Form 8 — 의료기기법 시행규칙
287877 · 2026-07-01
■ Enforcement Rules of the Medical Devices Act [Attachment Form No. 8] <Amended Apr. 1, 2025>
Request for Review of Medical Device Technical Documents, etc.
※ [ ]. (front)
───────────────┬───────┬────────┬────────────────────────
Receipt Number │Date of Receipt │Processing Date │Processing Period
│ │ │1. Newly Developed Medical Device
│ │ │ a. Subject to technical documentation review: 55 days
│ │ │ B. Clinical trial data subject to review: 70 days
│ │ │2. Medical devices other than newly developed medical devices
│ │ │ a. Technical documentation and clinical trial data subject to review: 70 days
│ │ │ B. Subject to Technical Document Review: 55 days
│ │ │3. Change Review
│ │ │ A. Review of changes to technical documentation and clinical trial data
│ │ │Target: 50 days
│ │ │ B. Technical Documentation Change Review Subject: 32 days
───────────────┴───────┴────────┴────────────────────────
─────────┬─────────────────────┬─────────────────────────
Client │Name │Date of Birth
├─────────────────────┴─────────────────────────
│Address
─────────┴───────────────────────────────────────────────
─────────┬─────────────────────┬─────────────────────────
Manufacture(Import) │Name (Trade Name) │Business Permit Number
Business ├─────────────────────┴─────────────────────────
│Location
─────────┴───────────────────────────────────────────────
─────────┬─────────────────────┬─────────────────────────
Manufacturer (Importer │Name (Trade Name) │Country of manufacture
or manufacturing process ├─────────────────────┴─────────────────────────
In the case of full consignment)│Location
─────────┴───────────────────────────────────────────────
─────────┬────┬──────────────────────────────────────────
Request for Examination │New Development │ [ ] Technical Documentation Review
Category │Medical Device│ [ ] Clinical Trial Data Review
├────┼──────────────────────────────────────────
│New Development │ [ ] Review of Technical Documentation and Clinical Trial Data [ ]Technical Documentation Review
│Medical Device│ [ ] Change Review ([ ]Submission of Technical Documentation and Clinical Trial Data [ ]Technical Document Submission)
│Other than │
│Medical Device│
─────────┴────┴──────────────────────────────────────────
─────────────────────────────────────────────────────────
Request for Review Item
───────────────┬─────────────────────────────────────────
Name(Product Name, Item Name, Model Name) │
───────────────┼─────────────────────────────────────────
Classification Number (Grade) │
───────────────┼─────────────────────────────────────────
Shape and Structure │
───────────────┼─────────────────────────────────────────
Raw Materials │
───────────────┼─────────────────────────────────────────
Manufacturing method │
───────────────┼─────────────────────────────────────────
Perfor- mance │
───────────────┼─────────────────────────────────────────
Purpose of use │
───────────────┼─────────────────────────────────────────
Usage Method │
───────────────┼─────────────────────────────────────────
Precautions for Use │
───────────────┼─────────────────────────────────────────
Packing Unit │
───────────────┼─────────────────────────────────────────
Storage Method and Usage Period │
───────────────┼─────────────────────────────────────────
Test standard │
───────────────┼─────────────────────────────────────────
PermissionConditions │
───────────────┼─────────────────────────────────────────
Remarks (Basis for Application) │
───────────────┴─────────────────────────────────────────
Pursuant to Article 9, Paragraphs 2 and 3 of the Enforcement Rule of the Medical Devices Act, we hereby request the examination of the medical device technical documents, etc. as stated above.
