍) Harmonic component] testing is possible │ │ │ ┃ ┃ │ │thing │ │ │ ┃ ┃ │ │3) Half-wave rectification[positive (+) of AC voltage or │ │ │ ┃ ┃ │ │Rectification using only one side of the negative terminal │ │ │ ┃ ┃ │ │method] By applying the generated current, DC and │ │ │ ┃ ┃ │ │Even harmonic testing will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. External Magnetic │1) Continuous magnetic induction │○ │ │ ┃ ┃ │Impact Tester │Test: Magnetomotive Force │ │ │ ┃ ┃ │ │will be 1000 AT │ │ │ ┃ ┃ │ │2) Power supply frequency magnetic field induction(magnetic │ │ │ ┃ ┃ │ │induction) Test: Magnetization Coil(magnetic │ │ │ ┃ ┃ │ │coil) Diameter 1 m, magnetomotive force 400 │ │ │ ┃ ┃ │ │AT will be │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. In case of overcurrent │Minimum current 5 kA or more and interval is │○ │ │ ┃ ┃ │Experience │1/2 cycle of the reference frequency or 0.5 s │ │ │ ┃ ┃ │ │Measurement will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Company. Power Analysis │1) Power Factor (0 ~ 1), Single-phase 2-wire │○ │ │ ┃ ┃ │START │(1P2W) ~ 3-phase 4-wire (3P4W) side │ │ │ ┃ ┃ │ │It will be possible to settle │ │ │ ┃ ┃ │ │2) Accuracy shall be within 0.1% │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Ah. Shockwave internal │Applied voltage 6 kV or higher, interval (10 ~ │○ │ │ ┃ ┃ │Voltage Tester │600) s possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Commercial Frequency │AC testing of at least 4 kV is │○ │ │ ┃ ┃ │Withstand voltage at │able thing │ │ │ ┃ ┃ │experience │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Insulation resistance │Apply 500 V DC to 100 MΩ │○ │ │ ┃ ┃ │series(insulation low │Measurement of the above will be possible │ │ │ ┃ ┃ │(Measuring Instrument) │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │K. Constant Temperature & Humidity │1) Temperature: (-40 ~ 80) °C ± 2 °C │○ │ │ ┃ ┃ │START │It will be within │ │ │ ┃ ┃ │ │Relative Humidity: (40 – 85) % ± 5 % │ │ │ ┃ ┃ │ │It must be within │ │ │ ┃ ┃ │ │3) Power │ │ │ ┃ ┃ │ │Be able to test the meter │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Acceleration Life │1) Temperature: (0 ~ 85) °C ± 2 °C This │○ │ │ ┃ ┃ │Test solution │Tomorrow's Thing │ │ │ ┃ ┃ │Temperature and Humidity Unit │2) Relative Humidity: (40 ~ 95) % ± 5 % │ │ │ ┃ ┃ │ │It must be within │ │ │ ┃ ┃ │ │3) Simultaneous testing of 30 electricity meters │ │ │ ┃ ┃ │ │able thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Wave. Salt Spray │1) The temperature of the salt spray test tank is (30 │○ │ │ ┃ ┃ │Tester │~ 40) °C possible │ │ │ ┃ ┃ │ │2) Salt solution concentration: Sodium chloride (NaCl) │ │ │ ┃ ┃ │ │, and in a dry state, impurities │ │ │ ┃ ┃ │ │is less than 0.3% by mass of the total, │ │ │ ┃ ┃ │ │Sodium iodide is 0.1% by mass. │ │ │ ┃ ┃ │ │미만인 것으로 사용할 것 │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │H. Dustproof Test │Dust Intrusion Protection Class 1 (IP1X) and up 5 │○ │ │ ┃ ┃ │START │Testable up to IP5X │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Waterproof test │IPX1 through IPX8 testable │○ │ │ ┃ ┃ │기 │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Vibration Test │Frequency range (10 ~ 150) Hz and │○ │ │ ┃ ┃ │Qa │Sweep cycles per axis (specified frequency │ │ │ ┃ ┃ │ │One round trip of the range) is 10(10 │ │ │ ┃ ┃ │ │Sweep cycle 75 min) will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │The. Spring │Moment of force (0.22 ± 0.05)Nㆍm │○ │ │ ┃ ┃ │Tester │Measurable │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Impact Test │300 m/s2 and 500 m/s2 tests possible │○ │ │ ┃ ┃ │START │Things to do │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Mer. Thermometer │(-10 ~ 100) oC The temperature in the range is 0.1 oC │○ │ │ ┃ ┃ │ │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Temperature Record │With a minimum of 4 channels, temperature measurement │○ │ │ ┃ ┃ │Total │The range shall be (-50 to +150) °C │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Calipers │with a minimum scale of 0.01 mm or less │○ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │uh. Glo Wire │1) Temperature measurement range: (0 ~ 1000) oC │○ │ │ ┃ ┃ │(glow wire) │(Adjustable) │ │ │ ┃ ┃ │Tester │2) Measurement time: (0 ~ 60) seconds (arbitrary group │ │ │ ┃ ┃ │ │clause possible) │ │ │ ┃ ┃ │ │3) Measurement Load: (1 ± 0.2) N │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Low. Sound Level Meter │Noise measurement is possible at a minimum of 30 dB or higher │○ │ │ ┃ ┃ │ │Things to do │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Lightfastness │Solar radiation is 8.96 1) per day cycle │○ │ │ ┃ ┃ │experience │kW/m2 │ │ │ ┃ ┃ │ │2) The internal temperature of the test chamber is at a constant temperature │ │ │ ┃ ┃ │ │, the cooling method is │ │ │ ┃ ┃ │ │Must be designed │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Care. Moisture·A │Approximately 20 mg/L of sulfur dioxide (SO2) │○ │ │ ┃ ┃ │Sulfuric Acid Test │It shall be a device to which START is supplied │ │ │ ┃ ┃ │Q44cde944fd0904d306842END │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Shielding Room │The shielding effect shall satisfy the following: │○ │ │ ┃ ┃ │ │1) In the low frequency range at 14 kHz │ │ │ ┃ ┃ │ │60 dB or higher, 80 dB at 200 kHz │ │ │ ┃ ┃ │ │Must be greater than or equal to │ │ │ ┃ ┃ │ │2) In the resonance region │ │ │ ┃ ┃ │ │is 100 dB or more at 30 MHz │ │ │ ┃ ┃ │ │thing │ │ │ ┃ ┃ │ │3) In the high-frequency range, 400 MHz, 1 │ │ │ ┃ ┃ │ │100 dB at 18 GHz and 18 GHz, respectively │ │ │ ┃ ┃ │ │Must be greater than or equal to │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │FUR. Electrostatic filter │Contact discharge 8 kV, standby discharge 15 kV │○ │ │ ┃ ┃ │Pre-test machine │able thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │He. Instantaneous voltage │Drop(100% and 50%), Period(1 │○ │ │ ┃ ┃ │Drop Tester │Measurement of cycle and 2 cycles), interval (10 s) │ │ │ ┃ ┃ │ │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Oscilloscope│Waveform observation up to 100 MHz is possible │○ │ │ ┃ ┃ │(oscilloscope │Things to do │ │ │ ┃ ┃ │: Voltage change │ │ │ │ ┃ ┃ │to screen │ │ │ │ ┃ ┃ │Showing the scene │ │ │ │ ┃ ┃ │value. │ │ │ │ ┃ ┃ │(The following is the same) │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │No. Grade and Overload │peak value (1 ~ 4) kV, │○ │ │ ┃ ┃ │Test Machine │rise time 5 ns, fall time 50 ns │ │ │ ┃ ┃ │(fast │pulse at 5 kHz (2.5 kHz for 4 kV) │ │ │ ┃ ┃ │transient │with a 300 ms cycle for 15 ms │ │ │ ┃ ┃ │burst │and be able to test │ │ │ ┃ ┃ │tester) │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Damped Vibration │3 ~ Peak value in the 6th period(peak │○ │ │ ┃ ┃ │Wave Tester │value) decreases to 50% 100 │ │ │ ┃ ┃ │ │kHz (repetition rate 40 Hz) and 1 MHz (half │ │ │ ┃ ┃ │ │complexity 400 Hz) (0.5 │ │ │ ┃ ┃ │ │~ Can be tested at 2.5 kV │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Conductivity │80% amplitude modulation with 1 kHz sine wave │○ │ │ ┃ ┃ │Frequency Interference │150 kHz ~ 80 MHz high-frequency interference │ │ │ ┃ ┃ │TestingMachine │Resistance testing will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Mo. High frequency │1) 80% Amplitude Modulation with 1 kHz Sine Wave │○ │ │ ┃ ┃ │Self-Test │High frequency (80 ~ 2000) MHz │ │ │ ┃ ┃ │START │Electromagnetic