Year month day
Client (Signature or seal)
Contact Person Name
Contact Person Phone Number
To the Minister of Food and Drug Safety (Head of Technical Document Review Agency)
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
210㎜ × 297㎜[White paper 80g/㎡ or medium paper 80g/㎡]
(rear)
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
─────────────────────────────────────────────┬──────────────────────────
Attached Documents │Fee
│
│Pursuant to Annex 10 of the Enforcement Rule of the Medical Devices Act
│Amount according to the following classification
├────────────┬─────────────
│Electronic Application │In-person or Postal Application
─────────────────────────────────────────────┼────────────┼─────────────
1. In the case of medical devices (excluding in vitro diagnostic medical devices), the documents specified in each of the following subparagraphs. However, already │(excluding newly developed medical devices │(excluding newly developed medical devices
A medical device that is substantially equivalent to an approved device in structure, principle, performance, purpose of use, and method of use │MedicalDevice) │MedicalDevice)
Among medical devices, those designated by the Minister of Food and Drug Safety as requiring submission regarding clinical trials │1. Technical documentation and clinical │1. Technical documentation and clinical
For medical devices other than those that the Minister of Food and Drug Safety has designated as requiring clinical trial data, submission of the documents under items (e) through (g) may be omitted. │Examination Data Review: │Examination Data Review:
. │1,338,000 won │1,488,000 won
a. Data compared with already approved products │2. Technical Documentation Review: │2. Technical Documentation Review:
B. Data regarding the purpose of use │561,000 won │624,000 Won
. Data regarding the operating principle │3. Technical documentation and clinical │3. Technical documentation and clinical
d. Test standards and the basis for their establishment as data for each of the following items to verify the performance and safety of the product │Review of Changes to Clinical Trial Data: │Review of Changes to Clinical Trial Data:
and actual measured values. However, if there are no domestic or foreign test standards, technical documents, etc. │888,000 won │987,000 won
inspection to verify the performance and safety of a product and its │4. Technical Documentation Change Review: │4. Technical Documentation Change Review:
Data regarding the basis and measured values │430,000 │478,000
1) Data on Electrical and Mechanical Safety │ │
2) Data on biological safety │(Newly Developed Medical Device) │(Newly Developed Medical Device)
3) Safety Data Regarding Radiation │1. Technical Documentation Review: │1. Technical Documentation Review:
4) Data on Electromagnetic Safety │55,240,000 won │55,240,000 won
5) Performance data │2. Clinical Trial Review: │2. Clinical Trial Review:
6) Data on physical and chemical properties │39,170,000 Won │39,170,000 won
7) Data regarding stability │ │
e. Data on origin, discovery and development history │ │
f. Clinical trial data │ │
g. Data on use in other countries │ │
│ │
2. In the case of in vitro diagnostic medical devices, the following documents │ │
A. Data on development history, measurement principles and methods, and current usage status at home and abroad │ │
B. Data on raw materials and manufacturing methods │ │
. Data regarding the purpose of use │ │
d. Data regarding storage method and period of use or validity │ │
E. The following data for verifying product performance │ │
1) Data on Analytical Performance Testing │ │
2) Data on clinical performance testing │ │
3) Data on Quality Control Testing │ │
4) Data on storage of standard materials and specimens, etc. │ │
f. Safety information for handlers of in vitro diagnostic medical devices │ │
g. Data comparing the device with already authorized products │ │
│ │
Note: Notwithstanding items 1 and 2, an applicant who has obtained prior review of manufacturing authorization or notification under Article 11 of the Medical Devices Act │ │
and submits the notice of the results of prior review under Article 25 (3) of its Enforcement Rules │ │
may omit the documents referred to in items 1 or 2 concerning matters │ │
found suitable in the prior review. │ │
─────────────────────────────────────────────┴────────────┴─────────────
━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━━
Processing Procedure
────────────────────────────────────────────────────────────────────────
┌──────┐ ┌───┐ ┌───┐ ┌───┐ ┌─────┐ ┌─────┐
│Request Form Creation │ ?│Receipt │ ?│Review │ ?│Approval │ ?│ResultNotification│ ?│Result Notification │
│ ├─ │ ├─ │ ├─ │ ├─ │Preparation ├─ │ │
│ │ │ │ │ │ │ │ │ │ │ │
└──────┘ └───┘ └───┘ └───┘ └─────┘ └─────┘
Client Ministry of Food and Drug Safety / Technical Document Review Agency
────────────────────────────────────────────────────────────────────────