immunity testing will be possible │ │ │ ┃ ┃ │ │2) Width and height each 3 m, length 7 m this │ │ │ ┃ ┃ │ │installed in the electromagnetic shielding room │ │ │ ┃ ┃ │ │Uniform within a range of 1.5 m in width and height │ │ │ ┃ ┃ │ │will be able to form an electromagnetic field │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. High frequency field │High-frequency measurements of 150 kHz ~ 2 GHz are │○ │ │ ┃ ┃ │Solver │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │so. Surge(surg │1) Voltage (1.2/50) μs, Current (8/20) │○ │ │ ┃ ┃ │e: Currentㆍ │μs waveform will be formed │ │ │ ┃ ┃ │voltage etc. │2) Surge voltage up to (0 ~ 4) kV │ │ │ ┃ ┃ │Transient waveform │able thing │ │ │ ┃ ┃ │refers to. │ │ │ │ ┃ ┃ │(The following is the same) │ │ │ │ ┃ ┃ │Tester │ │ │ │ ┃ ┠───────┼────────┼────────────────────┼──┼──┼──┨ ┃13. Armature │A. Laboratory │Temperature (23 ± 2) °C, relative humidity (40 ~ │○ │○ │○ ┃ ┃Driving │ │Maintaining % ± 10 % will be possible │ │ │ ┃ ┃Electric ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Reference Power │1) AC Charger: Charging Power Measurement │○ │○ │○ ┃ ┃ │Weight meter │Accuracy is three-phase │ │ │ ┃ ┃ │ │Power factor within 1.0 to 0.1%, interchangeable │ │ │ ┃ ┃ │ │Current voltage (0 ~ 300) V, AC current │ │ │ ┃ ┃ │ │(0 ~ 40) A, Phase (0 ~ 360)°, Primary │ │ │ ┃ ┃ │ │Capable of measuring wavenumbers (45 ~ 65) Hz │ │ │ ┃ ┃ │ │것, │ │ │ ┃ ┃ │ │2) DC Charger: Charging Power Measurement │ │ │ ┃ ┃ │ │With accuracy within 0.1%, DC voltage │ │ │ ┃ ┃ │ │(0 ~ 1000) V, direct current (0 ~ │ │ │ ┃ ┃ │ │120) A will be measurable │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Heo Bu-ha's foot │1) AC Charger: AC Voltage (0 ~ │○ │○ │○ ┃ ┃ │Production( Test │300) V, AC current (0 ~ 40) A, │ │ │ ┃ ┃ │Power supply for │Phase (0 ~ 360)°, Frequency (45 ~ │ │ │ ┃ ┃ │level device) │65) Hz power single-phase (single │ │ │ ┃ ┃ │ │phase) and three phases │ │ │ ┃ ┃ │ │May occur │ │ │ ┃ ┃ │ │2) DC Charger: DC Voltage (0 ~ │ │ │ ┃ ┃ │ │1000) V, DC current (0 ~ 120) A of │ │ │ ┃ ┃ │ │Can generate power │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Harmonic generator │1) Phase angle of 5th harmonic AC current │○ │ │ ┃ ┃ │animation │From 0° to (0 to 40)% of the rated, │ │ │ ┃ ┃ │ │voltage at a phase angle of 0° rated (0 │ │ │ ┃ ┃ │ │~ 10) To test including up to % │ │ │ ┃ ┃ │ │can be │ │ │ ┃ ┃ │ │2) AC current phase angles 90° and 270° │ │ │ ┃ ┃ │ │에서 발생되어 홀수고조파 전류 │ │ │ ┃ ┃ │ │testing shall be possible, during every 4 cycle │ │ │ ┃ ┃ │ │At current phase angle 0° every 2 cycles │ │ │ ┃ ┃ │ │all generated harmonics (less than 1 │ │ │ ┃ ┃ │ │Non-integer multiples of harmonic orders (Non-integer multiples of harmonic orders) │ │ │ ┃ ┃ │ │倍) Harmonic components) Testing possible │ │ │ ┃ ┃ │ │thing │ │ │ ┃ ┃ │ │3) Half-wave rectification (positive (+) of AC voltage also │ │ │ ┃ ┃ │ │uses only one side of the negative (-) │ │ │ ┃ ┃ │ │rectification method) by applying the rectified current directly │ │ │ ┃ ┃ │ │and even harmonic testing is possible │ │ │ ┃ ┃ │ │thing │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │E. External Magnetic │1) Continuous magnetic induction │○ │ │ ┃ ┃ │Impact Tester │Test: Magnetomotive Force │ │ │ ┃ ┃ │ │will be 1000 AT │ │ │ ┃ ┃ │ │2) Power supply frequency magnetic field induction(magnetic │ │ │ ┃ ┃ │ │induction) Test: Magnetization Coil(magnetic │ │ │ ┃ ┃ │ │coil) Diameter 1 m, magnetomotive force 400 │ │ │ ┃ ┃ │ │AT will be │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │B. Constant Temperature and Humidity │1) Temperature: (-30 ~ 130) °C ± 2 °C │○ │ │ ┃ ┃ │START │It will be within │ │ │ ┃ ┃ │ │Relative Humidity: (40 – 85) % ± 5 % │ │ │ ┃ ┃ │ │It will be within │ │ │ ┃ ┃ │ │3) Electric Vehicle Charger Error Test Site │ │ │ ┃ ┃ │ │Configured to test with START value │ │ │ ┃ ┃ │ │Things to do │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Temperature Recording │with at least 4 channels, temperature measurement │○ │ │ ┃ ┃ │Total │The range shall be (-50 to +150) °C │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Ah. Thermometer │(-10 ~ 100) °C range temperature │○ │ │ ┃ ┃ │ │Measurement down to 0.1 °C will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. Oscillos │Waveform observation up to 100 MHz is possible │○ │ │ ┃ ┃ │Corp │Things to do │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Car. Shielding Room │The shielding effect shall satisfy the following: │○ │ │ ┃ ┃ │ │1) In the low frequency range at 14 kHz │ │ │ ┃ ┃ │ │60 dB or higher, 80 dB at 200 kHz │ │ │ ┃ ┃ │ │Must be greater than or equal to │ │ │ ┃ ┃ │ │2) In the resonance region │ │ │ ┃ ┃ │ │is 100 dB or more at 30 MHz │ │ │ ┃ ┃ │ │thing │ │ │ ┃ ┃ │ │3) In the high-frequency range, 400 MHz, │ │ │ ┃ ┃ │ │100 dB at 1 GHz and 18 GHz │ │ │ ┃ ┃ │ │Must be greater than or equal to │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Car. Electrostatic discharge │Contact discharge 8 kV, standby discharge 15 kV │○ │ │ ┃ ┃ │Pre-test machine │able │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │T. Instantaneous voltage │Drop(100% and 50%), Period(1 │○ │ │ ┃ ┃ │Drop Tester │Measurement of cycle and 2 cycles), interval (10 s) │ │ │ ┃ ┃ │ │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │Part. Grade and also Bur │peak value (1 ~ 4) kV, phase │○ │ │ ┃ ┃ │ST Test Machine │Pulse with a rise time of 5 ns and a fall time of 50 ns │ │ │ ┃ ┃ │(fast │to 5 kHz (2.5 kHz for 4 kV) │ │ │ ┃ ┃ │transient │Apply with a 300 ms period for 15 ms │ │ │ ┃ ┃ │burst │Can test │ │ │ ┃ ┃ │tester) │ │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │H. Conductivity │80% amplitude modulation with 1 kHz sine wave │○ │ │ ┃ ┃ │Frequency Interference │150 kHz ~ 80 GHz high-frequency interference │ │ │ ┃ ┃ │TestingMachine │Resistance testing will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. High frequency │1) 80% Amplitude Modulation with 1 kHz Sine Wave │○ │ │ ┃ ┃ │Self-testing machine │High frequency (80 ~ 2700) MHz │ │ │ ┃ ┃ │ │Electromagnetic immunity testing will be possible │ │ │ ┃ ┃ │ │2) Width and height each 3 m, length 7 m this │ │ │ ┃ ┃ │ │installed in the electromagnetic shielding room │ │ │ ┃ ┃ │ │Uniform within a range of 1.5 m in width and height │ │ │ ┃ ┃ │ │will be able to form an electromagnetic field │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │. High frequency field │High-frequency measurements of 150 kHz ~ 2 GHz are │○ │ │ ┃ ┃ │Solver │will be possible │ │ │ ┃ ┃ ├────────┼────────────────────┼──┼──┼──┨ ┃ │The. Surge Test │1) Voltage (1.2/50) μs, Current (8/20) │○ │ │ ┃ ┃ │기 │μs waveform will be formed │ │ │ ┃ ┃ │ │2) Surge voltage up to (0 ~ 4) kV │ │ │ ┃ ┃ │ │able │ │ │ ┃ ┠───────┴────────┴────────────────────┴──┴──┴──┨ ┃Remark: Type approval agencies, inspection agencies, re-inspection performing agencies, and self-inspection business operators are measuring instruments ┃ ┃It is sufficient to be equipped with testing and verification facilities corresponding to the scope of work for each. However, for re-verification ┃ ┃The local government performing the task shall enter into a separate contract with the person possessing the testing and verification facilities ┃ ┃, external testing and verification facilities may be utilized. ┃ ┗━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━┛